K Number
K152555

Validate with FDA (Live)

Date Cleared
2016-02-11

(156 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Aria Bio's Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The Dental Implant System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Device Description

The Aria Bio Dental Implant System consists of two stage endosseous form dental implants, hexagonal and conical implants and hexagonal and conical abutments; cover screws and healing caps; abutment systems and superstructures and surgical instruments.

AI/ML Overview

The provided document describes a 510(k) premarket notification for Aria Bio's Dental Implant System, asserting its substantial equivalence to predicate devices. The document does not describe acceptance criteria for particular performance metrics nor does it present detailed study results proving a device meets specific acceptance criteria in the format requested.

Instead, the document focuses on demonstrating substantial equivalence through:

  • Comparison of indications for use and technological characteristics with predicate devices.
  • Reporting that a series of performance tests (fatigue, surface analysis, biocompatibility, sterilization, shelf life) were performed and that "All results are supporting Aria Bio's labeling claims in order to establish substantial equivalency." However, it does not provide the specific acceptance criteria for these tests, nor the detailed results.

Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance, or details regarding sample sizes for training/test sets, ground truth establishment, or specific study types like MRMC, as this information is not present in the provided text.

The document states:

  • "A series of performance tests and evaluations were performed to demonstrate that Aria Bio's Dental Implant System does not raise any new issues that may affect its substantial equivalency with its predicate devices. These evaluations include fatigue (ISO 14801), surface analysis by SEM/EDS and XPS, biocompatibility in accordance with ISO 10993-1, sterilization validation (ISO 11137-2, AAMI TIR 33, ISO 17665-1) and shelf life validation (ISO 11607-1) in order to assure maintaining of SAL 10° along the shelf life."
  • "All results are supporting Aria Bio's labeling claims in order to establish substantial equivalency."

This indicates that these tests were conducted, and their outcomes were deemed acceptable based on established industry standards (e.g., ISO 14801 for fatigue testing, ISO 10993-1 for biocompatibility). However, the specific quantitative acceptance criteria or the actual performance results (e.g., specific fatigue limits, exact sterilization validation parameters, detailed surface analysis findings) are not provided in the document beyond stating that they were performed and supported substantial equivalence.

Without these specific details, I cannot construct the requested table or answer most of the specific questions about the study design and results.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 11, 2016

AriaBio International, Inc. % Tali Hazan Regulatory Consultant Talmed, Ltd. Ramot Naftali, Upper Galilee Ramot Naftali, 13830 ISRAEL

Re: K152555

Trade/Device Name: Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: January 10, 2016 Received: January 13, 2016

Dear Tali Hazan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang -

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K152555

Device Name Dental Implant System

Aria Bio's Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The Dental Implant System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and complete

and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" in blue, with a green line underneath it. Below the line is the word "BIO" in blue. Underneath the word "BIO" are the words "implant science" in gray.

510(k) Summary for Aria Bio's Dental Implant and Abutments

1. Date Prepared: 07 February 2016

2. 510(k) Owner Name:

Aria Bio International, Inc. 2999 Overland Avenue, Suite 118, Los Angeles, California 90064 USA Phone: 855-373-1614; Fax 310-730-6241; Email: info@ariabio.com

Contact person:

Ms. Tali Hazan - RA Consultant Phone: +972-(0)50-5292-304 Fax: +972-(0)722448981 Email: tali.hazan@talmed.co.il

3. Device Name and Classification:

Common/Usual Name: Dental Implant System Proprietary/Trade name: Dental Implant System (Implants model names: Fixation, Versatile, Immediate, Solution2, Solution5) Classification: Aria Bio Dental Implants and Abutments have been classified as Class II devices under the following classification names:

Regulation DescriptionProduct Code21 CFR Ref.Panel
Endosseous dental implantPrimary: DZESecondary: NHA872.3640DAGRID,Dental

Aria Bio International, Inc., 2999 Overland Avenue, Suite 118, Los Angeles, California 90064 USA Phone: 855-373-1614 • Fax 310-730-6241 • Email: info@ariabio.comwww.ariabio.com

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" in large, bold, red letters above a green line. Below the green line is the word "BIO" in large, bold, blue letters. Below the word "BIO" is the phrase "implant science" in small, light gray letters.

