K Number
K152232

Validate with FDA (Live)

Date Cleared
2016-11-18

(469 days)

Product Code
Regulation Number
862.3250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Quest Diagnostics HairCheck-DT (Cocaine) test system utilizes an Enzyme Linked Immunosorbent Assay (ELISA) for the qualitative detection of cocaine in head hair samples through the measurement of cocaine and cocaine metabolites for concentrations at or above 300 pg/mg hair. This test system has not been evaluated for use with hair specimens from locations other than the head. It is an in vitro diagnostic device intended exclusively for in-house professional use and is not intended for sale to anyone.

The HairCheck-DT (Cocaine) test system was evaluated in two distinct study populations; individuals known to be chronic drug abusers, and individuals proclaiming to be drug-free.

The Quest Diagnostics HairCheck-DT (Cocaine) test system provides only a preliminary analytical test result. To confirm a presumptive screen positive result, a more specific alternate chemical method, such as gas chromatography-mass spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS), must be used. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

Device Description

The Quest Diagnostics Hair Check-DT (Cocaine) test system utilizes an Enzyme Linked Immunosorbent Assay (ELISA) for the qualitative detection of cocaine in head hair samples through the measurement of cocaine and cocaine metabolites for concentrations at or above 300 pg/mg hair. This test system has not been evaluated for use with hair specimens from locations other than the head. It is an in vitro diagnostic device intended exclusively for in-house professional use and is not intended for sale to anyone.
The ELISA Cocaine Kit is based on the competition for a limited number of antibody sites by unlabeled cocaine/cocaine metabolites and enzyme-labeled drug. The two will bind to the antibody in proportion to their concentration in solution.
Once accessioned in the lab, the aluminum foil is opened and the specimen is cut at approxima 3.9 cm from the root end. This specimen is cut into smaller lengths and mixed to ens homogeneity. Ten milligrams of the specimen is weighed out and placed into a properly labeled test tube. The specimen is then washed with methanol, decanted, and then placed in hot methanol containing 0.5% (v/v) trifluoroacetic acid for one hour forty-five minutes. The extracted methanol solution is then transferred to a new tube and evaporated under nitrogen. The tubes are reconstituted with phosphate buffer and assayed using the Cocaine ELISA Kit. This kit is a solid-phase microtiter plate immunoassay in which the microwells are coated with a high affinity capture antibody to cocaine. A hair sample extract is added to the well, followed by the horseradish peroxidase (HRP) enzyme conjugate. During this initial phase, the enzyme conjugate competes with the analyte in the sample for binding sites on the antibody-coated microwells. A wash solution (Tween-20 in phosphate buffered saline solution) is then applied to remove any unbound materials such as excess conjugate and residual sample. Enzyme substrate solution containing 3, 3', 5, 5'-tetramethylbenzidine (TMB) is then added for the final color development process. The reaction is stopped with 1N sulfuric acid and the absorbance is read at 450 nm, with a reference wavelength of 620 nm, using a plate reader. Color intensity is inversely proportional to the amount of analyte present in the sample. Therefore, samples that contain drug or analyte will inhibit binding of the enzyme conjugate to the antibody, resulting in little substrate binding and less color development than in the negative calibrator. For the screening assay an absorbance less than or equal to the absorbance of the 300 pg cocaine/mg hair cutoff calibrator is indicative of the presence of cocaine/cocaine metabolites.

AI/ML Overview

The Quest Diagnostics HairCheck-DT (Cocaine) test system is an in vitro diagnostic device that uses an Enzyme Linked Immunosorbent Assay (ELISA) for the qualitative detection of cocaine and cocaine metabolites in head hair samples at concentrations at or above 300 pg/mg hair. A positive result from this screening test requires confirmation using a more specific alternate chemical method, such as gas chromatography-mass spectrometry (GC/MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS).

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Study GoalReported Device Performance (Quest Diagnostics HairCheck-DT (Cocaine))
Precision/Reproducibility
Overall CV%Less than 10% for all spiked levels (25%, 50%, 100%, 125%, 150%, 175%, and 200%) of analyte.
Separation around cutoff (50% vs 100% and 100% vs 150%)The mean Ratio (sample OD/cutoff OD) minus 2 SD of the 50% samples was greater than the mean Ratio plus 2 SD of the 100% sample on each day. The mean Ratio minus 2 SD of the 100% sample was greater than the mean Ratio plus 2 SD of the 150% sample on each day. This indicates clear separation around the cutoff.
No crossover ±50% of cutoffDemonstrated no crossover ±50% of the cutoff concentration.
Cross-Reactivity
Structurally Related CompoundsCross-reactivity ranged from <0.1% to 200%. Compounds exhibiting highest cross-reactivity (benzoylecgonine, cocaethylene, meta-hydroxybenzoylecgonine) are cocaine metabolites, which is an expected and desired outcome.
Unrelated Compounds (Interference)136 out of 143 tested compounds (common medications, abused drugs, hair dyeing compounds) showed no significant reactivity/interference.
Limited Interference17 compounds (e.g., Thioridazine, Chlorpromazine, specific basic dyes) exhibited a limited degree of positive interference, primarily at very high concentrations (e.g., 75,000 pg/mg to 15,000,000 pg/mg equivalent in hair).
Method Comparison
Overall Accuracy% Agreement among positive: 100% (50/50)
% Agreement among negative: 92% (46/50)
Discrepant Results4 discrepant positive results out of 50 negative by GC-MS. One was due to pink hair dye interference, and three were near cutoff negative samples (155, 190, and 272 pg/mg) by GC-MS, which screened positive by ELISA. The 272 pg/mg sample also contained 80 pg/mg of benzoylecgonine.
Within-Run Specimen Extraction Reproducibility
Positive Sample ConsistencyEach replicate of five individual positive samples consistently tested positive by both ELISA and GC-MS.
Within-run CV (ELISA OD)<10% for all five hair specimens, with an overall mean CV of 3.9%.
Within-run CV (GC-MS)6.7% for cocaine.
Specimen Shipping Stability
Positive Sample Status28/28 positive samples remained positive by both ELISA and GC-MS after shipping.
Negative Sample Status28/28 negative samples remained negative by GC-MS after shipping.
Borderline Sample Status1 of 3 near cutoff negative samples screened positive by ELISA prior to shipping. 3 of 3 near cutoff negative samples screened positive by ELISA post-shipping, attributed to variability near the ELISA cutoff.
Cosmetic Hair Treatment Studies
Positive Hair (ELISA)Cocaine positive hair samples (60 samples) remained positive by ELISA after various treatments (shampoo, brown dye, bleach, perm, relaxer). Some treatments (bleach, relaxer) resulted in substantial loss of cocaine by GC-MS, and varied % change in ELISA OD.
Negative Hair (ELISA)Cocaine negative hair samples (60 samples) remained negative by ELISA after various treatments. Minimal variability in OD changes was observed for negative specimens.
Impact of TreatmentsSome treatments can reduce detected drug amounts; bleach and relaxer showed the most significant effect on positive specimens. None significantly impacted negative specimens.
Shelf Life (Real Time Stability)
Kit Shelf Life3 evaluation lots stable up to 7-8 months, establishing a 6-month shelf-life.
Calibrators and Controls3 lots stable up to 13 months, meeting QC criteria (+/-20% of target value).
Open Vial StabilityAll components showed stability after open vial conditions (e.g., microplates at room temp for 4 hours, then 30 days at 2-8°C; calibrators/controls at 2-8°C for 4 hours, then 30 days at -20°C).

2. Sample Size and Data Provenance for Test Set

  • Precision/Reproducibility: 9 samples/levels tested in quintuplicate (5 replicates) over 10 days, resulting in 50 measurements per level (9 levels: 0%, 25%, 50%, 75%, 100%, 125%, 150%, 175%, 200% of cutoff). Total of 450 measurements for precision. Data provenance is implied to be laboratory-prepared spiked samples, origin not specified.
  • Cross-Reactivity:
    • Structurally Related Compounds: 11 compounds tested. Serial dilutions were used to determine the concentration producing an OD akin to the cutoff. Specific number of test samples per compound not detailed, but implies multiple tests per compound across dilutions.
    • Interference (Unrelated Compounds): 143 compounds tested. Each spiked into low control (150 pg/mg) and high control (450 pg/mg) negative hair matrix. 10 hair dyeing compounds evaluated starting at 500 µg/mL. This suggests at least 2 tests per compound (low control spike + high control spike).
  • Method Comparison: 100 donor specimens (50 negative, 50 positive, with specific ranges near cutoff). Data provenance indicates these were "routine donor hair specimens provided by the Quest Diagnostics testing laboratory," previously confirmed by GC-MS, and eligible for discard. This implies retrospective clinical data from unspecified geographic origins.
  • Within-Run Specimen Extraction Reproducibility: 5 cocaine positive specimens, each tested in 3 replicates.
  • Specimen Shipping Stability: 56 hair samples (28 confirmed positive, including 3 near cutoff; 28 confirmed negative, including 3 near cutoff). These were shipped to 8 different geographical locations. This represents a prospective study with controlled shipping conditions.
  • Cosmetic Hair Treatment Studies: 120 hair samples (60 confirmed positive by GC-MS, 60 screened negative by ELISA). These were subjected to 5 different cosmetic hair treatments or left untreated. Specific breakdown by treatment group is 12 positive and 12 negative samples per treatment type (shampoo, brown dye, bleach, perm, relaxer, and untreated controls implicitly for the ELISA results where %change is noted relative to pre-treatment of the same sample).

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the number of experts or their qualifications for establishing ground truth. However, for the Method Comparison study and other stability/reproducibility studies comparing results against a definitive method, the ground truth relies on confirmation by GC-MS (Gas Chromatography-Mass Spectrometry). This is a highly sensitive and specific analytical technique commonly used as a gold standard in toxicology. The interpretation and operation of GC-MS would typically be performed by highly trained analytical chemists or toxicologists in a certified laboratory setting, but specific expert qualifications are not provided.

