K Number
K113246

Validate with FDA (Live)

Device Name
OCS-B
Date Cleared
2013-01-16

(441 days)

Product Code
Regulation Number
872.3930
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

OCS-BTM cancellous and cortical granules are recommended for:

  • Augmentation or reconstructive treatment of the alveolar ridge.
  • Filling of infrabony periodontal defects.
  • Filling of defects after root resection; apicoectorny, and cystectorny
  • Filling of extraction sockets to enhance preservation of the alveolar ridge
  • Elevation of maxillary sinus floor
  • Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
  • Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
Device Description

OCS-B™ is a sterile, porous bone mineral matrix produced by the removal of organic compounds from bovine bone. It is supplied as cancellous (spongiosa) or cortical granules in a single use container, packaged in a secondary thermoform blister, and sterilized by yirradiation.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

CriterionAcceptance Standard (as implied by comparison to predicate)OCS-B™ Reported Performance
Intended UseUsed as an adjective therapy in restoring bony defectsUsed as an adjective therapy in restoring bony defects
Target PopulationHuman Oral, PeriodontalHuman Oral, Periodontal
Dosage FormGranules contained in single use containerGranules contained in single use container
Granule Sizes0.25mm to 1.0mm or 1.0mm to 2.0 mm granules0.2mm to 1.0mm or 1.0mm to 2.0 mm granules
MaterialAnorganic derived osteoconductive hydroxyapatite bone mineralAnorganic derived osteoconductive hydroxyapatite bone mineral
Source BoneBovine boneBovine bone
Physical MorphologyTrabecular, interconnecting macro and micro poresTrabecular, interconnecting macro and micro pores
BiocompatibilityBiocompatible (as demonstrated in published literature)Biocompatible, demonstrated by a full range of tests (genotoxicity, intracutaneous reactivity, maximization/sensitization, pyrogen, acute systemic toxicity, cytotoxicity, implantation, preclinical safety/efficacy, clinical case studies)
Performance (Bone Formation)Bone formationBone formation (demonstrated in animal studies and clinical case series)
Compatibility with other devicesCan be used with GTR membraneCan be used with GTR membrane
Sterilization ProcessSterile by Gamma irradiationSterile by Gamma irradiation (SAL 1 x 10^-9 maintained)
Chemical CompositionSimilar to predicateSimilar to Bio-Oss® based on chemical analysis, XRD, FT-IR, and ICP analysis
Anatomical SitesOral, PeriodontalOral, Periodontal
Non-pyrogenicYesYes
Shelf LifeDetermined by Manufacturer (predicate)3 years
Risk FactorsControlled risks similar to predicateControlled risks demonstrated by TSE inactivation process validation, virus clearance study, analysis of residual solvent, risk analysis, cleaning validation

The study that proves the device meets the acceptance criteria is a compilation of data supporting substantial equivalence to the predicate device, Bio-Oss®. Specifically, the submission highlights:

  • Chemical and Physical Characterization: OCS-B™ and Bio-Oss® were found to be comparable in their physical and chemical characteristics.
  • Animal Studies: Multiple animal studies showed both products promoted new bone growth and were resorbed at similar rates.
  • Clinical Case Series: A summary of seven individual case studies demonstrated defect healing and new bone formation sufficient for dental implant placement using OCS-B™.
  • Biocompatibility Testing: OCS-B™ underwent a full range of biocompatibility tests according to FDA's "Class II Special Controls Guidance Document: Dental Bone Grafting Devices" and ISO 10993.
  • Risk Control Measures: Validation of TSE inactivation, virus clearance, analysis of residual solvent, risk analysis, and cleaning validation were conducted.

2. Sample Sizes and Data Provenance

  • Test Set Sample Size:
    • Clinical Case Studies: 7 individual case studies for OCS-B™.
    • Animal Studies: "Several animal studies" were conducted. The exact number of animals or studies is not specified.
    • Biocompatibility Testing: Not specified, but a full range of tests indicates multiple samples/groups would have been used for each test.
  • Data Provenance:
    • The document implies data was generated directly by the sponsor or their contracted labs for OCS-B™.
    • The origin (e.g., country) of the animal studies and biocompatibility testing is not specified.
    • The clinical case studies are presented. While the country isn't explicitly stated, the sponsor is based in Korea.
    • The data is retrospective for the clinical case series, as they are presented as a summary of existing cases rather than from a prospective clinical trial. The animal studies and biocompatibility testing would have been specifically conducted for the submission, making them prospective in nature.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

  • Clinical Case Studies:
    • Experts: Not explicitly stated how many experts established the ground truth. However, the report mentions "histological evaluation" and "radiographs," implying examination by qualified dental/medical professionals.
    • Qualifications: Implied to be dental/medical professionals capable of interpreting radiographs and performing histological evaluations for assessing bone growth and defect healing. Specific qualifications (e.g., years of experience, specific certifications) are not detailed in the summary.
  • Animal Studies: The experts involved in assessing bone growth and resorption in animal studies are not specified.