4. Predicate Devices:

Aria Bio's Dental Implants and Abutments are substantially equivalent to the following Predicate Devices:

  • 4.1 Primary Predicate Device: AB Dental Implants (by AB Dental Ltd) cleared under K132125; regulation number 872.3640 (Endosseous Dental Implant), product code: DZE (primary) and NHA (secondary).
  • 4.2 Reference Predicate Device: Alpha-Bio Tec Ltd Dental Implant System cleared under K063364; regulation number 872.3640 (Endosseous Dental Implant), product code: DZE.

5. Intended Use/Indications for Use:

Aria Bio's Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The Dental Implant System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

6. Device Description:

The Aria Bio Dental Implant System consists of two stage endosseous form dental implants, hexagonal and conical implants and hexagonal and conical abutments; cover screws and healing caps; abutment systems and superstructures and surgical instruments.

The Aria Bio Dental Implant System includes:

Versatile model is a platform switched, tapered implant designed for use in any bone type and is offered with either a deep conical or Morse tapered hex connection. It is available in lengths of 8, 10, 11.5, 13 and 16 mm except for the 4.2 mm diameter which is available in 8, 10, 11.5, 13, 16, 18 and 20, and in diameters of 3.5, 3.75, 4.2, 4.5, 5 and 6 mm. The Versatile implants are constructed of Ti-6ALAV-ELI (Grade 23). Immediate model is a platform switched root-form implant design. The Immediate suitable for both immediate load applications and insertion into fresh extraction sockets and is offered with either a deep conical or Morse tapered hex connection. It is available in lengths of 8, 10, 11.5, 13, 16 mm except for the 3.5 diameter model which is available in 10, 11.5, 13, 16 mm and in diameters of 3.5, 3.75, 4.2, 4.5, 5 and 6 mm. The immediate implants are constructed of Ti-6AL4V-ELI (Grade 23).

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" stacked on top of the word "BIO". There is a green line separating the two words. Underneath the word "BIO" is the phrase "implant science" in a smaller font.

Fixation model is a root form implant is appropriate for both immediate load applications and insertion into fresh extraction sockets. The Fixation is offered with a deep conical or prosthetic connection. The Fixation is available in lengths of 8.5, 10, 11.5, 13, 15, and 18 mm and in diameters of 3.5, 4.3, 5.0 and 6.0 mm. The Fixation implants are constructed of Ti-6AL4V-ELI (Grade 23).

Solution model is equipped with a specially designed, narrow, deep, conical connection and is designed for use in narrow bone volumes. The Solution5 is available in lengths of 10, 11.5, 13, 16 mm with diameter of 3.25 mm. The Solution2 is available with lengths of 8, 10, 11.5, 13, 16 mm and with a diameter of 3.25 mm. The Solution5 and Solution2 differ in that the Solution5 implants have an outer thread like the Immediate implant model, while the Solution2 implants have an outer thread like the Versatile implant model. The Solution implants are constructed of Ti-6AL4V-ELI (Grade 23).

Healing caps are available in 3 platform sizes: normal, narrow and wide and are made of Ti-6AL4V-ELI (Grade 23). The conical connection models are available in narrow, regular and wide platforms with lengths of 2, 3, 4, 5, 6, 7 mm and diameters of 3.0, 3.5 and 4.3 mm. The hex connection models are available in narrow emergence, regular platform with lengths of 2, 3, 4. 5. 6. 7 mm and a diameter of 3.75 mm and wide emergence regular platform with lengths of 3, 4, 5, and 6 mm and a diameter of 3.75 mm.

