K Number
K151770

Validate with FDA (Live)

Date Cleared
2015-11-20

(143 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Libra Pedicle Screw System is intended for immobilization of the posterior thoracic, lumbar and sacral/iliac spine (TI-S1/Ilium) in skeletally mature patients as an adjunct to the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

Device Description

The Libra Pedicle Screw System consists of longitudinal members (rods), anchors (screws) and fasteners in a variety of sizes to accommodate differing anatomic requirements.

AI/ML Overview

This document describes the FDA's clearance of the Libra Pedicle Screw System (K151770) based on substantial equivalence to predicate devices, rather than a study and acceptance criteria for a novel device's performance. Therefore, the requested information about device performance, sample sizes, ground truth establishment, expert adjudication, and MRMC studies is not directly available in this text.

However, I can extract information related to the performance data provided for substantial equivalence and the acceptance criteria implicitly met through comparison with predicates.

Implicit Acceptance Criteria and Reported Device Performance (Table)

Acceptance Criteria (Implied)Reported Device Performance
Mechanical Performance:
Static Compression Bending (as per ASTM F1717)The mechanical test results demonstrate that Libra Pedicle Screw System performance is substantially equivalent to the predicate devices. (Specifically, static compression bending was performed according to ASTM F1717).
Dynamic Compression Bending (as per ASTM F1717)The mechanical test results demonstrate that Libra Pedicle Screw System performance is substantially equivalent to the predicate devices. (Specifically, dynamic compression bending was performed according to ASTM F1717).
Static Torsion (as per ASTM F1717)The mechanical test results demonstrate that Libra Pedicle Screw System performance is substantially equivalent to the predicate devices. (Specifically, static torsion was performed according to ASTM F1717).
Biocompatibility:Not explicitly detailed in the provided text, but implied to be acceptable as part of regulatory requirements for Class III devices, likely based on material (Ti-6Al-4V ELI titanium alloy (ASTM F136)) and comparison to predicate devices utilizing similar materials.
Sterilization:Not explicitly detailed in the provided text, but implied to be acceptable as part of regulatory requirements for Class III devices.
Design Characteristics:The Libra Pedicle Screw System possesses the same technological characteristics as one or more of the predicate devices. These include: intended use, basic design (rod and screw configuration), material (titanium alloy), and sizes (dimensions are comparable to those offered by the predicate systems). The fundamental scientific technology is the same as previously cleared devices.
Intended Use & Indications: (Substantially Equivalent to Predicates)The Libra Pedicle Screw System is intended for immobilization and stabilization of the posterior thoracic, lumbar and sacral/iliac spine (T1-S1/Ilium) in skeletally mature patients as an adjunct to fusion for indications like degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudarthrosis, and/or failed previous fusion.

Since this is a 510(k) summary for a substantial equivalence determination for a medical device (pedicle screw system), the "study" is primarily focused on bench testing to demonstrate that the new device performs as safely and effectively as legally marketed predicate devices, not clinical studies with human performance metrics as one might see for AI/software devices. Therefore, many of your specific questions are not directly applicable or answerable from this document.

Here's an attempt to address your other points based on the provided text, acknowledging the nature of a 510(k) submission for a physical implantable device:


2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: The document does not specify the exact number of Libra Pedicle Screw System constructs tested for mechanical performance. It generally refers to "worst case Libra Pedicle Screw System constructs." For a 510(k) for a physical device, this typically involves a representative number of samples (e.g., typically 3 to 5 per test condition) to satisfy the standard (ASTM F1717). The specific number is usually detailed in the full 510(k) submission, not summarized at this level.
  • Data Provenance: The data is from bench testing (mechanical testing), conducted by the manufacturer, Spinal Balance Inc., to evaluate the physical components of the device. It is not clinical data, and thus concepts like "country of origin of the data" or "retrospective/prospective" do not apply in the same way they would for clinical or AI data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable in this context. For mechanical bench testing, "ground truth" is established by the specifications and pass/fail criteria of the referenced ASTM standards (ASTM F1717). Engineers and technicians conduct the tests and determine if the results meet the criteria, comparing them against the performance of similar predicate devices. There isn't "expert adjudication" of test results in the sense of clinical image review for diagnostic accuracy.

