The HS70A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-raginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
The HS70A is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The HS70A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS70A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
Here's an analysis of the provided text regarding the acceptance criteria and study proving device performance, based on the requirements you outlined.
It's important to note that the provided document is a 510(k) Premarket Notification from the FDA, specifically concerning the SAMSUNG MEDISON HS70A Diagnostic Ultrasound System. This type of submission focuses on demonstrating "substantial equivalence" to existing, legally marketed predicate devices, rather than proving novel clinical efficacy or undertaking broad clinical trials typical for new therapeutic devices or high-risk diagnostics. Therefore, much of the requested information regarding detailed clinical study performance, ground truth establishment for a test set, and multi-reader multi-case studies will likely not be present as it falls outside the scope of a 510(k) for an ultrasound system of this class.
Acceptance Criteria and Device Performance Study for SAMSUNG MEDISON HS70A Diagnostic Ultrasound System (K152962)
The provided document describes the substantial equivalence determination for the HS70A Diagnostic Ultrasound System. For this type of device and submission, the "acceptance criteria" are primarily based on demonstrating equivalence to predicate devices across various technological characteristics and meeting established medical device safety standards, rather than specific performance metrics from a clinical study for a new intended use.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the comparison to predicate devices and adherence to relevant standards. The performance is then demonstrated by showing that the HS70A meets or exceeds the capabilities of these predicate devices and complies with safety standards.
| Acceptance Criterion (Implicit) | Reported Device Performance (HS70A) |
|---|---|
| Intended Use Equivalence: Capable of diagnostic ultrasound imaging and fluid analysis for human body, across specified clinical applications. | Meets/Exceeds Predicates: Intended for diagnostic ultrasound imaging and fluid analysis of the human body. Clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel. (Identical or expanded list compared to predicates). |
| Technological Characteristics Equivalence: Comparable operational modes, scanhead types, frequency ranges, and features. | Meets/Exceeds Predicates: Operates in B mode, Color Doppler, Power Doppler (including S-Flow), PW Spectral Doppler, CW Spectral Doppler, Harmonic imaging (S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging (real time 4D imaging), Elastoscan Mode, or combinations. Supports Linear Array, Curved Linear Array, Endocavity, and Phased Array scanheads. Frequency range 1.0 ~ 20.0 MHz. Includes various advanced functionalities like MultiVision, Panoramic, Elastoscan, etc., some of which are new or enhanced compared to specific predicates. |
| Safety and EMC Compliance: Adheres to recognized medical device safety and electromagnetic compatibility standards. | Meets/Exceeds Predicates: Conforms to UL 60601-1, CSA C22.2 No. 601.1, IEC60601-2-37 (Diagnostic Ultrasound Safety), EN/IEC60601-1, EN/IEC60601-1-2 (EMC), NEMA UD-2 (Acoustic Output Measurement), NEMA UD-3 (Real Time Display of AI/TI). Biocompatibility tested to ISO10993-1. Application of risk management to ISO14971. Acoustic output display (Track 3) is present and compliant. |
| Patient Contact Materials: Biocompatibility shown. | Meets Criteria: Patient contact materials tested to ISO 10993-1. |
| Quality System Compliance: Manufactured under a quality management system. | Meets Criteria: Design, development, and quality process conforms with 21 CFR 820 and ISO 13485. |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "Summary of Clinical Tests: Not applicable. The subject of this submission. HS70A, did not require clinical studies to support substantial equivalence."
Therefore, there is no test set in the traditional sense of a clinical or retrospective imaging dataset used to evaluate diagnostic performance against a ground truth. The "proof" of performance relies on engineering verification and validation (bench testing, safety compliance) and comparison of technical specifications to legally marketed predicate devices, as detailed in the "Technological Characteristics Comparison Table" (pages 23-24).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Since no clinical test set was required or performed for diagnostic performance evaluation, this information is not applicable. The "ground truth" for a 510(k) submission of this nature refers to the established safety and effectiveness of the predicate devices.