Temporary abutment is for immediate loading if appropriate and is made of Ti-6AL4V-ELI (Grade 23). It is provided with a plastic cap and is available in 3 hardness levels from the hardest to the softest. The Temporary abutment is available with a conical connection in narrow, regular and wide platforms all with lengths of 1, 2, 3, 4, 5. 6 and 7 mm and diameters of 3.0. 3.5 and 4.3 mm. It is also available with a hex connection regular platform with lengths of 1, 2, 3, 4, 5, 6, 7 mm and a diameter of 3.75 mm.

Temporary abutment plastic caps are for temporary use in conjunction with the temporary abutments for up to 30 days and are constructed of Delrin. They are supplied soft, medium and hard.

Anti Rotation abutments are straight titanium (Ti-6AL4V-ELI (Grade 23)) abutments with hex designed for permanent restoration. The abutments are supplied as regular non-shouldered straight abutments with lengths of 5, 7, 9, 12 and 15 mm with a narrow conical platform of 3.0 mm width, a regular conical platform of 3.5 mm width a regular hex platform of 3.75 mm width and a wide conical platform of 4.3 mm. As a narrow emergence non-shouldered straight abutment with lengths of 5, 7 and 9 mm with a narrow conical platform of 3.0 mm width, a regular conical platform of 3.5 mm width a regular hex platform of 3.75 mm width and a wide conical platform of 4.3 mm. As a wide emergence non shouldered straight abutment with lengths of 9

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" in blue, with a green line underneath it. Below that is the word "BIO" in blue. Underneath that is the phrase "implant science" in a smaller, light gray font.

and 12 mm with a narrow conical platform of 3.0 mm width, a regular conical platform of 3.5 mm width a regular hex platform of 3.75 mm width and a wide conical platform of 4.3 mm.

15° and 25° angular abutment are made of Ti-6AL4V-ELI (Grade 23). 15° abutments are supplied as a conical connection narrow platform with a width of 3.0 mm and lengths of 9, 11 and 13 mm, a conical connection regular platform with a width of 3.5 mm and lengths of 9, 11 and 13 mm and a width of 3.75 mm with a length of 9 mm, a conical connection wide platform with a width of 4.3 mm and lengths of 9, 11 and 13 mm, a hex connection regular platform with a width of 3.75 mm and a length of 9 and 11 mm. 25° abutments are supplied as a conical connection narrow platform with a width of 3.0 mm and lengths of 9, 11 and 13 mm, a conical connection regular platform with a width of 4.3 mm and lengths of 9, 11 and 13 mm, a conical connection wide platform with a width of 3.5 mm and lengths of 9, 11 and 13 mm and , a hex connection regular platform with a width of 3.75 mm and lengths of 9, 11 and 13 mm. 15° and 25° narrow emergence abutments are supplied as a conical connection narrow platform with a width of 3.0 mm and length of 9.0 mm, as a conical connection regular platform with a width of 3.5 mm and length of 9.0 mm, as a conical connection wide platform with a width of 4.3 mm and length of 9.0 mm and as a hex connection regular platform with a width of 3.75 mm and length of 9.0 mm. Anatomic Straight Abutment are constructed of Ti-6ALAV-ELI (Grade 23). All

have a length of 7.5mm and shoulder heights of 1, 2 and 3 mm. They are available in the following configurations: Conical connection, narrow platform 3.0 mm width, Conical connection, regular platform 3.5 mm width, Conical connection, wide platform 4.3 mm width, Hex connection, regular platform 3.75 mm width, Wide emergence, conical connection, narrow platform width 3.0 mm, Wide emergence, conical connection, regular platform width 3.5 mm, Wide emergence, conical connection, wide platform width 4.3 mm, Wide emergence, hex connection, wide platform width 4.3 mm