4. Adjudication Method (Test Set)

  • Adjudication Method: Not specified for any of the studies mentioned. The clinical case series provides "histological and radiographic images demonstrate new bone growth," suggesting perhaps a single expert's assessment per case or an internal review without explicit mention of formal adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a MRMC comparative effectiveness study was not done. The submission focuses on demonstrating substantial equivalence to a predicate device through physical/chemical comparison, animal studies, and a clinical case series, rather than comparing human reader performance with and without AI assistance.

6. Standalone Performance Study (Algorithm Only)

  • No, a standalone (algorithm only) performance study was not done. The device (OCS-B™) is a physical bone grafting material, not an AI algorithm. Therefore, "algorithm only" performance is not applicable.

7. Type of Ground Truth Used

  • Clinical Case Studies:
    • Expert Consensus/Pathology/Imaging Data: Ground truth was established through a combination of histological evaluation (pathology) from core biopsies and interpretation of radiographs (imaging data), confirming new bone growth and defect healing.
  • Animal Studies: Ground truth for bone growth and resorption would have been based on similar methodologies, including histological analysis and potentially imaging or quantitative analysis of bone volume.
  • Biocompatibility Studies: Ground truth is based on established biological endpoints and validated assays (e.g., cell viability, immune response, genotoxicity, systemic toxicity, implantation site reaction).

8. Sample Size for the Training Set

  • Not Applicable. OCS-B™ is a physical bone grafting material, not a software algorithm that requires a "training set" in the context of machine learning. The studies described are for validation and characterization of the material itself.

9. How Ground Truth for the Training Set was Established

  • Not Applicable. As mentioned above, this is a physical medical device, not an AI/ML algorithm requiring a training set.

{0}------------------------------------------------

510(k) Summary

[as required by 807.92(c)]

Sponsor/Applicant

NIBEC Co., Ltd.

lwol electricity-electronic Agro-industrial Complex, 1127, Sinwol-ri,

lwol-myeon, Jincheon-gun, Chungcheongbuk-do, Korea

Phone : 82-10-2889-8590

Fax : 82-2-744-8732

Contact : Dr. Park, Yoon-Jeong

Date Prepared : JANUARY 07, 2013

Device Name and Identification

Proprietary Name : OCS-B™

Common / Usual Name : Bone Mineral Matrix

Anorganic Bovine Bone Grafting Material

Classification Name :

Bone Grafting Material

Animal Source Dental Bone Grafting Material

Predicate Device

Bio-Oss® bone grafting material (K871773, K952617, K970321, K033815)

Manufactured by :

Geistlich Pharma AG

Bahnhofstrasse 40

{1}------------------------------------------------

CH-6110 Wolhusen

Switzerland

Device Category/Class

Device Class : Class II

Regulation Number : 21 C.F.R. 872.3930

Product Code : NPM

Indication for use

OCS-BT® cancellous and cortical granules are recommended for:

  • Augmentation or reconstructive treatment of the alveolar ridge.
  • Filling of infrabony periodontal defects. -
  • Filling of defects after root resection; apicoectorny, and cystectorny -
  • Filling of extraction sockets to enhance preservation of the alveolar ridge -
  • -Elevation of maxillary sinus floor
  • Filling of periodontal defects in conjunction with products intended for Guided Tissue । Regeneration (GTR) and Guided Bone Regeneration (GBR)
    • Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

Device Description

OCS-B™ is a sterile, porous bone mineral matrix produced by the removal of organic compounds from bovine bone. It is supplied as cancellous (spongiosa) or cortical granules in a single use container, packaged in a secondary thermoform blister, and sterilized by yirradiation.

Basis for Substantial Equivalence

OCS-B™ and Bio-Oss® have a similar physical and chemical structure. Both are porous, biocompatible bone grafts that facilitate the formation and mineralization of new bone by the

{2}------------------------------------------------

osteoblast. As both products have same source of bone (bovine source) and similar process for removal of organic compounds, the product is substantially equivalent to Bio-Oss®

The following table summarizes the basis for the Sponsor's substantial equivalence determination :

