(89 days)
The Biomet Side Access Distal Femoral Expandable offers a treatment option for patients requiring distal femoral replacement who have not yet achieved full skeletal maturity (open epiphysis) or patients who require surgery who have significant residual leg length discrepancy. Indication for use of this device is most commonly tumor resection but could also involve osteoarthritis; rheumatoid arthritis; correction of deformity; and correction or revision of unsuccessful osteotomy, arthrodesis or previous joint replacement.
The devices are single use implants intended for implantation with bone cement or with Biomet Compress®.
The Biomet Side Access Distal Femoral Expandable Knee is an additional component to Biomet's Orthopedic Salvage System (OSS) that gives the surgeon the ability to expand the prosthesis as the patient grows. All expansion takes place where natural bone has been removed. The device does not lengthen existing bones. The expandable device is available in standard and Reduced Size (RS) and is compatible with Biomet OSS system components and Compress (CPS) devices.
The provided text describes the Biomet Side Access Distal Femoral Expandable knee implant and a summary of its non-clinical performance data. Here's an analysis of the available information regarding acceptance criteria and the study proving it:
Summary of Acceptance Criteria and Reported Device Performance:
The document states that cyclic fatigue testing was conducted. However, specific quantitative acceptance criteria for this test (e.g., number of cycles, load levels, maximum allowable deformation or failure) are not detailed in this summary.
The reported device performance is a general statement:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Adequate accuracy and performance for intended use | "Testing verified that the accuracy and performance of the system is adequate to perform as intended" |
| Substantial equivalence to predicate device | "and to demonstrate substantial equivalence to the predicate device." |
| Performance within intended use (based on mechanical testing) | "The results of mechanical testing indicated the devices performed within the intended use" |
| No new safety and efficacy issues | "did not raise any new safety and efficacy issues" |
| Substantial equivalence to predicate devices (based on mechanical testing) | "and were found to be substantially equivalent to the predicate devices." |
Detailed Information as Requested:
-
Table of acceptance criteria and the reported device performance:
(As provided above) It's important to note the lack of specific, quantitative acceptance criteria in the summary. The performance is stated in qualitative terms related to adequacy, intended use, and equivalence. -
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size: The document does not specify the sample size used for the cyclic fatigue testing or any other mechanical tests.
- Data Provenance: The document does not provide information on the country of origin of the data or whether the tests were retrospective or prospective. Given it's non-clinical, the concept of "prospective" or "retrospective" as it applies to human clinical trials is less relevant; these would be controlled laboratory studies.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not applicable to this submission. The tests described are "non-clinical" mechanical performance tests (cyclic fatigue). Ground truth, in the context of expert consensus, is typically established for diagnostic or clinical performance studies involving human readers or adjudicators, not for mechanical device testing. -
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable. Adjudication methods are relevant for clinical studies where subjective assessments (e.g., image interpretation, clinical outcomes) require expert review to establish a ground truth. Mechanical testing relies on objective measurements and established engineering standards, not adjudication. -
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. This submission describes a mechanical knee implant and its non-clinical performance testing. It does not involve AI, image analysis, human readers, or diagnostic performance evaluation. Therefore, an MRMC comparative effectiveness study is entirely outside the scope of this document. -
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable. As mentioned above, this device is a mechanical implant, not an algorithm or an AI-powered system. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for mechanical performance tests like cyclic fatigue is typically established by engineering standards and specifications. The device is tested against predefined load conditions, cycles, and failure criteria (e.g., no fracture, no significant deformation, maintaining functional integrity) as outlined in relevant ASTM or ISO standards for orthopedic implants. The document implies compliance with these standards through the statement "Testing verified that the accuracy and performance of the system is adequate to perform as intended." -
The sample size for the training set:
This information is not applicable. There is no "training set" for a mechanical device as it does not involve machine learning or AI. -
How the ground truth for the training set was established:
This information is not applicable. Since there is no training set, there's no ground truth to establish for it.
In conclusion, the provided document focuses solely on the non-clinical mechanical performance of a knee implant. It confirms that cyclic fatigue testing was performed, and the results indicated the device met its intended use and demonstrated substantial equivalence to predicate devices. However, it lacks specific quantitative acceptance criteria, sample sizes for testing, and details about the testing methodology beyond stating "cyclic fatigue was conducted" and "mechanical testing." Information related to clinical studies, AI, human readers, or expert ground truth establishment is not relevant to this type of non-clinical device submission.
{0}------------------------------------------------
. -
Image /page/0/Picture/2 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and connected, giving the word a solid, block-like appearance. A registered trademark symbol is present to the upper right of the letter "T".