15°, 25° and 30° Angled Anatomic Abutment with Shoulder, all have a length of 7.5 mm and are constructed of Ti-6AL4V-ELI (Grade 23). They all have shoulder heights of 1, 2 and 3 mm and are available as conical connection, narrow platform width 3.0 mm, conical connection, regular platform width 3.5 mm, conical connection, wide platform width 4.3 mm, hex connection, regular platform width 3.75 mm. Straight Adaptor are all constructed of Ti-6AL4V-ELI (Grade 23) and are supplied in lengths of 1, 2, 3, 4, 5 and 6 mm in the following models: conical connection, narrow platform diameter 3.0 mm, conical connection, regular platform diameter 3.5 mm, conical connection, wide platform diameter 4.3 mm, hex connection, regular platform diameter 3.75 mm.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" stacked on top of the word "BIO". There is a green line separating the two words. Below the word "BIO" are the words "implant science" in a smaller font.

Casting Sleeves for Straight Adaptor are available in gold and titanium, all with width of 4.5mm and a length of 10.0 mm. Both gold and titanium are available in the following configurations: Gold sleeve casting adaptor, conical, narrow, Gold sleeve casting adaptor, conical, regular, Gold sleeve casting adaptor, conical, wide, Gold sleeve casting adaptor, hex, regular.

Ball Attachment Abutments are all constructed of Ti-6AL4V-ELI (Grade 23). All provided with lengths of 1, 2, 3, 4, 5 mm and in the following configurations: Conical connection, narrow platform 3.0 mm diameter, Conical connection, regular platform 3.5 mm diameter, Conical connection, wide platform 4.3 mm diameter , Hex connection, regular platform 3.75 mm diameter.

Multi-Unit Abutment are all constructed of Ti-6AL4V-ELI (Grade 23) and are supplied with lengths of 1, 2, 3 and 4 mm. The following configurations are available in both narrow and wide platforms and in hex regular platform: Straight, Conical Connection 18°, Conical Connection 30°, Narrow and Wide Platform.

Cover Screw are all constructed of Ti-6AL4V-ELI (Grade 23) and are supplied with a conical connection as narrow, regular and wide and with a hex connection.

7. Substantial Equivalence:

The proposed Aria Bio Dental Implant System has similar indications for use, technological characteristics, mode of operation and, performance specification as the above identified predicate devices. The proposed device utilizes similar intended use as the predicates and is placed using the same methodology as all of the selected predicate devices. Both the proposed and predicate devices function in the same manner providing support for prosthetic devices in the upper or lower jaw, having similar range of dimensions and basic shape. The specific indications for use for the current device and two predicate devices are shown in the following table:

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" in large blue font with a green line underneath. Below that is the word "BIO" in a similar blue font. Underneath that is the phrase "implant science" in a smaller gray font.

510(k) NumberK063364 (Reference PD)K132125 (Primary PD)K152555 (Proposed)
DeviceAlpha-Bio Tec DentalImplant SystemA.B. Dental Devices DentalImplants SystemAria Bio DentalImplant System
Indications forUseThe Alpha-Bio DentalImplant System® isindicated for use insurgical and restorativeapplications forplacement in the bone ofthe upper or lower jaw toprovide support forprosthetic devices, suchas artificial teeth, in orderto restore the patient'schewing function. TheAlpha-Bio DentalImplant System®is indicated also forimmediate loading whengood primary stabilityis achieved and withappropriate occlusalloading.A.B.Dental Devices DentalImplants System is indicated foruse in surgical and restorativeapplications for placement in thebone of the upper or lower jaw toprovidesupport for prosthetic devices,such as artificial teeth, in order torestore the patient'schewing function. A.B. DentalImplants System is indicated alsofor immediate loading whengood primary stability isachieved and with appropriateocclusal loading.Two Stage Implants:12,15,16131.One Stage: 16, 16b, 168.One Stage & One-Piece 3.0 mmdiameter implants: 16, 168,16131, are intended forplacement at the mandibularcentral and lateral incisors andmaxillary and lateral incisors.Indicated also for denturestabilization using multipleimplants. One stage & One-Piece2.4 mm diameter implants fortemporary use or long term use:16, 16b, permit immediate splintstability and long term fixation ofnew or existing crown, bridgeand prosthesis. P1 4 AngulatedAbutment Adapter is to be usedwith implant diameter 4.2mmand higher.Aria Bio's DentalImplant System isindicated for use insurgical andrestorative applicationsfor placement in thebone of the upper orlower jaw to providesupport for prostheticdevices such asartificial teeth, in orderto restore the patient'schewing function. TheDental Implant Systemis indicated also forimmediate loadingwhen good primarystability is achievedwith appropriateocclusal loading.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" in blue with a green line underneath it. Below that is the word "BIO" in blue. Underneath that is the phrase "implant science" in a smaller, light gray font.

