K Number
K152621

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-10-14

(30 days)

Product Code
Regulation Number
888.3350
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

OSS INDICATIONS

  1. Painful and disabled joint resulting from avascular necrosis, osteoarthritis, or traumatic arthritis.
  2. Correction of varus, valgus, or posttraumatic deformity.
  3. Correction or revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
  4. Ligament deficiencies.
  5. Tumor resections.
  6. Treatment of non-unions, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. *
  7. Revision of previously failed total joint arthroplasty.
  8. Trauma.

These devices are to be used with bone cement unless composed of OsseoTi (titanium alloy, not licensed in Canada) or a proximal femur is indicated for use (USA).

Biomet OSS Reduced size (RS) components offers a variety of component options for treatment in small adults and adolescents (12-21 years) that require proximal femoral, total femur, or proximal tibial replacement as well as, resurfacing components for the proximal tibia and distal femur (USA).

*Not applicable to Regenerex Ultra Porous Construct titanium knee augment usage (not licensed in Canada), or any other knee component.

COMPRESS INDICATIONS
The Compress Segmental Femoral Replacement System is indicated for:

    1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
    1. Tumor resections.
    1. Revision of previously failed total joint arthroplasty.
    1. Trauma.

The Compress Segmental Femoral Replacement System components are intended for uncemented use. When components of the Orthopaedic Salvage System are used with Biomet's Compress Segmental Femoral Replacement System, the user should refer to the package insert contained with the Compress components for full prescription information.

EXPANDABLE INDICATIONS
The Biomet Side Access Distal Femoral Expandable offers a treatment requiring distal femoral replacement who have not yet achieved full skeletal maturity (open epiphysis) or patients who require surgery who have significant residual leg length discrepancy. Indication for use of this device is most commonly turnor resection but could also involve osteoarthritis; theumatoid arthritis; correction of deformity; and correction of unsuccessful osteotomy, arthrodesis or previous joint replacement.

The devices are single use implants intended for implantation with Biomet Compress.

Device Description

The OSS/Arcos IM Total Femur Rod is, in simplest terms, a metal rod that is inserted into the IM canal of a femur and connects a proximal femoral component to a distal femoral component. The aforementioned femoral components are secured to the rod by both a Morse taper and a screw; the proximal end features an "Arcos" taper while the distal end is an "OSS" taper. The Arcos taper is the same as that utilized by the devices in K090757 while the OSS taper is identical to that utilized by the predicate device.

The rods are slightly bowed to mimic the natural femur and find appropriate alignment. All rods have a diameter of 14 mm and are available in 1 cm length increments from 20 cm to 30 cm.

The device is compatible with Biomet's Arcos proximal bodies and OSS distal femoral components.

AI/ML Overview

Here's a breakdown of the requested information based on the provided document, noting that this document is a 510(k) summary for a medical device and not a study report on device performance with acceptance criteria in the typical sense of a diagnostic or AI algorithm:

This document is a 510(k) Premarket Notification for a medical device (OSS/Arcos IM Total Femur Rod), not a study demonstrating the performance of a diagnostic or AI algorithm against specific acceptance criteria. Therefore, most of the requested fields are not directly applicable in the terms usually associated with AI/diagnostic studies.

The "acceptance criteria" for a 510(k) submission are primarily demonstrating substantial equivalence to a legally marketed predicate device, meaning it is as safe and effective as the predicate and does not raise new questions of safety or effectiveness. The "study" here is a mechanical testing comparison, not a clinical trial.


1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (for Substantial Equivalence to Predicate)Reported Device Performance (Mechanical Testing)
Safety: Device is as safe as the predicate.Mechanical testing (previously supplied in K090757) on worst-case Arcos constructs deemed sufficient to show safety.
Effectiveness: Device is as effective as the predicate.Mechanical testing (previously supplied in K090757) on worst-case Arcos constructs deemed sufficient to show effectiveness.
No New Questions of Safety/Effectiveness: Design changes do not alter fundamental scientific technology.The proximal taper geometry was modified, and the proximal screw packaged with the device was changed. All other technological characteristics remained unchanged. The submitter asserts these changes do not present a new worst-case scenario.
Intended Use/Indications are Substantially Equivalent:The proposed device has the same intended use and indications for use as the predicate (K141331, K14050).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document refers to "mechanical testing" and "worst-case Arcos constructs" from a previous submission (K090757), but details on sample size for these mechanical tests are omitted. "Data provenance" such as country of origin or retrospective/prospective is not relevant to mechanical engineering tests of this nature.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable. This submission is for a physical orthopedic implant, not a diagnostic device requiring expert interpretation of images or clinical data for ground truth establishment. Mechanical testing relies on engineering standards and measurements, not expert consensus on qualitative data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable. Adjudication methods are used to resolve discrepancies in expert interpretation, which is not relevant for mechanical testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. This is not an AI or diagnostic device that involves human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for the assessment of this device is based on mechanical engineering standards and physical measurements obtained from testing "worst-case Arcos constructs" as performed in a previous submission (K090757). It's not based on biological or clinical "ground truth" like pathology or outcomes data in the usual sense for diagnostic devices.

