K Number
K151209

Validate with FDA (Live)

Device Name
InterOss
Manufacturer
Date Cleared
2015-11-06

(184 days)

Product Code
Regulation Number
872.3930
Panel
Dental
Age Range
2 - 12
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

InterOss® small granules are recommended for:

  • Augmentation or reconstructive treatment of the alveolar ridge .
  • . Filling of infrabony periodontal defects
  • . Filling of defects after root resection, apicocectomy, and cystectomy
  • . Filling of extraction sockets to enhance preservation of the alveolar ridge
  • . Elevation of the maxillary sinus floor
  • . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
  • . Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

InterOss® large granules are recommended for:

  • . Augmentation or reconstructive treatment of the alveolar ridge
  • . Elevation of the maxillary sinus floor
  • . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
Device Description

InterOss® is a hydroxyapatite material derived from Australian bovine bone. The osteoconductive mineral structure is produced from bone through a multi-step purification process. Following placement in bony voids or gaps, InterOss acts as an osteoconductive scaffold for the ingrowth of adjacent viable bone. InterOss® gradually resorbs and is replaced with bone during the healing process.

InterOss® is available in granule form and is packaged in vials or a syringe-like applicator.

InterOss®, in the vial form, will be available to the United States market in 8 versions: Filled with 0.25g, 0.5g, 1.0g, 2.0g, or 5.0g of small granules (0.25 - 1.0mm) or filled with 0.5g, 1.0g, or 2.0g of large granules (1.0 - 2.0mm).

InterOss , in the syringe-like applicator form, will be available to the United States market in 6 versions: Filled with 0.25cc, 0.5cc, or 1.0cc of small granules (0.25 - 1.0mm) or filled with 0.5cc, 1.0cc, or 1.5cc of large granules (1.0mm - 2.0mm).

The syringe-like applicator was designed to deliver InterOss® granules more precisely to the intended treatment site without having to use other sterile instruments. The InterOss® granules can be wetted with either the patient's blood or sterile physiological saline solution back and then pressing down on the plunger. A removable filter cap prevents granules from falling out of the syringe-like applicator during storage and wetting.

During the manufacturing process of InterOss®, the granules are placed into a polymer syringe-like applicator or a glass vial, the vial or syringe is then capped with a rubber cap (the vial is also sealed with an aluminum cap), packaged into a polyethylene terephthalate tray, covered with a Tyvek lid, sealed, and then sterilized by gamma irradiation. The sterilized device is place in a protective package (outer box) along with its Instructions for Use and doctors notes.

All InterOss® products are supplied sterile and are intended for single use only.

AI/ML Overview

The provided text describes the regulatory clearance of a bone grafting material called InterOss® and does not contain information about a medical device that relies on AI or offers performance metrics that would typically be described with acceptance criteria like sensitivity, specificity, or AUC, as commonly seen in AI/ML device descriptions.

Instead, the document focuses on demonstrating substantial equivalence to predicate devices through various tests and comparisons. The "acceptance criteria" here are implied by the standards for bone grafting materials and the equivalence to the predicate devices.

Here's an attempt to structure the information based on the request, reinterpreting "acceptance criteria" as the properties and test outcomes required for substantial equivalence in this context:

