(158 days)
The SiO2 Coated Syringe 3 mL is intended to be used to inject fluids into, or withdraw fluids from, the body.
SiO2 Coated Syringe 3 mL is a plastic disposable standard 3 mL piston syringe without needle is made of a cyclic olefin polymer (COP) and coated internally with silica glass layers. The layer is deposited by plasma enhanced chemical vapor deposition (PECVD) process.
This is a standard 3 mL single use, disposable syringe without needle.
This document describes the non-clinical testing for the SiO2 Coated Syringe 3 mL to demonstrate its substantial equivalence to a predicate device. Here's a breakdown of the requested information based on the provided text, focusing on the acceptance criteria and the study proving it:
1. Table of Acceptance Criteria and Reported Device Performance
The document lists numerous tests and standards to which the device conforms. Instead of specific numerical acceptance criteria (which are typically defined within the standards themselves and not fully detailed here), the document generally states "Complies" or describes the test performed and implies compliance by its inclusion in the summary of successful non-clinical testing.
| Acceptance Criteria (Standard & Section/Test) | Reported Device Performance |
|---|---|
| ISO 7886-1: Sterile hypodermic syringes for single use | |
| Section 5, Cleanliness | Complies (Test Performed) |
| Section 6 & Annex A, Limits for Acidity or Alkalinity | Complies (Test Performed) |
| Section 7 & Annex A, Limits for Extractable Metals | Complies (Test Performed) |
| Section 8, Lubricant | Complies (Test Performed) |
| Section 9, Tolerance on Graduated Capacity | Complies (Test Performed) |
| Section 10.1, Scale | Complies (Test Performed) |
| Section 10.2, Numbering of Scale | Complies (Test Performed) |
| Section 10.3, Overall Length of Scale to Nominal Capacity Line | Complies (Test Performed) |
| Section 10.4, Position of Scale | Complies (Test Performed) |
| Section 11.1, Dimensions | Syringe Dimensions (Test Performed) |
| Section 11.2, Finger Grips | Complies (Test Performed) |
| Section 12.1, Design | Complies (Test Performed) |
| Section 12.2, Fit of Piston in Barrel | Complies (Test Performed) |
| Annex G, Force to Operate | Plunger Withdraw: Fi=2.11 N, Fm=2.39 N; Plunger Deploy: Fi=2.35 N, Fm=0.58 N (Reported values) |
| Section 12.3, Fiducial Line | Complies (Test Performed) |
| Section 13.2, Position of Nozzle on End of Barrel | Complies (Test Performed) |
| Section 13.3, Nozzle Lumen | Complies (Test Performed) |
| Section 14.1 - Annex C, Dead Space | Complies (Test Performed) |
| Section 14.2 - Annex D, Liquid Leakage Compression | Complies (Test Performed) |
| Section 14.2 - Annex B, Air Leakage past Syringe Piston during Aspiration | Complies (Test Performed) |
| ISO 594 - 1 and -2: Conical fittings with a 6 % (Luer) taper | |
| Section 5.1, Gauging | Complies (Test Performed) |
| Section 5.2, Liquid Leakage | Complies (Test Performed) |
| Section 5.3, Air Leakage | Complies (Test Performed) |
| Section 5.4, Separation Force | Complies (Test Performed) |
| Section 5.5, Unscrewing Torque | Complies (Test Performed) |
| Section 5.6, Ease of Assembly | Complies (Test Performed) |
| Section 5.7, Resistance to Overriding | Complies (Test Performed) |
| Section 5.8, Stress Cracking | Complies (Test Performed) |
| Biocompatibility Testing (ISO 10993-1, G95-1) | |
| Cytotoxicity: Elution Test, Serum Supplemented MEM Extract | Materials found to be non-reactive (Test Performed) |
| Skin Sensitization Test Protocol, Guinea Pig Maximization Test | Materials found to be non-reactive (Test Performed) |
| Intracutaneous (Intradermal) Reactivity Test | Materials found to be non-reactive (Test Performed) |
| Systemic Toxicity: Saline and Vegetable Oil Extracts | Materials found to be non-reactive (Test Performed) |
| Hemolysis Test (NIH Method) Saline Extract | Materials found to be non-reactive (Test Performed) |
| Extractable and Leachables (ISO 7886-1 coated/uncoated) | Complies (Test Performed) |
| USP Rabbit Pyrogen Test (Material Mediated) | Complies (Test Performed) |
| Bacterial Endotoxin-Mediated Pyrogen test (Limulus Amebocyte Lysate Test) | Complies (Test Performed) |
| Other Non-Clinical Testing | |