8. The sample size for the training set

This is not applicable. This device is not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

This is not applicable. This device is not an AI/ML algorithm that requires a training set or its associated ground truth.


Summary of the "Study" (Mechanical Testing discussed):

The "study" cited in the document to support substantial equivalence is mechanical testing that was previously performed and supplied in a prior 510(k) submission (K090757). This testing was conducted on "worst-case Arcos constructs" to evaluate the performance of the Arcos taper and proximal screw. The current submission argues that since the OSS/Arcos IM Total Femur Rod does not introduce a new worst-case scenario compared to the previously tested Arcos constructs, this existing mechanical testing is sufficient to demonstrate the safety and effectiveness of the new device's modifications. No new mechanical testing data is presented in this specific 510(k) document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, possibly representing health and well-being. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 14, 2015

Biomet Incorporated Mr. Bryan McMahon Regulatory Affair Specialist 56 East Bell Drive Warsaw, Indiana 46581

Re: K152621

Trade/Device Name: OSS/Acros IM Total Femur Rod Regulation Number: 21 CFR 888.3350 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDI, KRO, JWH Dated: September 10, 2015 Received: September 14, 2015

Dear Mr. McMahon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K152621

Device Name OSS/Arcos IM Total Femur Rod

Indications for Use (Describe) OSS INDICATIONS

  1. Painful and disabled joint resulting from avascular necrosis, osteoarthritis, or traumatic arthritis.

  2. Correction of varus, valgus, or posttraumatic deformity.

  3. Correction or revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.

  4. Ligament deficiencies.

  5. Tumor resections.

  6. Treatment of non-unions, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. *

  7. Revision of previously failed total joint arthroplasty.

  8. Trauma.

These devices are to be used with bone cement unless composed of OsseoTi (titanium alloy, not licensed in Canada) or a proximal femur is indicated for use (USA).

Biomet OSS Reduced size (RS) components offers a variety of component options for treatment in small adults and adolescents (12-21 years) that require proximal femoral, total femur, or proximal tibial replacement as well as, resurfacing components for the proximal tibia and distal femur (USA).

*Not applicable to Regenerex Ultra Porous Construct titanium knee augment usage (not licensed in Canada), or any other knee component.

COMPRESS INDICATIONS

The Compress Segmental Femoral Replacement System is indicated for:

    1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
    1. Tumor resections.
    1. Revision of previously failed total joint arthroplasty.
    1. Trauma.

The Compress Segmental Femoral Replacement System components are intended for uncemented use. When components of the Orthopaedic Salvage System are used with Biomet's Compress Segmental Femoral Replacement System, the user should refer to the package insert contained with the Compress components for full prescription information.

EXPANDABLE INDICATIONS

The Biomet Side Access Distal Femoral Expandable offers a treatment requiring distal femoral replacement who have not yet achieved full skeletal maturity (open epiphysis) or patients who require surgery who have significant residual leg length discrepancy. Indication for use of this device is most commonly turnor resection but could also involve osteoarthritis; theumatoid arthritis; correction of deformity; and correction of unsuccessful osteotomy, arthrodesis or previous joint replacement.

The devices are single use implants intended for implantation with Biomet Compress.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The color of the text is a dark gray, and the background is white. The overall impression is clean and modern.

510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the OSS/Arcos IM Total Femur Rod 510(k) premarket notification.

Sponsor:Biomet Inc.56 East Bell DrivePO Box 587Warsaw, IN 46581Establishment Registration Number: 1825034
Contact:Bryan M. McMahonRegulatory Affairs Specialist
Date:October 12, 2015
Subject Device:Trade Name: OSS/Arcos IM Total Femur RodCommon Name: Oncology, Salvage Hip/KneeClassification Name:• JDI- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented(21 CFR 888.3350)• KRO- Prosthesis, Knee, Femorotibial, Constrained, Cemented,Metal/Polymer (21 CFR 888.3510)• IWH- Prosthesis, Knee, Patellofemorotibial, Semi-Constrained

Legally marketed devices to which substantial equivalence is claimed:

Cemented, Polymer/Metal/Polymer (21 CFR 888.3560)

Device Description

The OSS/Arcos IM Total Femur Rod is, in simplest terms, a metal rod that is inserted into the IM canal of a femur and connects a proximal femoral component to a distal femoral component. The aforementioned femoral components are secured to the rod by both a Morse taper and a screw; the proximal end features an "Arcos" taper while the distal end is an "OSS" taper. The Arcos taper is the same as that utilized by the devices in K090757 while the OSS taper is identical to that utilized by the predicate device.