Differences that were identified were addressed as follows:

Implants differences -

With the exception of the surface treatment the Aria Bio implants are similar as the two predicate implants. The predicate devices are treated with Calcium Fosfat-Hydroxyapatite while the Aria Bio implants are treated with HA blasting/Double acid etching. The slight difference in the surface treatment methods is not likely to affect the Titanium biocompatibility since both methods are valid and acceptable within the dental industry and biocompatibility was demonstrated by Aria Bio. Aria Bio products' list also do not include one-piece (one-stage) implant within this submission.

Abutments –

The difference between the Aria Bio abutments and the predicate abutments are in the dimensions offered. These difference are displayed in the following table (Abutments not shown do not differe from the predicate:

Product DescriptionAria Bio ProductPredicate Devices:
Diameter(mm)Length (mm)Diameter (mm)Length (mm)
AlphaBio K#063364
Healing abutment, NarrowPlatform, Conical3.02, 3, 4, 5, 6, 73.23,5,7
Healing abutment, RegularPlatform, Conical3.52, 3, 4, 5, 6, 73.63,5,7
Healing abutment, WidePlatform Conical4.32, 3, 4, 5, 6, 73.83,5,7
5.03,5,7
AB Dental K#132125
Temporary Abutment, NarrowPlatform, Conical3.01, 2, 3, 4 ,5, 6 ,73.751, 2, 3, 4 ,5, 6 ,7
Temporary Abutment,Regular Platform, Conical3.51, 2, 3, 4 ,5, 6 ,73.751, 2, 3, 4 ,5, 6 ,7
Temporary Abutment, WidePlatform, Conical4.31, 2, 3, 4 ,5, 6 ,73.751, 2, 3, 4 ,5, 6 ,7
Temporary Abutment,Regular Platform, Hex3.751, 2, 3, 4 ,5, 6 ,73.751, 2, 3, 4 ,5, 6 ,7
AB Dental K#132125
Anti-Rotation Abutment,Regular Non-ShoulderedStraight, Narrow, conical3.05, 7, 9, 12, 153.755, 7, 9, 12, 15
Anti-Rotation Abutment,Regular Non-ShoulderedStraight, Regular, conical3.55, 7, 9, 12, 153.755,7,9
Product DescriptionAria Bio ProductPredicate Devices:
Diameter(mm)Length (mm)Diameter (mm)Length (mm)
Anti-Rotation Abutment,Regular Non-ShoulderedStraight, Wide, conical4.35, 7, 9, 12, 153.757,9
Anti-Rotation Abutment,Narrow Emergence Non-Shouldered Straight, Narrow,Conical3.05,7,93.755, 7, 9, 12, 15
Anti-Rotation Abutment,Narrow Emergence Non-Shouldered Straight, Regular,Conical3.55,7,93.755,7,9
Anti-Rotation Abutment,Narrow Emergence Non-Shouldered Straight, Wide,conical4.35,7,93.757,9
Anti-Rotation Abutment WideEmergence Non-ShoulderedStraight Abutment, Narrow,Conical3.09, 123.755, 7, 9, 12, 15
Anti-Rotation Abutment WideEmergence Non-ShoulderedStraight, Regular, conical3.59,123.755,7,9
Anti-Rotation Abutment WideEmergence Non-ShoulderedStraight, Wide, conical4.39,123.757,9
Anti-Rotation Abutment,Regular Non-ShoulderedStraight, Regular, Hex3.755, 7, 9, 12, 153.755, 7, 9, 12, 15
Anti-Rotation Abutment,Narrow Emergence Non-Shouldered Straight, Regular,Hex3.755,7,93.755,7,9
Anti-Rotation Abutment,Wide Emergence Non-Shouldered Straight, Regular,Hex3.759, 123.757,9
AB Dental K#132125
15° Angled AnatomicAbutment with Shoulder,Conical Connection, NarrowPlatform3.01, 2, 33.751, 2, 3
Product DescriptionAria Bio ProductPredicate Devices:
Diameter(mm)Length (mm)Diameter (mm)Length (mm)
15° Angled AnatomicAbutment with Shoulder,Conical Connection, RegularPlatform3.51, 2, 33.751, 2, 3
15° Angled AnatomicAbutment with Shoulder,Conical Connection, WidePlatform4.31, 2, 33.751, 2, 3
25° Angled AnatomicAbutment with Shoulder,Conical Connection, NarrowPlatform3.01, 2, 33.751, 2, 3
30° Angled AnatomicAbutment with Shoulder,Conical Connection, NarrowPlatform3.013.751, 2, 3
25° Angled AnatomicAbutment with Shoulder,Conical Connection, RegularPlatform3.51, 2, 33.751, 2, 3
25° Angled AnatomicAbutment with Shoulder,Conical Connection, WidePlatform4.31, 2, 33.751, 2, 3
AB Dental K#132125
Straight Adaptor, ConicalConnection, Narrow Platform31, 2, 3, 4, 53.751, 2, 3, 4, 5
Straight Adaptor, ConicalConnection, Regular Platform3.51, 2, 3, 4, 53.751, 2, 3, 4, 5
Straight Adaptor, ConicalConnection, Wide Platform4.31, 2, 3, 4, 53.751, 2, 3, 4, 5
Straight Adaptor,HexConnection, Regular Platform3.751, 2, 3, 4, 53.751, 2, 3, 4, 5
AB Dental K#132125
Gold Sleeve Casting Adaptor,Conical, Narrow4.5104.510
Gold Sleeve Casting Adaptor,Conical, Regular4.5104.510
Gold Sleeve Casting Adaptor,Conical. Wide4.5104.510
Product DescriptionAria Bio ProductPredicate Devices:
Diameter(mm)Length (mm)Diameter (mm)Length (mm)
Titanium Sleeve CastingAdaptor, Conical, Narrow4.5104.510
Titanium Sleeve CastingAdaptor, Conical, Regular4.5104.510
Titanium Sleeve CastingAdaptor, Conical, Wide4.5104.510
AlphaBio K#063364
Multi-Unit Abutment:
Multi-Unit Abutment StraightAdaptor, Conical Connection,Narrow Platform1, 2, 3, 41.6, 2.5, 3.5, 4.5(Regular platform)
Multi-Unit Abutment 18°,Conical Connection, NarrowPlatform1, 2, 3, 42.5, 3.5 (Regularplatform)
Multi-Unit Abutment 30°,Conical Connection, NarrowPlatform1, 2, 3, 43.5, 4.5 (Regularplatform)
Multi-Unit Abutment StraightAdaptor, Conical Connection,Regular Platform1, 2, 3, 41.6, 2.5, 3.5, 4.5(Regular platform)
Multi-Unit Abutment 18°,Conical Connection, RegularPlatform1, 2, 3, 42.5, 3.5 (Regularplatform)
Multi-Unit Abutment 30°,Conical Connection, RegularPlatform1, 2, 3, 43.5, 4.5 (Regularplatform)
Multi-Unit Abutment StraightAdaptor, Conical Connection,Wide Platform1, 2, 3, 41.6, 2.5, 3.5, 4.5(Regular platform)
Multi-Unit Abutment 18°,Conical Connection, WidePlatform1, 2, 3, 42.5, 3.5 (Regularplatform)
Multi-Unit Abutment 30°,Conical Connection, WidePlatform1, 2, 3, 43.5, 4.5 (Regularplatform)
Multi-Unit Abutment StraightAdaptor, Hex Connection,Regular Platform1, 2, 3, 41.6, 2.5, 3.5, 4.5(Regular platform)
Multi-Unit Abutment 18°, HexConnection, Regular Platform1, 2, 3, 42.5, 3.5 (Regularplatform)
Product DescriptionAria Bio ProductPredicate Devices:
Diameter (mm)Length (mm)Diameter (mm)Length (mm)
Multi-Unit Abutment 30°, Hex Connection, Regular Platform1, 2, 3, 43.5, 4.5 (Regular platform)AB Dental K#132125
Ball Attachment Abutment Conical, Regular Platform3.51, 2, 3, 4, 5, 64.11, 2, 3, 4, 5, 6
Ball Attachment Abutment Conical, Wide Platform4.31, 2, 3, 4, 5, 64.11, 2, 3, 4, 5, 6
Ball Attachment Abutment Hex, Regular Platform3.751, 2, 3, 4, 5, 64.11, 2, 3, 4, 5, 6