For the cosmetic hair treatment study, "Cocaine positive hair is defined as a hair sample confirmed by GC-MS as having greater than 300 pg/mg cocaine." This again points to GC-MS as the ground truth.

4. Adjudication Method for Test Set

The document does not describe an adjudication method involving human experts for the test set. Instead, the device's performance is compared directly against the results obtained from a reference method, specifically GC-MS. Discrepancies (e.g., in the Method Comparison study) are analyzed and explained by the submitting company, not adjudicated by a panel of independent experts in the typical clinical sense. For example, specific discrepancies are attributed to pink hair dye interference or samples being near the cutoff.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study was mentioned. This device is an in vitro diagnostic (ELISA), not an imaging or assistive technology that human readers would use. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" does not apply.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire document describes the performance of the algorithm/device only (the HairCheck-DT (Cocaine) ELISA test system) without human-in-the-loop performance modifications beyond the standard operational procedures for running an ELISA and interpreting its results against a cutoff. The device provides "only a preliminary analytical test result," and "to confirm a presumptive screen positive result, a more specific alternate chemical method, such as gas chromatography-mass spectrometry (GC/MS) must be used." This clearly indicates its standalone screening role.

7. Type of Ground Truth Used

The primary ground truth used for evaluating the device's performance is quantitative Gas Chromatography-Mass Spectrometry (GC-MS) analysis. This is considered a highly reliable and definitive analytical method in toxicology for confirming the presence and concentration of substances. In some cases, outcomes data or specifically defined reference material concentrations act as the ground truth (e.g., for precision, cross-reactivity with spiked samples).

8. Sample Size for the Training Set

The document provided does not mention a "training set" or "training data" in the context of machine learning, as this device is a chemical immunoassay (ELISA), not an AI/ML algorithm that requires training data in that sense. The studies described are performance validation studies for the immunoassay.

9. How the Ground Truth for the Training Set Was Established

As noted in point 8, there is no "training set" in the machine learning sense. For the development and establishment of the immunoassay, the "ground truth" for calibrators and controls would be established by preparing solutions of known concentrations of cocaine and its metabolites using certified reference materials, and comparing against a reference method like GC-MS during the assay's development. The document mentions that "40X Calibrator and Control solutions are prepared from Cerilliant C-008, 1.0 mg/mL Cocaine in Acetonitrile Certified Reference Material." This demonstrates the rigorous process of establishing known concentrations for quality control and calibration purposes.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, line-art style. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

QUEST DIAGNOSTICS INCORPORATED November 18, 2016 LISA CHRISTO QUALITY ASSURANCE AND REGULATORY AFFAIRS MANAGER 10101 RENNER BLVD. LENEXA KS 66219-9275

Re: K152232

Trade/Device Name: Ouest Diagnostics HairCheck-DT (Cocaine) Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: II Product Code: DIO Dated: November 14, 2016 Received: November 15, 2016

Dear Ms. Christo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Katherine Serrano -S

For: Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152232

Device Name

Quest Diagnostics HairCheck-DT (Cocaine)

Indications for Use (Describe)

Intended Use

The Quest Diagnostics Hair Check-DT (Cocaine) test system utilizes an Enzyme Linked Immunosorbent Assay (ELISA) for the qualitative detection of cocaine in head hair samples through the measurement of cocaine metabolites for concentrations at or above 300 pg/mg hair. This test system has not been evaluated for use with hair specimens from locations other than the head. It is an in vitro diagnostic device intended exclusively for in-house professional use and is not intended for sale to anyone.

The Hair Check-DT (Cocaine) test system was evaluated in two distinct study populations; individuals known to be chronic drug abusers, and individuals proclaiming to be drug-free.

The Quest Diagnostics Hair Check-DT (Cocaine) test system provides only a preliminary analytical test result. To confirm a presumptive screen positive result, a more specific alternate chemical method, such as gas chromatography mass spectrometry (GC/MS) must be used. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Quest Diagnostics logo. The logo consists of a stylized "Q" symbol on the left and the words "Quest Diagnostics" on the right. The word "Quest" is in a larger, bolder font than the word "Diagnostics".

1.0 510(k) Summary

ApplicantQuest Diagnostics Incorporated10101 Renner Blvd.Lenexa, KS 66219-9275USA
Date Prepared11/17/2016
Primary ContactLisa ChristoTel 913-577-1784Lisa.x.Christo@questdiagnostics.com
Alternate ContactR. H. Barry Sample, Ph.D.Tel 770-800-9870Barry.x.Sample@questdiagnostics.com
Proprietary NameQuest Diagnostics HairCheck-DT (Cocaine)
Generic NameEnzyme Immunoassay, Cocaine And Cocaine Metabolites
US Product CodeDIO – Enzyme Immunoassay, Cocaine and Cocaine Metabolites
US Regulation Number862.3250
ClassificationUnited States Class II
CLIA ComplexityHigh
AnalyteCocaine
Special Instrument RequirementsThe device is for use with an automated microplate reader capable of measuring at 450 and 620 nm.
Specimen CollectionHead Hair
Predicate DeviceQuest Diagnostics HairCheck-DT (Cocaine) (K023626)
Reference MethodPresumptive positives are confirmed with Quantitative GC-MS
Device DescriptionThe Quest Diagnostics Hair Check-DT (Cocaine) test system utilizes an Enzyme LinkedImmunosorbent Assay (ELISA) for the qualitative detection of cocaine in head hair samples throughthe measurement of cocaine and cocaine metabolites for concentrations at or above 300 pg/mg hair.This test system has not been evaluated for use with hair specimens from locations other than thehead. It is an in vitro diagnostic device intended exclusively for in-house professional use and is notintended for sale to anyone.The ELISA Cocaine Kit is based on the competition for a limited number of antibody sites byunlabeled cocaine/cocaine metabolites and enzyme-labeled drug. The two will bind to the antibody inproportion to their concentration in solution.Once accessioned in the lab, the aluminum foil is opened and the specimen is cut at approxima3.9 cm from the root end. This specimen is cut into smaller lengths and mixed to enshomogeneity. Ten milligrams of the specimen is weighed out and placed into a properly labeled testtube. The specimen is then washed with methanol, decanted, and then placed in hot methanol
containing 0.5% (v/v) trifluoroacetic acid for one hour forty-five minutes. The extracted methanolsolution is then transferred to a new tube and evaporated under nitrogen. The tubes arereconstituted with phosphate buffer and assayed using the Cocaine ELISA Kit. This kit is a solid-phase microtiter plate immunoassay in which the microwells are coated with a high affinity captureantibody to cocaine. A hair sample extract is added to the well, followed by the horseradishperoxidase (HRP) enzyme conjugate. During this initial phase, the enzyme conjugate competes withthe analyte in the sample for binding sites on the antibody-coated microwells. A wash solution(Tween-20 in phosphate buffered saline solution) is then applied to remove any unbound materialssuch as excess conjugate and residual sample. Enzyme substrate solution containing 3, 3', 5, 5'-tetramethylbenzidine (TMB) is then added for the final color development process. The reaction isstopped with 1N sulfuric acid and the absorbance is read at 450 nm, with a reference wavelength of620 nm, using a plate reader. Color intensity is inversely proportional to the amount of analytepresent in the sample. Therefore, samples that contain drug or analyte will inhibit binding of theenzyme conjugate to the antibody, resulting in little substrate binding and less color developmentthan in the negative calibrator. For the screening assay an absorbance less than or equal to theabsorbance of the 300 pg cocaine/mg hair cutoff calibrator is indicative of the presence ofcocaine/cocaine metabolites.
DeviceBackgroundCocaine is one of the most potent of the naturally occurring central nervous system stimulants. Thecompound is found in the leaves of Erythroxylon coca, a South American shrub, in amounts of up to2% by weight. It was first isolated in pure form in 1855, and has been widely utilized in medicineas a local anesthetic and increasingly by drug abusers for its stimulant properties. For anestheticuses, cocaine is administered topically as the hydrochloride in 1-4% solutions for ophthalmologicprocedures and in 10-20% solutions for the membranes of the nose and throat. When self-administered, it is commonly taken as the hydrochloride by nasal insufflation or intravenousinjection or as the free base by smoking in doses of 10-120 mg.1
Cocaine is the primary analyte found in hair after cocaine ingestion, in spite of its very short half-lifein plasma. The major blood and urine metabolite, benzoylecgonine, whose concentration in plasmafar exceeds cocaine's, is present in hair at relatively lower concentrations. Additionally, the cocainemetabolite cocaethylene (ethylbenzoylecgonine) may be found in hair.
Intended UseThe Quest Diagnostics HairCheck-DT (Cocaine) test system utilizes an Enzyme LinkedImmunosorbent Assay (ELISA) for the qualitative detection of cocaine in head hair samples throughthe measurement of cocaine and cocaine metabolites for concentrations at or above 300 pg/mg hair.This test system has not been evaluated for use with hair specimens from locations other than thehead. It is an in vitro diagnostic device intended exclusively for in-house professional use and is notintended for sale to anyone.The HairCheck-DT (Cocaine) test system was evaluated in two distinct study populations; individualsknown to be chronic drug abusers, and individuals proclaiming to be drug-free.The Quest Diagnostics HairCheck-DT (Cocaine) test system provides only a preliminary analytical testresult. To confirm a presumptive screen positive result, a more specific alternate chemical method,such as gas chromatography-mass spectrometry (GC-MS) or Liquid Chromatography-Tandem MassSpectrometry (LC-MS/MS), must be used. Clinical consideration and professional judgment should beapplied to any drug of abuse test result, particularly when preliminary positive results are obtained.
IndicationsSee Intended Use Statement
StudiesPrecision/Reproducibility Cross reactivity (Structurally related compounds) Interferences (Unrelated compounds) Method Comparison Within-Run Specimen Extraction Reproducibility Cosmetic Hair Treatment Specimen Shipping Stability Study Accelerated Reagent Stability Shelf Life (Real Time) Stability

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a circular symbol with an arrow pointing downwards and to the left. To the right of the symbol is the text "Quest Diagnostics" in a sans-serif font. The word "Quest" is larger and bolder than the word "Diagnostics".

