K Number
K122249

Validate with FDA (Live)

Date Cleared
2012-08-28

(32 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TERUMO® SurGuard®3 Safety Needle device is intended for use in the aspiration and injection of fluids for medical purposes. The TERUMO SurGuard®3 Safety Needle is compatible for use with standard luer slip and luer lock syringes.

Additionally, after withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

Device Description

The Terumo SurGuard 3 Safety Hypodermic Needle consists of a hypodermic needle with a hinged safety sheath attached to the connector hub with or without an attached hypodermic syringe. This device features a hinged safety sheath attached to the needle hub. The safety sheath contains two locking mechanisms, the tooth-cannula and sheath-collar which are simultaneously activated when manually pressed over the needle immediately after use and just prior to disposal to minimize the possibility of sharps injury. The safety sheath is activated with a one-handed operation by pressing the sheath either by finger, thumb or surface activation.

The safety sheath has a finger guide area consisting of a circular dent (for thumb activation) and ramp (for finger activation) which gives the user a concrete confirmation that the user's finger is in the appropriate position in performing the activation. The ramp has steps to provide strong grasp when activating the sheath, There are two stoppers located at the end of the circular dent and ramp which prevent the user's finger from going towards the cannula during activation. Another method of activation is by manually pressing the safety sheath over the needle using a firm surface.

The locking mechanisms are positioned within the center and proximal end of the sheath. The hinge feature allows the medical practitioner to flexibly adjust the sheath to its desired position for use.

The Terumo SurGuard 3 Safety Hypodermic Needle will be individually packaged and sterilized by electron beam as a safety needle only or as a safety needle with attached Terumo syringe.

AI/ML Overview

The manufacturer, TERUMO (PHILIPPINES) CORPORATION, conducted multiple performance tests to demonstrate the substantial equivalence of their TERUMO® SurGuard®3 Safety Hypodermic Needle and with attached syringe. This submission primarily introduced new needle gauge sizes, lengths, and a 1cc/ml syringe size, along with a modified shorter safety sheath design, compared to their previously cleared device (K113422).

Here's a breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and the Reported Device Performance:

The document doesn't explicitly list numerical acceptance criteria values for each test. Instead, it states that the results of the verification testing "met the acceptance criteria of the functional performance tests of both devices" (referring to the proposed and predicate devices). The general acceptance criterion for these tests appears to be that the proposed device performs comparably to or as well as the predicate device, confirming its safety and effectiveness.