| ASTM D4169-14: Performance Testing of Shipping Containers and Systems | Complies (Test Performed) |
| ASTM F1980-07 (2011): Accelerated Aging of Sterile Barrier Systems | Complies (Test Performed) |
| Real-time aging (24 months) | Complies (Test Performed) |
| Oxygen Transmission Rate ("OTR") | Complies (Test Performed) |
| Container compatibility with certain drug products | Complies (Test Performed) |
| Mechanical stressor | Complies (Test Performed) |
| Environmental stressor - freeze-thaw study | Complies (Test Performed) |
| Ink adhesion and Wipe & Tape Test Report | Complies (Test Performed) |
| Physical Characteristics Test | Complies (Test Performed) |
| Syringe Dimensions | ID: 8.62mm, OD: 11.34mm (Reported values) |
| Plunger Force Test | Fi=2.11 N, Fm=2.39 N, Fi=2.35 N, Fm=0.58 N (Reported values) |
| FTIR Test method for coating chemistry | Complies (Test Performed) |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state the sample sizes for each of the non-clinical tests. It lists the tests performed, implying that appropriate sample sizes per the respective standards (ISO, ASTM, etc.) were used. The data provenance is non-clinical bench testing, conducted by or on behalf of SiO2 Medical Products. The document does not specify countries of origin for the data, but it is implied to be from the manufacturer's testing or accredited labs. All data presented appears to be prospective in the sense that it was generated for this specific submission to demonstrate conformance.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This section is not applicable to this submission. The tests are non-clinical engineering and biocompatibility tests against established international standards (ISO, ASTM, USP). Ground truth is established by the methods defined in these standards, not by expert consensus on clinical observations. The "experts" in this context would be laboratory technicians and engineers qualified to perform these specific tests according to the standard operating procedures.
4. Adjudication Method for the Test Set
This section is not applicable. Adjudication methods like 2+1 or 3+1 typically refer to clinical or diagnostic studies involving human interpretation. For non-clinical bench testing, results are typically determined by measurements against predefined acceptance criteria from the relevant standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
This section is not applicable. This submission is for a physical medical device (syringe) and does not involve AI, human readers, or diagnostic interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This section is not applicable. This submission is for a physical medical device and does not involve an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the non-clinical tests listed is defined by the acceptance criteria specified within the referenced international standards (e.g., ISO 7886-1, ISO 594, ISO 10993-1, ASTM D4169-14, ASTM F1980-07). For biocompatibility, this includes outcomes of biological assays (e.g., cytotoxicity, sensitization, systemic toxicity, hemolysis, pyrogenicity). For physical tests, it refers to quantitative measurements (e.g., dimensions, forces, leakage rates) meeting specified ranges or thresholds.
8. The Sample Size for the Training Set
This section is not applicable. This submission is for a physical medical device and does not involve a training set for an algorithm.
9. How the Ground Truth for the Training Set Was Established
This section is not applicable. This submission is for a physical medical device and does not involve a training set for an algorithm.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular pattern.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 17, 2016
SiO2 Medical Products c/o Mr. Paul Dryden Consultant 350 Enterprise Drive Auburn, Alabama 36830
Re: K153553
Trade/Device Name: SiO2 Coated Syringe 3 mL Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: April 15, 2016 Received: April 18, 2016
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
K153553
Device Name
SiO2 Coated Syringe 3 mL
Indications for Use (Describe)
The SiO2 Coated Syringe 3 mL is intended to be used to inject fluids into, or withdraw fluids from, the body.