4. Adjudication method for the test set

  • Not applicable. See point 3. Testing against an industry standard (ASTM F1717) provides objective pass/fail criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a physical spinal implant, not an AI software. Therefore, MRMC studies and the concept of "human readers improving with AI assistance" are irrelevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used

  • For mechanical testing, the "ground truth" is adherence to the performance standards and specifications outlined in ASTM F1717, and the comparative performance to predicate devices as demonstrated by their own cleared mechanical testing data. The goal is to show equivalent, safe, and effective mechanical properties.

8. The sample size for the training set

  • Not applicable. This is a physical medical device, not a machine learning model, so there is no training set in the AI sense.

9. How the ground truth for the training set was established

  • Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 20, 2015

Spinal Balance, Incorporated % Karen E. Warden, Ph.D. BackRoads Consulting, Incorporated P.O. Box 566 Chesterland, OH 44026

Re: K151770

Trade/Device Name: Libra Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH Dated: October 27, 2015 Received: October 28, 2015

Dear Dr. Warden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Karen E. Warden, Ph.D.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K151770

Device Name Libra Pedicle Screw System

Indications for Use (Describe)

The Libra Pedicle Screw System is intended for immobilization of the posterior thoracic, lumbar and sacral/iliac spine (TI-S1/Ilium) in skeletally mature patients as an adjunct to the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Date:29 June 2015
Sponsor:Spinal Balance Inc.
Nitschke Technology Commercialization Complex (NTCC)1510 N. Westwood Ave
Toledo OH 43606
440.530.5940
Sponsor Contact:Anand Agarwal, MD, President and CEO
510(k) Contact:Karen E. Warden, PhD
BackRoads Consulting Inc.
PO Box 566
Chesterland, OH 44026Office: 440.729.8457
Proposed Trade Name:Libra Pedicle Screw System
Common Name:Posterior pedicle screw system
Device Classification:Class III
Classification Name:Pedicle screw spinal system
Regulation:888.3070
Device Product Code:MNI, MNH, NKB
Device Description:The Libra Pedicle Screw System consists of longitudinal members (rods),anchors (screws) and fasteners in a variety of sizes to accommodatediffering anatomic requirements.
Indications for Use:The Libra Pedicle Screw System is intended for immobilization andstabilization of the posterior thoracic, lumbar and sacral/iliac spine (TI-S1/Ilium) in skeletally mature patients as an adjunct to fusion for thefollowing indications: degenerative disc disease (defined as back pain ofdiscogenic origin with degeneration of the disc confirmed by history andradiographic studies), spondylolisthesis, trauma (i.e., fracture ordislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, orlordosis), tumor, pseudarthrosis, and/or failed previous fusion.
Materials:Libra components are manufactured from Ti-6Al-4V ELI titanium alloy(ASTM F136)
Predicate Devices:Primary: Tiger® Spine System (Corelink, LLC - K133369)Additional: CD Horizon® Spinal System (Medtronic Sofamor Danek -K143569), Radius® Spinal System (Stryker Spine - K101144) andFoundation Spinal System (Skeletal Design Partnership - K120074)
Reference Devices:Moss Miami™ Spinal System (DePuy AcroMed, Inc. - K022623), Optima™Spinal System (U&I Corporation - K051971)
Performance Data:Mechanical testing of worst case Libra Pedicle Screw System constructsincluded static and dynamic compression bending and static torsionaccording to ASTM F1717.The mechanical test results demonstrate that Libra Pedicle Screw Systemperformance is substantially equivalent to the predicate devices.
TechnologicalCharacteristics:The Libra Pedicle Screw System possesses the same technologicalcharacteristics as one or more of the predicate devices. These include:intended use (as described above)basic design (rod and screw configuration),material (titanium alloy),sizes (dimensions are comparable to those offered by the predicatesystems) andThe fundamental scientific technology of the Libra Pedicle Screw System isthe same as previously cleared devices.
Conclusion:The Libra Pedicle Screw System possesses the same intended use andtechnological characteristics as the predicate devices. Therefore LibraPedicle Screw System is substantially equivalent for its intended use.

{5}------------------------------------------------

N/A