K Number
K152362

Validate with FDA (Live)

Date Cleared
2015-10-16

(57 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Terumo SurGuard®3 Safety Hypodermic Needle is intended for use in the aspiration and injection of fluids for medical purposes. The Terumo Safety Hypodermic Needle is compatible for use with standard luer lock syringes. Additionally, after withdrawal of the needle from the body, the attached needle safety sheath can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

Device Description

The Terumo SurGuard®3 Safety Hypodermic Needle [hereinafter referred to as "SurGuard 3 (Short Sheath 2)"] consists of a hypodermic needle with a hinged safety sheath attached to the connector hub which can be attached to either a luer slip or luer lock syringe. This device features a hinged safety sheath attached to the needle hub. The safety sheath contains two locking mechanisms, the tooth-cannula and sheathcollar which are simultaneously activated when manually pressed over the needle immediately after use and just prior to disposal to minimize the possibility of sharps injury. The safety sheath is activated with a one-handed operation, using the finger, thumb, or surface activation.

AI/ML Overview

The document describes the Terumo SurGuard®3 Safety Hypodermic Needle and its modifications. However, the provided text does not contain detailed information about acceptance criteria and a study that comprehensively proves the device meets those criteria in the format requested.

Specifically, the document states:

  • "No deviations from recognized consensus ISO standards were identified, except where the design of the device resulted in a modified method or acceptance criterion." (page 10)
  • "The device complies with the acceptance criteria established based on the predicate, as shown in the table below: Table 5.6: Performance Testing per Internal Standards" (page 10)
  • "Performance testing demonstrates that the new Terumo SurGuard®3 Safety Hypodermic Needle conforms to the recognized consensus ISO standards (ISO 7864 and ISO 23908) and meets internal standards acceptance criteria." (page 10)

This indicates that acceptance criteria were established and met, but the specific numerical criteria for each test and the detailed results demonstrating conformance are not provided. The results are summarized as "Meets acceptance criteria" or "Meets standard."

Additionally, the document explicitly states:

  • "This 510(k) does not include data from clinical tests." (page 11)

Therefore, much of the requested information regarding study details (sample size, data provenance, ground truth establishment, expert involvement, MRMC studies) is not available in the provided text.

Here's a summary of what can be extracted or inferred:

1. Table of acceptance criteria and the reported device performance

Acceptance Criteria (Implied)Reported Device Performance
ISO Standards
Cleanliness(Not explicitly stated)
Limits for Acidity or AlkalinityN/A (Not applicable)
Conical FittingMeets standard (ISO 7864)
Bond Between Hub and Needle TubeMeets standard
Adhesive HoldMeets standard
Exposed cannula lengthMeets standard
Needle Access to the Sharp in Safe Mode TestMeets standard (ISO 23908)
Internal Standards
Sheath ActivationMeets acceptance criteria
Sheath DeactivationMeets acceptance criteria
Needle PenetrationMeets acceptance criteria
Manual Sheath ActivationMeets acceptance criteria
MeasurementsMeets acceptance criteria
Simulated Use StudyPassed

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified for any of the performance tests.
  • Data Provenance: Not specified, but generally, such performance testing is conducted by the manufacturer (Terumo (Philippines) Corporation). The nature of the tests (mechanical, sterilization, biocompatibility) suggests laboratory testing rather than human subject data. Retrospective/Prospective is not applicable in this context.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable as the testing involves physical/mechanical performance and adherence to ISO standards and internal criteria, not expert interpretation of outputs.

4. Adjudication method for the test set

  • Not applicable as the testing involves physical/mechanical performance and adherence to ISO standards and internal criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This is a medical device (hypodermic needle) and not an AI/imaging device. The document explicitly states: "This 510(k) does not include data from clinical tests." (page 11)

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This is a medical device (hypodermic needle) and not an AI/imaging device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the physical and mechanical performance tests, the "ground truth" would be the established specifications, engineering tolerances, and requirements set forth by ISO standards (ISO 7864, ISO 23908, ISO 11137-1, ISO 14971) and the manufacturer's internal standards.