4. Adjudication Method for the Test Set
As no clinical test set was used for diagnostic performance evaluation against an established ground truth, adjudication methods are not applicable.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The submission explicitly states that clinical studies were not required to support substantial equivalence.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The HS70A is a diagnostic ultrasound system. While it incorporates advanced features/algorithms (like Tissue Harmonic Imaging, 3D imaging, MultiVision, Elastoscan), these are integrated functionalities of the device rather than standalone AI algorithms being submitted for independent performance claims. The performance is assessed as part of the overall system's technical specifications and safety profile, in comparison to predicates. Therefore, a "standalone algorithm performance" study as typically understood for AI/ML devices is not applicable in this context.
7. The Type of Ground Truth Used
The "ground truth" in this 510(k) pertains to the established safety, effectiveness, and performance of the predicate devices (WS80A, RS80A, A35, ACCUVIX XG). The HS70A demonstrates substantial equivalence by showing that its technological characteristics and performance meet or exceed those of these already cleared devices, and that it complies with all relevant safety standards. No new disease diagnosis performance metrics against an independent ground truth (e.g., pathology, outcomes data) were required or presented for this specific submission given its regulatory pathway.
8. The Sample Size for the Training Set
Since this is a diagnostic ultrasound system submission under the 510(k) pathway, and not a de novo AI/ML algorithm submission, there is no mention or requirement for a "training set" in the context of machine learning model development. The development of the device itself would involve internal engineering and design verification/validation, but not typically a "training set" for an explicit algorithm performance evaluation against clinical ground truth.
9. How the Ground Truth for the Training Set Was Established
As there is no mention of a "training set" in the context of an AI/ML algorithm, this information is not applicable. The "ground truth" for the predicate devices' safety and effectiveness was established through their own prior clearances and clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2015
SAMSUNG MEDISON CO., LTD. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW BUFFALO MN 55313
Re: K152962
Trade/Device Name: HS70A Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: October 6, 2015 Received: October 7, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name HS70A Diagnostic Ultrasound System
Indications for Use (Describe)
The HS70A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-raginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
510(k) No.:
Device Name: HS70A Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | N | N | N | N | N | Note 1 | Notes 2, 4, 5, 6, 7, 8, 9, 11 | |
| Abdominal (See Note 10) | N | N | N | N | N | Note 1 | Notes 2, 4, 6, 7, 8, 9, 11, 12, 13 | |
| Intra-operative (See Note 6) | N | N | N | N | N | Note 1 | Notes 7, 9, 11 | |
| Intra-operative (Neuro.) | N | N | N | N | N | Note 1 | Notes 7, 9, 11 | |
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | N | N | N | N | N | Note 1 | Notes 2, 4, 6, 7, 8, 9, 11 | |
| Small Organ (See Note 5) | N | N | N | N | N | Note 1 | Note 2, 5, 6, 7, 8, 9, 11, 12 | |
| Neonatal Cephalic | N | N | N | N | N | Note 1 | Notes 7, 8, 9, 11 | |
| Adult Cephalic | N | N | N | N | N | Note 1 | Notes 7 | |
| Trans-rectal | N | N | N | N | N | Note 1 | Notes 2, 7, 8, 9, 12 | |
| Trans-vaginal | N | N | N | N | N | Note 1 | Notes 2, 7, 8, 9, 12 | |
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | N | N | N | N | N | Note 1 | Notes 2, 5, 6, 7, 8, 9, 11 | |
| Musculo-skel. (Superfic.) | N | N | N | N | N | Note 1 | Notes 2, 5, 6, 7, 8, 9, 11 | |
| Intra-luminal | ||||||||
| Other (See Note 13) | N | N | N | N | N | Note 1 | Notes 2, 7, 8, 9, 12 | |
| Cardiac Adult | N | N | N | N | N | Note 1 | Notes 4, 7 | |
| Cardiac | Cardiac Pediatric | N | N | N | N | N | Note 1 | Notes 4, 7 |
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | N | N | N | N | N | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 |
| Other (spec.) |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
{4}------------------------------------------------
510(k) No.:
Device Name: CA1-7A for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | ||
| (Track I only) | (Tracks I & III) | |||||||||
| Ophthalmic | Ophthalmic | |||||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Notes 2, 4, 7, 8, 9, 11 | |||
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11, 13 | |||
| Intra-operative (See Note 6) | ||||||||||
| Intra-operative (Neuro.) | ||||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||||
| Pediatric | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |||
| Small Organ (See Note 5) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Cardiac) | ||||||||||
| Musculo-skel. (Convent.) | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |||
| Musculo-skel. (Superfic.) | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |||
| Intra-luminal | ||||||||||
| Other (See Note 13) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac | Cardiac Pediatric | |||||||||
| Trans-esophageal (Cardiac) | ||||||||||
| Other (spec.) | ||||||||||
| PeripheralVessel | Peripheral vessel | |||||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+CW, B+C+PW, B+C+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{5}------------------------------------------------
510(k) No.:
Device Name: CA2-8A for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Notes 2, 4, 7, 8, 9, 11 | |
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Notes 2, 4, 7, 8, 9, 11 | |
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+CW, B+C+PW, B+C+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
{6}------------------------------------------------
510(k) No.:
Device Name: CF4-9 for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | |
| (Track I only) | (Tracks I & III) | ||||||||
| Ophthalmic | Ophthalmic | ||||||||
| Fetal/Obstetrics (See Note 3) | |||||||||
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Notes 8, 9, 11 | ||
| Intra-operative (See Note 6) | |||||||||
| Intra-operative (Neuro.) | |||||||||
| Fetal Imaging& Other | Laparoscopic | ||||||||
| Pediatric | P | P | P | P | P | Note 1 | Notes 8, 9, 11 | ||
| Small Organ (See Note 5) | |||||||||
| Neonatal Cephalic | P | P | P | P | P | Note 1 | Notes 8, 9, 11 | ||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Cardiac) | |||||||||
| Musculo-skel. (Convent.) | |||||||||
| Musculo-skel. (Superfic.) | |||||||||
| Intra-luminal | |||||||||
| Other (See Note 13) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac | Cardiac Pediatric | ||||||||
| Trans-esophageal (Cardiac) | |||||||||
| Other (spec.) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | Notes 8, 9, 11 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 4: Color M-mode
Note 8: 3D imaging
{7}------------------------------------------------
510(k) No.:
Device Name: E3-12A for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | ||
| Ophthalmic | Ophthalmic | |||||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Notes 7, 8, 9 | |||
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Notes 7, 8, 9, 12 | |||
| Intra-operative (See Note 6) | ||||||||||
| Intra-operative (Neuro.) | ||||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||||
| Pediatric | ||||||||||
| Small Organ (See Note 5) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Trans-rectal | P | P | P | P | Note 1 | Notes 7, 8, 9, 12 | ||||
| Trans-vaginal | P | P | P | P | Note 1 | Notes 7, 8, 9, 12 | ||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Cardiac) | ||||||||||
| Musculo-skel. (Convent.) | ||||||||||
| Musculo-skel. (Superfic.) | ||||||||||
| Intra-luminal | ||||||||||
| Other (See Note 13) | P | P | P | P | Note 1 | Notes 7, 8, 9, 12 | ||||
| Cardiac Adult | ||||||||||
| Cardiac | Cardiac Pediatric | |||||||||
| Trans-esophageal (Cardiac) | ||||||||||
| Other (spec.) | ||||||||||
| PeripheralVessel | Peripheral vessel | |||||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
{8}------------------------------------------------
510(k) No.:
Device Name: VR5-9 for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Note 2, 7, 8, 9 | |
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | |
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | P | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | |
| Trans-vaginal | P | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | |
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 12 | |
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+CW, B+C+PW, B+C+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
{9}------------------------------------------------
510(k) No.:
Device Name: L3-12A for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 9, 11 | ||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Notes 2, 4, 6, 7, 9, 11 | ||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | P | P | P | P | Note 1 | Notes 2, 4, 6, 7, 9, 11 | ||
| Small Organ (See Note 5) | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 9, 11, 12 | ||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 9, 11 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 9, 11 | ||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 9, 11 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+CW, B+C+PW, B+C+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{10}------------------------------------------------
510(k) No.:
Device Name: LA3-16A for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | |
| (Track I only) | (Tracks I & III) | ||||||||
| Ophthalmic | Ophthalmic | ||||||||
| Fetal (See Note 3) | |||||||||
| Abdominal (See Note 10) | |||||||||
| Intra-operative (See Note 6) | |||||||||
| Intra-operative (Neuro.) | |||||||||
| Fetal Imaging& Other | Laparoscopic | ||||||||
| Pediatric | |||||||||
| Small Organ (See Note 5) | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 9, 11, 12 | ||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Cardiac) | |||||||||
| Musculo-skel. (Convent.) | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 9, 11 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 9, 11 | ||
| Intra-luminal | |||||||||
| Other (See Note 13) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac | Cardiac Pediatric | ||||||||
| Trans-esophageal (Cardiac) | |||||||||
| Other (spec.) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 9, 11 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
Note 12: ElastoScan
{11}------------------------------------------------
510(k) No.:
Device Name: LA3-16AI for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal (See Note 3) | ||||||||
| Abdominal | ||||||||
| Intra-operative (See Note 6) | P | P | P | P | Note 1 | Note 7, 9, 11 b) | ||
| Intra-operative (Neuro.) | P | P | P | P | Note 1 | Note 7, 9, 11 b) | ||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | P | Note 1 | Note 7, 9, 11 a) | |
| Musculo-skel. (Superfic.) | P | P | P | P | P | Note 1 | Note 7, 9, 11 a) | |
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | Note 7, 9, 11 a) |
| Other (spec.) |
N= new indication; P= previously cleared by FDA K1516639 and K103397 9) (LS5-13); E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
{12}------------------------------------------------
510(k) No.:
Device Name: LA2-9A for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal/Obstetrics (See Note 3) | |||||||||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | |||
| Intra-operative (See Note 6) | |||||||||
| Intra-operative (Neuro.) | |||||||||
| Fetal Imaging& Other | Laparoscopic | ||||||||
| Pediatric | |||||||||
| Small Organ (See Note 5) | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | |||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Cardiac) | |||||||||
| Musculo-skel. (Convent.) | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | |||
| Musculo-skel. (Superfic.) | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | |||
| Intra-luminal | |||||||||
| Other (See Note 13) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac | Cardiac Pediatric | ||||||||
| Trans-esophageal (Cardiac) | |||||||||
| Other (spec.) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | ||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K151663; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
-
Note 2: Includes imaging for guidance of biopsy
-
Note 3: Includes infertility monitoring of follicle development
-
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
-
Note 6: Abdominal organs and peripheral vessel
-
Note 7: Tissue Harmonic Imaging (THI)
-
Note 8: 3D imaging
-
Note 9: MultiVision (Spatial Compound Imaging)
-
Note 10: Includes Renal, Gynecology/Pelvis
-
Note 11: Panoramic imaging
-
Note 12: ElastoScan
-
Note 13: Includes Urology/Prostate
Note 4: Color M-mode
{13}------------------------------------------------
510(k) No.:
Device Name: PE2-4 for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PWD | CWD | Color | Combined* | Other | ||
| (Track I only) | (Tracks I & III) | Doppler* | (Spec.) | (Spec.) | ||||||
| Ophthalmic | Ophthalmic | |||||||||
| Fetal/Obstetrics (See Note 3) | ||||||||||
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 7 | |||
| Intra-operative (See Note 6) | ||||||||||