Performance Test CategoryAcceptance Criterion (Implicit)Reported Device Performance
Functional PerformanceThe device's safety features should activate and deactivate correctly, the sheath should remain attached and provide protection, and the needle's physical integrity should be maintained. The manual activation mechanisms (finger, thumb, surface) should function as intended."Results of the verification testing performed on both the predicate device and the proposed device met the acceptance criteria of the functional performance tests of both devices." This covers: Sheath Activation, Sheath Deactivation, Sheath Removal, Sheath Radial, Sheath Retraction, Protector Fitting, Collar removal, Sheath Puncture, Adhesive Hold, Needle Penetration, Manual Sheath Activation (Finger, Thumb, Surface), and Measurement Testing. The "Simulated Use Study" further concluded that "the proposed Terumo SurGuard 3 Safety Hypodermic Needle met its intended use, and the Instructions for Use were adequate for safe and effective activation of the safety feature."
Shelf LifeThe device's functional performance should remain acceptable throughout its claimed shelf life (5 years in this case)."Results of verification testing indicated that the functional performance of the proposed Terumo SurGuard®3 Safety Hypodermic Needle and with attached syringe (1cc/ml) is acceptable throughout a 5-year (60months) shelf life." This was based on accelerated aging tests, with real-time testing to confirm. The specific functional tests re-evaluated for shelf life included: Sheath Activation, Sheath Deactivation, Sheath Removal, Sheath Radial, Sheath Retraction, Protector Fitting, Collar removal, Sheath Puncture, Adhesive Hold, Needle Penetration, Manual Sheath Activation (Finger, Thumb, Surface), and Package Seal Strength.
BiocompatibilityThe device's blood-contacting materials should not cause adverse biological reactions.Biocompatibility was established based on the predicate device (K113422) as the blood-contacting materials are identical. No re-testing was deemed necessary. The predicate device's materials were tested "in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO - 10993, 'Biological Evaluation of Medical Devices part 1: Evaluation and Testing'." The results demonstrated the materials are biocompatible.
Risk AnalysisThe device should not introduce new safety or effectiveness issues."Additionally a product risk analysis was conducted and there were no new issues of safety and effectiveness."
Standard ComplianceThe device should comply with relevant international standards for sharps injury protection, sterile hypodermic syringes, conical fittings, and needle color coding. (Note: Exception for patency of lumen for ultra-thin wall cannulas on specific gauges).The device "complies with the requirements specified in the following standards: ISO 23908:2011, ISO 7886-1:1993, ISO 594-1:1986, ISO 6009:1992/Cor:1:2008." It also complies with ISO 7864:1993, with an explicitly stated exception for Section 13.2a Patency of Lumen for 26G, 27G, and 30G ultra-thin wall cannulas, as no specifications are given for these types.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: The document does not specify the exact sample sizes (number of devices) used for each individual performance test (Sheath Activation, Puncture, etc.). It mentions that the "Simulated Use Study" involved "healthcare workers with varied amounts of experience," but doesn't quantify the number of participants or devices used in that study.
  • Data Provenance: The studies were conducted by TERUMO (PHILIPPINES) CORPORATION. The country of origin for the actual testing data would therefore be the Philippines. The studies appear to be prospective in nature, as they involve testing newly manufactured devices to verify their performance against specifications and predicate devices. Accelerated aging for shelf-life testing also falls under a prospective study design.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • The document does not mention the use of experts to establish a "ground truth" in the typical sense for a diagnostic device. The performance tests are largely objective, measuring mechanical and functional aspects of the device against predefined specifications or the performance of a predicate device.
  • For the "Simulated Use Study," it states that "healthcare workers with varied amounts of experience" performed the activations. Their "experience" implies a level of expertise, but specific qualifications or a "ground truth" establishment process by experts as one might find in clinical image analysis are not described. Their role was to assess if the device could be "safely activated" and if the "Instructions for Use were adequate."

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Given the nature of the performance tests (functional, mechanical, physical measurements), an adjudication method like "2+1" or "3+1" (common in human-read diagnostic studies) is generally not applicable. These tests typically involve objective measurement or observation against a pass/fail criterion.
  • For the "Simulated Use Study," the report states it "was concluded that the proposed Terumo SurGuard 3 Safety Hypodermic Needle met its intended use, and the Instructions for Use were adequate." This implies an overall assessment of the outcomes rather than an adjudication of individual interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices (especially those involving human interpretation of images or data, often with AI assistance). The TERUMO® SurGuard®3 Safety Hypodermic Needle is a physical medical device, not a diagnostic tool, and therefore MRMC studies are not applicable to its evaluation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • The concept of a "standalone algorithm" is not applicable here as the TERUMO® SurGuard®3 Safety Hypodermic Needle is a physical medical device and not an AI algorithm. Its performance is inherent in its physical design and manufacturing.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the functional and physical performance tests, the "ground truth" is implied by the pre-defined engineering specifications, international standards (ISO), and the performance of the legally marketed predicate device (K113422). The device is essentially compared against these objective benchmarks.
  • For the "Simulated Use Study," the "ground truth" was whether the device could be "safely activated" as intended, considering the Instructions for Use. This is verified through direct observation of actions by healthcare workers rather than an established "ground truth" from external sources like pathology or outcomes data.

