K Number
K151680

Validate with FDA (Live)

Date Cleared
2016-02-12

(235 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Nitrile films form a barrier to body fluids and bloodborne Pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

AI/ML Overview

The provided document describes the acceptance criteria and the study results for the "Powder Free Nitrile Patient Examination Gloves, Blue Color" (subject device) which is seeking substantial equivalence to a predicate device.

Here's a breakdown of the requested information:

1. Table of acceptance criteria and the reported device performance:

CharacteristicsStandard (Acceptance Criteria)Device Performance (Subject Device)
DimensionASTM standard D 6319-10.
Length≥230mm234-243mm
Width (Small)70-90 mm78-85mm
Width (Medium)85-105mm95-99mm
Width (Large)100-120mm108-113mm
Width (X large)110-130 mm117-126mm
Thickness (Fingertip)≥0.05mm0.08-0.11 mm
Thickness (Palm)≥0.05mm0.08-0.12 mm
Physical PropertiesASTM standard D 6319-10.
Tensile strength (Before aging)≥14MPa (based on predicate comparison)20-23MPa
Tensile strength (After aging)≥14MPa18-21 MPa
Elongated rate (Before aging)≥500%560-600%
Elongated rate (After aging)≥400%530-580%
Freedom from pinholes• 21 CFR 800.20• ASTM standard D 6319-10.• Test method in accordance with ASTM D5151-06(Reapproved 2011)Passed Standard Acceptance Criteria (Tested in accordance with ASTM D5151 (Reapproved 2011) with acceptable results)
Powder ResidualASTM standard D 6319-10.Test method in accordance with D6124-06 (Reaffirmation 2011)below 2mg of residual powder (based on predicate comparison)0.1mg/glove (Results generated values below 2mg of residual powder)
BiocompatibilityPrimary Skin Irritation in rabbits (ISO 10993-10: Third Edition 2010-08-01)Under the conditions of the study, the device was a non-irritant.
Dermal sensitization in the guinea pig (ISO 10993-10: Third Edition 2010-08-01)Under the conditions of the study, the device was a non-sensitizer.

2. Sample size used for the test set and the data provenance:

The document does not explicitly state the sample sizes used for each specific test (e.g., number of gloves tested for dimensions, tensile strength, pinholes). However, it implies testing was conducted on samples derived from the "Powder Free Nitrile Patient Examination Gloves, Blue Color."

The data provenance is from China, as the submitter, Suqian Xingye Glove Co,Ltd, is located in Suqian City, Jiangsu Province, China. The studies appear to be prospective as they were conducted to obtain data for this 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This is not applicable to this submission. The "ground truth" for glove performance is established by standardized testing methods (ASTM, CFR, ISO) rather than expert consensus on medical images or diagnoses.

4. Adjudication method for the test set:

This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used for studies involving human interpretation (e.g., diagnostic image reading) where there might be disagreement among experts. For performance testing of medical gloves, the results are determined by objective measurements against established standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. MRMC studies are relevant for AI-powered diagnostic tools or decision support systems. This submission concerns the physical and biological properties of patient examination gloves, not an AI system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This is not applicable. There is no AI algorithm involved in the performance of patient examination gloves. The device's performance is standalone in the sense that its physical properties are measured directly.

7. The type of ground truth used:

The ground truth used for these studies are:

  • Standardized specifications: Primarily ASTM standards (e.g., D6319-10, D5151-06, D6124-06) and FDA regulations (21 CFR 800.20) for physical properties.
  • Biological safety standards: ISO 10993-10: Third Edition 2010-08-01 for biocompatibility.
  • These standards define the acceptable range or threshold for each characteristic (e.g., minimum length, maximum powder residual, non-irritant/non-sensitizer).

8. The sample size for the training set:

This is not applicable. "Training set" refers to data used to train machine learning models. This submission is for a physical medical device (gloves) and does not involve AI or machine learning.