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized "Q" symbol on the left, with the words "Quest" and "Diagnostics" to the right of the symbol. The "Q" symbol is a circular shape with an arrow pointing inward, and the text is in a sans-serif font.

2.0 Classification and Regulatory Status

U.S. Classification 2.1

Class II

3.0 Intended Use and Indications

3.1 Intended Use - Quest Diagnostics HairCheck-DT (Cocaine)

The Quest Diagnostics HairCheck-DT (Cocaine) test system utilizes an Enzyme Linked Immunosorbent Assay (ELISA) for the qualitative detection of cocaine in head hair samples through the measurement of cocaine and cocaine metabolites for concentrations at or above 300 pg/mg hair. This test system has not been evaluated for use with hair specimens from locations other than the head. It is an in vitro diagnostic device intended exclusively for in-house professional use and is not intended for sale to anyone.

The HairCheck-DT (Cocaine) test system was evaluated in two distinct study populations; individuals known to be chronic drug abusers, and individuals proclaiming to be drug-free.

The Quest Diagnostics HairCheck-DT (Cocaine) test system provides only a preliminary analytical test result. To confirm a presumptive screen positive result, a more specific alternate chemical method, such as gas chromatography-mass spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS), must be used. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

3.2 Indications

See Intended Use Above

3.3 Reactivity and Corresponding Follow-up

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Quest Diagnostics logo. The logo consists of a stylized letter Q on the left and the words "Quest Diagnostics" on the right. The letter Q is a circular shape with a break in the upper right quadrant, and an arrow pointing inward. The word "Quest" is above the word "Diagnostics".

Reactivity and Corresponding Follow-up
ResultFollow-up
Cocaine Detected/NotdetectedThe Quest Diagnostics HairCheck-DT (Cocaine) ELISA provides only apreliminary result. To obtain confirmed analytical results a morespecific alternate method, such as gas chromatography/-massspectrometry (GC/-MS) or Liquid Chromatography-Tandem MassSpectrometry (LC/-MS/MS), must be used. Clinical consideration andprofessional judgment must be applied to any drug of abuse testresult, particularly in evaluating a preliminary positive result.A positive test result does not always mean a person took illegal drugsand a negative test result does not always mean a person did not takeillegal drugs. There are a number of factors that influence thereliability of drug tests.
Invalid ResultRepeat testing with a newly prepared sample.

Major Risks and Benefits vs. Result 3.4

There are no major risks. The Quest Diagnostics Hair Check-DT (Cocaine) test system is a screening test for the presence of Cocaine and all presumptively positive specimens should be tested with a more specific alternate chemical method, such as gas chromatography-mass spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS), to obtain a confirmed result.

Comment [MJ2]: Font size

Device Description 3.5

The Quest Diagnostics Hair Check-DT (Cocaine) test system utilizes an Enzyme Linked Immunosorbent Assay (ELISA) for the qualitative detection of cocaine in head hair samples through the measurement of cocaine and cocaine metabolites for concentrations at or above 300 pg/mg hair. This test system has not been evaluated for use with hair specimens from locations other than the head. It is an in vitro diagnostic device intended exclusively for in-house professional use and is not intended for sale to anyone.

The test consists of two parts; a pre-analytical hair treatment procedure (to convert the solid matrix of hair to a measurable liquid matrix) and the screening assay.

The ELISA screening assay consists of micro strip plates coated with mouse anti-Cocaine monoclonal antibody, enzyme concentrate conjugate (horseradish peroxidase conjugated to cocaine), enzyme diluent, substrate containing tetramethylbenzidine (TMB), concentrated wash solution, acid stop solution containing 1 N H2SO4.

Pre-Analytical:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the Quest Diagnostics logo. The logo consists of a circular graphic on the left and the words "Quest Diagnostics" on the right. The word "Quest" is on the top line and the word "Diagnostics" is on the bottom line.

A sample of hair (approximately 120 strands) should be cut as close as possible to the scalp, preferably from the back of the crown (vertex). This kit has not been evaluated for other types of hair. The amount of hair collected in this manner is such that a 3.9 cm long sample should weigh approximately 100 – 120 mg. The hair is placed in the V-shaped aluminum foil, ensuring that the root ends are aligned with the pointed end of the foil. The aluminum foil is pinched tightly around the length of hair (DO NOT BEND IN HALF). The foil is placed in a sample acquisition card, root end left towards the arrow. The card is then sealed lengthwise with the box seal. The card is placed in the Toxicology Specimen Baqqie and sealed. Keep hair specimens at ambient temperature until they are shipped, without refrigeration, to the laboratory. No special handling is needed.

Unknown specimens are prepared by weighing out ten (10) milligrams of the hair that has been cut into a fine mince (0.2-0.5 cm) and placing it into a properly labeled test tube. Wash specimens: add 2 mL of methanol to each specimen and let stand for 5 minutes at room temperature. Swirl and discard the methanol with a disposable pipette. Take care not to aspirate any hair into the pipette. Extract specimens: Add 3 mL of acidified methanol (0.5% (v/v) Trifluoroacetic acid (TFA) in methanol) and place in heated water bath (62°C) for 1 hour 45 minutes. Remove the samples from the water bath and transfer them to a heated sonicator set at 55°C. Sonicate the samples for 15 minutes. Dry the specimens: Transfer the methanol solution to a new tube and evaporate until dry under nitrogen. Reconstitute dried specimens with 0.025 mL of acetonitrile and 0.3 mL of phosphate buffer and vortex.

Test Principle:

The ELISA Cocaine Kit is based on the competition for a limited number of antibody sites by unlabeled cocaine/cocaine metabolites and enzyme-labeled drug. The two will bind to the antibody in proportion to their concentration in solution.

Once accessioned in the lab the aluminum foil is opened and the specimen is cut at approximately 3.9 cm from the root end. This specimen is cut into smaller lengths and mixed to ensure homogeneity. Ten (10) milligrams of the specimen is weighed out and placed into a properly labeled test tube. The specimen is then washed with methanol, decanted, and then placed in hot methanol for two hours. The methanol is then transferred to a new tube and evaporated under nitrogen. The tubes are reconstituted with phosphate buffer and assayed using the ELISA Cocaine Kit. This kit is a solid-phase microtiter plate immunoassay in which the microwells are coated with a high affinity capture antibody to cocaine. A hair sample extract is added to the well, followed by the horseradish peroxidase (HRP) enzyme conjugate. During this initial phase, the enzyme conjugate competes with the analyte in the sample for binding sites on the antibody-coated microwells. A wash solution (Tween-20 in phosphate buffered saline solution) is then applied to remove any unbound materials such as excess conjugate and residual sample. Enzyme substrate solution containing 3,3', 5,5'tetramethylbenzidine (TMB) is then added for the final color development process. The reaction is stopped with 1N sulfuric acid and the absorbance is read at 450 nm with a reference wavelength of

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized letter "Q" on the left and the words "Quest Diagnostics" on the right. The "Q" is a circular shape with an arrow pointing inward, and the words "Quest Diagnostics" are in a sans-serif font.

620 nm using a plate reader. Color intensity is inversely proportional to the amount of analyte present in the sample. Therefore, samples that contain drug or analyte will inhibit binding of the enzyme conjugate to the antibody, resulting in little substrate binding and less color development than in the negative calibrator. For the screening assay an absorbance less than or equal to the absorbance of the 300 pg cocaine/mg hair cut-off calibrator is indicative of the presence of cocaine/cocaine metabolites.

3.6 Kit Components

Each kit Quest Diagnostics HairCheck-DT (Cocaine) contains enough reagents to make 4,800 determinations

Reagents

50 x 96 well micro strip plates (12 x 8), coated with mouse anti-Cocaine monoclonal antibody.

2 x 4 mL of enzyme concentrate conjugate (horseradish peroxidase conjugated to cocaine) in a proprietary buffer containing stabilizing agents and thimerosal.

2 x 400 mL of enzyme diluent as a proprietary buffer containing stabilizing agents and thimerosal.

1 x 500 mL of substrate containing tetramethylbenzidine (TMB) and hydrogen peroxide in a citrate buffer containing stabilizers.

1 x 1000 mL of concentrated wash solution with surfactants and thimerosal as a preservative; Dilute 1:10 with deionized water prior to use.

1 x 500 mL of acid stop solution containing 1 N H2SO4.

Calibrators and Controls for Screening Assay:

100mL of 1X HairCheck-DT (Cocaine) Negative Containing 0 pg cocaine/mg hair (No need to dilute, provided at working concentration)

5mL of 40X HairCheck-DT (Cocaine) Low Control containing 6,000 pg cocaine/mg hair (Dilute 1:40 with Negative Control to reach 1X Low Control working concentration of 150 pg cocaine/mg hair)

5mL of 40X HairCheck-DT (Cocaine) Cutoff Calibrator containing 12,000 pg cocaine/mg hair (Dilute 1:40 with Neqative Control to reach 1X calibrator working concentration of 300 pg cocaine/mg hair/

5mL of 40X HairCheck-DT (Cocaine) High Control containing 18,000 pg cocaine/mg hair (Dilute 1:40 with Negative Control to reach 1X High Control working concentration of 450 pg cocaine/mg hair

40X Calibrator and Control solutions are prepared from Cerilliant C-008, 1.0 mg/mL Cocaine in Acetonitrile Certified Reference Material. Calibrator and Control solutions are prepared by diluting commercial reference materials with water and methanol to achieve the target drug concentration; they are prepared in a similar manner however the Calibrator is made from different reference stock material than the Controls and are prepared at different concentrations. The shelf life of these solutions is 1 year at -20° C.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the Quest Diagnostics logo. The logo consists of a stylized "Q" symbol on the left, followed by the text "Quest" on the top line and "Diagnostics" on the bottom line. The "Q" symbol is a circular shape with an arrow pointing inward.