Table 1 Substantial Equivalence Comparison

ITEMOCS-BTMBio-Oss®
Intended UseUsed as an adjective therapy in restoring bony defectsUsed as an adjective therapy in restoring bony defects
Target populationHuman Oral, PeriodontalHuman Oral, Periodontal
Dosage formGranules contained in single use containerGranules contained in single use container
Granule sizes0.2mm to 1.0mm or 1.0mm to 2.0 mm granules0.25mm to 1.0mm or 1.0mm to 2.0 mm granules
MaterialAnorganic derived osteoconductive hydroxyapatite bone mineralAnorganic derived osteoconductive hydroxyapatite bone mineral
Source boneBovine boneBovine bone
PhysicalMorphologyTrabecular, interconnecting macro and micro poresTrabecular, interconnecting macro and micro pores
BiocompatibilityBiocompatible, as demonstrated by :- Genotoxicity testing ( In vitro, In vivo )- Intracutaneous reactivity testing- Maximization and sensitization testing- Pyrogen testing- Acute systemic toxicity testing- Cytotoxicity testing- Implantation testing- Preclinical safety and efficacy testing- Clinical case studiesBiocompatible (as demonstrated in published literature)
PerformanceBone formationBone formation
Compatibilityw/other devicesCan be used with GTR membraneCan be used with GTR membrane
SterilizationProcessSterile by Gamma irradiationSterile by Gamma irradiation
ChemicalCompositionSimilar to Bio-Oss® based onchemical analysis, XRD, FT-IR andICP analysisSimilar to OCS-BTM based onchemical analysis, XRD, FT-IR andICP analysis
Anatomical sitesOral, PeriodontalOral, Periodontal
Non-pyrogenicYesYes
Shelf life3 yearsDetermined by Manufacturer
RiskNon-risk, as demonstrated by :
- TSE inactivation Process Validation
- Virus Clearance study
- Analysis of residual solvent
- Risk analysis
- Cleaning Validation

{3}------------------------------------------------

Brief Summary of Data Submitted

The Sponsor evaluated the performance characteristics of OCS-B™ and Bio-Oss® with a thorough chemical and physical characterization. The physical and chemical characteristics of the products were found to be comparable. Further, in several animal studies, both products were found to grow new bone and be subsequently resorbed at similar rates. Finally, in a clinical case series, use of OCS-BTM resulted in defect healing and formation of new bone of sufficient quality to obtain dental implant placement. The submission includes a summary of seven individual case studies of OCS-BTM. The patients were treated for intrabony periodontal defects. For each case study, the report includes baseline radiographs, radiographs at various time point, and core biopsy for histological evaluation. Histological and radiographic images demonstrate new bone growth.

OCS-B™ was the subject of the full range of biocompatibility tests recommended in the FDA's "Class II . Special Controls Guidance Document : Dental Bone Grafting Devices" and in accordance with ISO 10993. Organic material has been removed from the product, and product specifications have been established to limit protein content. Throughout the risk analysis for each production step, for example, cleaning validation, the removal of organic solvent, the risk control was conducted during the manufacturing process. In addition, the

{4}------------------------------------------------

TSE inactivation validation as well as virus inactivation study result was conducted. Further, the product is sterilized to achieve a sterility assurance level SAL 1 x 10-9

Based on the information presented herein, it has been demonstrated that OCS-BM is substantially equivalent to Bio-Oss®.

Conclusion

The OCS-B™ presents the same types of potential risks to consumers as the predicate device Bio-Oss®, and has controlled these risks in a similar manner. And biocompatibility tests and compatibility test show that the device meets the requirements of those standards. Literatures and post market experience show that the device is substantially equivalent. Comparison with the predicate device shows that the device has similar specification and performance.

Therefore, it is concluded that OCS-B™ are substantially equivalent to the predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 16, 2013

Nibec Company, Limited C/O Mr. Daniel Nam General Manager Pats Corporation 4568 West 181 Street, Suite 104 LOS ANGELES CA 90004

Re: K113246

Trade/Device Name: OCS-BTM Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPM Dated: January 7, 2013 Received: January 14, 2013

Dear Mr. Nam:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm1.

Sincerely vours.

Kwame O. Ulmer

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): K (13246

Device Name: OCS-B™

Indications for Use:

OCS-BTM cancellous and cortical granules are recommended for:

Augmentation or reconstructive treatment of the alveolar ridge. -

  • Filling of infrabony periodontal defects. -
  • Filling of defects after root resection, apicoectorny, and cystectorny -
  • Filling of extraction sockets to enhance preservation of the alveolar ridge -
  • Elevation of maxillary sinus floor
  • Filling of periodontal defects in conjunction with products intended for Guided Tissue . Regeneration (GTR) and Guided Bone Regeneration (GBR)
    • Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
Prescription UseXAND/OR
Subpart D)(21 CFR 801 Subpart C)
Over-The-Counter Use(Part 21 CFR 801

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runner DDS, MA2013.01.15
15:02:04 -05'00'

(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices510(k) Number;

§ 872.3930 Bone grafting material.

(a)
Identification. Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.(b)
Classification. (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.