The rods are slightly bowed to mimic the natural femur and find appropriate alignment. All rods have a diameter of 14 mm and are available in 1 cm length increments from 20 cm to 30 cm.

Image /page/3/Picture/12 description: In the image, a bone is shown with metal attachments on both ends. The bone is long and slender, with a slightly curved shape. The metal attachments are silver and appear to be connected to the bone with screws or bolts. The bone and metal attachments are likely part of a medical device or implant, possibly for orthopedic surgery. The image is well-lit and the details of the bone and metal attachments are clearly visible.

Figure 7.1: An example OSS/Arcos IM Total Femur Rod construct inside the IM canal of a femur.

{4}------------------------------------------------

The device is compatible with Biomet's Arcos proximal bodies and OSS distal femoral components.

Intended Use and Indications for Use

OSS INDICATIONS

    1. Painful and disabled joint resulting from avascular necrosis, osteoarthritis, rheumatoid arthritis, or traumatic arthritis.
  • Correction of varus, valgus, or posttraumatic deformity. 2.
    1. Correction or revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
  • Ligament deficiencies. 4.
    1. Tumor resections.
  • Treatment of non-unions, femoral neck fracture, and trochanteric fractures of the 6. proximal femur with head involvement, unmanageable using other techniques. *
    1. Revision of previously failed total joint arthroplasty.
    1. Trauma.

These devices are to be used with bone cement unless composed of OsseoTi (titanium alloy, not licensed in Canada) or a proximal femur is indicated for use (USA).

Biomet OSS Reduced size (RS) components offers a variety of component options for treatment in small adults and adolescents (12-21 years) that require proximal femoral, total femur, or proximal tibial replacement as well as, resurfacing components for the proximal tibia and distal femur (USA).

*Not applicable to Regenerex Ultra Porous Construct titanium knee augment usage (not licensed in Canada), or any other knee component.

COMPRESS INDICATIONS

The Compress Segmental Femoral Replacement System is indicated for:

    1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
    1. Tumor resections.
  • Revision of previously failed total joint arthroplasty. 3.
    1. Trauma.

The Compress Segmental Femoral Replacement System components are intended for uncemented use.

When components of the Orthopaedic Salvage System are used with Biomet's Compress Segmental Femoral Replacement System, the user should refer to the package insert contained with the Compress components for full prescription information.

EXPANDABLE INDICATIONS

The Biomet Side Access Distal Femoral Expandable offers a treatment option for patients requiring distal femoral replacement who have not yet achieved full skeletal maturity (open

{5}------------------------------------------------

epiphysis) or patients who require surgery who have significant residual leg length discrepancy. Indication for use of this device is most commonly tumor resection but could also involve osteoarthritis; rheumatoid arthritis; correction of deformity; and correction or revision of unsuccessful osteotomy, arthrodesis or previous joint replacement.

The devices are single use implants intended for implantation with bone cement or with Biomet Compress.

Summary of Technological Characteristics

The new devices are a modification to the predicate device (K033871, K123501) in the following ways:

  • The proximal taper geometry is being modified
  • The proximal screw packaged with the device is changing

All other technological characteristics remain unchanged.

Summary of Performance Data

Mechanical testing in support of the Arcos taper and proximal screw was previously supplied in K090757. These tests were performed on the worst-case Arcos constructs; as the OSS/Arcos IM Total Femur Rod does not present a new worst-case, this testing is sufficient show the subject device is as safe and effective as the predicate.

Substantial Equivalence Conclusion

The proposed and predicate IM Total Femur (K033871, K123501) devices have the same intended use and indications for use (K141331, K14050) and Taper Connection References (K912712, K090757). The proposed design changes do not alter the fundamental scientific technology shared by both the proposed and predicate IM Total Femur devices. The information provided in this submission demonstrates that the proposed device is as safe and effective as the predicate and does not raise different questions of safety and effectiveness than the predicate. Therefore, the proposed OSS IM Total Femur Rod is substantially equivalent to the cited predicate device(s).

§ 888.3350 Hip joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.