| Intra-operative (Neuro.) | ||||||||||
| Fetal Imaging | Laparoscopic | |||||||||
| &Other | Pediatric | |||||||||
| Small Organ (See Note 5) | ||||||||||
| Neonatal Cephalic | P | P | P | P | P | Note 1 | Note 7 | |||
| Adult Cephalic | P | P | P | P | P | Note 1 | Note 7 | |||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Cardiac) | ||||||||||
| Musculo-skel. (Convent.) | ||||||||||
| Musculo-skel. (Superfic.) | ||||||||||
| Intra-luminal | ||||||||||
| Other (See Note 13) | ||||||||||
| Cardiac Adult | P | P | P | P | P | Note 1 | Note 4, 7 | |||
| Cardiac | Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 4, 7 | ||
| Trans-esophageal (Cardiac) | ||||||||||
| Other (spec.) | ||||||||||
| Peripheral | Peripheral vessel | |||||||||
| Vessel | Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
-
Note 2: Includes imaging for guidance of biopsy
-
Note 3: Includes infertility monitoring of follicle development
-
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
-
Note 6: Abdominal organs and peripheral vessel
-
Note 7: Tissue Harmonic Imaging (THI)
-
Note 8: 3D imaging
-
Note 9: MultiVision (Spatial Compound Imaging)
-
Note 10: Includes Renal, Gynecology/Pelvis
-
Note 11: Panoramic imaging
-
Note 12: ElastoScan
-
Note 13: Includes Urology/Prostate
Note 4: Color M-mode
{14}------------------------------------------------
510(k) No.:
Device Name: PA3-8B for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal Imaging& Other | Fetal/Obstetrics (See Note 3) | |||||||
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 7 | |
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | P | P | P | P | P | Note 1 | Note 7 | |
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | N | N | N | N | N | Note 1 | Note 7 | |
| Adult Cephalic | P | P | P | P | P | Note 1 | Note 7 | |
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac | Cardiac Adult | P | P | P | P | P | Note 1 | Note 4, 7 |
| Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 4, 7 | |
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K151663; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
-
Note 2: Includes imaging for guidance of biopsy
-
Note 3: Includes infertility monitoring of follicle development
-
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
-
Note 6: Abdominal organs and peripheral vessel
-
Note 7: Tissue Harmonic Imaging (THI)
-
Note 8: 3D imaging
-
Note 9: MultiVision (Spatial Compound Imaging)
-
Note 10: Includes Renal, Gynecology/Pelvis
-
Note 11: Panoramic imaging
-
Note 12: ElastoScan
-
Note 13: Includes Urology/Prostate
Note 4: Color M-mode
{15}------------------------------------------------
510(k) No.:
Device Name: CV1-8A for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | Note 1 | Note 2, 4, 7, 8, 9, 11 | ||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Note 2, 7, 8, 11 | ||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
{16}------------------------------------------------
510(k) No.:
Device Name: LV3-14A for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal/Obstetrics (See Note 3) | |||||||||
| Abdominal (See Note 10) | |||||||||
| Intra-operative (See Note 6) | |||||||||
| Intra-operative (Neuro.) | |||||||||
| Fetal Imaging& Other | Laparoscopic | ||||||||
| Pediatric | |||||||||
| Small Organ (See Note 5) | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | ||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Cardiac) | |||||||||
| Musculo-skel. (Convent.) | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | ||
| Intra-luminal | |||||||||
| Other (See Note 13) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac | Cardiac Pediatric | ||||||||
| Trans-esophageal (Cardiac) | |||||||||
| Other (spec.) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
{17}------------------------------------------------
510(k) No.:
Device Name: V5-9 for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | Note 1 | Note 2, 7, 8, 9 | ||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Note 2, 7, 8, 9 | ||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | P | P | P | P | Note 1 | Note 2, 7, 8, 9 | ||
| Trans-vaginal | P | P | P | P | Note 1 | Note 2, 7, 8, 9 | ||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9 | ||
| Cardiac | Cardiac Adult | |||||||
| Cardiac Pediatric | ||||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{18}------------------------------------------------
510(k) No.:
Device Name: DP2B for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal Imaging& Other | Fetal/Obstetrics (See Note 3) | |||||||