Predicate Device Information

3.7 Comparison to Predicate

Similarities

SimilaritiesSubject DevicePredicate
Quest Diagnostics HairCheck-DT (Cocaine)(K152232)Quest Diagnostics HairCheck-DT (Cocaine)(K023626) September 29, 2003
Intended UseSameSame
Sample TypeHead HairHead Hair
Collection DeviceSameSame
Low Control150 pg cocaine/mg hair150 pg cocaine/mg hair
Cutoff300 pg cocaine/mg hair300 pg cocaine/mg hair
Operating PrincipleQualitative ImmunoassayQualitative Immunoassay
CompetitiveEnzyme-conjugateHRP - CocaineHRP - Cocaine
CLIA ComplexityHighHigh
SubstrateTMBTMB

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the Quest Diagnostics logo. The logo consists of a stylized letter "Q" in black, followed by the words "Quest" and "Diagnostics" in gray. The word "Quest" is larger and bolder than the word "Diagnostics", and there is a trademark symbol after the word "Diagnostics".

Quest Diagnostics HAIRCHECK-DT (Cocaine) 510(k) Summary Page 8 of 27

Differences

DifferenceSubject DevicePredicate
Quest Diagnostics HairCheck-DT(Cocaine) (K152232)Previously Cleared 510K Device (K023626)September 29, 2003)
AntibodyMouse anti-Cocaine monoclonal antibodyRabbit anti-Cocaine polyclonal antibody
Extraction MethodAcidified Methanol (0.5% Trifluoroaceticacid) (TFA)Methanol
High ControlHigh Control modified to 450 pg/mgequivalent concentration in Hair or +50% ofcutoff concentration.High Control - 600 pg/mg equivalentconcentration in Hair or +100% of cutoffconcentration.
Sample SizeSample Preparation modified to 10 mg ofHair used for extraction then reconstitutedwith 0.3 mL phosphate bufferSample Preparation 20 mg of Hair used forextraction then reconstituted with 0.6 mLphosphate buffer
MeasurementWavelengthAssay absorbance measured at 450 nm witha reference wavelength of 620 nm.Assay absorbance was measured at 450 nm with areference wavelength of 630 nm.
KitConfigurationKit Configuration modified to = 50microplate kitKit Configuration = 5 microplate kit

4.0 Studies

Precision Studies/Cutoff Characterization 4.1

Precision

Precision/Reproducibility aliquots were tested in quintuplicate (i.e. 5-replicates) on each of ten (10) different days using the Quest Diagnostics Cocaine specific ELISA kit (Lot # EVAL1). The same nine (9) samples/levels were analyzed in replicates of five (5) over the course of ten (10) days, alternating the order of the levels on the microplates each day. Calibrator and control materials were analyzed in wells located prior to and after the specimens on the microplates each day of testing.

Within-Run Precision

A summary of data for the 10 days of testing for between-run and within-run precision (reproducibility) is shown in the tables below.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized "Q" symbol on the left and the words "Quest Diagnostics" on the right. The "Q" symbol is a circular shape with an arrow pointing inward, and the words "Quest Diagnostics" are written in a simple, sans-serif font.

Quest Diagnostics HAIRCHECK-DT (Cocaine) 510(k) Summary Page 9 of 27

Table 1: Precision Performance at the Cutoff and Overall (Between-Run) Precision
0%25%50%75%100%125%150%175%200%
Target ng/mL0.02.55.07.510.012.515.017.520.0
Target pg/mg0.075.0150.0225.0300.0375.0450.0525.0600.0
Level123456789
Count505050505050505050
Grand Mean2.15281.67171.37521.15601.00080.90250.79480.72240.6484
Within run SD0.07300.05260.03590.03390.03800.03000.02850.02930.0270
Within run CV3.39%3.15%2.61%2.94%3.80%3.32%3.59%4.05%4.16%
Overall SD0.18860.13210.09650.06670.08050.06730.05400.05030.0464
Overall CV8.76%7.90%7.02%5.77%8.04%7.46%6.79%6.96%7.16%
Within-Device SD0.18440.12870.09490.07510.07590.06430.04950.04470.0412
8.57%7.70%6.90%6.50%7.59%7.13%6.23%6.18%6.36%

The overall (between-run) precision (CV %) was less than 10% for all spiked (25%, 50%, 100%, 125%, 150%, 175%, and 200%) levels of analyte.

Evaluation of separation around the cutoff:

The Ratio was calculated (sample OD/cutoff OD) for each of the test samples. The mean Ratio and standard deviation (SD) of all pools were calculated for each day of analysis. The mean Ratio minus 2 SD of the 50% samples was greater than the mean Ratio plus 2 SD of the 100% sample on each day of analysis (Table 3, highlighted in blue). The mean Ratio minus 2 SD of the 100% sample was greater than the mean Ratio plus 2 SD of the 150% sample on each day of analysis (Table 3, highlighted in green).

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized letter "Q" on the left and the words "Quest Diagnostics" on the right. The "Q" is a circular shape with a break in the circle and an arrow pointing inward. The word "Quest" is above the word "Diagnostics" and the trademark symbol is to the right of the word "Diagnostics".

Table 30%25%50%75%100%125%150%175%200%
Target ng/mL0.002.505.007.5010.0012.5015.0017.5020.00
Target pg/mg075150225300375450525600
Day 1Ratio Mean1.84831.54421.21231.05720.88570.80790.72100.68630.6073
CV%1.21531.98441.38432.50013.80271.06482.42844.87122.3754
SD0.02250.03060.01680.02640.03370.00860.01750.03340.0144
Mean - 2SD1.80341.48291.17871.00440.81840.79070.68600.61950.5785
Mean + 2SD1.89321.60541.24581.11010.95310.82510.75600.75320.6362
Day 2Ratio Mean2.01591.57131.30471.09660.94790.85630.74670.69140.5956
CV%3.47554.09951.67742.91734.10826.55462.60963.96066.2692
SD0.07010.06440.02190.03200.03890.05610.01950.02740.0373
Mean - 2SD1.87581.44251.26091.03260.87000.74410.70770.63660.5209
Mean + 2SD2.15601.70021.34851.16061.02580.96860.78560.74610.6703
Day 3Ratio Mean2.12341.63851.40561.18251.00040.93420.81440.78130.6627
CV%4.79813.21861.94183.53665.95732.98662.29943.86515.1202
SD0.10190.05270.02730.04180.05960.02790.01870.03020.0339
Mean - 2SD1.91961.53311.35101.09890.88120.87840.77690.72090.5948
Mean + 2SD2.32711.74401.46021.26621.11950.99000.85180.84170.7306
Day 4Ratio Mean2.32941.90861.48021.21611.04870.97240.89930.72250.7066
CV%2.48991.90872.15441.82564.14691.55541.85222.33101.2972
SD0.05800.03640.03190.02220.04350.01510.01670.01680.0092
Mean - 2SD2.21341.83571.41641.17170.96170.94210.86600.68880.6883
Mean + 2SD2.44541.98141.54391.26051.13571.00260.93270.75620.7249
Day 5Ratio Mean2.25681.81081.51291.24341.10160.95560.83570.76730.6618
CV%4.33894.41071.85351.84995.11496.05515.85374.25414.4263
SD0.09790.07990.02800.02300.05630.05790.04890.03260.0293
Mean - 2SD2.06091.65111.45681.19740.98890.83990.73780.70210.6032
Mean + 2SD2.45261.97051.56901.28941.21431.07140.93350.83260.7204
Day 6Ratio Mean1.94741.57771.29311.11260.92260.88310.77330.66020.6027
CV%3.02623.88112.35404.29443.95411.62893.11154.55804.7650
SD0.05890.06120.03040.04780.03650.01440.02410.03010.0287
Mean - 2SD0.11791.45521.23221.01700.84960.85430.72520.60000.5452
Mean + 2SD2.06531.70021.35401.20820.99560.91190.82140.72040.6601
Day 7Ratio Mean2.18051.57731.32101.13830.94250.81950.77660.69430.6220
CV%1.54862.15302.20881.89231.89452.63573.16005.06062.9375
SD0.03380.03400.02920.02150.01790.02160.02450.03510.0183
Mean - 2SD2.11301.50941.26271.09520.90680.77630.72750.62400.5854
Mean + 2SD2.24801.64521.37941.18140.97830.86270.82560.76450.6585
Day 8Ratio Mean2.13341.61341.35861.11370.99080.88040.77590.69090.6381
CV%5.35364.16623.62103.78702.51061.14165.21413.94326.0450
SD0.11420.06720.04920.04220.02490.01010.04050.02720.0386
Mean - 2SD1.90501.47891.26021.02930.94100.86030.69500.63640.5610
Mean + 2SD2.36181.74781.45701.19801.04060.90050.85680.74540.7153
Day 9Ratio Mean2.21691.82771.46541.17401.08880.91580.80560.74300.6868
CV%2.84571.72832.85983.04491.86432.08442.67473.36392.5813
SD0.06310.03160.04190.03570.02030.01910.02150.02500.0177
Mean - 2SD2.09071.76451.59111.10251.04820.87760.76260.69300.6513
Mean + 2SD2.34311.89091.54921.24551.12940.95400.84870.79300.7223
Day 10Ratio Mean2.47591.64721.39791.22571.07861.00000.79910.78660.7000
CV%2.18562.47654.30442.85902.07871.66424.21573.88893.5696
SD0.05410.04080.06020.03500.02240.01660.03370.03060.0250
Mean - 2SD2.36761.56561.27761.15571.03370.96670.73170.72540.6500

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized letter "Q" on the left and the words "Quest Diagnostics" on the right. The letter "Q" is a black circle with a white arrow pointing inward. The words "Quest Diagnostics" are in a sans-serif font, with "Quest" in a larger font than "Diagnostics".