1. Table of Acceptance Criteria and Reported Device Performance

Parameter / Acceptance Criteria (Implied by Equivalence to Predicate & Standards)InterOss® Reported Performance
Physical/Chemical Properties
Crystallinity ratio106%
Phase purity100% hydroxyapatite
Ca/P ratio1.67
MorphologyGranular and porous
Average granule size (small)568 µm (0.25 - 1.0mm range)
Average granule size (large)1625 µm (1.0 - 2.0mm range)
Pore interconnectivityPositive
Absence of heavy metalsConfirmed
Soluble Ca elution (small granules)0.533 ± 0.016 (3 days), 0.322 ± 0.012 (7 days), 0.366 ± 0.024 (14 days)
Soluble P elution (small granules)3.723 ± 0.004 (3 days), 6.277 ± 0.079 (7 days), 5.865 ± 0.019 (14 days)
Soluble Ca elution (large granules)0.513 ± 0.018 (3 days), 0.315 ± 0.009 (7 days), 0.351 ± 0.021 (14 days)
Soluble P elution (large granules)3.640 ± 0.085 (3 days), 6.241 ± 0.041 (7 days), 5.847 ± 0.019 (14 days)
Pore size (small granules)149.1660 Å
Pore size (large granules)47.3236 Å
Surface area (small granules)114.4452 m²/g
Surface area (large granules)109.0581 m²/g
Solubility (small granules)74.274 ± 13.599 µg/mm³
Solubility (large granules)70.247 ± 12.612 µg/mm³
Average pH (small granules)7.86 ± 0.05
Average pH (large granules)7.83 ± 0.05
Compressive strength (small granules)1.00 MPa
Compressive strength (large granules)1.17 MPa
Physiochemical properties (comparison to Bio-Oss®)Comparable (pore structure, microstructure, phase structure, chemical composition, residual organic substance)
Biocompatibility
Skin SensitizationNon-sensitizing (Stimulation Index = 1.03±0.26)
MutagenicityNon-mutagenic (Micronucleus and AMES test)
Intracutaneous ReactivityNo skin reactions
PyrogenicityNon-pyrogenic
HemolysisNo hemolytic reaction
Systemic Toxicity / CytotoxicityNon-toxic and non-cytotoxic (Acute Systemic Injection Test and Cytotoxicity Test)
Systemic Toxicity (leachables)Negative
SterilityPassed (SAL 1 X10⁻⁶)
Bacterial Endotoxin (LAL)Passed (<0.005 EU/mL or <0.232 EU/device)
Extractables from syringeAcceptable levels (e.g., <0.097 mg/device ethyl ether, no cell lysis, no intraplasmic granules in MEM test)
In-Vivo Performance (Animal Model)
Biocompatibility and healingIndistinguishable from predicate (Bio-Oss®) and in some cases superior to untreated defects. Low inflammation. No necrosis.
Graft resorption and boney integration (radiographic)Similar mean scores to predicate
Percent mineralized volume and density (micro-CT)No statistical differences between test article and predicate. Statistically greater amounts of bone in defect sites compared to empty controls.
Percent residual implant and bone marrow values (histomorphometry)No statistically significant differences between test article and predicate.
Microscopic examination (implant slides)Equivalent performance to predicate; similar average residual implant material; InterOss® had a higher mean percent bone value compared to Bio-Oss® (not statistically significant).

2. Sample size(s) used for the test set and data provenance:

  • Animal study (in-vivo performance): 4 treatment groups of beagle dogs. Three groups of nine dogs, and one group of six dogs. This refers to the study conducted on the device itself, not an AI test set.
  • Data Provenance: The animal study was conducted by SigmaGraft, Inc. (the manufacturer) or on their behalf, in the USA (presumably, as the FDA submission is from a US company). It is a prospective animal study as described (surgical procedures, observations, termination points).

3. Number of experts used to establish the ground truth for the test set and their qualifications:

  • Animal study (histology/pathology): Histological analysis was performed by Alizée Pathology, LLC. The specific number and qualifications of experts (e.g., veterinary pathologists with X years of experience) are not detailed in this summary.

4. Adjudication method for the test set:

  • The document does not describe an adjudication method for human reader interpretations. The evaluation methods for the animal study (micro-CT, semi-quantitative histology scoring, histomorphometry) are objective measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This document describes a medical device (bone grafting material), not an AI/ML software device. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was performed or is relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No. This document is not about an AI/ML algorithm.

7. The type of ground truth used:

  • Animal study (in-vivo performance): A combination of imaging (micro-CT) for mineralized volume/density and histopathology (undecalcified histology with semi-quantitative scoring and histomorphometry) for biocompatibility, tissue response, residual implant, and bone formation. This serves as the "ground truth" for the device's biological performance.

8. The sample size for the training set:

  • N/A. This is not an AI/ML device where a "training set" would be applicable. The closest analogy would be prior knowledge and predicate device data used in the design and development of InterOss®.