8. The sample size for the training set

  • This device is not an AI algorithm or a diagnostic tool that requires a "training set" in the context of machine learning. Therefore, the concept of a training set sample size is not applicable.

9. How the ground truth for the training set was established

  • As the device is not an AI algorithm, there is no "training set" and thus no "ground truth for the training set" to be established.

{0}------------------------------------------------

K1224'9

page 1 of 10

AUG 2 8 2012

510(k) SUMMARY

Manufacturer/Owner: TERUMO (PHILIPPINES) CORPORATION124 East Main Avenue, Laguna Technopark,Biñan, Laguna, Philippines
Prepared by: Ma. Cristina Faderagao – Supervisor Regulatory AffairsTel. (+63) 49 541 2111Fax (+63) 49 541 2121 :
Date prepared: June 22, 2012
Contact person: Phillip Michael LesterRegulatory Affairs SpecialistTerumo Medical Corporation950 Elkton BoulevardElkton, MD 21921Phone #: (410) 392-7321Fax #: (410) 398-6079

E-mail: philmichael.lester@terumomedical.com

··

.

・ ・・

.

.

·

14

·

:

{1}------------------------------------------------

K122249

A. DEVICE NAME

Proprietary Name

TERUMO® SurGuard®3 Safety Hypodermic Needle TERUMO® SurGuard®3 Hypodermic Syringe with Safety Needle

Classification Name

Hypodermic Single Lumen Needle (880.5570) with antistick

Product Code: 80FMI / 80MEG

Classification: Class II

Common Name

Hypodermic Needle with Safety Sheath or Needle with needle protection device.

B. INTENDED USE

The Terumo SurGuard®3 Safety Hypodermic Needle device is intended for use in the aspiration and injection of fluids for medical purposes. The Terumo SurGuard 3 Safety Needle is compatible for use with standard luer slip and luer lock syringes.

Additionally, after withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

Note: This is the same intended use as the predicate device, Terumo SurGuard®3 Safety Hypodermic Needle.

C. DEVICE DESCRIPTION

The Terumo SurGuard 3 Safety Hypodermic Needle consists of a hypodermic needle with a hinged safety sheath attached to the connector hub with or without an attached hypodermic syringe. This device features a hinged safety sheath attached to the needle hub. The safety sheath contains two locking mechanisms, the tooth-cannula and sheath-collar which are simultaneously activated when

{2}------------------------------------------------

K12249

manually pressed over the needle immediately after use and just prior to disposal to minimize the possibility of sharps injury. The safety sheath is activated with a one-handed operation by pressing the sheath either by finger, thumb or surface activation.

The safety sheath has a finger guide area consisting of a circular dent (for thumb activation) and ramp (for finger activation) which gives the user a concrete confirmation that the user's finger is in the appropriate position in performing the activation. The ramp has steps to provide strong grasp when activating the sheath, There are two stoppers located at the end of the circular dent and ramp which prevent the user's finger from going towards the cannula during activation. Another method of activation is by manually pressing the safety sheath over the needle using a firm surface.

The locking mechanisms are positioned within the center and proximal end of the sheath. The hinge feature allows the medical practitioner to flexibly adjust the sheath to its desired position for use.

The Terumo SurGuard 3 Safety Hypodermic Needle will be individually packaged and sterilized by electron beam as a safety needle only or as a safety needle with attached Terumo syringe.

D. SUBSTANTIAL EQUIVALENCE

The Terumo SurGuard®3 Safety Hypodermic Needle and with attached syringe submitted in this 510(k) and manufactured by Terumo (Philippines) Corporation is substantially equivalent to:

    1. K113422 TERUMO® SurGuard® 3 Safety Hypodermic Needle with or without syringe manufactured by Terumo (Philippines) Corporation.

E. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS

Principle of Operation and Technology

The Terumo SurGuard 3 Safety Hypodermic Needle device with and without syringe manufactured by Terumo (Philippines) Corporation, subject of this 510(k), and the predicate devices covered under K113422 are operated manually.

