K Number
K151306

Validate with FDA (Live)

Device Name
Solea
Date Cleared
2015-09-23

(128 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solea system is indicated for the following:

  • Ablation of hard tissue for caries removal and cavity preparation
  • Incision, Excision, Vaporization, Coagulation and Hemostasis of soft tissue in the oral cavity
  • Cutting, shaving, contouring and resection of oral osseous tissue (bone)
Device Description

The Solea system is a dental laser device previously cleared by the FDA for soft tissue dental indications (K123494) and hard tissue indications (K130420). The only changes from the previously cleared device are the addition of the osseous tissue indications. There have been no software changes or graphic changes to the Solea system. The osseous tissue settings are the same as the previously cleared device (K130420). The osseous settings have substantially equivalent fluence and irradiation as the hard tissue predicate devices. There are no other hardware or software changes to the Solea system device pending herein when compared to the device cleared under K123494 and K130420.

The Solea system is a mobile, cart-based dental treatment system that uses pulsed laser energy to cut and ablate hard tissue and to cut soft tissue in the oral cavity. The Solea system utilizes advanced CO2 laser technology with a wavelength of 9.25μm to safely and effectively perform ablation, incision, excision, vaporization, coagulation and hemostasis procedures.

AI/ML Overview

The provided text does not contain detailed acceptance criteria or a study that directly proves the device meets specific acceptance criteria in a quantitative manner as typically expected for assessing diagnostic accuracy (e.g., sensitivity, specificity).

However, it outlines a series of non-clinical tests that demonstrate the device's substantial equivalence to predicate devices in terms of safety and performance, indirectly serving as "acceptance criteria" for regulatory clearance. Since no clinical testing was performed or required due to substantial equivalence, metrics like sensitivity, specificity, or reader improvement with AI assistance are not applicable in this context.

Here's an interpretation of the non-clinical testing and how it relates to acceptance criteria fulfillment:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Electrical Safety (compliance with IEC 60601-1)The system passed electrical safety testing in accordance with requirements for IEC 60601-1.
Electromagnetic Compatibility (EMC) (compliance with IEC 60601-1-2)The system passed electromagnetic compatibility (EMC) testing to meet requirements for IEC 60601-1-2.
Laser Safety (compliance with IEC 60601-2-22 and IEC 60825-1)The system passed particular requirements for IEC 60601-2-22 and IEC 60825-1 for the safety of diagnostic and therapeutic laser equipment.
Cleaning and Sterilization (Sterility Assurance Level (SAL) of at least 10^-6, conventional autoclave cycles qualified to ANSI/AAMI ST79)The handpieces of the Solea system passed cleaning and sterilization validations for reusable medical devices based on the overkill approach to demonstrate sterilization cycle lethality as described in AAMI TIR12 to achieve a Sterility Assurance Level (SAL) of at least 10^-6<. The Solea system handpieces are designed for sterilization by exposure to moist heat under conventional autoclave cycles qualified to ANSI/AAMI ST79.
Software Verification and ValidationVerification and validation testing was conducted on the Solea software. All tests were completed successfully with respect to stated pass/fail criteria thereby deeming the device and software appropriate for its intended use.
Usability (as described in IEC 62366, acceptance criteria for user design validation met)Usability testing was conducted on the Solea system as described in IEC 62366. Based on the participant feedback and ratings of usability of the Solea system, all of the acceptance criteria for the user design validation have been met for the intended use.
Hard Tissue Performance (Thermal Effects) (substantially equivalent results to predicate systems)Performance data was collected from Bench Testing for hard tissue. Results show that hard tissue thermal effects are equivalent. The results show substantially equivalent results for the Solea system and predicate systems. This implies that the thermal effects caused by the Solea system on hard tissue are within acceptable limits and comparable to those of already approved predicate devices.
Overall Design, BiocompatibilityThe document states, "The Solea system meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety." While detailed results are not provided for overall design and biocompatibility in the same way as other tests, the statement indicates these requirements were met.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: The document does not provide specific sample sizes (e.g., number of handpieces for sterilization, number of samples for hard tissue testing, number of participants for usability). It generally describes "testing."
  • Data Provenance: The testing was "non-clinical testing" and "bench testing." This implies laboratory or simulated environments, rather than human clinical data. The country of origin for the data is not specified, but the submission is to the U.S. FDA, suggesting the testing was conducted under standards acceptable for U.S. regulatory review. The data is retrospective in the sense that it's reported after the tests were performed to support the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This is not applicable to the non-clinical testing described. The "ground truth" for these tests are objective measurements against established engineering, safety, and performance standards (e.g., IEC standards for electrical safety, AAMI standards for sterilization validation). Expert consensus or qualifications for ground truth establishment are not relevant for these types of tests.

