K Number
K152271

Validate with FDA (Live)

Date Cleared
2016-02-09

(181 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Nitrile Patient Examination Gloves, Black Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Nitrile films form a barrier to body fluids and bloodborne Pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

AI/ML Overview

The provided document is a 510(k) summary for "Powder Free Nitrile Patient Examination Gloves, Black Color". It details the device's characteristics and compares it to a predicate device to establish substantial equivalence.

Here's the information about the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

CharacteristicsStandard (Acceptance Criteria)Device Performance
DimensionASTM standard D 6319-10 (e.g., Length ≥230mm min for all sizes)Meets (e.g., Length 230mm min for all sizes, Width values are provided and specific)
Physical PropertiesASTM standard D 6319-10 (e.g., Elongation ≥500%, Tensile Strength ≥14MPa)Meets (e.g., Elongation: 520-620%, Tensile Strength: 20-36 MPa)
Freedom from Pinholes21 CFR 800.20, ASTM D 5151-06 (Reapproved 2011) (Inspection Level I, AQL 2.5)Meets (Holes at Inspection Level I, AQL 2.5)
Powder ResidualASTM standard D 6319-10 and D6124-06 (Reapproved 2011) (< 2mg/glove)Meets (< 2mg/glove; Results generated values below 2mg of residual powder)
BiocompatibilityPrimary Skin Irritation in rabbits (ISO 10993-10: Third Edition 2010-08-01)Passes (non-irritant)
BiocompatibilityDermal sensitization in the guinea pig (ISO 10993-10: Third Edition 2010-08-01)Passes (non-sensitizer)

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample size used for each test. It refers to compliance with ASTM standards, which would implicitly define sample sizes for those tests.

The data provenance is from China as the submitter, Suqian Xingye Glove Co, Ltd, is located in Suqian City, Jiangsu Province, China. The studies are retrospective, as they involve testing of a manufactured device according to established standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The "ground truth" for these types of devices is objective measurement against specified standards, rather than expert interpretation. For biocompatibility, it's typically conducted by trained laboratory personnel following standard protocols.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Adjudication methods such as 2+1 or 3+1 are typically used in clinical studies involving human interpretation or subjective assessment. This document describes objective performance testing against established ASTM and ISO standards for physical properties, barrier integrity, and chemical content. Therefore, no adjudication method as described would be applicable or necessary. The standard itself defines the pass/fail criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices involving human interpretation, especially with AI assistance. The device in question is a patient examination glove, and its effectiveness is determined by its physical and chemical properties, not by human reader performance or AI collaboration.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not applicable to the device. The device is a physical medical glove, not an algorithm or software. Its performance is evaluated through physical and chemical laboratory testing as described by the standards.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the test set is established by objective, quantitative standards and specified test procedures defined by:

  • ASTM standard D 6319-10 (for Dimension and Physical Properties)
  • 21 CFR 800.20 and ASTM D 5151-06 (Reapproved 2011) (for Freedom from Pinholes)
  • ASTM standard D 6319-10 and D6124-06 (Reapproved 2011) (for Powder Residual)
  • ISO 10993-10: Third Edition 2010-08-01 (for Biocompatibility – Primary Skin Irritation and Dermal Sensitization)

These standards specify the methodology and acceptance limits for each characteristic.

8. The sample size for the training set

This information is not applicable. The device is a medical glove, not an AI or machine learning model that requires a training set. The manufacturing process for gloves is subject to quality control and process validation, not a "training set" in the context of data-driven systems.

9. How the ground truth for the training set was established

This information is not applicable, as there is no "training set" for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 9, 2016

Suqian Xingye Glove Co, Ltd % Chu Xiaoan Official Correspondent Beijing Easylink Co. Ltd Rm. F302 Bldg. 41 Jing Cheng Ya Ju Courtyard 6 of Southern Dou Ge Zhuang Beijing, 100121 CHINA

Re: K152271

Trade/Device Name: Nitrile Powder Free Patient Examination Gloves, Black Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: January 03, 2016 Received: January 11, 2016

Dear Chu Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Chu Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin Keith, M.S. Division Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K152271

Device Name

Powder Free Nitrile Patient Examination Gloves, Black Color

Indications for Use (Describe)

Powder Free Nitrile Patient Examination Gloves, Black Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 6 510(k) Summary

510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: K152271 =

Premarket Notification [510(k)] Summary

1.0 Submitter:

Submitter's name :Suqian Xingye Glove Co,Ltd
Submitter's address :Dongwu Road,Economic DevelopmentZone,Suqian City,JiangsuProvince,223800,China
Phone number :0086-527-82860533
Fax number :0086-527-82860080
Name of contact person:Jian Zhong Deng
Date of preparation :2016-01-03

Name of the Device 2.0

Device Name:Powder Free Nitrile Patient ExaminationGloves, Black Color
Proprietary/Trade name:Powder Free Nitrile Patient ExaminationGloves, Black Color
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LZA

3.0 Predicate device

Device Name:Black Nitrile Powder Free Exam Glove
Company name:Shandong Yuanhang Medical Products Co., Ltd.
510(K) Number:K133307

4.0 Device Description:

How the device functions: 4.1 Nitrile films form a barrier to body fluids and bloodborne Pathogens

{4}------------------------------------------------

Scientific concepts that form the basis for the device 4.2

The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

Physical and performance characteristics such as design, materials and 4.3 physical properties:

Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

5.0 Device Intended Use (Indication for use):

Powder Free Nitrile Patient Examination Gloves, Black Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6.0 Summary of the Technological Characteristics of the Device:

The Powder Free Nitrile Patient Examination Gloves, Black Color, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 6319-10.Meets
Physical PropertiesASTM standard D 6319-10.Meets
Freedom from pinholes21 CFR 800.20 ASTM D 5151-06 (Reapproved 2011)Meets
Powder ResidualASTM standard D 6319-10 and D6124-06(Reapproved 2011).Meets<2mg/glove
BiocompatibilityPrimary Skin Irritation in rabbitsISO 10993-10: Third Edition2010-08-01.PassesUnder the conditions ofthis study, the test article wasa non-irritant
Dermal sensitization in theguinea pigISO 10993-10: Third Edition2010-08-01.PassesUnder the conditions ofthis study, the test article wasa non-sensitizer

{5}------------------------------------------------

7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:

Powder Free Nitrile Patient Examination Gloves, Black Color, meet requirements per ASTM D6319-10.per ASTM D6124-06(Reapproved 2011), per 21 CFR 800.20 and ISO 10993-10: Third Edition 2010-08-01.

The performance test data of the non clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data is not needed for the subject device.

9.0 Substantial Equivalence Comparison:

Features &DescriptionPredicate DeviceSubject DeviceResult ofComparison
CompanyShandong YuanhangMedical Products Co.,Ltd.Suqian Xingye GloveCo,Ltd--
510(K) NumberK133307
Product nameBlack Nitrile PowderFree Exam GlovePowder Free NitrilePatient ExaminationGloves, Black Colorsame
Product CodeLZALZAsame
SizeSmall/ Medium/Large/X largeSmall/ Medium/Large/X largesame
Indications foruseA disposable deviceintended for medicalpurposes that is wornon the examiner's handor finger to preventcontamination betweenpatient and examiner.Powder Free NitrilePatient ExaminationGloves, Black Color is adisposable deviceintended for medicalpurposes that is worn onthe examiner's hand orfinger to preventcontamination betweenpatient and examiner.Substantiallyequivalent
DeviceDescription andSpecificationsMeets ASTMD6319-10Meets ASTMD6319-10Substantiallyequivalent
Dimensions--LengthMeets ASTMD6319-10$≥230$ mm min230mm min for all sizesSubstantiallyequivalent
Dimensions-- WidthMeets ASTMD6319-10Substantiallyequivalent
Small 70-90 mmMedium 85-105mmLarge 100-120mmXlarge 110-130 mmSmall 82-86 mmMedium 94-98mmLarge 105-109mmX large 114-120 mm
DimensionsMeets ASTMSubstantially
--ThicknessD6319-10Thickness (mm) min.equivalent
Finger 0.05mm min.Finger 0.10-0.12
Palm 0.05mm min.Palm0.06-0.09
PhysicalMeets ASTM DSubstantially
PropertiesD6319-10equivalent
Before aging/after agingBefore aging/after aging
Elongation ≥500%Elongation :520-620%
Tensile Strength≥14MPaTensile Strength:20-36 MPa
FreedomfromMeets●Meets ASTMD5151-06Substantiallyequivalent
Pinholes21 CFR 800.20ASTM D6319-10●(Reapproved 2011)
ASTM D 5151-06●
(Reapproved 2011)Holes at
Inspection Level IAQL2.5
ResidualMeets ASTMMeets ASTMSubstantially
PowderD 6124-06D 6124-06equivalent
(Reapproved 2011)(Reapproved 2011)
below 2mg of residualResults generated values
powderbelow 2mg of residual
powder
Materials usedNitrileNitrileSubstantially
to fabricate theequivalent
devicesDusting orPUPU-120CSubstantially
Donningequivalent
Powder:
Dusting orSurface Coating AgentSurface Coating AgentSubstantially
DonningPowder: nameequivalent
CompareMeetsMeetsSubstantially
performanceASTM D5151-06ASTM D5151-06equivalent
data supporting(Reapproved 2011)(Reapproved 2011)
substantialequivalenceASTM D6319-10ASTM D6124-06ASTM D6319-10ASTM D6124-06
(Reapproved 2011)(Reapproved 2011)
Single Patient UseSingle Patient UseSingle Patient UseSubstantially
equivalent
BiocompatibilityUnder the conditions ofthis study, the test articleUnder the conditions ofthis study, the test articleSubstantially
was a non- irritant or non-was a non- irritant or non-equivalent
sensitizer.sensitizer.
SKIN IRRITATIONIRRITATIONSKIN
DERMAL andDERMALand
SENSITIZATIONSENSITIZATION
STUDIES Meets ISOSTUDIES MeetsISO
10993-10:2002/Amd.1:200610993-10: Third Edition2010-08-01.
Labeling for the-Powder Free-Powder FreeSubstantially
legally marketed-Patient Examination-Patient Examinationequivalent
device to whichGloveGlove
substantial-Single Use Only-Single Use Only
equivalence isclaimed.- Manufactured For:- Lot- Manufactured For:- Lot
-Black color-Black color
- Non sterile- Non sterile

{6}------------------------------------------------

{7}------------------------------------------------

10.0 Substantial Equivalence Comparison:

It can be concluded that the Powder Free Nitrile Patient Examination Gloves, Black Color meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims.

It can be concluded that the Powder Free Nitrile Patient Examination Gloves, Black Color is as safe, as effective, and performs as well as the predicate device, Black Nitrile Powder Free Exam Glove, Shandong Yuanhang Medical Products Co., Ltd. K133307.

The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.