ા ર

{3}------------------------------------------------

K122249
page 4 of 10

Materials

All materials for the Terumo SurGuard®3 Safety Hypodermic Needle and with attached syringe are the same as the materials of the SurGuard3 as submitted in the 510(k), K113422. The 1 cc/ml syringe plunger is made up of polystyrene, while the 3, 5, and 10 cc/ml syringes utilize polypropylene for the plunger; however, both polystyrene and polypropylene were cleared for use in the plunger under K113422.

Specifications

The new needle gauge sizes to be added under this 510(k) are 26G, 27G, 30G and the new needle lengths are 3/8", 5/8", and 1/2". The syringe size to be added is 1cc/ml, as K113422 only provided for 3, 5, and 10 cc/ml syringe and needle combinations. This submission also provides for new needle gauge and length combinations for 23G and 25G needles and 1" needle length which were cleared under K113422. Finally, this submission includes the shorter safety sheath with modified design for all new gauge and needle length combinations.

The proposed Terumo SurGuard®3 Safety Hypodermic Needle comes with a shorter safety sheath. The inside rib design of the sheath was modified, specifically extending rib 2 towards the inside wall of the sheath and modifying rib 3 to be symmetrical. This design allows the cannula to slide towards the center of the sheath during the activation.

For further clarification of the reason for this submission and differences from the previous 510(k), Tables 1 and 2 on the following page are included here. Table 1 shows the needle gauges x needle length combinations both new for this submission and those cleared under previous 510(k), K113422. Table 2 demonstrates gauge and needle length combinations and the syringe size specific to each needle gauge x needle length. "X" demonstrates clearance under previous 510(k), K113422, while items new for this submission are denoted by "New."

{4}------------------------------------------------

K122244

page

ਸਿੰ

. .

.

്ര

Terumo (Philippines) Corporation
SurGiaard3 Safety Needle
Special 510k
Section II 510k
Section II 510(k) Summary

TABLE 1Needle Length
3/8"יי2/15/8"] **1 1/2"2"
(9mm)(13mm)(16mm)(25mm)(38mm)(51mm)
18GXX
19GXX
20GXXX
21GXX
Gauge22GXX
Size23GNewX
25GNewNewX
26GNewNew
27GNew
30GNew

Table 1 – Needle Gauges x Needle Length

Table 2 - Needle Length and Gauge x Syringe

TABLE 2Syringe Size
1 cc/ml3 cc/ml5 cc/ml10 cc/ml
Needle Gauge xNeedle Length18G x 1"
18G x 1 ½"
19G x 1"
19G x 1 ½"
20G x 1"XXX
20G x 1 ½"XXX
20G x 2"
21G x 1"XX
21G x 1 ½"XX
22G x 1"X
22G x 1 ½"X
23G x 1"New
23G x 1 ½"
25G x 5/8"NewNew
25G x 1"New
25G x 1 ½"
26G x 3/8"New

.

{5}------------------------------------------------

TABLE 2Syringe Size
1 cc/ml3 cc/ml5 cc/ml10 cc/ml
Needle Gauge xNeedle Length26G x ½"
27G x ½"New
30G x ½"

1777499

ਹੰਦਿ

IO

The above cleared devices (indicated as "X") serve as predicate for the devices that are subject of this 510(k). '