| Abdominal (See Note 12) | ||||||||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | P | |||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac | Cardiac Adult | P | ||||||
| Cardiac Pediatric | P | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | ||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K150478; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
{19}------------------------------------------------
510(k) No.:
Device Name: CW4.0 for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal Imaging& Other | Fetal/Obstetrics (See Note 3) | ||||||||
| Abdominal (See Note 12) | |||||||||
| Intra-operative (See Note 6) | |||||||||
| Intra-operative (Neuro.) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | P | ||||||||
| Small Organ (See Note 5) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | P | ||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Cardiac) | |||||||||
| Musculo-skel. (Convent.) | |||||||||
| Musculo-skel. (Superfic.) | |||||||||
| Intra-luminal | |||||||||
| Other (See Note 13) | |||||||||
| Cardiac | Cardiac Adult | P | |||||||
| Cardiac Pediatric | P | ||||||||
| Trans-esophageal (Cardiac) | |||||||||
| Other (spec.) | |||||||||
| PeripheralVessel | Peripheral vessel | P | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K150478; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
-
Note 2: Includes imaging for guidance of biopsy
-
Note 3: Includes infertility monitoring of follicle development
-
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
-
Note 6: Abdominal organs and peripheral vessel
-
Note 7: Tissue Harmonic Imaging (THI)
-
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
Note 12: ElastoScan
Note 4: Color M-mode
{20}------------------------------------------------
510(k) No.:
Device Name: CW6.0 for use with HS70A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal Imaging& Other | Fetal/Obstetrics ( See Note 3 ) | |||||||
| Abdominal ( See Note 12 ) | ||||||||
| Intra-operative ( See Note 6 ) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | P | |||||||
| Small Organ ( See Note 5 ) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | P | |||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other ( See Note 13 ) | ||||||||
| Cardiac | Cardiac Adult | P | ||||||
| Cardiac Pediatric | P | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | ||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K150478; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
{21}------------------------------------------------
510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
1. Submitter's Information: 21 CFR 807.92(a)(1)
SAMSUNG MEDISON CO., LTD. 42, Teheran-ro 108-gil, Gangnam-gu, Seoul, Korea
Contact Person:
Kyeong-Mi, Park Regulatory Affairs Manager
| Telephone: | 82.2.2194.1373 |
|---|---|
| Facsimile: | 82.2.556.3974 |
Data Prepared: August 21, 2015
2. Name of the device:
Common/Usual Name: Diagnostic Ultrasound System and Accessories Proprietary Name: HS70A Diagnostic Ultrasound System Classification Names: FR Number Product Code Ultrasonic Pulsed Doppler Imaging System 892.1550 IYN Ultrasound Pulsed Echo Imaging System 892.1560 IYO Diagnostic Ultrasound Transducer 892.1570 ITX
3. Identification of the predicate or legally marketed device:
- WS80A Diagnostic Ultrasound System (K143089) -
- RS80A Diagnostic Ultrasound System (K151663) -
- -A35 Diagnostic Ultrasound System (K150478)
- ACCUVIX XG Diagnostic Ultrasound System (K103397) -
{22}------------------------------------------------
4. Device Description:
The HS70A is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The HS70A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS70A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
5. Intended Uses:
The HS70A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
6. Technological Characteristics:
The HS70A is substantially equivalent with respect to safety, effectiveness, and functionality to the WS80A Diagnostic Ultrasound System (K143089), RS80A Diagnostic Ultrasound System (K151663), A35 Diagnostic Ultrasound System (K150478) and ACCUVIX XG Diagnostic Ultrasound System (K103397).
All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations.
These are described in detail in the technological characteristics comparison table as below.
{23}------------------------------------------------
| Feature /Characteristics | The subject deviceHS70A | The predicate device | |||
|---|---|---|---|---|---|
| WS80A(K143089) | RS80A(K151663) | A35(K150478) | ACCUVIX XG(K103397) | ||