4. Adjudication method for the test set

Not applicable. Testing against objective engineering and safety standards typically involves direct measurement and comparison to pass/fail criteria, not consensus-based adjudication as would be used for image interpretation.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The Solea system is a dental laser device for ablation, incision, excision, etc., not an AI-assisted diagnostic imaging device. Therefore, MRMC studies and the concept of "human readers improving with AI assistance" are not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. The Solea system is a physical device used by a human operator, not a standalone algorithm.

7. The type of ground truth used

The "ground truth" for the non-clinical tests were:

  • Established engineering standards and regulatory requirements: For electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and laser safety (IEC 60601-2-22, IEC 60825-1).
  • Established sterilization validation protocols: For cleaning and sterilization (AAMI TIR12, ANSI/AAMI ST79).
  • Software verification and validation criteria: For software performance.
  • Usability evaluation criteria: For usability (IEC 62366), presumably objective metrics and participant feedback aligned with the standard.
  • Comparative performance data against predicate devices: For hard tissue performance (thermal effects were shown to be "substantially equivalent" to predicate systems).

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI/machine learning algorithm requiring a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set mentioned or implied for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles a stylized human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Convergent Dental Incorporated % Ms. Carrie Hetrick Emergo 816 Congress Avenue, Suite 1400 Austin, Texas 78701

September 23, 2015

Re: K151306 Trade/Device Name: Solea Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: May 14, 2015 Received: May 18, 2015

Dear Ms. Hetrick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Joshua C. Nipper -S

For I

Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151306

Device Name Solea

Indications for Use (Describe)

The Solea system is indicated for the following:

  • Ablation of hard tissue for caries removal and cavity preparation
  • Incision, Excision, Vaporization, Coagulation and Hemostasis of soft tissue in the oral cavity
  • Cutting, shaving, contouring and resection of oral osseous tissue (bone)

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

for

Solea

1. Submission Sponsor

Convergent Dental, Inc. 2 Vision Drive Natick, MA 01760 USA Phone: (508) 500.5656 Contact: Mike Manor, Director QA/RA

2. Submission Correspondent

Emergo 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (720) 838.4113 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Carrie Hetrick, Senior Consultant, RA Email: project.management@emergogroup.com

3. Date Prepared

August 14, 2015

4. Device Identification

Trade/Proprietary Name:Solea
Common/Usual Name:Powered laser surgical instrument
Classification Name:Laser surgical instrument for use in general and plastic surgeryand dermatology
Classification Regulation:21 CFR §878.4810
Product Code:GEX
Device Class:Class II
Classification Panel:General and Plastic Surgery

5. Legally Marketed Predicate Device(s)

Convergent Dental, Inc., Solea, 510(k) Numbers: K123494, K130420 Biolase Technology, Inc., Waterlase MD, 510(k) Number: K091746 Fotona d.d., LightWalker Laser System, 510(k) Number: K101817

6. Device Description

{4}------------------------------------------------

The Solea system is a dental laser device previously cleared by the FDA for soft tissue dental indications (K123494) and hard tissue indications (K130420). The only changes from the previously cleared device are the addition of the osseous tissue indications. There have been no software changes or graphic changes to the Solea system. The osseous tissue settings are the same as the previously cleared device (K130420). The osseous settings have substantially equivalent fluence and irradiation as the hard tissue predicate devices. There are no other hardware or software changes to the Solea system device pending herein when compared to the device cleared under K123494 and K130420.

The Solea system is a mobile, cart-based dental treatment system that uses pulsed laser energy to cut and ablate hard tissue and to cut soft tissue in the oral cavity. The Solea system utilizes advanced CO2 laser technology with a wavelength of 9.25μm to safely and effectively perform ablation, incision, excision, vaporization, coagulation and hemostasis procedures.

7. Indication for Use Statement

The Solea system is indicated for the following:

  • । Ablation of hard tissue for caries removal and cavity preparation
  • Incision, Excision, Vaporization, Coagulation and Hemostasis of soft tissue in the oral ၊ cavity
  • -Cutting, shaving, contouring and resection of oral osseous tissue (bone)

8. Substantial Equivalence Discussion

The following table compares the Solea to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

The Convergent Dental, Inc. Solea system is a dental laser previously cleared by the FDA for hard tissue indications (K130420) and soft tissue indications (123494).