Type of Use (Select one or both, as applicable) XX Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
EF PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
510(k) Summary Page 1 of 4 16-May-16
| SiO2 Medical Products350 Enterprise DriveAuburn, Alabama 36830 | Tel – 334-321-5060Email: randy.crenshaw@sio2med.com |
|---|---|
| Official Contact: | Randy CrenshawDirector of Quality and Regulatory |
| Proprietary or Trade Name: | SiO2 Coated Syringe 3 mL |
| Common/Usual Name: | Syringe, Piston |
| Classification Name/Code: | FMF - Syringe, Piston21 CFR 880.5860Class II |
| Device: | SiO2 Coated Syringe 3 mL |
| Predicate Device: | K110771 – Becton Dickinson – Single Use syringe |
| Reference Device: | K111091 - Merit Medical Syringe |
Device Description:
SiO2 Coated Syringe 3 mL is a plastic disposable standard 3 mL piston syringe without needle is made of a cyclic olefin polymer (COP) and coated internally with silica glass layers. The layer is deposited by plasma enhanced chemical vapor deposition (PECVD) process.
This is a standard 3 mL single use, disposable syringe without needle.
Indications for Use:
The SiO2 Coated Syringe 3 mL is intended to be used to inject fluids into, or withdraw fluids from, the body.
Contraindications
- Do not use this syringe with sesame oil-based drugs.
- . Do not use this syringe with non-polar solvents.
Discussion of Substantial Equivalence
| Element ofComparison | Subject Device | Becton Dickinson Single Use SyringeK110771 |
|---|---|---|
| Syringe Type | Piston Syringe - FMF | Piston Syringe - FMF |
| Intended Use(s) | The SiO2 Coated Syringe 3 mL is intended tobe used to inject fluids into, or withdrawfluids from, the body. | For use by health care professionals forgeneral purpose fluid aspiration/injection. |
| Length | 82.77mm | 79.3 – 80.7mm |
| Barrel diameter | ID: 8.62mmOD: 11.34mm | ID: 8.5 – 8.7mmOD: 10.0 – 10.2mm |
Table 1 - Substantial Equivalence Comparative Table
{4}------------------------------------------------
510(k) Summary Page 2 of 4 16-May-16
| Element ofComparison | Subject Device | Becton Dickinson Single Use SyringeK110771 |
|---|---|---|
| Finger Grip Size andShape | Length: 24.64 – 25.14mmWidth: 14.43 - 14.93mmThickness: 1.85-2.11mm | Length: 24.33 - 24.59mmWidth: 12.57 – 12.83mmThickness: 1.85-2.11mm |
| Tip type | Centric - male luer lockComplying with ISO 594-1, -2 | Centric - male luer lockComplying with ISO 594-1,-2 |
| Volume | 3 mL | 3 mLPlus various sizes |
| Barrel marking specs | Complies with ISO 7886-1 | Complies with ISO 7886-1 |
| Graduations legibility | Legible | Legible |
| Lubricant type | Silicone | Silicone |
| Plunger Operation(Fi and Fm Forces) | Plunger Withdraw:Fi=2.11 N, Fm=2.39 NPlunger Deploy:Fi=2.35 N, Fm=0.58 N | Plunger Withdraw:Fi=1.18 N, Fm=1.48 NPlunger Deploy:Fi=2.21 N, Fm=1.68 N |
| Barrel transparency | Transparent and Clear | Transparent and Clear |
| Delivery accuracy | Complies with ISO 7886-1ISO 7886-1 | Complies with ISO 7886-1ISO 7886-1 |
| Labeling | ISO 7886-1ISO 15223-121 CFR Part 801 | 21 CFR Part 801 |
| Sterilization | E-beam Irradiation sterilization 25KGy;achieving 10-6 SAL | Ethylene Oxide or Irradiation sterilization10-6 SAL |
| Shelf-life andAge testing | 12 months empty | |
| Mechanical andenvironmentalstressors | Mechanical forceFreeze / thaw testing | |
| Verification of coatingchemistry | Using FTIR Microscopy Test Method |
The subject device is viewed as substantially equivalent to the predicate device because:
Indications for Use – The proposed indications for use are similar, in that they are to be used to inject fluids into, or withdraw fluids from, the body, also referred to as general purpose fluid aspiration/injection.
Discussion - The indications for use are similar for the subject device and the predicate -K110771 – BD single use syringe.
Technology – The basic design of the subject device is similar as the subject device complies with ISO 7886-1.
Discussion - The basic technology of a plunger style syringe is identical to the predicate. The technology of applying a glass coating to the inside of the barrel is different than the predicate but we have demonstrated that its performance is equivalent and complies with ISO 7886-1 as does the predicate - K110771 - BD Single Use Syringe. The difference in technology has been evaluated and do not raise any new questions related to risk, safety, or effectiveness.