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" in a large font, with a green line underneath it. Below the green line is the word "BIO" in a large blue font. Underneath the word "BIO" is the phrase "implant science" in a smaller, light gray font.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" in a large font, with a green line underneath it. Below the green line is the word "BIO" in a large blue font. Underneath the word "BIO" is the phrase "implant science" in a smaller, light gray font.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" on top in blue, with a green line underneath. Below that is the word "BIO" in a larger blue font. Underneath that is the phrase "implant science" in a smaller grey font.

{13}------------------------------------------------

As can be shown in the comparison table, all the abutments (Aria Bio Dental, Alpha-Bio, AB Dental), consist of many variations which are basically very similar to each other. Any differences between the dimensions of the predicates and the predicate abutments fall well within the dimensions of abutments widely used in the dental field. One abutment angulation of Aria Bio Angled Anatomic Abutment 30° exceeds the predicate maximal angle 25°, yet this difference was evaluated through performance data (fatigue) with successful results. Therefore it was concluded that the dimensional differences do not affect the device substantial equivalency.

In the light of the above, we believe that the overall evaluation of similarities and differences between our device and the selected predicate devices, showed that substantial equivalency of our device with its predicate devices was demonstrated.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Aria Bio. The logo has the word "ARIA" stacked on top of the word "BIO". There is a green line separating the two words. Below the word "BIO" is the phrase "implant science" in a smaller, light gray font.

8. Performance Testing:

A series of performance tests and evaluations were performed to demonstrate that Aria Bio's Dental Implant System does not raise any new issues that may affect its substantial equivalency with its predicate devices. These evaluations include fatigue (ISO 14801), surface analysis by SEM/EDS and XPS, biocompatibility in accordance with ISO 10993-1, sterilization validation (ISO 11137-2, AAMI TIR 33, ISO 17665-1) and shelf life validation (ISO 11607-1) in order to assure maintaining of SAL 10° along the shelf life.

All results are supporting Aria Bio's labeling claims in order to establish substantial equivalency.

9. Summary and Conclusion:

We believe that Aria Bio Dental Implants products, which are the subject of this 510(k) submission, are substantially equivalent to the predicate devices cited above. The device constitutes a reliable medical device, meeting all declared requirements of its intended use and the device does not introduce new risks to patients when used as intended.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.