The Solea system included herein is the same device that was previously cleared by the FDA in K130420 (Solea, expanded hard tissue indications) and K123494 (Solea®, indications for soft tissue indications). There are no unique applications, materials or specifications presented herein. All the presented indications for use retain the same meaning as their equivalent indications cleared by the FDA.

The subject Solea system is substantially equivalent to the Biolase Technology, Inc. Waterlase MD (K091746) and the Fotona d.d., LightWalker Laser System (K101817) for the osseous tissue indications in terms of fluence, measured comparative performance data, and all features that affect safety or effectiveness.

The following table compares the subject Convergent Dental, Inc. Solea system to the Biolase Technology Inc. Waterlase MD, Fotona d.d. LightWalker Laser System Family, and the previously cleared Solea systems, with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence for intended uses.

{5}------------------------------------------------

ConvergentDental, Inc.ConvergentDental, Inc.ConvergentDental, Inc.BiolaseTechnology, Inc.Fotona d.d.
Trade NameSoleaSoleaSoleaWaterlase® MDLightWalkerLaser SystemFamily
510(k)NumberK152306K130420K123494K091746K101817
CommonnamePoweredLaser surgicalinstrumentPoweredLaser surgicalinstrumentPoweredLaser surgicalinstrumentPowered LasersurgicalinstrumentPowered Lasersurgicalinstrument
FDAClassificationNamesPoweredLaser surgicalinstrumentPoweredLaser surgicalinstrumentPoweredLaser surgicalinstrumentPowered LasersurgicalinstrumentDrill, bone,poweredSystem, Dental,Hydrokinetic,Carries Removal& CavityPreparationPowered Lasersurgicalinstrument
FDAClassificationcodesGEX, a ClassII deviceGEX, a ClassII deviceGEX, a Class IIdeviceGEX, MXF, DZI, aClass II deviceGEX, a Class IIdevice
PredicatescitedK130420 andK123494 forhard and softtissueindicationsK091746,K101817 forosseoustissueindicationsK091746,K101817 forhard tissueindicationsK123494 forsoft tissueindicationsK091320 forsoft tissueindicationsK031140,K071363,K090181 forhard tissueindicationsK093162 forhard tissueindications
Target UserGeneralpractitionerdentists andspecialistsGeneralpractitionerdentists andspecialistsGeneralpractitionerdentists andspecialistsGeneralpractitionerdentists andspecialistsGeneralpractitionerdentists andspecialists
Indicationsfor UseThe Soleasystem isindicated forthefollowing:- Ablationof hardtissueforcariesremovalandcavitypreparation- Incision,The Soleasystem isindicated forthefollowing:- Ablationof hardtissueforcariesremovalandcavitypreparation.- Incision,The Soleasystem isindicated forthe following:- Incision,- Excision,- Vaporization,- Coagulationand- Hemostasis of softtissue inthe oralcavity.GeneralIndications*Class I, II, III, IVand V cavitypreparationCaries removalHard tissuesurfaceroughening oretchingEnameloplasty,excavation ofpits and fissuresfor placement ofsealants** For use onThe LightWalkerEr:YAG laser,and itsaccessories, areintended foruse in dentistry,dermatologyand othersurgical areas inthe followingprocedures:In dentistry, for:- Intra-oralsoft tissuesurgery(incision,
ConvergentConvergentConvergentBiolaseFotona d.d.
Dental, Inc.Dental, Inc.Dental, Inc.Technology, Inc.
VaporizaVaporizapediatricablation
tion,tion,patientscoagulation
CoagulatCoagulatRoot Canal Hard)
ion andion andTissueLeukoplaki
HemostaHemostaIndicationsa
sis ofsis ofTooth- Pulpotomy
softsoftpreparation toas adjunct
tissue intissue inobtain access toto root
the oralthe oralroot canalcanal
cavitycavity.Root canalretreatmen
- Cutting,preparationt
shaving,including- Pulp
contourienlargementextirpation
ng andRoot canal- Removal of
resectiodebridementfibroma
n of oraland cleaning- Removal of
osseousRoot Canalgranulated
tissueDisinfectiontissue
(bone)Laser root canal- Caries
disinfection afterremoval,
endodonticcavity
instrumentationpreparatio
Bone Surgicaln, enamel
Indicationsroughening
Cutting, shaving,- Sulcular
contouring anddebrideme
resection of oralnt
osseous tissues- Tooth
(bone)preparatio
Osteotomyn to obtain
Endodonticaccess to
Surgery (Rootroot canal,
Amputation)root canal
Indicationsdebrideme
Flap preparationnt and
– incision of softcleaning,
tissue to prepareroot
a flap and- canal
expose thepreparatio
bone.n including
Cutting bone toenlargeme
prepare ant
window access- Cutting,
to the apexshaving,
(apices) of thecontouring
root(s).and
Apicoectomy -resection
amputation ofof oral
the root end.osseous
Root endtissue
preparation for(bone)
retrofillOsteotomy
ConvergentDental, Inc.ConvergentDental, Inc.ConvergentDental, Inc.BiolaseTechnology, Inc.Fotona d.d.