8. The sample size for the training set

  • Not applicable as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

  • Not applicable as this is not an AI/machine learning device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a group of human profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 16, 2015

Terumo (Philippines) Corporation c/o Ms. Phebe Varghese Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Avenue, Suite 320 Somerset, New Jersey 08873

Re: K152362

Trade/Device Name: Terumo SurGuard®3 Safety Hypodermic Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: August 19, 2015 Received: August 21, 2015

Dear Ms. Varghese:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang
-s

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152362

Device Name Terumo SurGuard®3 Safety Hypodermic Needle

Indications for Use (Describe)

The Terumo SurGuard®3 Safety Hypodermic Needle is intended for use in the aspiration and injection of fluids for medical purposes. The Terumo Safety Hypodermic Needle is compatible for use with standard luer lock syringes. Additionally, after withdrawal of the needle from the body, the attached needle safety sheath can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

A. SUBMITTER INFORMATION (807.92(a)(1))

Prepared by:Ma.Cristina Faderagao
Regulatory Affairs Assistant Manager
Terumo (Philippines) Corporation
Tel. +63 (49) 5412111
Fax +63 (49) 5412121

Prepared for: Owner/Operator Terumo (Philippines) Corporation 124 East Main Avenue, Laguna Technopark Binan, Laguna, Philippines Owner/Operator Number: 10033966

Manufacturer and Sterilization Facility (Applicant)

Terumo (Philippines) Corporation 124 East Main Avenue, Laguna Technopark Binan, Laguna, Philippines Registration Number: 3003902955

Contact Person: Phebe Varghese Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Avenue, Suite 320 Somerset, NJ 08873 Tel. (732) 412-4166 Fax (732) 302-4905 E-mail: phebe.varghese@terumomedical.com

Date prepared: October 14, 2015

{4}------------------------------------------------

B. DEVICE NAME (807.92(a)(2))

Proprietary Name:Terumo SurGuard®3 Safety Hypodermic Needle
Common Name:Hypodermic needle with safety sheath or needle withneedle protection device
Classification Name:Hypodermic single lumen needle
Classification Panel:General Hospital
Regulation:21 CFR 880.5570
Product Code:FMI
Classification:Class II

C. PREDICATE DEVICE (807.92(a)(3))

The legally marketed device to which substantial equivalence is claimed is the current device, K122249 - Terumo SurGuard®3 Safety Hypodermic Needle, manufactured by Terumo (Philippines) Corporation.

D. REFERENCE DEVICE

Biocompatibility data from K121607 - Terumo Hypodermic Needle, manufactured by Terumo (Philippines) Corporation, was levered for information regarding hub colorants.

E. REASON FOR 510(k) SUBMISSION

This premarket notification (Special 510(k)) is being submitted for the Terumo SurGuard 3 Safety Hypodermic Needle, manufactured by Terumo (Philippines) Corporation, for the following modifications to the current device: new needle length (19mm), new gauge sizes (18, 19, 20, and 22) for 13 and 16mm needle lengths, cleared under K122249, and a modified safety sheath design for all new gauge and needle length combinations.

F. DEVICE DESCRIPTION (807.92(a)(4))

Principle of Operation Technology

The Terumo SurGuard®3 Safety Hypodermic Needle is operated manually or by manual process.

{5}------------------------------------------------

Design/Construction

The Terumo SurGuard®3 Safety Hypodermic Needle [hereinafter referred to as "SurGuard 3 (Short Sheath 2)"] consists of a hypodermic needle with a hinged safety sheath attached to the connector hub which can be attached to either a luer slip or luer lock syringe. This device features a hinged safety sheath attached to the needle hub. The safety sheath contains two locking mechanisms, the tooth-cannula and sheathcollar which are simultaneously activated when manually pressed over the needle immediately after use and just prior to disposal to minimize the possibility of sharps injury. The safety sheath is activated with a one-handed operation, using the finger, thumb, or surface activation.

The safety sheath has a finger guide area consisting of a circular dent (for thumb activation; see Figure A) and ramp (for finger activation; see Figure B) which provides tactile confirmation that the user's finger is in the appropriate position for performing the activation. The ramp has steps to provide strong grasp when activating the sheath. There are two stoppers which is located at the end of the circular dent and ramp which prevent the user's finger from going towards the cannula during activation. Another method of activation is by manually pressing the safety sheath over the needle using a firm surface.

Image /page/5/Figure/4 description: The image shows a close-up of a mechanical part with a circular dent on its surface. A black arrow points to the dent, labeled as "Circular Dent." A red arrow points to a component on the side of the part, labeled as "Stopper." The image provides a detailed view of the part's features, highlighting the dent and the stopper.

Figure A: Circular dent for thumb activation

Image /page/5/Figure/6 description: The image shows a close-up of a gray mechanical part with two stoppers. The stoppers are labeled with red arrows pointing to them. The part has a ridged design with multiple horizontal lines.

