K Number
K152625

Validate with FDA (Live)

Date Cleared
2016-03-09

(176 days)

Product Code
Regulation Number
870.1210
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MMS Guide Catheter is intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents into the peripheral vasculature.

Device Description

The MMS Guide Catheter is a small diameter tubular device. It is inserted into the vasculature using standard access techniques and advanced over a guidewire to the physician desired location. The MMS Guide Catheter can be used as a pathway for other diagnostic or therapeutic devices to enter the vasculature. The distal end has a standard luer fitting.

AI/ML Overview

This appears to be a 510(k) summary for the MMS Guide Catheter, making a case for substantial equivalence to a predicate device (Marksman Catheter, K111490). The document describes non-clinical testing to demonstrate performance, not diagnostic performance of an AI/ML device. Therefore, a direct answer to the request regarding "acceptance criteria and reported device performance" in the context of an AI/ML device is not applicable for this document.

However, I can extract the general acceptance criteria and performance reported for the MMS Guide Catheter as presented in the document, treating "acceptance criteria" as the planned tests and "reported device performance" as the outcome of those tests.

Here's an analysis based on the provided text, adapted to the closest interpretations:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Test)Reported Device Performance
Biocompatibility:
CytotoxicityFound to be biocompatible
SensitizationFound to be biocompatible
Acute Systemic ToxicityFound to be biocompatible
Hemolysis ExtractFound to be biocompatible
Hemolysis Direct ContactFound to be biocompatible
PyrogenFound to be biocompatible
Complement ActivationFound to be biocompatible
Canine ThrombogenicityFound to be biocompatible
Animal Testing (Simulated Clinical Performance):Performed to specification
Cadaver Experience (Simulated Clinical Use):Performed to specification
Non-Clinical Testing:
Dimensional verificationMet all acceptance criteria
Surface ReviewMet all acceptance criteria
LeakMet all acceptance criteria
TrackingMet all acceptance criteria
FlowMet all acceptance criteria
Flex/KinkMet all acceptance criteria
TorqueMet all acceptance criteria
TensileMet all acceptance criteria
PackagingMet all acceptance criteria
Overall Performance Data:All necessary verification and validation testing has been performed to assure substantial equivalence to the predicate devices.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

This document does not specify sample sizes for mechanical or biocompatibility tests. It briefly mentions "Animal testing" and "Cadaver testing" but provides no details on the number of animals or cadavers used, let alone their provenance or whether the studies were retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This device is a guide catheter, not a diagnostic AI/ML system requiring expert ground truth for its performance evaluation (e.g., image interpretation). The "ground truth" for this device's performance would be engineering specifications and biological reactions, evaluated through laboratory and preclinical tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not a study involving human interpretation or adjudication. Performance is assessed against predetermined engineering and biological specifications.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device, nor does it involve human readers in a diagnostic context. This is a medical device clearance submission for a guide catheter.

6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

Not applicable. This is a physical medical device (catheter), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the MMS Guide Catheter, the "ground truth" for demonstrating its performance relies on:

  • Engineering Specifications: Defined physical and mechanical properties (e.g., dimensions, tracking, flex/kink, torque, tensile strength).
  • Biocompatibility Standards (ISO 10993): Established tests for biological safety.
  • Simulated Clinical Performance: Performance in animal models and cadaveric vasculature against expected functionality.
  • Predicate Device Comparison: Establishing substantial equivalence by comparing technological characteristics and intended use to an already cleared device.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI/ML system that requires a "training set."

9. How the ground truth for the training set was established

Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a human figure, with three faces in profile. The figure is stylized and appears to be in motion.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 9, 2016

Micro Medical Solutions, Inc. Mr. Gregory Mathison Vice President, Regulatory, Clinical & Quality 790 Willard Street, #209 Quincy, MA 02169

Re: K152625 Trade/Device Name: MMS Guide Catheter Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: January 29, 2016 Received: February 3. 2016

Dear Mr. Mathison,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K152625

Device Name MMS Guide Catheter

Indications for Use (Describe)

The MMS Guide Catheter is intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents into the peripheral vasculature.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Micro Medical Solutions. The logo has a red DNA-like strand above the company name. The company name is in gray, with "micro medical" in a larger font than "solutions."

510(k) Summary

This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

General Information

Applicant:Micro Medical Solutions, Inc.240 Andover StreetWilmington, MA 01887Tel: 978.909.3045
Trade Name:MMS Guide Catheter
Common Name:Guide Catheter
Classification Name:Catheter, continuous flush
21CFR Number:870.1210
Device Classification:Class II
Product Code:KRA
Predicate Devices:Marksman Catheter (K111490)
Contact:Greg MathisonVP Regulatory, Clinical & Quality
DATE:March 1, 2016

Substantially Equivalent to:

The MMS Guide Catheter is equivalent in intended use, principal of operation and technological characteristics to the Marksman Catheter (K111490).

