K Number
K151423

Validate with FDA (Live)

Date Cleared
2016-02-19

(267 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APEX Transfer Set is intended to be used with the APEX Compounding System. The administration sets are to be used in a sterile compounding environment within a pharmacy to provide a fluid path for the compounding of ingredients from multiple source containers into a final solution.

Device Description

The Apex Compounding System Transfer Set consists of a valved manifold with micro, macro and flex source lines. The proximal end of the source line consists of a luer lock connector which allows for connection to a transfer spike or connector depending upon the source container being used. The outlet tubing connects each valve, through the manifold body and consists of two sections of pump tubing just before ending at a final container connector, a luer adapter, to connect to a final container bag for eventual patient administration. There are 26 source lines on the manifold which correspond to the device's ability to support 26 source containers. Each individual line is labeled with a unique numeric identifier and barcode for traceability and scanning.

AI/ML Overview

The document provided describes the APEX Compounding System Transfer Set, an intravascular administration set (Class II, Product Code: LHI, NEP). It primarily focuses on demonstrating substantial equivalence to a predicate device through performance testing. The information requested aligns with studies typically conducted for AI/machine learning medical devices, which this product is not. However, I can extract the relevant information regarding acceptance criteria and performance testing conducted for this traditional medical device.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Visual InspectionFree of cracks, damage, kinked tubes, disconnected components, valves seated within 5% of alignment.
LeakageNo bubbles observed at joints with 10 psi air pressure (positive pressure). Vacuum pressure shall not fall in 30s while applying 28 in Hg (negative pressure).
Joint Strength (Static Load Tensile Force per ISO 8536-4)Peak pull force joints shall be > 5 lbf pulled to failure at 14 in/min (7 lbf for macro pump tube joints).
Valve TorqueMicro valve torque shall not exceed 14.2 oz in. Macro valve torque shall not exceed 42.5 oz in.
Vacuum TestVacuum pressure shall not fall in 30s while applying 28 in Hg.
Occlusion TestBubbles shall be observed from end after applying 0.5 psi air pressure while immersed.
Luer Gauging (ISO 594-2)(Performance not explicitly detailed, but implied meeting standard)
Barcode ScanScan shall match human readable number.
Magnetic Flux DensityShall be > 40 Gauss at a distance of 0.28".
Life Cycle Testing (Spallation and Particulate Contamination per USP 788 after simulated use)(Performance not explicitly detailed, but implied meeting standard)
Biocompatibility (ISO 10993-1, including Cytotoxicity, Delayed-type Hypersensitivity, Intracutaneous Reactivity, Acute Systemic Toxicity, Hemocompatibility, Material-Mediated Pyrogenicity)(Performance not explicitly detailed, but implied meeting standard)
Associate Device (APEX Compounding System) - Delivery AccuracyDextrose 70% at 20-25°C shall be ± 10% at 0.2 mL, ± 3% above 1 mL. Dextrose 70% at 15-30℃ shall be ± 10% at 0.2 mL, ± 6% above 1 mL. Other solutions at 15-30°C shall be ± 10% at 0.2 mL, ± 3% above 1 mL.
Associate Device (APEX Compounding System) - FitSet shall be able to be attached and removed with audible and tactile feedback and without damage to components.
Associate Device (B. Braun Spikes and Fluid Connector, B. Braun Final Containers) - Leakage (Source lines)No bubbles observed at connections after applying air pressure while immersed. Vacuum pressure shall not fall in 30s while applying 28 in Hg.
Associate Device (B. Braun Spikes and Fluid Connector, B. Braun Final Containers) - Leakage (Final container connector)No leakage when subjected to water pressure source of 20 psi.
Comparative Testing: Specific Gravity InterchangeMean concentration of unintended ingredient during compounding for the subject device shall be < mean concentration of unintended ingredient during compounding for the predicate device. Reported: Results demonstrate that the APEX Transfer Set is substantially equivalent to the predicate device. (This implies the criterion was met. Specific numerical improvement is not quantified in the provided text.)

Regarding AI/ML device specific information:

The provided document is for a traditional medical device, an "Intravascular administration set," and does not appear to involve any AI or machine learning components. Therefore, many of the requested items (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set size, ground truth for training set) are not applicable to this type of device and are not present in the given text.

Since this is not an AI/ML device, the following points are not directly addressed by the provided text:

  1. Sample sizes used for the test set and the data provenance: Not applicable for a non-AI/ML device. Performance testing for this device would typically involve a specific number of units tested per batch or according to a sampling plan for verification and validation, but these details are not provided in the summary.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for this device involves defined physical and functional criteria, not expert consensus on interpretations.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For this device, the "ground truth" is defined by the physical/functional specifications and performance standards (e.g., "no bubbles observed," "> 5 lbf pull force," "± 10% delivery accuracy").
  7. The sample size for the training set: Not applicable as there is no training set for a non-AI/ML device.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 25, 2017