9. How the ground truth for the training set was established:

This is not applicable, as there is no training set for a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 12, 2016

Suqian Xingye Glove Co., Ltd. c/o Mr. Chu Xiaoan Beijing Easylink Co., Ltd. Room F302 Bldg. 41. Jing Cheng Ya Ju Courtyard 6 of Southern Dou Ge Zhuang Beijing 100121 CHINA

Re: K151680

Trade/Device Name: Powder Free Nitrile Patient Examination Gloves, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: January 3, 2016 Received: January 11, 2016

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151680

Device Name

Powder Free Nitrile Patient Examination Gloves, Blue Color

Indications for Use (Describe)

Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 6 510(k) Summary

510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

" "The assigned 510(k) number is: K 151680

Premarket Notification [510(k)] Summary

1.0 Submitter:

Submitter's name :Suqian Xingye Glove Co,Ltd
Submitter's address :Dongwu Road,Economic DevelopmentZone,Suqian City,JiangsuProvince,223800,China
Phone number :0086-527-82860533
Fax number :0086-527-82860080
Name of contact person:Jian Zhong Deng
Date of preparation :2015-11-07

2.0 Name of the Device

Device Name:Powder Free Nitrile Patient ExaminationGloves, Blue Color
Proprietary/Trade name:Powder Free Nitrile Patient ExaminationGloves, Blue Color
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LZA

3.0 Predicate device

Device Name:Nitrile Powder Free Patient ExaminationGloves, Cloured (Blue)
Company name:Tangshan Zhonghong Pulin Plastic Co.,Ltd.
510(K) Number:K120970

4.0 Device Description:

  • 4.1 How the device functions: Nitrile films form a barrier to body fluids and bloodborne Pathogens

4.2 Scientific concepts that form the basis for the device

{4}------------------------------------------------

The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

Physical and performance characteristics such as design, 4.3 materials and physical properties:

Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

5.0 Device Intended Use (Indication for use):

Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6.0 Summary of the Technological Characteristics of the Device:

The Powder Free Nitrile Patient Examination Gloves, Blue Color, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 6319-10.
Length≥230mm234-243mm
WidthSmall 70-90 mm78-85mm
Medium 85-105mm95-99mm
Large 100-120mm108-113mm
X large 110-130 mm117-126mm
ThicknessFingertip ≥0.05mm0.08-0.11 mm
Palm ≥0.05mm0.08-0.12 mm
PhysicalPropertiesASTM standard D 6319-10.
Tensile strength(Before aging)20-23MPa
Tensile strength(After aging)≥14MPa18-21 MPa
Elongated rate(Before aging)≥500%560-600%
Elongated rate(After aging)≥400%530-580%
Freedom frompinholes• 21 CFR 800.20Passed StandardAcceptance Criteria
• ASTM standard D 6319-10.
Test method in accordance withASTM D5151-06(Reapproved 2011)
Powder ResidualASTM standard D 6319-10.0.1mg/glove
Test method in accordance withD6124-06(Reaffirmation 2011)
BiocompatibilityPrimary Skin Irritation in rabbitsUnder the conditionsof the study, thedevice was anon-irritant.
ISO 10993-10: Third Edition2010-08-01
Dermal sensitization in the guinea pigUnder the conditionsof the study, thedevice was anon-sensitizer
ISO 10993-10: Third Edition2010-08-01

{5}------------------------------------------------

7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:

The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. The Powder Free Nitrile Patient Examination Gloves, Blue Color made by Sugian Xingye Glove Co,Ltd met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing.

  • . Dimension per ASTM D6319-10,
  • Tensile strength (Before aging/ After aging) and Elongated rate (Before aging/ . After aging) per ASTM D6319-10,
  • Water leak test on pinhole per ASTM D6319-10 and per 21 CFR 800.20, ●
  • . Powder Residual tests per ASTM D6319-10(Reapproved 2011)
  • Biocompatibility test per ISO 10993-10: Third Edition 2010-08-01.

Substantial Equivalent Based on Assessment of Clinical Performance Data: 8.0 -

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

So determination of substantial equivalence is not based on an assessment of clinical performance data.