| Indication for Use | |||||
| - Fetal/Obstetrics | √ | √ | √ | √ | √ |
| - Abdominal | √ | √ | √ | √ | √ |
| - Gynecology | √ | √ | √ | √ | √ |
| - Pediatric | √ | √ | √ | √ | √ |
| - Small Organ | √ | √ | √ | √ | √ |
| - Neonatal Cephalic | √ | √ | √ | √ | √ |
| - Adult Cephalic | √ | √ | √ | √ | √ |
| - Trans-rectal | √ | √ | √ | √ | √ |
| - Trans-vaginal | √ | √ | √ | √ | √ |
| - Musculo-skeletal(Conventional) | √ | √ | √ | √ | √ |
| - Musculo-skeletal(Superficial) | √ | √ | √ | √ | √ |
| - Urology | √ | √ | √ | √ | √ |
| - Cardiac Adult | √ | √ | √ | √ | √ |
| - Cardiac Pediatric | √ | √ | √ | √ | √ |
| - Peripheral vessel | √ | √ | √ | √ | √ |
| - Intra-operative | √ | √ | √ | √ | √ |
| Scanhead Types | |||||
| - Linear Array | √ | √ | √ | √ | √ |
| - Curved Linear Array | √ | √ | √ | √ | √ |
| - Endocavity | √ | √ | √ | √ | √ |
| - Phased Array | √ | √ | √ | √ | √ |
| - Static Probes | √ | √ | √ | √ | √ |
| Scanhead Frequency | |||||
| 1.0 ~ 20.0 MHz | √ | √ | √ | √ | √ |
| Modes of Operation | |||||
| - B-mode | √ | √ | √ | √ | √ |
| - M-mode | √ | √ | √ | √ | √ |
| - Pulsed wave (PW)Doppler | √ | √ | √ | √ | √ |
| - Continuous wave(CW) Doppler | √ | √ | √ | √ | √ |
| - Color Doppler | √ | √ | √ | √ | √ |
| - Power AmplitudeDoppler | √ | √ | √ | √ | √ |
| - Tissue HarmonicImaging | √ | √ | √ | √ | √ |
| - 3D/4D imagingmode | √ | √ | √ | √ | √ |
| - Combined modes | √ | √ | √ | √ | √ |
| Safety & EMCCompliance | |||||
| - IEC 60601-1 | √ | √ | √ | √ | √ |
| Feature /Characteristics | The subject device | The predicate device | |||
| HS70A | WS80A(K143089) | RS80A(K151663) | A35(K150478) | ACCUVIX XG(K103397) | |
| - CSA C22.2No.601.1 | √ | √ | √ | √ | |
| - IEC 60601-2-37 | √ | √ | √ | √ | √ |
| - IEC 60601-1-2 | √ | √ | √ | √ | √ |
| Acoustic Output DisplayStandard | |||||
| Track 3 | √ | √ | √ | √ | √ |
| Patient ContactMaterials | |||||
| Tested to ISO 10993-1 | √ | √ | √ | √ | √ |
| Functionality | |||||
| - Q Scan | √ | √ | √ | √ | √ |
| - ClearVision | √ | √ | √ | √1) | √1) |
| - MultiVision | √ | √ | √ | √2) | √2) |
| - Panoramic | √ | √ | √ | √ | |
| - Needle Mate | √ | √ | |||
| - 2D NT | √ | √ | |||
| - Auto IMT+(Auto IMT) | √ | √ | √ | √ | √ |
| - Strain+ | √ | √ | √ | ||
| - Stress Echo | √ | √ | √ | ||
| - Elastoscan | √ | √ | √ | √ | √ |
| - E-Thyroid | √ | √ | |||
| - E-Breast | √ | √ | |||
| - S-Detect | √ | √ | |||
| - ADVR | √ | √ | √ | √ | |
| - 3D Imaging(Volume DataAcquisition) | √ | √ | √ | √ | √ |
| - 3D Imagingpresentation | |||||
| 3D Cine/4D Cine | √ | √ | √ | √ | √ |
| - 3D RenderingMPR | √ | √ | √ | √ | √ |
| - 3D XIMSV/Oblique View | √ | √ | √ | √ | √ |
| - 3D MXIVolume Slice/Mirror View | √ | √ | √ | √ | √ |
| - Volume CT | √ | √ | √ | √ | √ |
| - 3D MagiCut | √ | √ | √ | √ | √ |
| - Volume Calculation(VOCAL, XI VOCAL) | √ | √ | √ | √ | √ |
| - XI STIC | √ | √ | √ | √ | √ |
| - Realistic Vue | √ | √ | √ | √ | |
| - 5D Follicle | √ | √ |
{24}------------------------------------------------
-
The old name of ClearVision is DMR+ or Dynamic MR Plus.
-
The old name of MultiVison is Spatial Compound Imaging.
{25}------------------------------------------------
7. A brief discussion of the bench and non-clinical tests conducted on the subject device
The device has been evaluated for acoustic output, biocompatibility effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform to applicable medical device safety standards.
The HS70A and its application comply with voluntary standards as below:
- UL 60601-1, Safety requirements for Medical Equipment
- CSA C22.2 No. 601.1, Safety requirements for Medical Equipment
- IEC60601-2-37, Diagnostic Ultrasound Safety Standards
- EN/IEC60601-1, Safety requirements for Medical Equipment
- EN/IEC60601-1-2, EMC requirements for Medical Equipment
- NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- ISO10993-1, Biocompatibility
- ISO14971, Application of risk management to medical devices
Summary of Clinical Tests:
Not applicable. The subject of this submission. HS70A, did not require clinical studies to support substantial equivalence.
8. Conclusion
Intended uses and other key features are consistent with traditional clinical practices and FDA guidelines. The design, development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The device is designed to conform to applicable medical device safety standards and compliance. Therefore, SAMSUNG MEDISON CO., LTD. considers the HS70A to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
END of 510(K) Summary