Figure B: Ramp for finger activation

The locking mechanisms are positioned within the center and proximal ends of the sheath. The hinge feature allows the medical practitioner to adjust the sheath to its desired position for use.

The SurGuard 3 (Short Sheath 2) will have two kinds of needle bevel alignment; zero degree bevel alignment and 90 degree bevel alignment. The SurGuard 3 (Short Sheath 2) will be individually packaged and sterilized by electron beam.

{6}------------------------------------------------

Materials

The materials for all SurGuard 3 (Short Sheath 2) sizes are provided in the table below.

Part NameMaterial NameManufacturer /Supplier
Cannula*Stainless SteelKaneko Medix / TerumoCorporation
Needle Hub*PolypropyleneTerumo Corporation
Hub Colorant BaseResin*Polypropylene + PigmentTerumo Coporation
Pink*Quinacridone RedPerylene RedTerumo Corporation
Cream*Perylene RedCondenzed Azo YellowTerumo Corporation
Yellow*Perylene RedCondenzed Azo YellowTitanium DioxideTerumo Corporation
Black*Carbon BlackTerumo Corporation
Needle ProtectorPolypropyleneTerumo Corporation
Safety SheathPolypropyleneTerumo Corporation
CollarPolypropyleneTerumo Corporation
GlueAdhesive (Bond 927-T)Terumo Corporation
Needle Lubricant*Reactive SiliconeTerumo Corporation

*Body contacting material

Specifications

The new needle gauge sizes to be added under this 510(k) are: 18G, 19G, 20G, and 22G; additionally, a new needle length of 19mm will be added. This submission also provides new gauge and length combinations for the 13mm and 16mm needle lengths, which were cleared under K122249. Finally, this submission includes a modified design for the short sheath for all new gauge and needle length combinations.

The SurGuard 3 (Short Sheath 2) features a modified short sheath designed to allow activation of the safety sheath with larger gauge sizes of short needle lengths. Because the needle is captured and locked inside the tooth cannula part of the safety sheath, the short sheath internal clearances and tooth thicknesses were modified to

{7}------------------------------------------------

accommodate these kinds of needles.

The following table shows the needle gauge and needle combinations that are new for this submission and those that were cleared under K122249. "X" demonstrates combinations clearance under K122249, and "New" demonstrates combinations proposed for this submission.

CombinationNeedle Length
3/8"(9mm)1/2"(13mm)5/8"(16mm)3/4"(19mm)1"(25mm)
GaugeSize18GNewNew19GNewNew20GNewNew21G22GNewNewSize23GX25GXX26GXX27GX30G
GaugeSize18GNewNew
19GNewNew
20GNewNew
21G
22GNewNew
Size23GX
25GXX
26GXX
27GX
30GX

Table 5.2: Gauge/Needle Length Combination

G. INDICATIONS FOR USE (807.92(a)(5))

The Terumo SurGuard®3 Safety Hypodermic Needle is intended for use in the aspiration and injection of fluids for medical purposes. The Terumo Safety Hypodermic Needle is compatible for use with standard luer slip and luer lock syringes. Additionally, after withdrawal of the needle from the body, the attached needle safety sheath can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

Note: The indications for use are identical to the predicate device, Terumo SurGuard 3 Safety Hypodermic Needle (K122249).

H. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))

The SurGuard 3 (Short Sheath 2), the subject of this Special 510(k), is substantially equivalent in its intended use/indications for use, technology/principal of operation, materials, and performance to the Terumo SurGuard®3 Safety Hypodermic Needle, manufactured by Terumo (Philippines) Corporation (K122249).

{8}------------------------------------------------

A comparison of the technological characteristics is summarized in the table below.

DeviceCharacteristicSubject Device:SurGuard 3 (Short Sheath 2)Predicate Device:SurGuard 3 (K122249)
ManufacturerTerumo (Philippines) CorporationSame
Intended Use /Indications forUseThe Terumo SurGuard3 SafetyHypodermic Needle device isintended for use in the aspirationand injection of fluids for medicalpurposes. The Terumo SurGuard3Safety Hypodermic Needle iscompatible for use with standardluer slip and luer lock syringes.Same
OperationPrincipleManualSame
Design /ConstructionHypodermic needle with integratedsafety feature without syringeattachedSame
MaterialsStainless steel cannula Polypropylene Hub Polypropylene Needle Protector Polypropylene safety featureSame
PackageBlister Shelf box Shipping boxSame
SpecificationsNeedle gauges:18G, 19G, 20G, and 22GNeedle length:13mm, 16mm, and 19mmSafety Sheath:Short sheath 2Needle gauges:23G, 25G, 26G, and 27GNeedle length:13mm, 16mm, and 25mmSafety Sheath:Short sheath
SterilizationE-beam radiation (validated inaccordance with ISO 11137-1 toachieve SAI 10-6)Same