510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92 .
| Submitter Information | |
|---|---|
| Name | Biomet Manufacturing LLC |
| Address | 56 East Bell DriveWarsaw, IN 46581-0857 |
| Phone number | (574) 372-1761 |
| Fax number | (574) 372-1683 |
| EstablishmentRegistration Number | 1825034 |
| Name of contact person | Tracy Bickel Johnson, RAC |
| Date prepared | February 2014 |
| Name of device | |
| Trade or proprietaryname | Biomet Side Access Distal Femoral Expandable |
| Common or usualname | Constrained knee implant |
| Classification name /Regulation | Prosthesis, knee, femorotibial, constrained, cemented,metal/polymer (888.3510) |
| Classification panel | Orthopedic |
| Product Code(s) | KRO |
| Name of accessories | |
| Trade or proprietaryname | Biomet Side Access Expandable Revolution CounterBiomet Side Access Expandable 5mm Straight Hex and ModularBall Hex |
| Common or usualname | Instrument Specific Instrument- Surgical, Orthopedic AccessoryScrewdriver |
| Classification name /Regulation | Implant specific Instrument- Prosthesis, knee, femorotibial,constrained, cemented, metal/polymer (888.3510)Screwdriver (888.4540) |
| Classification panel | Orthopedic |
| Product Code(s) | KROHXX |
| Legally marketed device(s)to which equivalence isclaimed | Biomet Expandable Knee (Biomet Manufacturing), K020381 |
| Reason for 510(k) submission | New device |
| Device description | The Biomet Side Access Distal Femoral Expandable Knee is an additional component to Biomet's Orthopedic Salvage System (OSS) that gives the surgeon the ability to expand the prosthesis as the patient grows. All expansion takes place where natural bone has been removed. The device does not lengthen existing bones. The expandable device is available in standard and Reduced Size (RS) and is compatible with Biomet OSS system components and Compress (CPS) devices. |
| Intended use of the device | Knee replacement |
| Indications for use | The Biomet Side Access Distal Femoral Expandable offers a treatment option for patients requiring distal femoral replacement who have not yet achieved full skeletal maturity (open epiphysis) or patients who require surgery who have significant residual leg length discrepancy. Indication for use of this device is most commonly tumor resection but could also involve osteoarthritis; rheumatoid arthritis; correction of deformity; and correction or revision of unsuccessful osteotomy, arthrodesis or previous joint replacement.The devices are single use implants intended for implantation with bone cement or with Biomet Compress®. |
| Summary of the Technologies | The expandable device is available in standard and Reduced Size (RS) and is compatible with all Biomet OSS system components and Compress (CPS) devices. The Biomet Side Access Distal Femoral Expandable knee functions by means of an inner telescoping tube, also called the intercalary segment, which is moved by an adjusting screw that is coaxial with the tube. The expansion is performed in a manner simpler for the patient and the surgeon. |
| PERFORMANCE DATA | |
| SUMMARY OF NON-CLINICAL TESTS | |
| Performance Test Summary-New Device | |
| Cyclic fatigue was conducted. | |
| Testing verified that the accuracy and performance of the system is adequate to perform as intended and to demonstrate substantial equivalence to the predicate device. | |
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION | |
| Clinical Performance Data/Information: N/A | |
| CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA | |
| No clinical testing was necessary for a determination of substantial equivalence. |
Mailing Address: P.O. Box 587 Warsaw, IN 46581-0587 Toll Free: 800.348.9500 Office: 574.267.6639 Main Fax: 574.267.8137 www.biomet.com
Shipping Address: 56 East Bell Drive Warsaw, IN 46582
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and connected to each other. There is a trademark symbol to the right of the letter "T".
.
The results of mechanical testing indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the snake and staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 28, 2014
Biomet Manufacturing LLC Ms. Tracy Bickel Johnson Regulatory Global Project Manager 56 East Bell Drive Warsaw. Indiana 46581-0857
Re: K140509
Trade/Device Name: Biomet Side Access Distal Femoral Expandable Regulation Number: 21 CFR 888.3510 Regulation Name: Knee Joint Femorotibial Metal/Polymer Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KRO Dated: February 25, 2014 Received: February 28, 2014
Dear Ms. Bickel Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act than I Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{3}------------------------------------------------
Page 2 - Ms. Tracy Bickel Johnson
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcestorYou/Industry/default.htm.
Sincerely yours,
Lori A. Wigqins
- for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
K140509 510(k) Number:
Device Name: Biomet Side Access Distal Femoral Expandable
Indications For Use:
The Biomet Side Access Distal Femoral Expandable offers a treatment option for patients requiring distal femoral replacement who have not yet achieved full skeletal maturity (open epiphysis) or patients who require surgery who have significant residual leg length discrepancy. Indication for use of this device is most commonly tumor resection but could also involve osteoarthritis; rheumatoid arthritis; correction of deformity; and correction or revision of unsuccessful osteotomy, arthrodesis or previous joint replacement.
The devices are single use implants intended for implantation with bone cement or with Biomet Compress®.
× Prescription Use_ (Per 21 CFR 801 Subpart D)
.
AND/OR Over-the-Counter_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
Division of Orthopedic Devices
Page 1 of 1
Riomet Side Access Distal Femoral Expanduble Traditional 510(k) ነገስ እንደተለ
Biomet Manufacturing, LLC
N/A