9. How the ground truth for the training set was established:

  • N/A. As above, this is not an AI/ML device. The "ground truth" for the predicate devices and general scientific understanding of bone grafting materials is based on decades of medical research, clinical experience, and regulatory standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three faces in profile, representing health, services, and people. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the top and left side of the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 6, 2015

SigmaGraft, Inc. Megan Holden R&D Manager 335 North Puente Street, Unit A Brea, California 92821

Re: K151209

Trade/Device Name: InterOss® Regulation Number: 21 CFR 872.3930 Regulation Name: Bone grafting material Regulatory Class: Class II Product Code: NPM Dated: October 29, 2015 Received: October 29, 2015

Dear Megan Holden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Erin I. Keith -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indication for Use:

InterOss® small granules are recommended for:

  • Augmentation or reconstructive treatment of the alveolar ridge .
  • . Filling of infrabony periodontal defects
  • . Filling of defects after root resection, apicocectomy, and cystectomy
  • . Filling of extraction sockets to enhance preservation of the alveolar ridge
  • . Elevation of the maxillary sinus floor
  • . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
  • . Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

InterOss® large granules are recommended for:

  • . Augmentation or reconstructive treatment of the alveolar ridge
  • . Elevation of the maxillary sinus floor
  • . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
Prescription UseAND/OROver-The-Counter
(Part 21 CFR 801 Subpart D)(Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

{3}------------------------------------------------

510(k) Summary

Submitter and Official Correspondent

SigmaGraft, Inc. Megan Holden 335 N. Puente St. Unit A. Brea, CA 92821 USA Email: meganholden@sigmagraft.com Phone: 714-525-0114 Fax: 714-525-0116

Device Information

Trade Name: InterOss® Common Name: Bone Grafting Material Classification Name: Bone Grafting Material Regulation Number: 21 CFR 872.3930 Product Code: NPM Device Class: Class II Date of Submission: 10/30/2015

Predicate Devices

The subject device is substantially equivalent to the following predicate device:

  • Primary Predicate Device o
    • . Bio-Oss® (K970321) manufactured by Geistlich Pharma AG.
  • Reference Predicate Device (Vial) o
    • OCS-B® (K113246) manufactured by Nibec Company, Limited. .
  • O Reference Predicate Device (Syringe)
    • . GEISTLICH Bio-Oss Pen (K120601) manufactured by Geistlich Pharma AG.

General Description

InterOss® is a hydroxyapatite material derived from Australian bovine bone. The osteoconductive mineral structure is produced from bone through a multi-step purification process. Following placement in bony voids or gaps, InterOss acts as an osteoconductive scaffold for the ingrowth of adjacent viable bone. InterOss® gradually resorbs and is replaced with bone during the healing process.

InterOss® is available in granule form and is packaged in vials or a syringe-like applicator.

{4}------------------------------------------------

InterOss®, in the vial form, will be available to the United States market in 8 versions: Filled with 0.25g, 0.5g, 1.0g, 2.0g, or 5.0g of small granules (0.25 - 1.0mm) or filled with 0.5g, 1.0g, or 2.0g of large granules (1.0 - 2.0mm).

InterOss , in the syringe-like applicator form, will be available to the United States market in 6 versions: Filled with 0.25cc, 0.5cc, or 1.0cc of small granules (0.25 - 1.0mm) or filled with 0.5cc, 1.0cc, or 1.5cc of large granules (1.0mm - 2.0mm).

The syringe-like applicator was designed to deliver InterOss® granules more precisely to the intended treatment site without having to use other sterile instruments. The InterOss® granules can be wetted with either the patient's blood or sterile physiological saline solution back and then pressing down on the plunger. A removable filter cap prevents granules from falling out of the syringe-like applicator during storage and wetting.

During the manufacturing process of InterOss®, the granules are placed into a polymer syringe-like applicator or a glass vial, the vial or syringe is then capped with a rubber cap (the vial is also sealed with an aluminum cap), packaged into a polyethylene terephthalate tray, covered with a Tyvek lid, sealed, and then sterilized by gamma irradiation. The sterilized device is place in a protective package (outer box) along with its Instructions for Use and doctors notes.

All InterOss® products are supplied sterile and are intended for single use only.

Indications for Use

InterOss® small granules are recommended for:

  • . Augmentation or reconstructive treatment of the alveolar ridge
  • . Filling of infrabony periodontal defects
  • . Filling of defects after root resection, apicocectomy, and cystectomy
  • Filling of extraction sockets to enhance preservation of the alveolar ridge .
  • . Elevation of the maxillary sinus floor
  • Filling of periodontal defects in conjunction with products intended for Guided Tissue . Regeneration (GTR) and Guided Bone Regeneration (GBR)

{5}------------------------------------------------

  • . Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
    InterOss® large granules are recommended for:

  • Augmentation or reconstructive treatment of the alveolar ridge .