Product CodeDescription
SG3-232523gauge x 1" (25mm) safety hypodermic needle
SG3-251625gauge x 5/8" (16mm) safety hypodermicneedle
SG3-252525gauge x 1" (25mm) safety hypodermic needle
SG3-260926gauge x 3/8" (9mm) safety hypodermic needle
SG3-261326gauge x 1/2" (13mm) safety hypodermicneedle
SG3-271327gauge x 1/2" (13mm) safety hypodermicneedle
SG3-301330 gauge x 1/2" (13mm) safety hypodermicneedle
SG3-01T251625gauge x 5/8" (16mm) safety hypodermicneedle, with 1cc/ml syringe
SG3-01T260926gauge x 3/8" (9mm) safety hypodermic needle,with 1cc/ml syringe
SG3-01T271327gauge x 1/2" (13mm) safety hypodermicneedle, with 1cc/ml syringe
SG3-03L232523gauge x 1" (25mm) safety hypodermic needle,with 3cc/ml syringe
SG3-03L251625gauge x 5/8" (16mm) safety hypodermicneedle with 3cc/ml syringe
SG3-03L252525gauge x 1" (25mm) safety hypodermic needlewith 3cc/ml syringe

Specifications for product proposed in this 510(k)

{6}------------------------------------------------

Performance

The following tests were performed on the Terumo SurGuard®3 Safety Hypodermic Needle and with attached syringe.

Performance Testing

These tests were performed on both predicate device K113422 and the proposed device:

  • . Sheath Activation
  • Sheath Deactivation .
  • Sheath Removal �
  • Sheath Radial .
  • Sheath Retraction .
  • Protector Fitting .
  • Collar removal .
  • Sheath Puncture �
  • Adhesive Hold .
  • Needle Penetration �
  • Manual Sheath Activation
    • -Finger Activation
    • Thumb Activation -
    • Surface Activation .
  • Measurement Testing .
  • Manual Sheath Activation by analytical balance .
  • . Simulated Use Study

This study was performed to confirm that the safety feature of the proposed device can be safely activated by following the Instructions For Use of the device. The study was conducted in a simulated clinical environment and the safety feature activation was performed by healthcare workers with varied amounts of experience. It was concluded that the proposed Terumo SurGuard 3 Safety Hypodermic Needle met its intended use, and the Instructions for Use were adequate for safe and effective activation of the safety feature.

Results of the verification testing performed on both the predicate device and the proposed device met the acceptance criteria of the functional performance tests of both devices.

20

{7}------------------------------------------------

Shelf Life Testing

Shelf life tests both on the predicate device (K113422) and the proposed Terumo SurGuard 3 Safety Hypodermic Needle and with attached syringe (1cc/ml) were performed using accelerated aged samples to simulate a 5-year (60 months) shelf life for this 510(k) submission. The accelerated aging condition was performed in accordance with ASTM F-1980.

  • Sheath Activation .
  • Sheath Deactivation .
  • Sheath Removal .
  • . Sheath Radial
  • . Sheath Retraction
  • Protector Fitting .
  • . Collar removal
  • Sheath Puncture .
  • Adhesive Hold .
  • Needle Penetration .
  • Manual Sheath Activation
    • Finger Activation -
    • Thumb Activation
    • Surface Activation -
  • . Package Seal Strength

Results of verification testing indicated that the functional performance of the proposed Terumo SurGuard®3 Safety Hypodermic Needle and with attached syringe (1cc/ml) is acceptable throughout a 5-year (60months) shelf life. The shelf-life will be confirmed with real time testing.

Biocompatibility Testing

Biocompatibility was established for the Terumo SurGuard®3 Safety Hypodermic Needle manufactured by Terumo (Philippines) Corporation (K113422) which is comprised of blood contacting materials identical to the Terumo SurGuard®3 Safety Hypodermic Needle (26G, 27G and 30G) and with attached syringe (1cc/ml) device. No changes have been made since 510(k) clearance that would require re-testing; therefore, biocompatibility established under K113422 is still applicable.

{8}------------------------------------------------

The Terumo SurGuard®3 Safety Hypodermic Needle is classified as Externally Communicating Device, Circulating Blood, Limited Duration of Contact (≤24hr.). The device's blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO - 10993, "Biological Evaluation of Medical Devices part 1: Evaluation and Testing". Results of the testing demonstrate that the blood contacting materials are biocompatible.