Quest Diagnostics HAIRCHECK-DT (Cocaine) 510(k) Summary Page 11 of 27

% Relative to Cutoff0%25%50%75%100%125%150%175%200%
Target pg/mg075150225300375450525600
Negative Count50505050253000
Positive Count00002547505050
% Negative100%100%100%100%50%6%0%0%0%
% Positive0%0%0%0%50%94%100%100%100%

Table 4: % Positive and Negative

Table 4 also demonstrates separation around the cutoff. There is no crossover ±50% of the cutoff.

4.2 Cross-Reactivity

Structurally Related Compounds

Cocaine, structurally-related compounds, pharmacologically-related compounds, and metabolites were tested for cross-reactivity in the assay. Eleven (11) compounds, structurally related to cocaine and known cocaine metabolites, were selected for the study (see table below for the list). The cross-reactant solutions were prepared by adding the compounds to negative hair matrix. The concentrations listed below produced a result approximately equal to the cutoff calibrator.

Serial dilutions of each compound were prepared and analyzed. If the OD response of the spiked sample was positive then the spiked sample was diluted further with finer dilutions until a positive result was obtained with an OD within 5% of the cutoff value (sample OD/Cutoff OD). Cross reactivity was calculated as: (Cutoff Concentration/Lowest Cross Reactant Concentration with a Positive Result) x 100.

CompoundCrossReactivity (%)Tested Concentration inNegative Hair Matrix(ng/mL)Concentration of compound(pg/mg hair) needed toproduce results equivalent to300 pg/mg of Cocaine
Anhydroecgonine<0.1%10,000>300,000
Anhydroecgonine methyl ester<0.1%10,000>300,000
Atropine<0.1%10,000>300,000
Benzoylecgonine143%7210
Cocatheylene125%8240
Ecgonine<0.1%10,000>300,000
Ecgonine methyl ester<0.1%10,000>300,000
Meta-hydroxybenzoylecgonine200%5150
Norcocaine1%1,00030,000
Tropacocaine8%1253,750
Cocaine100%10300
Cocaine N-oxide HCI11%902,700
Scopolamine hydrobromide<0.1%10,000>300,000
Norbenzoylecgonine1%90027,000
m-hydroxycocaine67%15105
Benzoylecgonine Isopropyl Ester182%5.5165

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a circular graphic on the left and the words "Quest Diagnostics" on the right. The word "Quest" is in a larger font than the word "Diagnostics", and there is a trademark symbol after the word "Diagnostics".

Conclusions

Cross-reactivity for tested compounds ranged from <0.1% to 200%. Several of the compounds are metabolites of cocaine and cross-reactivity is not an undesired outcome. The compounds exhibiting the highest cross-reactivity – e.g. benzoylecgonine, and cocaethylene – would also be expected to produce a presumptive positive result for the hair cocaine screen, and therefore, would not elicit an undue number of additional confirmation tests and would also be a desired outcome for those users requesting the reporting of these additional cocaine metabolites

4.3 Interference

Various common over-the-counter and prescription medications, as well as abused drugs, were tested for cross-reactivity in the assay. One hundred and forty three (143) structurally unrelated compounds were added to 1:1 methanol/water at a concentration of 10,000 ng/mL and were subsequently spiked into both low control (equivalent to 150 pg/mg) negative hair matrix and also to high control (equivalent to 450 pg/mg) negative hair matrix. Additionally ten (10) hair dyeing compounds were evaluated in the same manner starting at a concentration of 500 µg/mL (500,000 ng/mL). Based on the OD of the spiked low control producing results above the OD of the cutoff calibrator and the OD of the spiked high control producing results below or equal to the OD of the cutoff calibrator, it is concluded that one hundred and thirty six (136) of the compounds display no significant reactivity with the assay:

Compounds That Do Not Interfere with the Assay:

(-)-11-nor-9-Carboxy-Δ9-THCCannabinolHydrochlorthiazideNormenperidinic Acid (4-Phenylpiperidine-4-carboxylic acidhydrochloride)
(+)-11-Nor-Δ9-THC-9-carboxylicacid glucuronideCatharanthine(Ergoloid)HydrocodoneNormorphine
(-) CotinineChlordiazepoxideHydrocortisone(Cortisol)Noroxymorphone
(-) MethamphetamineCimaterolHydromorphoneO-Desmethyl-cis-tramadol HCl
(+) IsoproterenolClenbuterolIbuprofenOxazepam
(+) MethamphetamineClonazepamImipramineOxymorphone
(+) NorpseudoephedrineCodeineLAMPAPCP
(±) KetamineCorticosteroneLevorphanolPenicillin G Sodium salt
(±) MDA [(±)-3,4-Methylenedioxyamphetamine]CortisoneLidocainePentazocine
(±) MDEA [(±)-3,4-Methylenedioxyethylamphetamine](-)-Δ8-THCLorazepamphenothiazine
(±) MDMA [(±)-3,4-Methylenedioxymethamphetamine](-)-Δ9-THCLSDPhentermine
(±) PropanololDesalkylflurazepamMeperdinePhenylbutazone
(±)-N-EthylamphetamineDesipramineMetaproterenolhemisulfateP-Hydroxymethamphetamine (Isodrin)
(±)-N-PropylamphetamineDesmethyldoxepin(cis/trans)Methedrone(Methoxyephedrine)Progesterone
(±)-Phenylpropanolamine(Norephedrine)DexamethasoneMethoxyphenamine HCIPromethazine HCI
(±)-MethadoneDextromethorphanMethylergometrinemaleateProprionyl Promazine HCI
Nandrolone (19-Nortesterone)DiazepamMethylphenidateR(-) Phenylephrine
1R,2S(-)-EphedrineDihydrocodeineMonensin Sodium SaltR(-)-Amphetamine
1S,2R(+)-EphedrineDihydroergotamineMesylateMorphineR(+) Methcathinone
2-Oxo-3-hydroxy-LSD (2-Oxo-3-hydroxy-lysergic aciddiethylamide)DihydromorphineMorphine-3-β-D-glucuronideR,R(-)-Pseudoephedrine
Acetaminophen (4-Acetoamidophenol)Doxepin (cis/trans)Morphine-6-β-D-glucuronideS(-)-Nicotine
4-MeO-PCP HCI (4-Methoxyphencyclidine HCI)DoxylaminesuccinateNadololS(+)-Amphetamine
6-Acetylmorphine (6-Monoacetylmorphine)Venlafaxinehydrochloride(Effexor)NalbuphineSalbutamol (Albuterol)
Acebutolol HCIErythromycinNalorphineS, S(+)Pseudoephedrine
p-Acetophenetidin (Phenacetine)EthylmorphineNaltrexoneStanozolol
Acetylsalicylic AcidFenfluramineN-DesmethyodramadolStreptomycin Sulfate
7-AminoflunitrazepamFentanylNeomycinSufentanil
AmoxicillinFlumethasoneN-EthylcathioneTetracycline HCI
BetamethasoneFlunitrazepam(±)-4-Methyl-N-ethyl-norephedrine HCI (N-Ethyl nor ephedrine)Theophylline
BuprenorphineFlurazepamNorbuprenorphinecis-Tramadol
CaffeineFurosamideNorcodeineTriazolam
CannabidiolHeroinNordiazepamTriflupromazine HCL
HMMA (4-Hydroxy-3-Methoxymandelic Acid)NormeperidineTylosin Tartrate

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image contains the logo for Quest Diagnostics. The logo consists of a stylized "Q" symbol on the left and the words "Quest Diagnostics" on the right. The "Q" symbol is a circular shape with an arrow pointing inward. The text "Quest Diagnostics" is in a sans-serif font, with "Quest" in a larger, bolder font than "Diagnostics".

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image contains the logo for Quest Diagnostics. The logo consists of a stylized letter "Q" on the left and the words "Quest Diagnostics" on the right. The "Q" is a black circle with a white arrow pointing inward. The words "Quest Diagnostics" are in a sans-serif font, with "Quest" in a larger font than "Diagnostics".

Seventeen (17) compounds exhibited a limited degree of reactivity in the immunoassay, as listed below.

CompoundPositive Interferenceobserved at or above:Equivalent Concentration in Hair(pg/mg)
Thioridazine2,500 ng/mL75,000
Trifluperazine HCl5,000 ng/mL150,000
Chlorpromazine10,000 ng/mL300,000
Trimerperazine10,000 ng/mL300,000
Prochlorperazine dimaleate2,500 ng/mL75,000
Fluphenazine dihydrochloride5,000 ng/mL150,000
Iso-LSD1,250 ng/mL37,500
Basic Yellow 407,812.5 ng/mL234,375
Methylene Blue15,625 ng/mL468,750
Basic Violet 1631,250 ng/mL937,500
Basic Blue 9962,500 ng/mL1,875,000
Basic Brown 1662,500 ng/mL1,875,000
Ethyl Violet62,500 ng/mL1,875,000
Basic Brown 1762,500 ng/mL1,875,000
Basic Red 51125,000 ng/mL3,750,000
Safranin O125,000 ng/mL3,750,000
Basic Yellow 87500,000 ng/mL15,000,000

The following compounds exhibited positive interference at the following levels:

No negative interference was observed in this study.

Method Comparison 4.4

The tested group of specimens consisted of a total of 100 donor specimens. This study involved analysis of routine donor hair specimens provided by the Quest Diagnostics testing laboratory. These are specimens that screened negative for cocaine or have previously been confirmed positive for cocaine and/or cocaine metabolites, and were eligible for discard. Per protocol, a minimum of 50 negative samples (five (5) between 150 and 300 pg cocaine/mg hair) and 50 positive samples (five (5) between 300 and 450 pg cocaine/mg hair) were analyzed. Specimens were first confirmed by GC-MS randomized and then screened with the Quest Diagnostics HairCheck –DT (Cocaine) ELISA.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a circular shape with an arrow pointing to the right, followed by the text "Quest" in a bold, sans-serif font. Below the word "Quest" is the word "Diagnostics" in a smaller, lighter font. The logo is simple and modern, and the colors are black and white.

