The Reform Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).
The Reform Pedicle Screw System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The Reform Pedicle Screw System is also intended for non-cervical pedicle screw fixation (T1-S1/ilium) for the following indications: degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); spinal tumor; pseudoarthrosis; and failed previous fusion.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Reform Pedicle Screw System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Reform Pedicle Screw System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The Reform System is a top-loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods, cross-connectors, locking cap screws, hooks, domino connectors, and lateral offset connectors. All of the components are available in a variety of sizes to match more closely to the patient's anatomy. All components are made from medical grade stainless steel, cobalt chromium alloys, titanium or titanium alloy described by such standards as ASTM F138, ASTM F1537, ISO 5832-12, ASTM F136 or ISO 5832-3.
The provided text is a 510(k) summary for a medical device (Reform Pedicle Screw System) and does not describe acceptance criteria or a study proving that an AI/device meets acceptance criteria. Instead, it details the indications for use, technical characteristics, and a performance testing summary related to demonstrating substantial equivalence to a predicate device.
Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving device performance in the context of an AI device. The document describes a physical medical device (pedicle screw system) and its regulatory clearance process based on substantial equivalence, not the performance of an AI algorithm against defined acceptance criteria.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected and appear to be a single, continuous line. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 7, 2015
Precision Spine. Incorporated % Mr. Kenneth C. Maxwell II Regulatory and Quality Specialist Empirical Consulting, LLC 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K150856
Trade/Device Name: Reform Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNH, MNI, KWP Dated: June 4, 2015 Received: June 8, 2015
Dear Mr. Maxwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Mr. Kenneth C. Maxwell II
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K150856 Device Name Reform Pedicle Screw System
Indications for Use (Describe)
The Reform Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).
The Reform Pedicle Screw System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The Reform Pedicle Screw System is also intended for non-cervical pedicle screw fixation (T1-S1/ilium) for the following indications: degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); spinal tumor; pseudoarthrosis; and failed previous fusion.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Reform Pedicle Screw System is indicated as an adjunct to treat adolescent idiopathic scoliosis. The Reform Pedicle Screw System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Type of Use (Select one or both, as applicable)
മ Prescription Use (Part 21 CFR 801 Subpart D) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY |
|---|
| ------------------ |
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(K) SUMMARY | |
|---|---|
| Submitter's Name: | Precision Spine |
| Submitter's Address: | 2050 Executive DrivePearl, MS 39208 |
| Submitter's Telephone: | 973-455-7150 |
| Contact Person: | Kenneth C Maxwell IIEmpirical Consulting LLC719.291.6874 |
| Date: | June 4th, 2015 |
| Trade or Proprietary Name: | Reform Pedicle Screw System |
| Common or Usual Name: | Orthosis, Spinal Pedicle FixationOrthosis, Spondylolisthesis Spinal FixationOrthosis, Spinal Pedicle Fixation, For Degenerative Disc DiseaseAppliance, Fixation, Spinal Interlaminal |
| Classification: | Class III per 21 CFR §888.3070 (Pedicle screw spinal system)888.3050 (Spinal interlaminal fixation orthosis) |
| Product Code: | NKB, OSH, MNI, MNH, KWP |
| Classification Panel: | 87 Orthopedic Panel |
510(K) SUMMARY
DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION
The Reform System is a top-loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods, cross-connectors, locking cap screws, hooks, domino connectors, and lateral offset connectors. All of the components are available in a variety of sizes to match more closely to the patient's anatomy. All components are made from medical grade stainless steel, cobalt chromium alloys, titanium or titanium alloy described by such standards as ASTM F138, ASTM F1537, ISO 5832-12, ASTM F136 or ISO 5832-3.
CHANGE FROM PREDICATE
The subject device modifications allow the device to be assembled by the surgeon during implantation with the addition of modular pedicle screws.
INDICATIONS FOR USE
The Reform Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).
The Reform Pedicle Screw System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The Reform Pedicle Screw System is also intended for non-cervical pedicle screw fixation (T1
{5}------------------------------------------------
S1/ilium) for the following indications: degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); spinal tumor; pseudoarthrosis; and failed previous fusion.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Reform Pedicle Screw System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Reform Pedicle Screw System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
TECHNICAL CHARACTERISTICS
The subject device modifications allow the device to be assembled by the surgeon during implantation, but do not differ in regards to intended use, material, or technological function when compared to the predicate devices.
| 510k Number | Trade or Proprietary orModel Name | Manufacturer | Primary orAdditional |
|---|---|---|---|
| K143248 | PSS System (Reform PedicleScrew System) | Precision Spine(Spinal USA) | Primary |
| K131343, K130279, K121172,K092128, K090033, K073240,and K071438 | PSS System (Reform PedicleScrew System) | Precision Spine(Spinal USA) | Additional |
| K031175, K022949, K992739 | Click'X Pedicle ScrewSystem | Depuy Synthes | Additional |
| Table 5-1 Predicate Devices | |
|---|---|
| ----------------------------- | -- |
PERFORMANCE TESTING SUMMARY
Analysis was performed to show that the subject devices are substantially equivalent to the predicate devices. Custom static pull-off testing was performed to determine substantial equivalence.
CONCLUSION
The overall technology characteristics and mechanical engineering analysis lead to the conclusion that the Reform Pedicle Screw System is substantially equivalent to the predicate device.