  • . Elevation of the maxillary sinus floor

  • . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)

Product Testing

  • a. Physical
    The crystallinity ratio (106%), phase purity (100% hydroxyapatite), and Ca/P ratio (1.67) of InterOss® was assessed using X-ray diffraction. Additionally, phase analysis of InterOss® using Fourier Transform Infrared spectroscopy confirmed the presence of hydroxyapatite. The morphology (granular and porous), average granule size (568 um and 1625 um for small and large granules respectively), and pore interconnectivity (positive for pore interconnectivity) of InterOss was assessed using scanning electron microscopy. Inductively coupled plasma analysis was used to confirm the absence of heavy metals in InterOss . the elution of Ca (0.533 ± 0.016 at 3 days, 0.322 ± 0.012 at 7 days, and 0.366 ± 0.024 at 14 days) and P (3.723 ± 0.004 at 3 days, 6.277 ± 0.079 at 7 days, and 5.865 ± 0.019 at 14 days) from InterOss® small granules, and the elution of Ca (0.513 ± 0.018 at 3 days, 0.315 ± 0.009 at 7 days, and 0.351 ± 0.021 at 14 days) and P (3.640 ± 0.085 at 3 days, 6.241 ± 0.041 at 7 days, and 5.847 ± 0.019 at 14 days) from InterOss® large granules. The Barrett-Jonhner-Halenda method was used to determine the pore size (149.1660 Å for small granules and 47.3236 Å for large granules) of InterOss®. The surface area of InterOss® small (114.4452 m²/g) and large (109.0581 m²/g) granules was determined using the Brunauer-Emmett-Teller method. The solubility of InterOss® was found to be 74.274 ± 13.599 µg/mm3 for small granules and 70.247 ± 12.612 µg/mm3 for large granules and the average pH of InterOss® was found to be 7.86 ± 0.05 for small granules and 7.83 ± 0.05 for large granules. Lastly, the average compressive strength of InterOss® was found to be 1.00 MPa for small granules and 1.17 MPa for large granules.

{6}------------------------------------------------

Additionally, the physiochemical properties (pore structure, microstructure, phase structure, chemical composition, and residual organic substance) of InterOss® and Bio-Oss® were found to be comparable.

Biocompatibility b.

None of the tested animals exhibited abnormal clinical signs indicative of toxicity during test period for the Skin Sensitization test (LLNA; Stimulation Index = 1.03±0.26). InterOss® was determined to be non-mutagenic via the Micronucleus and AMES test. No skin reactions to InterOss® were observed for the Intracutaneous Reactivity Test. InterOss® was determined to be non-pyrogenic. The hemolysis test indicates that InterOss® does not cause a hemolytic reaction. InterOss was determined to be non-toxic and non-cytotoxic based upon the results of the Acute Systemic Injection Test and Cytotoxicity Test, respectively. Inteross® was also considered negative for signs of systemic toxicity due to leachable compounds. Additionally, InterOss® passed sterility test and Bacterial Endotoxin (LAL) tests.

Lastly, we performed an extractables test ( <0.097 mg/device ethyl ether, 0.04 mg/device bis(2ethylhexyl) phthalate, <0.00002 mg/device Chromium, <0.000038 mg/device Copper, 0.08 mg/device Silica, and 0.0008 mg/device Zinc), a second LAL test (<0.005 EU/mL or <0.232 EU/device), and a Minimum Essential Media test on extractables from the syringe (No cell lysis, intraplasmic granules) to evaluate the biocompatibility of the syringe like adapter.

  • c. Animal Testing
    This study was conducted for SigmaGraft, Inc., to evaluate the in-vivo performance of SigmaGraft's bone void filler, InterOss®, in a dental application. The relative performance of InterOss® was compared against the predicate device, Bio-Oss®, and an empty control in a welldocumented critical sized supra-alveolar ridge defect study in beagles. Four treatment groups of beagle dogs, three groups of nine and one group of six, underwent a surgical procedure to implant the predicate or test device into critical sized defects of the right and left mandible (P3 and P4 regions) or were left empty after drilling into the remaining bone immediately after tooth extraction. The predicate or test device was implanted into the defect site on Day 0. The animals remained on study for 4, 8, or 12 weeks.