Method Comparison Data:

CandidateDeviceResultNegative (Less thanhalf the cutoffconcentration) byconfirmatoryanalysisNear Cutoff/BorderlineNegative (Between50% below thecutoff and thecutoffconcentration) byconfirmatoryanalysisNear Cutoff/Borderline Positive(Between the cutoffand 50% above thecutoffconcentration) byconfirmatoryanalysisHigh Positive(greater than 50%above the cutoffconcentration) byconfirmatoryanalysis
Conc. byGCMS<150 pg/mg150-299 pg/mg300-450 pg/mg>450 pg/mg
Negative44200
Positive13545
Overall Accuracy
% Agreement among positive100% (50/50)
% Agreement among negative92% (46/50)

Definitions:

Ratio = specimen OD/mean OD of cutoff calibrator Negative by ELISA: Ratio >1.0 Negative by GC-MS: < 150 pg/mg Near Cutoff Negative by ELISA: Ratio falls between cutoff and low control value Near Cutoff Negative by GCMS: 150-299 pg/mg Near Cutoff Positive by ELISA: Ratio falls between high control and cutoff value Near Cutoff Positive by GCMS: 300-450 pg/mg Positive by ELISA : Ratio ≤1.0 High Positive by GC-MS: >450 pg/mg

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image contains the logo for Quest Diagnostics. The logo consists of a stylized "Q" symbol on the left, followed by the text "Quest" on the top line and "Diagnostics" on the bottom line. The "Q" symbol is a circular shape with an arrow pointing inward, and the text is in a sans-serif font.

Discrepant Results

UniqueIdentifierGC-MS pg/mgCocaineEq. TotalELISARatioODDeviceResultReferenceNumber
BECOCCENORP/N
MCC 06000000.038POS1
MCC 470155001550.807POS2
MCC 490190001900.704POS3
MCC 5080272003860.560POS4

BE-Benzoylecgonine

COC-Cocaine

CE-Cocaethylene

NC-Norcocaine

(1) This specimen confirmed negative by GC-MS for cocaine and metabolites, yet screened strongly positive by the HairCheck-DT ELISA device. The hair specimen was dyed pink, likely using Manic Panic Semi-Permanent Cotton Candy Pink Hair Color, or similar hair care product. A deep pink color was encountered during the extraction and re-suspension process and the specimen was also robustly pink when pipetted onto the ELISA plates. Exposing the specimen to a black light resulted in intense pink fluorescence emission. The cited hair care product contains three cationic dyes, Basic Violet 16, Safranin O and Methylene Blue. All three dyes are fluorescent, emitting in the vicinity of 583-595 nm, the red region of the visible light spectrum. Further investigation into the effect of semi-permanent dyes on the assay has provided evidence that the binding of cocaine-HRP-conjugate to the antibody is disrupted by the presence of high concentrations of these dyes (relative to the concentration of cocaine) in sample extracts. This effect however, is not specific to the cocaine antibody since the offending hair specimen also produced false positive results in the HairCheck-DT (Opiates) ELISA and the HairCheck-DT (Methamphetamine) ELISA. Semi-permanent hair care products contain various cationic dyes that are readily extracted along with the cocaine during the pre-analytical sample preparation. The high concentration of extracted dyes, leads to positive interference through interaction with one or more steps of the fundamental immunoassay procedure.

(2) This specimen confirmed negative by GC-MS with a quantitative value of 155 pg cocaine/mg hair.

(3) This specimen confirmed negative by GC-MS with a quantitative value of 190 pg cocaine/mg hair.

(4) This specimen confirmed negative by GC-MS with a quantitative value of 272 pg cocaine/mg hair, but also contained 80 pg/mg of benzoylecgonine.

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the Quest Diagnostics logo. The logo consists of a stylized letter "Q" with an arrow incorporated into its design, followed by the text "Quest Diagnostics". The text is in a simple, sans-serif font, with "Quest" appearing above "Diagnostics".

4.5 Within-Run Specimen Extraction Reproducibility

This study assessed the within-run reproducibility of hair specimen extraction by the ELISA and GC-MS method by evaluating three (3) replicates of five (5) cocaine positive specimens on the HairCheck-DT (Cocaine) ELISA as well as the GC-MS confirmatory method.

Unique IDNumber[Cocaine](pg/mg)ReplicateNumberMean ODCoeff. ofVar.(CV%)ELISAPos/Neg
156030.8110.6 %3/0
21807830.0252.3 %3/0
3292630.1167.8 %3/0
4769830.0593.5 %3/0
5232030.2315.2 %3/0

Reproducibility Results

Conclusions

These results demonstrate the reproducibility of the specimen extraction and result for both the HairCheck-DT (Cocaine) ELISA as well as the GC-MS confirmatory method. Each replicate of each individual sample was positive for cocaine (100% agreement) by the HairCheck-DT (Cocaine) ELISA immunoassay and positive for cocaine by the GC-MS confirmation method. The within-run coefficient of variation for the GC-MS assay for cocaine was 6.7%. The within-run coefficients of variation for each of the five hair specimens by ELISA immunoassay (OD) was less than 10%, with an overall mean coefficient of variation of 3.9%.

Specimen Shipping Stability 4.6

In order to demonstrate the stability of cocaine in hair specimens during the shipping process, 56 hair samples- twenty eight (28) confirmed positive cocaine hair specimens, of which three specimens (3) have pre-shipping quantitative GC-MS results in the near cutoff positive concentration range (300 to 449 pg/mg Cocaine) and twenty eight (28) confirmed negative cocaine hair specimens, of which three (3) have pre-shipping quantitative GC-MS results in the near cutoff negative concentration range (150 to 299 pg/mg Cocaine) were shipped to eight different geographical locations and the temperatures tracked during shipping. By enclosing an electronic temperature sensor within each shipment of hair samples, we were able to retrieve a range of temperatures collected every 20 minutes during the shipping process. In order to mimic potential shipping extreme temperatures, the hair was first cold shocked at -15°C for 15 hours and then heat shocked at +47°C for 6 hours prior to shipping.

After the boxes were returned to facility, ELISA screening and GC-MS confirmation were performed on each sample and compared to the non-shipped results for the same samples. The cocaine negative samples that were shipped maintained their neqative status both for ELISA screening and GC-MS with respect to cocaine levels. One of the borderline negative samples originally screened positive prior to

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized "Q" with an arrow pointing downwards, followed by the text "Quest" in a bold font and "Diagnostics" in a smaller, lighter font. The logo is simple and modern, with a focus on the company's name and brand.

shipping and all three of the borderline negative samples screened positive after shipping due to variability near the cutoff of the ELISA screening method. The cocaine positive samples that were shipped maintained their positive status for both the ELISA screening and the GC-MS confirmation with respect to cocaine levels.

Summary of results for positive specimens

Pre-ShippingPost-Shipping
GC-MSELISAGC-MSELISA
Positive28282828
Negative0000

Summary of results for negative specimens

Pre-ShippingPost-Shipping
GC-MSELISAGC-MSELISA
Positive01*03**
Negative28272825
  • One of the three Near Cutoff Negative specimens (as determined by GC-MS pre-shipping) screened positive by ELISA prior to shipping.

** Three of three Near Cutoff Negative specimens (as determined by GC-MS pre-shipping) screened positive by ELISA post-shipping.

Conclusions

This study agrees with published literature that hair specimens containing cocaine are stable to various shipping regimes. No changes in positive or negative status were produced upon shipment of the hair other than borderline specimens which could be expected due to variability near the cutoff of the ELISA screening assay. The hair samples were exposed to a range of temperatures that exceeded what would normally be encountered in a standard shipping process, with no changes in reportable status.

Cosmetic Hair Treatment Studies 4.7

This study is designed to assess the contribution of various common cosmetic hair treatments; shampooing, dyeing, bleaching, perming, and relaxing, on the detection of cocaine using the HairCheck DT Cocaine ELISA. This data is used by end-users (customers) to assess the degree to which various cosmetic hair treatments may interfere with Cocaine detection in their donor population.

This study involves the testing of a panel of sixty (60) confirmed positive cocaine hair samples and sixty (60) screened negative cocaine hair samples that are treated with one of five (5) cosmetic hair treatments or left untreated. Cocaine positive hair is defined as a hair sample confirmed by GC-MS as

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image shows the Quest Diagnostics logo. The logo consists of a stylized "Q" with an arrow incorporated into the design, followed by the text "Quest Diagnostics" in a simple, sans-serif font. The "Q" is in black, while the text is in gray. The logo is clean and professional, conveying a sense of reliability and expertise.

having greater than 300 pg/mg cocaine. Absorbance readings after treatment were compared to absorbance readings prior to treatment, with the resulting change and direction of change noted. The resulting changes in ELISA test results (if any) are also noted, and are included in the table(s) below. (Absorbance values are normalized to the absorbance value of the cutoff calibrator absorbance value.)

ELISA Results for Cocaine Positive Hair

Pre-TreatmentPost-TreatmentPre to Post
Number ofSamples Positiveby GC-MSNumber ofSamples Positiveby ELISANumber ofSamples Positive byGC-MSNumber ofSamples Positive byELISARange of% change in OD
Shampoo12121212(-48%, +56%)
Brown Dye12121212(-59%, +174%)
Bleach12121212(-31%, +278%)
Perm12121212(-60%, +61%)
Relaxer12121212(+12%, +197%)

Cocaine positive hair remained positive by ELISA post-treatment, though substantial loss of cocaine was observed by GC-MS with bleach and relaxer.