None of the data raises any new issues of safety and effectiveness.

Product Risk Analysis

Additionally a product risk analysis was conducted and there were no new issues of safety and effectiveness.

Conclusion: The performance of the TERUMO SurGuard®3 Safety Needle submitted in this 510(k) is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the legally marketed predicate devices. The proposed devices perform as well as the predicate devices.

F. COMPARISON TO STANDARDS

The Terumo SurGuard®3 Safety Hypodermic Needle complies with the requirements specified in the following standards:

  • ISO 23908:2011 Sharps Injury Protection, requirements and test methods . - sharps protection features for single hypodermic needles, introducers for catheters, and needles used for blood sampling
  • ISO 7886-1:1993 Sterile Hypodermic Syringes for Single Use .
  • ISO 594-1:1986 Conical Fittings with a 6% taper on syringes, needles, . and certain other medical equipment
  • ISO 6009:1992/Cor:1:2008 Hypodermic Needles for Single Use color . coding for identification

The Terumo SurGuard 3 Safety Hypodermic Needle also complies with the requirements specified in ISO 7864:1993 - Sterile Hypodermic Needles for Single

{9}------------------------------------------------

K122249
page 10 of 10

Use, with the exception of:

Section 13.2a Patency of Lumen: no specifications are given for 26G, 27G and 30G having an ultra thin wall cannula. As a result, these needles cannot be compared to this requirement.

G. ADDITIONAL SAFETY INFORMATION

Manufacturing controls include visual, functional and sterility tests.

The sterility of the device is assured using a sterilization method validated in accordance with ANSI/AAMI/ISO 11137 - Medical Devices - Validation and Routine Control of Radiation Sterilization. The Terumo SurGuard®3 Safety Hypodermic Needle is sterilized to provide a Sterility Assurance Level (SAL) of 10-6.

A specific sterilization dose to provide Sterility Assurance Level (SAL) of 100 is set using bioburden information obtained.

H. CONCLUSION

The performance of the TERUMO® SurGuard®3 Safety Needle submitted in this 510(k) is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the legally marketed predicate device which is:

    1. K113422 TERUMO® SurGuard®3 Safety Hypodermic Needle with or without syringe manufactured by Terumo (Philippines) Corporation.
      Differences among the devices do not raise any significant issues of safety or effectiveness.

Terumo's statement of substantial equivalence is done solely to comply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended whatsoever to be the basis for a patent infringement action.

{10}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

Terumo (Philippines) Corporation C/O Mr. Phillip Michael Lester Regulatory Affairs Specialist Terumo Medical Corporation 950 Elkton Boulevard Elkton, Maryland 21921

AUG 28 2012

Re: K122249

Trade/Device Names: TERUMO® SurGuard®3 Safety Needle, 23G, 25G, 26G, 27G and 30G; TERUMO® SurGuard® 3 Hypodermic Syringe with Safety Needle, Icc/mL Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI, MEG Dated: July 25, 2012 Received: July 30, 2012

Dear Mr. Lester:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{11}------------------------------------------------

Page 2- Mr. Lester

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For. S. Smith D.

Anthony B. Watson, BS, MS, MBA Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{12}------------------------------------------------

22249

Indications for Use

510(k) Number (if known):

Device Name: TERUMO® SurGuard®3 Safety Needle, 23G, 25G, 26G, 27G and 30G TERUMO® SurGuard®3 Hypodermic Syringe with Safety Needle, 1cc/mL

Indications for Use:

The TERUMO® SurGuard®3 Safety Needle device is intended for use in the aspiration and injection of fluids for medical purposes. The TERUMO SurGuard®3 Safety Needle is compatible for use with standard luer slip and luer lock syringes.

Additionally, after withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Alid Chapman 8/27/12

(Division Sign-Off) (Division of Anesthesiology, General Hospital nfection Control, Dental Devices

510(k) Number: K122249

Page 1 of 1

4

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).