{5}------------------------------------------------
510(k) Summary Page 3 of 4 16-May-16
Materials – The materials in fluid contact have been evaluated in accordance to the suggestions of ISO 7886-1 and ISO 10993-1 and G95-1 which consider the patient contact as Externally communicating, Circulating blood with Limited duration (<24 hours).
We evaluated the subject device utilizing the following tests:
- Cytotoxicity: Elution Test, Serum Supplemented MEM Extract ●
- Skin Sensitization Test Protocol, Guinea Pig Maximization Test ●
- Intracutaneous (Intradermal) Reactivity Test
- Systemic Toxicity: Saline and Vegetable Oil Extracts
- . Hemolysis Test (NIH Method) Saline Extract
- Extractable and Leachables as described in ISO 7886-1 coated / uncoated .
- . USP Rabbit Pyrogen Test (Material Mediated)
- . Bacterial Endotoxin-Mediated Pyrogen test (Limulus Amebocyte Lysate Test)
Discussion - We performed the applicable testing for biocompatibility and the materials were found to be non-reactive. We have made reference to the Merit Medical Syringe K111091 which has similar indications for use and the barrel is made of Cyclo-olefin polymer. We included this reference device to support that the Cyclo-olefin polymer is not a new material for use in syringes.
Non-clinical Testing
We performed bench testing to demonstrate that the subject device meets the ISO 7886-1 and other applicable standards. Testing included:
| Standard and Tests Performed |
|---|
| ISO 7886-1- Sterile hypodermic syringes for single use -- Part 1: Syringes for manual use |
| Section 5, Cleanliness |
| Section 6 & Annex A, Limits for Acidity or Alkalinity |
| Section 7 & Annex A, Limits for Extractable Metals |
| Section 8, Lubricant |
| Section 9, Tolerance on Graduated Capacity |
| Section 10.1, Scale |
| Section 10.2, Numbering of Scale |
| Section 10.3, Overall Length of Scale to Nominal Capacity Line |
| Section 10.4, Position of Scale |
| Section 11.1, Dimensions |
| Section 11.2, Finger Grips |
| Section 12.1, Design |
| Section 12.2, Fit of Piston in Barrel |
| Annex G, Force to Operate |
| Section 12.3, Fiducial Line |
| Section 13.2, Position of Nozzle on End of Barrel |
| Section 13.3, Nozzle Lumen |
| Section 14.1 - Annex C, Dead Space |
| Section 14.2 - Annex D, Liquid Leakage Compression |
| Section 14.2 - Annex B, Air Leakage past Syringe Piston during Aspiration |
| ISO 594 - 1 and -2 - Conical fittings with a 6 % (Luer) taper for syringes, needles and certain othermedical equipment -- Part 1 and Part 2 |
| Section 5.1, Gauging |
| Section 5.2, Liquid Leakage |
{6}------------------------------------------------
510(k) Summary Page 4 of 4 16-May-16
| Section 5.3, Air Leakage |
|---|
| Section 5.4, Separation Force |
| Section 5.5, Unscrewing Torque |
| Section 5.6, Ease of Assembly |
| Section 5.7, Resistance to Overriding |
| Section 5.8, Stress Cracking |
| ASTM D4169-14; Standard Practice for Performance Testing of Shipping Containers and Systems |
| Manual handling – drop test and second Drop Sequence |
| Vehicle stacking |
| Loose Load Vibration |
| Vehicle Vibration |
| Concentrated Impact |
- ASTM F1980-07 (2011); Standard Guide for Accelerated Aging of Sterile Barrier ● Systems for Medical Devices
- Real-time aging 24 months ●
- Oxygen Transmission Rate ("OTR") ●
- Container compatibility with certain drug products ●
- Mechanical stressor
- Environmental stressor - freeze-thaw study Summary
- . Ink adhesion and Wipe & Tape Test Report
- . Plunger Force Test
- . Physical Characteristics Test
- . Syringe Dimensions
- Plunger Force Test .
- . FTIR Test method for coating chemistry
Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the SiO2 Coated Syringe 3 mL and is substantially equivalent to the predicate device.
§ 880.5860 Piston syringe.
(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).