Positive Specimens Individual Results

Pre-Cosmetic TreatmentPost-Cosmetic Treatment
TreatmentUniqueIDCocaineGC-MSpg/mgELISARaw ODCal.ODELISAP/NELISARaw ODCal.ODELISAP/N%Change(ELISARaw OD)
ShampooP18090.7721.732POS0.9761.732POS26%
P213910.5051.732POS0.5041.732POS0%
P342090.1641.732POS0.1651.732POS1%
P459270.0891.732POS0.1151.732POS29%
P58040.6641.732POS0.8491.732POS28%
P69290.4751.732POS0.7391.732POS56%
P77930.8171.732POS0.6981.732POS-15%
P89840.2751.732POS0.3161.732POS15%
P913360.4171.732POS0.2221.732POS-47%
P1014710.6171.732POS0.3221.732POS-48%
P1120640.2291.732POS0.1431.732POS-38%
P128120.4371.732POS0.31.732POS-31%
P139510.7221.732POS0.6231.732POS-14%
P149680.7391.732POS0.5441.732POS-26%
P1515530.4291.732POS0.3681.732POS-14%
Brown DyeP1615960.3661.732POS0.4141.732POS13%
P1735360.1991.732POS0.0821.732POS-59%
P1813060.1661.732POS0.4551.732POS174%
P1914800.571.732POS0.3891.732POS-32%

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image contains the logo for Quest Diagnostics. The logo consists of a stylized "Q" symbol on the left, followed by the text "Quest" and "Diagnostics" on the right. The "Q" symbol is a circular shape with an arrow-like cutout, and the text is in a simple, sans-serif font.

P2018520.3221.732POS0.2591.732POS-20%
P219190.4991.732POS0.4671.732POS-6%
P229680.1251.732POS0.2621.732POS110%
P2343750.1591.732POS0.1661.732POS4%
P2469480.0471.732POS0.0431.732POS-9%
P258100.5161.732POS0.8991.732POS74%
P2613150.2641.732POS0.3791.732POS44%
P2736050.1181.732POS0.161.732POS36%
P2851580.0431.732POS0.061.732POS40%
P2957960.0631.732POS0.0961.732POS52%
BleachBlondeP3068120.0731.732POS0.0671.732POS-8%
P31227700.0341.732POS0.0281.732POS-18%
P32397690.0161.732POS0.0161.732POS0%
P337500.3891.732POS0.6271.732POS61%
P3469200.0491.732POS0.1851.732POS278%
P3514400.3151.732POS0.5111.732POS62%
P3616070.3451.732POS0.2391.732POS-31%
Pre-Cosmetic TreatmentPost-Cosmetic Treatment
TreatmentUniqueIDCocaineGC-MSpg/mgELISARaw ODCal.ODELISAP/NELISARaw ODCal. ODELISAP/N%Change(ELISARaw OD)
PermP3716620.1811.480POS0.231.480POS27%
P3817790.2491.480POS0.3651.480POS47%
P3922980.1121.480POS0.1231.480POS10%
P4034970.11.480POS0.0881.480POS-12%
P4141420.0541.480POS0.0871.480POS61%
P4245220.111.480POS0.1071.480POS-3%
P4354610.0981.480POS0.1041.480POS6%
P4456980.0931.480POS0.1121.480POS20%
P4559930.0341.480POS0.0391.480POS15%
P4663760.0721.480POS0.0811.480POS13%
P4767860.0821.480POS0.1021.480POS24%
P4883150.2191.480POS0.0871.480POS-60%
RelaxerP4992030.0491.480POS0.0671.480POS37%
P50200000.0211.480POS0.0261.480POS24%
P518560.6191.480POS0.8331.480POS35%
P5214550.1891.480POS0.5091.480POS169%

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized letter "Q" on the left and the words "Quest Diagnostics" on the right. The "Q" is black, and the words "Quest Diagnostics" are gray.

P5315550.3541.480POS0.4091.480POS16%
P5415970.0651.480POS0.1071.480POS65%
P5520390.2451.480POS0.2941.480POS20%
P5634620.1091.480POS0.1221.480POS12%
P5744910.1291.480POS0.1871.480POS45%
P5881470.0241.480POS0.0431.480POS79%
P59110240.031.480POS0.0891.480POS197%
P60200000.0151.480POS0.0311.480POS107%

ELISA Results for Cocaine Negative Hair

Pre-TreatmentPost-TreatmentPre to Post
Number ofSamples Negativeby GC-MSNumber ofSamples Negativeby ELISANumber of SamplesNegative by GC-MSNumber of SamplesNegative by ELISARange of% change in OD
Shampoo12121212(-8%, +10%)
Brown Dye12121212(-16%, +4%)
Bleach12121212(-9%, +5%)
Perm12121212(-7%, +6%)
Relaxer12121212(-16%, +4%)

Cocaine negative hair remained negative by ELISA post-treatment.

Negative Specimen Individual Results

Pre-Cosmetic TreatmentPost-Cosmetic Treatment
TreatmentUniqueIDCocaineGC-MSpg/mgELISARaw ODCal.ODELISAP/NELISARaw ODCal.ODELISAP/N%Change(ELISARaw OD)
N102.651.562NEG2.6981.562NEG2%
N202.7851.562NEG2.5591.562NEG-8%
N302.8281.562NEG2.9281.562NEG4%
N402.7671.562NEG2.851.562NEG3%
N502.7951.562NEG2.7671.562NEG-1%
N602.5531.562NEG2.8021.562NEG10%
ShampooN702.51.562NEG2.641.562NEG6%
N802.8371.562NEG2.7631.562NEG-3%
N902.5621.562NEG2.4061.562NEG-6%
N1002.5721.562NEG2.6631.562NEG4%
N1102.6481.562NEG2.7051.562NEG2%
N1202.6791.562NEG2.6471.562NEG-1%
N1302.7031.562NEG2.4711.562NEG-9%
Brown DyeN1402.7561.562NEG2.5831.562NEG-6%
N1502.411.562NEG2.51.562NEG4%
N1602.6181.562NEG2.3261.562NEG-11%
N1702.6781.562NEG2.571.562NEG-4%
N1802.5591.562NEG2.3751.562NEG-7%
N1902.811.562NEG2.4551.562NEG-13%
N2002.941.562NEG2.4681.562NEG-16%
N2102.7311.562NEG2.6051.562NEG-5%
N2202.7651.562NEG2.4271.562NEG-12%
N2302.7721.562NEG2.5151.562NEG-9%
N2402.5281.562NEG2.5261.562NEG0%
Pre-Cosmetic TreatmentPost-Cosmetic Treatment
TreatmentUniqueIDCocaineGC-MSpg/mgELISARaw ODCal.ODELISAP/NELISARaw ODCal.ODELISAP/N%Change(ELISARaw OD)
N2502.721.562NEG2.4841.562NEG-9%
N2602.4871.562NEG2.6111.562NEG5%
N2702.5681.562NEG2.5121.562NEG-2%
N2802.7541.562NEG2.7111.562NEG-2%
N2902.8531.562NEG2.8751.562NEG1%
BleachBlondeN3002.7441.562NEG2.8621.562NEG4%
N3102.7121.562NEG2.6441.562NEG-3%
N3202.9111.562NEG2.7851.562NEG-4%
N3302.4721.562NEG2.5791.562NEG4%
N3402.7371.562NEG2.7181.562NEG-1%
N3502.6741.562NEG2.6541.562NEG-1%
N3602.7091.562NEG2.7751.562NEG2%
N3702.4691.348NEG2.3271.348NEG-6%
N3802.4861.348NEG2.41.348NEG-3%
N3902.4451.348NEG2.5041.348NEG2%
N4002.4391.348NEG2.561.348NEG5%
N4102.4361.348NEG2.2711.348NEG-7%
PermN4202.5741.348NEG2.5191.348NEG-2%
N4302.5781.348NEG2.5731.348NEG0%
N4402.6721.348NEG2.521.348NEG-6%
N4502.5741.348NEG2.4961.348NEG-3%
N4602.4521.348NEG2.3031.348NEG-6%
N4702.5781.348NEG2.5311.348NEG-2%
N4802.4461.348NEG2.5891.348NEG6%
N4902.4811.348NEG2.3931.348NEG-4%
RelaxerN5002.3381.348NEG2.2511.348NEG-4%
N5102.5431.348NEG2.31.348NEG-10%
N5202.3811.348NEG2.3551.348NEG-1%
N5302.5761.348NEG2.1611.348NEG-16%
N5402.5831.348NEG2.2811.348NEG-12%
N5502.5281.348NEG2.3231.348NEG-8%
N5602.2941.348NEG2.1391.348NEG-7%
N5702.2231.348NEG2.3171.348NEG4%
N5802.3651.348NEG2.2481.348NEG-5%
N5902.4751.348NEG2.2281.348NEG-10%
N6002.3591.348NEG2.3961.348NEG2%

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image contains the logo for Quest Diagnostics. The logo consists of a stylized letter "Q" on the left, followed by the words "Quest" and "Diagnostics" on the right. The letter "Q" is designed with a circular shape and a break in the line, creating a sense of movement. The word "Quest" is larger and bolder than the word "Diagnostics", which is placed below it.

{25}------------------------------------------------

Image /page/25/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized "Q" symbol on the left, followed by the text "Quest" and "Diagnostics" on the right. The "Q" symbol is a circular shape with an arrow pointing inward, and the text is in a simple, sans-serif font.

Quest Diagnostics HAIRCHECK-DT (Cocaine) 510(k) Summary Page 23 of 27

Conclusions

The results of this study agree with the extensive published literature15-24 that the use of various cosmetic hair treatments can potentially reduce the amount of drug and drug metabolites detected in hair specimens. In this study, certain cosmetic treatments did lead to a substantial change in raw optical density values obtained in the ELISA, i.e. - relaxing and bleaching, due to loss of cocaine during the treatment. It is possible that a cosmetic hair treatment could cause a negative hair sample to read positive hair sample to read negative. Bleaching and relaxing had the most significant effect, decreasing the extent of positivity of positive specimens. None of the treatments tested significantly impacted the negative specimens.

Variance in how much drug and/or drug metabolites are incorporated into hair as a result of drug use is dependent upon the characteristics of the donor and hair specimen (e.g. age, texture, shape (thickness), density, melanin content, and segment of hair tested). Analogously, cosmetic treatments may also have a varying degree of effect on the drug concentration measured in hair specimens for the same reasons i.e., the effect of the treatment is impacted by the individual characteristics of each hair specimen (e.g. degree of penetration into the hair, effect on hair structure/matrix). The variability in the effect of cosmetic treatments can be seen in the Percent Change (ELISA OD) column of the tables above with bleaching, dyeing, and relaxing exhibiting the most variable effect on the optical density (i.e. the detected cocaine concentration or how positive the specimen was after treatment). The effect of each cosmetic treatment on negative specimens demonstrated minimal variability between specimens. Additionally, despite the hair specimens being minced and mixed in this study prior to assignment into the Treatment groups, differences in drug concentration between the hair segments, due to inconsistent drug use, cannot be discounted as a contributor to the variability observed between specimens.

