K Number
K123494

Validate with FDA (Live)

Device Name
SOLEA
Date Cleared
2013-02-11

(90 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solea system is indicated for the following:

  • Incision .
  • . Excision
  • . Vaporization
  • . Coagulation
  • Hemostasis .
    of soft tissue in the oral cavity.
Device Description

The Solea system is a mobile, cart-based dental treatment system that uses pulsed laser energy to cut soft tissue in the oral cavity. The Solea system utilizes advanced CO2 laser technology with a wavelength of 9.25μm to safely and effectively perform incision, excision, vaporization, coagulation, and hemostasis.

AI/ML Overview

The provided text describes a 510(k) submission for the Solea system, a CO2 laser for dental treatment. The submission aims to establish substantial equivalence to a predicate device, the Lutronic Corporation Spectra DENTA II Laser System (K091320), rather than demonstrating a new level of performance or effectiveness through a clinical trial. Therefore, the information typically associated with acceptance criteria for device performance and a study proving those criteria (as specified in your request) is not explicitly detailed in the context of this 510(k) summary.

However, based on the provided text, we can infer some "acceptance criteria" through the non-clinical testing performed to establish substantial equivalence in safety and performance. The study described is a bench testing comparison against the predicate device.

Here's an analysis of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a substantial equivalence submission, the "acceptance criteria" are implicitly that the Solea system's performance should be substantially equivalent to the predicate device in terms of safety and performance characteristics for soft tissue cutting. The reported device performance is that these criteria were met.

Acceptance Criteria (Implied for Substantial Equivalence to Predicate)Reported Solea System Performance
Electrical Safety (compliance with medical electrical equipment requirements)Passed electrical safety testing
Electromagnetic Compatibility (EMC) (compliance with medical electrical equipment requirements)Passed EMC testing
Laser Safety (compliance with particular requirements for diagnostic and therapeutic laser equipment)Passed laser safety requirements
Cleaning and Sterilization (validations for reusable medical devices)Handpieces passed cleaning and sterilization validations
Software Functionality (meet stated pass/fail criteria for intended use)All tests completed successfully, deemed appropriate for intended use
Usability (meet acceptance criteria for user design validation)All acceptance criteria for user design validation met based on participant feedback and ratings
Soft Tissue Performance (depth, width, length of cuts, histological assessment, effect on surrounding tissue) should be substantially equivalent to predicate deviceShowed substantially equivalent results to the predicate device in terms of depth, width, and length of cuts, histological changes, and effects on surrounding tissue.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Bench Testing (Soft Tissue): Not explicitly stated. The text only mentions "Performance data was collected using the Solea system and the predicate device" and "a quantitative analysis was conducted." It does not provide the number of samples, tissue types, or experimental repetitions.
  • Data Provenance: The bench testing was conducted by Convergent Dental, Inc. (the submitter) as non-clinical testing to support the 510(k) submission. The exact location or origin of the tissue samples is not specified. It is an internal, retrospective study for regulatory purposes.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • This type of information is generally not applicable for a bench test comparing physical characteristics of a laser.
  • For the histological assessment, implied in "histological assessment of changes in sample tissue," it would typically involve a pathologist. However, the number of pathologists or their qualifications are not specified in the document.

4. Adjudication Method for the Test Set

  • Not applicable as this was a bench test with objective measurements and comparative histological assessment, not a diagnostic imaging study requiring expert adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This question is not applicable as the Solea system is a physical laser device, not an AI algorithm. Its "standalone performance" refers to its physical output and effect on tissue, which was assessed in the bench testing. No human-in-the-loop element beyond operating the device is relevant here in the context of device performance.

7. The Type of Ground Truth Used

  • For the bench testing, the "ground truth" was established through direct physical measurement (depth, width, length of cuts) and histological assessment (microscopic analysis of tissue changes). These are objective measurements and expert pathological interpretation, respectively.

8. The Sample Size for the Training Set

  • Not applicable. This device is a laser system, not an AI/machine learning algorithm that requires a "training set."

9. How the Ground Truth for the Training Set was Established

  • Not applicable for the same reason as above.

{0}------------------------------------------------

K123 494 1/4

510(k) Summary

1. Submission Sponsor

FEB 1 1 2013

Convergent Dental, Inc. 2 Vision Drive Natick, MA 01760 USA Phone: (508) 500-5656 Contact: Jon Quillard, VP of Regulatory Affairs

2. Submission Correspondent

Emergo Group 611 West 5th Street, Third Floor Austin, TX 78701 Cell Phone: (720) 838.4113 Office Phone: (512) 327 9997 Fax: (512) 327.9998 Contact: Carrie Hetrick, Senior Consultant Email: project.management@emergogroup.com

3. Date Prepared

23 October 2012

4. Device Identification

Trade/Proprietary Name:Solea
Common/Usual Name:CO2 laser
Classification Name:Laser surgical instrument for use in general and plastic surgery and dermatology
Classification Regulation:878.4810
Product Code:GEX
Device Class:Class II
Classification Panel:General and Plastic Surgery

5. Predicate Devices

Lutronic Corporation Spectra DENTA II Laser System, 510(k) Number: K091320

6. Device Description

The Solea system is a mobile, cart-based dental treatment system that uses pulsed laser energy to cut soft tissue in the oral cavity. The Solea system utilizes advanced CO2 laser technology with a wavelength of 9.25μm to safely and effectively perform incision, excision, vaporization, coagulation, and hemostasis.

{1}------------------------------------------------

K123499 2/4

7. Intended Use

The Solea system is indicated for the following:

  • . Incision
  • . Excision
  • . Vaporization
  • . Coagulation
  • Hemostasis

of soft tissue in the oral cavity.