Features & DescriptionPredicate DeviceSubject DeviceResult of Comparison
CompanyTangshan ZhonghongPulin Plastic Co., Ltd.Suqian Xingye GloveCo,Ltd--
510(K)NumberK120970K151680
Product nameNitrile Powder FreePatient ExaminationGloves, Cloured (Blue)Powder Free NitrilePatient ExaminationGloves, Blue Colorsame
Product CodeLZALZAsame
SizeSmall/ Medium/Large/X largeSmall/ Medium/Large/X largesame
Intend for useNitrile Powder FreePatient ExaminationGloves, Cloured (Blue) isa disposable deviceintended for medicalpurposes that is worn onthe examiner's hand orfinger to preventcontamination betweenpatient and examiner.Powder Free NitrilePatient ExaminationGloves, Blue Color isa disposable deviceintended for medicalpurposes that is wornon the examiner'shand or finger toprevent contaminationbetween patient andexaminer.Substantiallyequivalent
Device DescriptionandSpecificationsMeets ASTM D6319-10Meets ASTMD6319-10Substantiallyequivalent

9.0 Substantial Equivalence Comparison:

{6}------------------------------------------------

DimensionsMeets ASTMD6319-10234 mm min for allsizesSubstantiallyequivalent
Dimensions--Length≥230mm min
Dimensions-- WidthMeets ASTM D6319-10Substantiallyequivalent
Small70-90 mmSmall 78-85 mm
Medium85-105mmMedium 95-99 mm
Large100-120mmLarge 108-113mm
X large110-130 mmX large 117-126 mm
Dimensions--ThicknessMeets ASTM D6319-10Substantiallyequivalent
Finger 0.05mm min.Palm 0.05mm min.Thickness (mm) min.Finger 0.08Palm 0.08
PhysicalPropertiesMeets ASTM D D6319-10Aging Before AfterSubstantiallyequivalent
Before aging/after agingElongation ≥500%Tensile Strength≥ 14MPaElongation(%)560-600530-580
TensileStrength(MPa)20-2318-21
Freedom fromPinholesMeets• 21 CFR 800.20• ASTM D6319-10Meets• 21 CFR 800.20• ASTM D6319-10Substantiallyequivalent
Test method in accordancewith ASTM D5151-06(Reapproved 2011)Tested in accordancewith ASTM D5151(Reapproved 2011)with acceptable results
ResidualPowderMeets ASTMD 6124-06(Reapproved 2011)Meets ASTMD 6124-06(Reapproved 2011)Substantiallyequivalent
below 2mg of residualpowderResults generatedvalues below 2mg ofresidual powder
Materials usedto fabricate thedevicesNitrileNitrileSubstantiallyequivalent
Dusting orDonningPowder:PUPU-120CSubstantiallyequivalent
Dusting orDonningPowder: nameSurface Coating AgentSurface Coating AgentSubstantiallyequivalent
Compareperformancedata supportingsubstantialequivalenceMeetsASTM D5151-06(Reapproved 2011)ASTM D6319-10ASTM D6124-06(Reapproved 2011)MeetsASTM D5151-06(Reapproved 2011)ASTM D6319-10ASTM D6124-06(Reapproved 2011)Substantiallyequivalent
Single PatientUseSingle Patient UseSingle Patient UseSubstantiallyequivalent
BiocompatibilityUnder the conditions of thisstudy, the test article was anon- irritant or non-sensitizerUnder the conditions ofthis study, the testarticle was a non-irritant or non-Substantiallyequivalent

{7}------------------------------------------------

sensitizer.
SKIN IRRITATION
DERMAL andSKIN IRRITATION
SENSITIZATIONDERMALand
STUDIES Meets ISOSENSITIZATION
10993-10:2002/Amd.1:2006STUDIES Meets ISO
10993-10: Third
Edition 2010-08-01.
Labeling for-Powder Free-Powder FreeSubstantially
the legally-Patient Examination-Patient Examinationequivalent
marketedGloveGlove
device to-Single Use Only-Single Use Only
which- Manufactured For:- Manufactured For:
substantial- Lot- Lot
equivalence is-Blue color-Blue color
claimed.- Non sterile- Non sterile

10.0 Conclusion:

It can be concluded that the Powder Free Nitrile Patient Examination Gloves, Blue Color is as safe, as effective, and performs as well as the predicate device, Nitrile Powder Free Patient Examination Gloves, Cloured (Blue), Tangshan Zhonghong Pulin Plastic Co., Ltd., K120970.

The conclusions drawn from the nonclinical tests demonstrate that the subject device is substantially equivalent to the predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.