4.8 Shelf Life

Shelf life was determined using (3) kits lots on real time stability.

Evaluation Lots used to support 6 month kit shelf life claim

ValidationLotsConjugate DOM i.e.shortest datedcomponent in kitKitExpirationDateDate of Day 0TestingFinal Interval TestDates
-----------------------------------------------------------------------------------------------------------------------------------------------------------------

{26}------------------------------------------------

Image /page/26/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized "Q" symbol on the left, followed by the words "Quest" and "Diagnostics" on the right. The "Q" symbol is a circular shape with a tail extending from the bottom, creating a sense of movement. The word "Quest" is in a bold, sans-serif font, while "Diagnostics" is in a smaller, lighter font below it.

Evaluation lot405/10/201511/10/201505/27/201512/29/2015
Evaluation lot505/09/201511/09/201505/28/201511/30/2015
Evaluation lot605/10/201511/10/201505/29/201512/29/2015

Real Time Stability ELISA – this study was conducted monthly for ~7 months for Evaluation Lot 5, and out to ~8 months for Evaluation lots 4 and 6.

Cutoff calibrator and controls stability validation of cutoff calibrator / controls concentration(s) relative to target is performed every 3 months with 5 replicates via GC-MS. 3 lots of HairCheck Cocaine kits and 3 separate "batches" of cutoff calibrator and low and high controls have been evaluated to 13 months.

Open Vial - Stability was performed for all components including the cutoff calibrator and controls. The microplates were held at room temperature for 4 hours and then returned to 2-8°C and QC was performed 30 days later. For the cutoff calibrator and controls they are stored at -20°C so moved to 2-8°C for 4 hours then returned to -20°C and held for 30 days.

Conclusion

All 3 Stability Validation lots are stable up to ~7 and ~8 months beyond the Date of Manufacture. All 3 lots of calibrator and controls have been tested at the 13 month time point and passed required QC criteria +/-20% of target value. The accelerated stability equivalent for this product is 7.6 months. Real-time stability demonstrated for (3) lots - up to 7 months, establishing the shelf-life at 6 months.

{27}------------------------------------------------

Image /page/27/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized "Q" symbol on the left, followed by the text "Quest" on the top line and "Diagnostics" on the bottom line. The "Q" symbol is a circular shape with a break in the upper right quadrant, and an arrow pointing inward.

6.0 Standards and Guidance Documents

Design Control Guidance for Medical Device Manufacturers (March 11, 1997)

Deciding When to Submit a 510(k) for a Change to an Existing Device (K97-1) (January 10, 1997)

Guidance for Industry and FDA Staff. Format for Traditional and Abbreviated 510(k)s (August 12, 2005)

Guidance for Industry and FDA Staff, eCopy Program for Medical Device Submissions (December 3, 2015)

Guidance for Industry and FDA Staff, Refuse to Accept Policy for 501(k)s (August 4, 2015)

Guidance for Industry and FDA Staff, Use of Symbols on Labeling of for In Vitro Diagnostic Devices Intended for Professional Use (November 30, 2004)

Draft Guidance for Industry and FDA Staff, Premarket Submission and Labeling Recommendations for Drugs of Abuse Screening Tests (December 2, 2003) (Withdrawn)

Draft Guidance for Industry And FDA Staff, Guidance for Prescription Use Drugs of Abuse Assays Premarket Notifications (November 14, 2000)

7.0 References

    1. Bourland, J. A., Hayes, E. F., Kelly, R.C., Sweeney, S.A., and Hatab, M.M., Quantitation of cocaine, benzoylecgonine, cocaethylene, methylecgonine in human hair by positive ion chemical ionization (pici) gas chromatography -tandem mass spectrometry, Journal of Analytical Toxicology Vol 24, No. 7: 489-495, 2000
  • Sweeney, S. A., Kelly, R. C., Bourland, J. A., Johnson, T., Brown, W. C., Lee, H. and Lewis, E., Amphetamines in 2. hair by enzyme-linked immunosorbent assay, Journal of Analytical Toxicology Vol 22, No. 6: 418-424, 1998
    1. Baselt, R.C., and Cravey, R.H., Disposition of Toxic Drugs and Chemicals in Man, Fifth Edition, Chemical Toxicology Institute, Foster City, CA (2000), 676-679
    1. Clarke's Isolation and Identification of Drugs, 2nd ed. Moffat A.C., Jackson JV, Moss MS, Widdop B, eds. London: The Pharmaceutical Press. 1986
  • Analytical Aspects of Drug Testing, Deutsch DG. NEW YORK: Wiley-Interscience, 1989. 5.
  • Harkey, M.R., Anatomy and physiology of hair, Forensic Science International 63 (1993), 9-18 6.
    1. Henderson, G.L., Mechanisms of drug incorporation into hair, Forensic Science International 63 (1993), 19-29
    1. Kalasinsky, K.S., Magluilo, J. Jr., and Schaefer, T., Study of Drug Distribution in Hair by Infrared Microscopy Visualization, Journal of Analytical Toxicology Vol 18, October 1994, 337-341
  • Kintz, P., Tracqui, A., and Mangin. P., Opiate Concentration in Human Head, Axillary, and Pubic Hair, Journal of 9. Forensic Sciences, JFSCA. Vol. 38. No. 3. May 1993, pp. 657-662
    1. Sniegoski, L.T., and Welch, M.J., Interlaboratory Studies on the Analysis of Hair for Drugs of Abuse: Results from the fourth Exercise, Journal of Analytical Toxicology Vol 20, July/August 1996, 242-247
    1. Kintz, P., Ludes, B., and Mangin, P., Detection of Drugs in Human Hair Using Abbott ADx, with Confirmation by Gas Chromatography/Mass Spectrometry (GC/MS), Journal of Forensic Sciences, JFSCA. Vol. 37. No. 1. Jan 1992. pp 328-331.
    1. Kintz, P., Tracqui, A., and Mangin. P., Detection of drugs in human hair for clinical and forensic applications, International Journal of Legal Medicine, (1992) 105:1-4

Page -25

{28}------------------------------------------------

Image /page/28/Picture/0 description: The image shows the Quest Diagnostics logo. The logo consists of a stylized "Q" symbol on the left, followed by the words "Quest" and "Diagnostics" on the right. The "Q" symbol is a circular shape with a curved line extending from the top, resembling a stylized letter Q. The word "Quest" is in a bold, sans-serif font, while "Diagnostics" is in a smaller, lighter font below it.

    1. Martinez, F., Poet, T.S., Pillai, R., Erickson, J., Estrada, A.L. and Watson, R.R., Cocaine Metabolite (Benzoylecgonine) in Hair and Urine of Drug Users, , Journal of Analytical Toxicology Vol 17, May/June 1993, 138-142
    1. Mieczkowski, T., A research note: the outcome of GC/MS/MS confirmation of hair assays on 93 cannabinoid (+) cases, Forensic Science International 79 (1995), 83-91
    1. Welch, M.J., Sniegoski, L.T., Allgood, C.C., Habram, M. (1993) Hair analysis for drugs of abuse-evaluation of analytical methods, environmental issues, and development of reference materials. Journal of Analytical Toxicology, 17, 389-398.
    1. Rohrich, J., Zorntlein, J., Pötsch, L., Skopp, G., Becker, J. (2000) Effect of the shampoo Ultra Clean on drug concentrations in human hair. International Journal of Legal Medicine, 113, 102–106.
    1. Skopp, G., Pötsch, L., Moeller, M.R. (1997) On cosmetically treated hair-aspects and pitfalls of interpretation. Forensic Science International, 84, 43-52.
    1. Takayama, N., Tanaka, S., Kizu, R., Hayakawa, K. (1999) High performance liquid chromatography study on effects of permanent wave, dye and decolorant treatments on methamphetamine in hair. Biomedical Chromatography, 13, 257-314.
    1. Pötsch, L., Skopp, G. (1996) Stability of opiates in hair fibers after exposure to cosmetic treatment. Forensic Science International, 81, 95-102.
    1. Baeck, S., Han, E., Chung, H., Pyo, M. (2011) Effects of repeated hair washing and a single hair dyeing on concentrations of methamphetamine in human hairs. Forensic Science International, 206, 77– 80.
    1. Agius R (2014) Utility of coloured hair for the detection of drugs and alcohol. Drug Test Anal. 6 Suppl 1:110-119.
    1. Martins, L.F., Yegles, M., Thieme, D., Wennig, R. (2008) Influence of bleaching on the enatiomeric disposition of amphetamine-type stimulants in hair. Forensic Science International, 176, 38–41.
    1. Pritchett JS, Phinney KW (2015). Influence of chemical straightening on the stability of drugs of abuse in hair. J Anal Toxicol. 39:13-16.
    1. Jurado C, Kintz P, Menéndez M, Repetto M. (1997) "Influence of the cosmetic treatment of hair on drug testing." Int J Legal Med. 110 (3):159-63

7.0 Conclusion

The modified subject device Quest Diagnostics HairCheck-DT (Cocaine) assay submission number K152232 is substantially equivalent to the predicate device Quest Diagnostics HairCheck-DT (Cocaine) assay submission number K023626

{29}------------------------------------------------

Image /page/29/Picture/0 description: The image shows the logo for Quest Diagnostics. The logo consists of a stylized "Q" symbol on the left, followed by the text "Quest" in a bold, sans-serif font. Below "Quest" is the word "Diagnostics" in a smaller, lighter font. The logo is simple and modern, with a focus on clarity and readability.

This Page is Intentionally Left Blank

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).