8. Comparison of Technological Characteristics

The following table compares the Solea to the predicate device with respect to the intended use, technological characteristics, and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

Convergent Dental, Inc.Lutronic Corporation
Trade NameSoleaSpectra DENTA II
510(k)NumberPendingK091320
CommonNameCO2 LaserCO2 Laser
ClassificationNames (FDA)Powered Laser Surgical InstrumentPowered Laser Surgical Instrument
FDAClassificationCodesGEXGEX
PredicatescitedK091320K050254, K030147, K935563, K963229,K991297, K021508, K081181
Target UserGeneral practitioner dentists andspecialistsGeneral practitioner dentists and specialists
Indicationsfor useThe Solea system is indicated for thefollowing:• Incision• Excision• Vaporization• Coagulation• Hemostasisof soft tissue in the oral cavity.The Spectra DENTA II Laser System isindicated for use in soft tissue dentalindications including periodontic proceduressuch as, but not limited to, removal ofdiseased or inflamed soft tissue in theperiodontal pocket (sulcular debridement),vaporization, gingivectomy-removal ofhyperplasias, gingivoplasty, papillectomy,vestibuloplasty, fibroma (nonmalignanttumor, mucosa, tongue), epulis, incision andexcision, removal of soft tissue, cysts, andtumors, and laser assisted new attachmentprocedure (cementum mediated periodontalligament new attachment to the root surfacein the absence of long junctional epithelium);Oral surgery such as frenectomy, frenumrelease, drainage (abscess), flap surgery,incisional and excisional biopsy, incision and

Table 5A – Comparison of Characteristics

{2}------------------------------------------------

K123494 3/4

Convergent Dental, Inc.Lutronic Corporation
Trade NameSoleaSpectra DENTA II
excision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity tumors and hemangiomas, salivary gland pathologies, preprosthetic gum preparation, leukoplakia; partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening.
LaserclassificationClass 4 (IV) Laser ProductClass 4 (IV) Laser Product
Type of LaserCO₂ (Carbon Dioxide)CO₂ (Carbon Dioxide)
Wavelength9.25µm (9250 nm)10.6µm (10600 nm)
Fluence:Energy permm²1.13J/mm²1.08J/mm²
Irradiance:Power permm²112.80W/mm²194.80W/mm²
OperatingModesAblation laser: PulsedAiming laser: ContinuousAblation laser: PulsedAiming laser: Continuous
BeamDeliveryArticulating Arm (Free Space)Articulating Arm (Free Space)
SterilizationMethodsSteam AutoclaveSteam Autoclave
RF emissionsCISPR 11 Group 1CISPR 11 Group 1
EMCcomplianceCISPR 11 Class ACISPR 11 Class A
HarmonicemissionsIEC 61000-3-2 Class AIEC 61000-3-2 Class A
Voltagefluctuations/flickeremissionsIEC 61000-3-3IEC 61000-3-3 Class A

9. Non-Clinical Performance Data

The Solea system meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety. The results of the non-clinical testing confirm the output meets the design inputs and specifications. Bench testing was performed to demonstrate substantial equivalence to the predicate device in terms of safety and performance. The following non-clinical testing was performed:

. Electrical Safety Testing:

The system passed electrical safety testing in accordance with requirements for medical electrical equipment.

Electromagnetic Compatibility: .

The system passed electromagnetic compatibility (EMC) testing to meet requirements for medical electrical equipment.

{3}------------------------------------------------

Laser Safety: .

The system passed particular requirements for the safety of diagnostic and therapeutic laser equipment.

. Cleaning and Sterilization:

The handpieces of the Solea system passed cleaning and sterilization validations for reusable medical devices.

. Software:

Verification testing was conducted on the Solea software. All tests were completed successfully with respect to stated pass/fail criteria thereby deeming the device and software appropriate for its intended use.

. Usability:

Usability testing was conducted on the Solea system. Based on the participant feedback and ratings of usability of the Solea system, all of the acceptance criteria for the user design validation have been met for the intended use.

. Bench Testing: Solea Soft Tissue Testing:

Performance data was collected using the Solea system and the predicate device. A quantitative analysis was conducted evaluating the depth, width and length of cuts, a histological assessment of changes in sample tissue, and effect on tissue surrounding the incision areas. The results show substantially equivalent results for both systems.

10. Clinical Testing

There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the use of the device. The nonclinical testing detailed in this submission supports the substantial equivalence of the device.

11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. The Solea device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.

It has been shown in this 510(k) submission that the difference between the Solea system and the predicate devices do not raise any questions regarding its safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the Solea system is substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, sterilization, biocompatibility, performance characteristics, and intended use. Solea, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Convergent Dental, Incorporated % Emergo Group. Incorporated Ms. Carrie Hetrick Senior Regulatory Consultant 611 West 5th Street Third Floor Austin, Texas 78701

February 11, 2013

Re: K123494

Trade/Device Name: Solea Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 05, 2012 Received: November 13, 2012

Dear Mr. Hetrick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{5}------------------------------------------------

Page 2 - Ms. Carrie Hetrick

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

FOR ·

Peter Di Rumm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K123494

Device Name: Solea

Indications for Use:

The Solea system is indicated for the following:

  • Incision .
  • . Excision
  • . Vaporization
  • . Coagulation
  • Hemostasis .

of soft tissue in the oral cavity.

Prescription Use ×

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R Ogden 2013.02.11 ,1 3:12:15 -05'00'

. .

(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ K123494

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.