amalgam orcomposite.Removal ofpathologicaltissues (i.e.,cysts, neoplasmor abscess) andhyperplastictissues (i.e.,granulationtissue) fromaround the apexNOTE: Any tissuegrowth (i.e.,cyst, neoplasmor other lesions)must besubmitted to aqualifiedlaboratory forhistopathological evaluation.Soft TissueIndicationsincluding PulpalTissues*Incision,excision,vaporization,ablation andcoagulation oforal soft tissues,including:Excisional andincisionalbiopsiesExposure ofunerupted teethFibroma removalFlap preparation– incision of softtissue to preparea flap andexpose thebone.Flap preparation– incision of softtissue to preparea flap andexposeunerupted teethosseouscrownlengthening,osteoplastyApicectomysurgery.Removal ofsubgingivalcalculi inperiodontalpockets withperiodontitis byclosed or opencurettage
ConvergentConvergentConvergentBiolaseFotona d.d.
Dental, Inc.Dental, Inc.Dental, Inc.Technology, Inc.
tissue
impactions)
Frenectomy and
frenotomy
Gingival
troughing for
crown
impressions
Gingivectomy
Gingival incision
and excision
Hemostasis
Implant recovery
Incision and
drainage of
abscesses
Laser soft tissue
curettage of the
post-extraction
tooth sockets
and the
periapical area
during apical
surgery
Leukoplakia
Operculectomy
Oral
papillectomies
Pulpotomy
Pulp extirpation
Pulpotomy as an
adjunct to root
canal therapy
Root canal
debridement
and cleaning
Reduction of
gingival
hypertrophy
Removal of
pathological
tissues (i.e.,
cysts, neoplasm
or abscess) and
hyperplastic
tissues (i.e.,
granulation
tissue) from
around the apex
NOTE: Any tissue
growth (i.e.,
cyst, neoplasm
ConvergentConvergentConvergentBiolaseFotona d.d.
Dental, Inc.Dental, Inc.Dental, Inc.Technology, Inc.
or other lesions)must besubmitted to aqualifiedlaboratory forhistopathological evaluation.Soft tissuecrownlengtheningTreatment ofcanker sores,herpetic andaphthous ulcersof the oralmucosaVestibuloplastyLaserPeriodontalProceduresFull thicknessflapPartial thicknessflapSplit thicknessflapLaser soft tissuecurettageLaser removal ofdiseased,infected,inflamed andnecrosed softtissue within theperiodontalpocketRemoval ofhighly inflamededematoustissue affectedby bacteriapenetration ofthe pocket liningjunctionalepitheliumRemoval ofgranulationtissue from bonydefectsSulculardebridement(removal of
ConvergentConvergentConvergentBiolaseFotona d.d.
Dental, Inc.Dental, Inc.Dental, Inc.Technology, Inc.
diseased,
infected,
inflamed or
necrosed soft
tissue in the
periodontal
pocket to
improve clinical
indices including
gingival index,
gingival bleeding
index, probe
depth,
attachment loss
and tooth
mobility)
Osteoplasty and
osseous
recontouring
(removal of
bone to correct
osseous defects
and create
physiologic
osseous
contours)
Ostectomy
(resection of
bone to restore
bony
architecture,
resection of
bone for
grafting, etc.)
Osseous crown
lengthening
Laser assisted
new attachment
procedure
(cementum-
mediated
periodontal
ligament new-
attachment to
the root surface
in the absence of
long junctional
epithelium).
Removal of
subgingival
calculi in
periodontal
ConvergentDental, Inc.ConvergentDental, Inc.ConvergentDental, Inc.BiolaseTechnology, Inc.pockets withperiodontitis byclosed or opencurettage.Fotona d.d.
LaserclassificationClass 4 (IV)LaserProductClass 4 (IV)LaserProductClass 4 (IV)Laser ProductClass 4 (IV) LaserProductClass 4 (IV)Laser Product
Type ofLaserCO2 (CarbonDioxide)CO2 (CarbonDioxide)CO2 (CarbonDioxide)Er,Cr:YSGG(Erbium,Chromium:Yttrium,Scandium,Gallium, Garnet)Er:YAG (Erbium:Yttrium,Aluminum,Garnet)Nd:YAG(Neodymium-doped: Yttrium,Aluminum,Garnet)
Wavelength9.25μm(9250nm)9.25μm(9250nm)9.25μm(9250nm)2.78μm(2780nm)Er:YAG =2.94μm(2940nm)Nd:YAG =1.064μm(1064nm)
Fluence:Energy permm20.39J/mm²(hard tissue)1.13J/mm²(soft tissue)0.39J/mm²(hard tissue)1.13J/mm²(soft tissue)1.13J/mm²0.33J/mm²0.25J/mm²
Irradiance:Power permm26W/mm²(hard tissue)112.8W/mm2(soft tissue)6W/mm²(hard tissue)112.8W/mm2(soft tissue)112.80W/mm210W/mm²13W/mm²
OperatingModesAblationlaser: PulsedAiming laser:ContinuousAblationlaser: PulsedAiming laser:ContinuousAblation laser:PulsedAiming laser:ContinuousAblation laser:PulsedAiming laser:ContinuousAblation laser:PulsedAiming laser:Continuous
BeamDeliveryArticulatingArm (FreeSpace)ArticulatingArm (FreeSpace)ArticulatingArm (FreeSpace)FiberArticulatingArm (FreeSpace)
SterilizationSteamSteamSteamSteam AutoclaveSteam
MethodAutoclaveAutoclaveAutoclaveAutoclave
RF emissionsCISPR 11Group 1CISPR 11Group 1CISPR 11Group 1CISPR 11 Group1CISPR 11 Group1
EMCcomplianceCISPR 11Class ACISPR 11Class ACISPR 11 ClassACISPR 11 Class ACISPR 11 Class A