Description of the device subject to premarket notification

The MMS Guide Catheter is a small diameter tubular device. It is inserted into the vasculature using standard access techniques and advanced over a guidewire to the physician desired location. The MMS Guide Catheter can be used as a pathway for other

{5}------------------------------------------------

diagnostic or therapeutic devices to enter the vasculature. The distal end has a standard luer fitting.

Indications for Use

The MMS Guide Catheter is intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents into the peripheral vasculature.

Materials

All materials used in the manufacture of the MMS Guide Catheter are suitable for this use and have been used in numerous previously cleared products. The MMS Guide Catheter materials were tested per ISO10993 and found to be biocompatible. Testing included the following:

  • . Cytotoxicity
  • Sensitization
  • Acute Systemic Toxicity ●
  • Hemolysis Extract ●
  • Hemolysis Direct Contact
  • Pyrogen
  • Complement Activation ●
  • Canine Thrombogenicity ●

Animal Testing

Animal testing was conducted to assess the simulated clinical performance of the Guide Catheter. The product performed to specification.

Cadaver Experience

Cadaver testing was conducted using the vasculature below the knee to assess the simulated clinical use of the product. The product performed to specification.

Non-Clinical Testing

Product testing was completed and met all of the acceptance criteria. Testing was conducted on sterile final product. Testing was performed on baseline (non-aged) and aged products. Testing included:

  • Dimensional verification
  • Surface Review ●
  • Leak
  • Tracking ●
  • Flow ●
  • Flex/Kink ●
  • Torque ●
  • Tensile ●
  • Packaging ●

{6}------------------------------------------------

Performance Data

All necessary verification and validation testing has been performed for the MMS Guide Catheter to assure substantial equivalence to the predicate devices.

Basis for Determination of Substantial Equivalence

Upon reviewing the safety information provided in this submission and comparing intended use, principle of operation and overall technological characteristics, the MMS Guide Catheter is determined to be substantially equivalent to existing legally marketed devices.

Trade nameMMS Guide CatheterMarksman CatheterSE Discussion
Product codeKRAKRASame product code
510(k) numberK152625K111490
21CFR870.1210870.1210Same CFR number
DeviceClassificationIIIISame - Class II
DevicedescriptionThe MMS Guide Catheter is asmall diameter tubular device.It is inserted into thevasculature using standardaccess techniques andadvanced over a guidewire tothe physician desired location.The MMS Guide Catheter canbe used as a pathway for otherdiagnostic or therapeuticdevices to enter thevasculature. The distal endhas a standard luer fitting.The Marksman Catheter is avariable stiffness, singlelumen catheter designed toaccess small, tortuous vascularareas. The outer surface of thecatheter's distal segment iscoated with a hydrophilicmaterial to provide lubricityduring use. The catheter alsoincorporates a PTFE liner tofacilitate movement ofintroduction devices passedthrough its lumen. TheMarksman Catheter has aradiopaque marker at thedistal tip to facilitatefluoroscopic visualization.The distal tip of the catheter isshapeable.The Marksman Catheter isprovided with various workinglengths. The MarksmanCatheter is for single use only.The device description is thesame.

Comparison of Product Features

{7}------------------------------------------------

Intended UseThe MMS Guide Catheter isintended for the introductionof interventional devices andinfusion of diagnostic ortherapeutic agents into theperipheral vasculature.The Marksman Catheter isindicated for the introductionof interventional devices andinfusion of diagnostic ortherapeutic agents into theneuro, peripheral and coronaryvasculature.Same Indications for Use
Length40 cm working length105cmEquivalent
Diameter3.2 Fr0.027"Equivalent
Method ofvisualizationFluoro / UltrasoundFluoroEquivalent
SterilizationYes - ETOYes - ETOBoth devices are suppliedsterile.
Single useYYSame
Shelf life22 months after productionNot indicated on submissionThis will be on the productlabel
PackagingTyvek / Poly pouchUnknownTyvek / Poly heat sealedpouches are commonpackaging for sterile products
MaterialsBiocompatibleBiocompatibleBiocompatible

Conclusion

The products are substantially equivalent as the indications for use are the same, the clinical application is the same, the materials are equivalent, the dimensions are equivalent and the tested product performance attributes are equivalent.

§ 870.1210 Continuous flush catheter.

(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).