B. Braun Medical Inc. Ms. Tracy Maddock Sr. Regulatory Affairs Specialist 901 Marcon Blvd. Allentown, Pennsylvania 18109

Re: K151423

Trade/Device Name: APEX Compounding System Transfer Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: LHI. NEP Dated: January 13, 2016 Received: January 14, 2016

Dear Ms. Tracy Maddock:

This letter corrects our substantially equivalent letter of February 19, 2016.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device comples with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Ryan -S

for Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151423

Device Name

APEX Compounding System Transfer Set

Indications for Use (Describe)

The APEX Transfer Set is intended to be used with the APEX Compounding System. The administration sets are to be used in a sterile compounding environment within a pharmacy to provide a fluid path for the compounding of ingredients from multiple source containers into a final solution.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

5. 510(k) SummaryK151423
SUBMITTER:B. Braun Medical Inc.901 Marcon BoulevardAllentown, PA 18109-9341610-266-0500
Contact: Tracy Maddock, RACSr. Regulatory Affairs SpecialistPhone: (610) 596-2545Fax: (610) 266-4962E-mail: Tracy.Maddock@bbraun.com
DATE:February 18, 2016
DEVICE NAME:APEX Compounding System Transfer Set
COMMON ORUSUAL NAME:IV Fluid Transfer Set
DEVICECLASSIFICATION:Class II per 21 CFR 880.5440 Intravascular administration setProduct Code: LHI, NEPClassification Panel: General Hospital
PREDICATE DEVICE:Exacta-Mix 2400 Compounding System Administration set, BaxaCorporation K002705, Class II, LHI, 880.5440.

DESCRIPTION:

The Apex Compounding System Transfer Set consists of a valved manifold with micro, macro and flex source lines. The proximal end of the source line consists of a luer lock connector which allows for connection to a transfer spike or connector depending upon the source container being used. The outlet tubing connects each valve, through the manifold body and consists of two sections of pump tubing just before ending at a final container connector, a luer adapter, to connect to a final container bag for eventual patient administration. There are 26 source lines on the manifold which correspond to the device's ability to support 26 source containers. Each individual line is labeled with a unique numeric identifier and barcode for traceability and scanning.

{5}------------------------------------------------

INDICATIONS FOR USE:

The APEX Transfer Set is intended to be used with the APEX Compounding System. The administration sets are to be used in a sterile compounding environment within a pharmacy to provide a fluid path for the compounding of ingredients from multiple source containers into a final solution.

SUBSTANTIAL EQUIVALENCE:

The APEX Compounding System Transfer Set is substantially equivalent to the Baxa Exacta-Mix 2400 Compounding System Administration Set (K002705).

Comparison of Technological Characteristics with the Predicate Device

The subject device has the same indications for use (to provide a fluid path for the compounding of ingredients from multiple source containers into a final solution) and the same principle of operation as the predicate device. The subject Transfer Set is similar in design with respect to components and materials of construction. Both devices are provided as sterile, nonpyrogenic, individually packaged devices to be used with their respective compounding devices.

A table summarizing the comparison between the APEX Transfer Set and the predicate device is provided below.

SUBJECT DEVICE:APEX COMPOUNDING SYSTEM TRANSFER SETPREDICATE DEVICE:EXACTA-MIX 2400 COMPOUNDING SYSTEMADMINISTRATION SET (K002705)
INDICATIONS FOR USE
The APEX Transfer Set is intended to be usedwith the APEX Compounding System. Theadministration sets are to be used in a sterilecompounding environment within a pharmacy toprovide a fluid path for the compounding ofingredients from multiple source containers into afinal solution.The Exacta-Mix 2400 compounding systemadministration sets, manufactured by BaxaCorporation, are ancillary devices used inconjunction with the Exacta-Mix 2400 compounddevice. The administration sets are used in thepharmacy to provide the fluid path for multiplesource ingredients into one final solution.
DESCRIPTION
Transfer set is a sterile, non-pyrogenic,disposable, individually packaged devicedesigned to work in conjunction with the APEXcompounding system to quickly and accuratelycompound multi-ingredient sterile solutions bywithdrawing requested amounts of sourceingredients from their containers in a userspecified sequence into a final container.Transfer set is a sterile, non-pyrogenic, disposable,individually packaged device designed to work inconjunction with the Baxa EM-2400 compoundingsystem to quickly and accurately compound multi-ingredient sterile solutions by withdrawing requestedamounts of source ingredients from their containersin a user specified sequence into a final container.
SUBJECT DEVICE:APEX COMPOUNDING SYSTEM TRANSFER SETPREDICATE DEVICE:EXACTA-MIX 2400 COMPOUNDING SYSTEMADMINISTRATION SET (K002705)
TECHNOLOGY
✓ Twenty-six (26) inlet sources✓ Twenty- four (24) inlet sources
✓ Micro or macro volume fluid transfer✓ Micro or macro volume fluid transfer
✓ 25 L max use, micro side / 300 L max use, macro side✓ 150 L max use
✓ Set requires attachment of transfer spikes or connector✓ Set requires attachment of inlet tubing sets
COMPONENTS AND MATERIALS
Manifold Body: PolycarbonateValve Set Body: Polycarbonate
Valve: TPVValve Core: Santoprene
Pump Tube: SiliconeOutlet Pump Tube: Silicone
Final Container Adapter: ABSLarge-Bore Connector: ABS
Final Container Adapter Cap: ABSConnector Cap: HDPE
Inlet Male Luer Lock Connector: ABS, AcrylicInlet Connector: ABS
Connector Cap: HDPECap: Polycarbonate
Source Line Tubing: PVCInlet tubing: PVC
Outlet Connector: Thermoplastic Elastomer
TESTING
Visual inspection Leakage Static Load Joint tensile strength Valve torque Associate device (Delivery accuracy, Specific gravity interchange, Leakage, Fit) Life cycle testing - Spallation and Particulate contaminationVacuum Occlusion Luer gauging Barcode scan Magnetic flux densityAccuracyFluid/air LeakagePrecipitant
STERILIZATION
Method: Ethylene oxide (EO)Method: Gamma
Sterility Assurance Level (SAL): 10-6Sterility Assurance Level (SAL): 10-6