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

9. Non-Clinical Testing

The Solea system meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety. The results of the non-clinical testing confirm the output meets the design inputs and specifications. Bench testing was performed to

{12}------------------------------------------------

demonstrate substantial equivalence to the predicate devices in terms of safety and performance. The following non-clinical testing was performed:

. Electrical Safety Testing:

The system passed electrical safety testing in accordance with requirements for IEC 60601-1 medical electrical equipment.

● Electromagnetic Compatibility:

The system passed electromagnetic compatibility (EMC) testing to meet requirements for IEC 60601-1-2 medical electrical equipment.

● Laser Safety:

The system passed particular requirements for IEC 60601-2-22 and IEC 60825-1 for the safety of diagnostic and therapeutic laser equipment.

. Cleaning and Sterilization:

The handpieces of the Solea system passed cleaning and sterilization validations for reusable medical devices based on the overkill approach to demonstrate sterilization cycle lethality as described in AAMI TIR12 to achieve a Sterility Assurance Level (SAL) of at least 10-6. The Solea system handpieces are designed for sterilization by exposure to moist heat under conventional autoclave cycles qualified to ANSI/AAMI ST79.

● Software:

Verification and validation testing was conducted on the Solea software. All tests were completed successfully with respect to stated pass/fail criteria thereby deeming the device and software appropriate for its intended use.

● Usability:

Usability testing was conducted on the Solea system as described in IEC 62366. During the usability evaluation, dentists used the system to perform procedures on simulated tissues in a laboratory environment that replicates the intended deployment environment of the dental office. Based on the participant feedback and ratings of usability of the Solea system, all of the acceptance criteria for the user design validation have been met for the intended use.

. Bench Testing: Solea Hard Tissue Testing:

Performance data was collected from Bench Testing for hard tissue. Results show that hard tissue thermal effects are equivalent. The results show substantially equivalent results for the Solea system and predicate systems.

10. Clinical Testing

There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate devices. These types of devices, including the predicate devices, have been on the market for many years have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device.

The Solea system has the same or similar intended use, indications, principles of operation,

{13}------------------------------------------------

and technological characteristics as the predicate devices. Solea, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices in terms of intended use, design, materials, and function.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.