Table 5.3: Comparison Between Subject and Predicate Device

The colorant biocompatibility data from the reference device, Terumo Hypodermic Needle (K121607), were levered for the subject device. A comparison of hub colorant

{9}------------------------------------------------

information is summarized in the table below.

Hub ColorSubject Device:SurGuard 3 (Short Sheath 2)Reference Device:Terumo Hypodermic Needle(K121607)
Pink*Quinacridone RedSame
Pink*Perylene RedSame
Cream*Perylene RedSame
Cream*Condensed Azo YellowSame
Yellow*Perylene RedSame
Yellow*Condensed Azo YellowSame
Black*Titanium DioxideSame
Black*Carbon BlackSame
Table 5.4: Hub Colorant Information Table

*Body contacting material

Colorants present on the subject device and K121607 are supplied by Terumo Corporation.

I. NON CLINICAL TESTS (807.92(b)(1))

Performance

Performance testing was conducted to ensure sterility and functionality of the SurGuard 3 (Short Sheath 2) throughout the labeled shelf life, verify conformity to the applicable parts of ISO standards, and demonstrate substantial equivalence to the predicate device. The following performance tests were performed on the SurGuard 3 (Short Sheath 2):

TestStandardResult
Cleanliness
Limits for Acidity or AlkalinityN/A
Conical FittingISO 7864
Bond Between Hub and Needle Tube
Adhesive HoldMeets standard
Exposed cannula length
Needle Access to the Sharp in Safe Mode TestISO 23908Meets standard

Table 5.5: Performance Testing per ISO Standards

{10}------------------------------------------------

No deviations from recognized consensus ISO standards were identified, except where the design of the device resulted in a modified method or acceptance criterion.

Additionally, performance testing other than to the above ISO Standards was performed on the device to verify the modified device against the currently cleared device. The device complies with the acceptance criteria established based on the predicate, as shown in the table below:

Performance TestResults
Sheath ActivationMeets acceptance criteria
Sheath DeactivationMeets acceptance criteria
Needle PenetrationMeets acceptance criteria
Manual Sheath ActivationMeets acceptance criteria
MeasurementsMeets acceptance criteria
Simulated Use StudyPassed

Table 5.6: Performance Testing per Internal Standards

Performance testing demonstrates that the new Terumo SurGuard®3 Safety Hypodermic Needle conforms to the recognized consensus ISO standards (ISO 7864 and ISO 23908) and meets internal standards acceptance criteria. Therefore, it is substantially equivalent to the predicate device and is acceptable for clinical use throughout the shelf life.

Biocompatibility

In accordance with ISO 10993-1, the SurGuard 3 (Short Sheath 2) is classified as: Externally Communicating Device, Circulating Blood, Limited Duration of Contact (<24hr). This classification was the same as the predicate and reference devices (K122249 and K121607).

All of the subject device's materials are the same as the predicate and reference devices (K122249 and K121607). These devices have the same intended use, body contact, and contact duration classification based on ISO 10993-1:2009. We conclude, therefore, that the subject device is biocompatible for its intended use.

{11}------------------------------------------------

Sterilization

The sterility of the device is assured using a sterilization method validated in accordance with ISO 11137-1:2006, Sterilization of Health Care Products – Radiation – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. The Terumo SurGuard® 3 Safety Hypodermic Needle is sterilized to provide a Sterility Assurance Level (SAL) of 10 °

Risk Analysis

A Product Risk Analysis was conducted in accordance with ISO 14971, taking into account the modifications to the previous device, and it was determined that there were no new issues of safety or effectiveness.

J. CLINICAL TESTS (807.92(b)(2))

This 510(k) does not include data from clinical tests.

K. CONCLUSION (807.92(b)(3))

In summary, the SurGuard 3 (Short Sheath 2), subject of this Special 510(k), is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to the current device:

K122249 - Terumo SurGuard 3 Safety Hypodermic Needle, manufactured by Terumo (Philippines) Corporation

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).