{6}------------------------------------------------

The differences between the predicate device and the subject device include:

  • The subject device has pre-attached source lines with male luer lock connectors and barcoded flags indicating the line number. The predicate device consists of a valve set and separate inlet lines to be added by the user. Additionally, the user must attach barcode labels to the inlet lines.

{7}------------------------------------------------

  • The subject device contains twenty-six (26) source lines giving the user the ability to compound solutions using 26 source containers. The predicate device has the potential for twenty-four (24) source lines.

  • The subject device and the predicate device allow the compounding of both micro and macro volumes. The predicate device has one fluid channel through the valve set. The subject device has two (2) channels, a micro channel and a macro channel.

  • The sterilization method for the subject device is ethylene oxide whereas the sterilization method for the predicate device is gamma irradiation.

The differences, between subject device and predicate device, do not raise any new issues of safety and effectiveness.

Performance Testing

The proposed APEX Transfer sets were subjected to functional and performance testing to demonstrate that the sets perform as intended.

The following testing was performed on the subject device to the performance criteria listed or included in the referenced standard.

  • Visual inspection - free of cracks, damage, kinked tubes, disconnected components, valves seated within 5% of alignment
  • Leakage no bubbles observed at joints with 10 psi air pressure (positive pressure), vacuum . pressure shall not fall in 30s while applying 28 in Hg (negative pressure)
  • Joint strength
    • Static load tensile force (ISO 8536-4) O
    • Peak pull force joints shall be > 5 lbf pulled to failure at 14 in/min (7 lbf for macro O pump tube joints)
  • Valve torque micro valve torque shall not exceed 14.2 oz in, macro valve torque shall not ● exceed 42.5 oz in
  • Vacuum test vacuum pressure shall not fall in 30s while applying 28 in Hg ●
  • Occlusion test - bubbles shall be observed from end after applying 0.5 psi air pressure while immersed
  • Luer gauging (ISO 594-2) ●
  • . Barcode Scan - scan shall match human readable number
  • Magnetic flux density - shall be > 40 Gauss at a distance of 0.28"
  • Life cycle testing (Spallation and Particulate contamination) USP 788 after simulated use ●
  • Biocompatibility (ISO 10993-1)
    • Cytotoxicity (MEM Elution, ISO 10993-5) O
    • Delayed-type hypersensitivity (Guinea Pig Maximization, ISO 10993-10) O
    • Intracutaneous Reactivity (ISO 10993-10) O
    • Acute Systemic Toxicity (ISO 10993-11) O
    • Hemocompatibility (Hemolysis, ASTM F756, ISO 10993-4) O

{8}------------------------------------------------

  • Material-Mediated Pyrogenicity (Rabbit Pyrogen, USP <151>, ISO 10993-11) o
  • Associate device (APEX Compounding System) ●
    • Delivery accuracy o

Dextrose 70% at 20-25°C shall be ± 10% at 0.2 mL, ± 3% above 1 mL. Dextrose 70% at 15-30℃ shall be ± 10% at 0.2 mL, ± 6% above 1 mL. Other solutions at 15-30°C shall be ± 10% at 0.2 mL, ± 3% above 1 mL.

  • Fit -set shall be able to be attached and removed with audible and tactile feedback o and without damage to components
  • Associate device (B. Braun Spikes and Fluid Connector, B. Braun Final Containers) ●
    • Leakage o
      • Source lines - no bubbles observed at connections after applying air pressure while immersed, vacuum pressure shall not fall in 30s while applying 28 in He
      • . Final container connector - no leakage when subjected to water pressure source of 20 psi

The following comparative testing was performed on the subject and predicate device:

  • . Specific Gravity Interchange – mean concentration of unintended ingredient during compounding for the subject device shall be < mean concentration of unintended ingredient during compounding for the predicate device
    No clinical testing was performed as the transfer set does not require clinical studies to demonstrate substantial equivalence with the predicate device.

CONCLUSION:

Results of functional and performance testing conducted on the subject device demonstrate that the APEX Transfer Set is substantially equivalent to the predicate device.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.