Observations for morbidity, mortality, injury, and the availability of food and water were

{7}------------------------------------------------

conducted twice daily for all animals. Clinical observations were conducted daily during Week 1 and weekly thereafter. Body weights were measured and recorded prior to surgery and weekly. Physical examinations were conducted pretest to ensure that animals were healthy prior to being placed in the study. With the exception of animal numbers 109, 129 and 131 which underwent repair surgery due to exposure of the implant site resulting from the sutures tearing out through the gingiva, there were no unexpected or noteworthy clinical observations noted during the course of this study. All animals survived to study termination in good health.

At study termination, Faxitron images were obtained and necropsy examinations were performed. The right and left hemi-mandibles (along with the implantation sites and surrounding bone) were collected and analyzed using micro-computed tomography to determine percent mineralized volume and mineralized densities of the samples. Samples were then processed for undecalcified histology and analyzed microscopically by Alizée Pathology, LLC using semi-quantitative histology scoring and histomorphometry.

By all the parameters assessed in this twelve week study, biocompatibility and healing of InterOss® treated defects were indistinguishable from those treated with the Predicate, Bio Oss® and in some cases, superior to untreated defects. Radiographic comparison of graft resorption and boney integration demonstrated similar mean scores for both the Predicate and Test Articles. Likewise, no statistical differences were observed between the Test Article and Predicate with respect to percent mineralized volume and density. Moreover, when compared to the critical sized empty controls, both InterOss® and Bio-Oss® showed statistically greater amounts of bone present within the defect sites and appeared to help preserve the mesial and distal alveolar walls of the defect. Although some migration of the Test and Predicate materials were observed, the overall loss of this material did not appear to compromise the performance of the Test Article or Predicate as compared to leaving the defect unfilled. Histomorphometry also supported the similarity in performance of the Test Article to the Predicate as no statistically significant differences were observed with regards to percent residual implant and percent bone marrow values.

Microscopic examination of the implant slides also demonstrated equivalent performance between the Test Article and Predicate. In general, the average amount of residual implant material remaining within the defect site and surrounding soft tissue was similar for both the

{8}------------------------------------------------

Predicate and Test Article with a slight temporal decrease in value from 4 to 12 weeks (Group 3), with the exception of a slight increase in implant material at 12 weeks in the Test Article treatment over the amount found at 8 weeks. While not significantly different, InterOss® had a higher mean percent bone value as compared to Bio-Oss® at 4, 8 and 12 weeks. At 4, 8, and 12 weeks (Group 3), rare instances of polymorphonuclear cells (neutrophils) were noted and, more commonly, macrophages and multinucleated giant cells were found surrounding the implant material in the Test Article and Predicate Control treatments. The Negative Controls had a low incidence of neutrophils at 4 weeks, macrophages at 4, 8, and 12 weeks (Group 3), and a low incidence of multinucleated giant cells at 12 weeks (Group 3). The amount of inflammation present was on the low end of the spectrum and is indicative of a good biocompatibility response that would not interfere with healing. No evidence of necrosis was found in any of the treatment groups at 4, 8 or 12 weeks. Fibrosis scores tended to be equivalent or slightly higher, but not significant, in the Test Article treated sites as compared to the Predicate. No incidences of fatty infiltrate associated with fibrosis were observed for any treatment and neovascularization tended to be variable between the treatments and time points with no major differences or trends.

Basis for Substantial Equivalence

InterOss and predicates have similar physical and chemical structures. Both are porous, biocompatible bone grafts that facilitate the formation and mineralization of new bone by the osteoblast. As both products have same source of bone (bovine source) and similar process for removal of organic compounds, the product is substantially equivalent to the predicates.

ItemInterOss®Bio-Oss®OCS-B®
Intended useIdentical to the predicateUsed as an adjectivetherapy in restoring bonydefectsUsed as an adjectivetherapy in restoring bonydefects
TargetPopulationIdentical to the predicateHuman Oral, PeriodontalHuman Oral, Periodontal
Dosage formIdentical to the predicateGranule contained in singleGranule contained in
use containersingle use container
Granule Sizes0.25mm to 1mm or 1.0mmto 2.0mm granules0.25mm to 1mm or 1.0mmto 2.0mm granules0.2mm to 1mm or 1.0mmto 2.0mm granules
MaterialsIdentical to the predicateAnorganic derivedosteoconductivehydroxyapatite bonemineralAnorganic derivedosteoconductivehydroxyapatite bonemineral
Source BoneIdentical to the predicateBovine BoneBovine Bone
PhysicalMorphologyIdentical to the predicateTrabecular, interconnectingmacro and micro poresTrabecular,interconnecting macro andmicro pores
BiocompatibilityGenotoxicity Intractaneous reactivity Maximization and sensitization Pyrogen (LAL) Acute systemic toxicity Cytotoxicity Implantation SubchronicBiocompatibleBiocompatible
PerformanceIdentical to the predicateBone formationBone formation
Compatibilityw/ other devicesIdentical to the predicateCan be used with GTRmembraneCan be used with GTRmembrane
SterilizationProcessIdentical to the predicateSterile by GammaIrradiationSterile by GammaIrradiation
Anatomical sitesIdentical to the predicateSimilar based on chemicalanalysis, RD, RT-IR andICP analysisSimilar based on chemicalanalysis, RD, RT-IR andICP analysis
Non-pyrogenicIdentical to the predicateYesYes
AnatomicalSitesIdentical to the predicateOral, PeriodontalOral, Periodontal
Shelf LifeIdentical to the predicateDetermined byManufactures3 Years
RiskNon-risk as demonstratedby :-Virus inactivation processvalidation- Risk analysis- Cleaning Validation--

Table 1. Substantial Equivalence Comparison for vials

{9}------------------------------------------------

{10}------------------------------------------------

Table 2. Substantial Equivalence Comparison for Syringe-like Applicator

Subject DevicePredicate Device
510(k) #N/AK120601
Device NameInterOssGeistlich Bio-Oss Pen®
ManufacturerSigmaGraft, IncGeistlich Parma AG
Package TypeVial and Syringe-likeApplicatorSyringe-like Applicator
Materials forSyringe-likeApplicatorIdentical to the predicateGranules are placed in the polymer syringe-like applicator,packaged in a polyethylene terephthalate tray and coveredwith a tyvek lid, sealed and then sterilized by gammairradiation.
Intended UseIdentical to the predicateGeistlich Bio-Oss Pen®is intended for the following uses:- Augmentation or reconstructive treatment of thealveolar ridge- Filling of infrabony periodontal defects- Filling of defects after root resection, apicoectomy,and cystectomy- Filling of extraction sockets to enhance preservationof the alveolar ridge- Elevation of the maxillary sinus floor- Filling of periodontal defects in conjunction withproducts intended for Guided Tissue Regeneration(GTR) and Guided Bone Regeneration (GBR)- Filling of peri-implant defects in conjunction withproducts intended for Guided Bone Regeneration (GBR)

{11}------------------------------------------------

Syringe-likeapplicatorversion0.25cc, 0.5cc, or 1.0cc ofsmall granules (0.25 -1.0mm) or0.5cc, 1.0cc, or 1.5cc oflarge granules (1.0 -2.0mm)0.25g, 0.5g, or 0.7g of small granules (0.25-1.0mm) orfilled with 0.5g of large granules (1.0 - 2.0mm)
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Summary of Data to Support Substantial Equivalence

InterOss was the subjected to the full range of biocompatibility test recommended in the FDA's "Class II Special Controls Guidance Documents: Dental Bone Grafting Devices" and in accordance with ISO 10993. Test results confirmed product safety. Organic material has been removed from the product and product specifications have been established to limit protein content. InterOss® passed the TSE inactivation validation and virus inactivation test. Furthermore, the product is sterilized to achieve a sterility assurance level SAL 1 X10-°.

The results of these studies confirm the substantial equivalence of InterOss® to its predicate devices.

Conclusion

InterOss® presents the same types of potential risks to consumers as the predicate device Bio-Oss® and has controlled these risks in a similar manner. Biocompatibility test show that the device meets the requirements of those standards. Chemical and physical characterization tests, as well as pre-clinical data, show that the device is substantially equivalent. Comparison with the predicate device shows that the device has similar specification and performance. Therefore, it is concluded that InterOss® is substantially equivalent to the predicate device.

§ 872.3930 Bone grafting material.

(a)
Identification. Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.(b)
Classification. (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.