K Number
K150385

Validate with FDA (Live)

Date Cleared
2015-04-30

(71 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NaturalVue (etafilcon A) Sphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.

NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

The lenses are intended for single-use disposable wear.

Device Description

The NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are visibility-tinted, containing a UV blocker and are available in a spherical, aspheric, toric, multifocal, and multifocal toric designs. They are manufactured by cast molding method. The device is an ionic hydrogel lens derived from etafilcon A which is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. The copolymer consists of 42% etafilcon A and 58% water by weight when immersed in normal buffered saline solution. The lenses are tinted blue using C.I. Reactive Blue 19 to make them more visible for handling. The lenses contain a UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of the lenses are less than 5% in the UVB range of 280-315nm and less than 30% in the UVA range of 316-380mm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution.

AI/ML Overview

The provided text is a 510(k) summary for the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses. This document does not describe a study that uses AI or machine learning to determine performance against acceptance criteria. Instead, it describes non-clinical tests to demonstrate the safety and effectiveness of the contact lenses.

Therefore, many of the requested elements (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size) are not applicable as they relate to AI/ML device evaluations.

However, I can extract the acceptance criteria and reported device performance from the non-clinical studies described in the document.


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are based on demonstrating substantial equivalence to a predicate device (K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses) and compliance with the May 1994 FDA Guidance Document for Daily Wear Contact Lenses. Where specific numerical acceptance criteria are not explicitly stated, the equivalence to the predicate device serves as the implicit acceptance criterion.

CharacteristicAcceptance Criteria (Implicitly: Equivalent to Predicate K140025 or regulatory standards)Reported Device Performance (NaturalVue (etafilcon A))Meets Acceptance Criteria?
Material Properties
Production MethodCast-Molded (per predicate)Cast-MoldedYes
USAN Nameetafilcon A (per predicate)etafilcon AYes
Material ClassificationGroup 4 high water ionic (per predicate)Group 4 high water ionicYes
Water Content (%)58% (per predicate)58%Yes
Refractive Index1.4023 (per predicate)1.4023Yes
Oxygen Permeability19.3 x 10-11 (cm²/sec)(ml O2/ml-mmHg) @ 35°C (per predicate)19.3 x 10-11 (cm²/sec)(ml O2/ml-mmHg) @ 35°CYes
% Transmittance at 593nm> 95% (per predicate)> 95%Yes
% Transmittance at 380-315nm< 30% (per predicate)< 30%Yes
% Transmittance at 315-280nm< 5% (per predicate)< 5%Yes
BiocompatibilityNegative responses for cytotoxicity, maximization, and ocular irritation (per predicate). Validity of blister packaging (passing standard extraction tests).Negative responses for all tests. Blister packaging validity demonstrated.Yes
MicrobiologyMinimum SAL of 10-9 for steam sterilization. Sterility maintained through expiration.Validated to deliver SAL of 10-9. Shelf-life stability supports sterility.Yes
Bacteriostatic ValidationUniformity of temperature in sterilizer; Tested microorganisms killed.Demonstrated uniformity of temperature; Tested microorganisms killed.Yes
Shelf-LifeLenses remain sterile and packages remain tight for claimed expiration.Lenses remained sterilized with no microbial growth for 4 years (accelerated conditions). Seal remained tight for 4 years (accelerated conditions).Yes
LeachabilityNo leachable monomers and additive residues at detection levels.No leachable monomers and additive residues at detection levels.Yes

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in the context of an AI/ML device. For non-clinical material property testing, the sample sizes are not explicitly stated. The biocompatibility, microbiology, and leachability studies were conducted on the predicate device (K140025) and applied to the current device due to substantial equivalence, so the specifics of their sample sizes are not in this document.

Data Provenance: The document refers to the predicate device (K140025), which was submitted by Pegavision Corporation, Taiwan. The current device is from Visioneering Technologies, Inc., and the non-clinical tests were performed to demonstrate equivalence. It's a non-clinical, retrospective comparison to previously established data and standards.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is not an AI/ML device, and no "ground truth" established by experts for a test set is mentioned.


4. Adjudication method for the test set

Not applicable. No test set requiring expert adjudication is described for this type of device and study.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done

Not applicable. This is not an AI/ML device, and thus, no MRMC study was conducted.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device or an algorithm.


7. The type of ground truth used

Not applicable in the context of AI/ML. For the non-clinical tests, "ground truth" would be the established scientific and regulatory standards for contact lens material properties, biocompatibility, sterility, and leachability. For example:

  • Material Properties: Direct measurements using laboratory techniques.
  • Biocompatibility: Standardized in vitro and in vivo (animal, if applicable, though not specified here) tests.
  • Microbiology: Validation of sterilization processes and sterility testing.
  • Leachability: Analytical chemistry techniques.

8. The sample size for the training set

Not applicable. This is not an AI/ML device, so there is no "training set."


9. How the ground truth for the training set was established

Not applicable. No training set is described.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 30, 2015

Visioneering Technologies, Inc. c/o Ms. Penny Northcutt, RAC, FRAPS, CQA Regulatory Consultant and Official Correspondent 174 Watercolor Way Suite 103-403 Santa Rosa Beach, Florida 32459

Re: K150385

Trade/Device Name: Naturalvue (etafilcon A) Daily Disposable Soft (hydrophilic) Contact Lenses NaturalVue (etafilcon A) SPHERE and ASPHERE NaturalVue (etafilcon A) Toric NaturalVue (etafilcon A) Multifocal NaturalVue (etafilcon A) Multifocal Toric

Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: March 9, 2015 Received: March 10, 2015

Dear Ms. Northcutt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"

(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Y. Alexander -A

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150385

Device Name

NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses

Indications for Use (Describe)

Spherical and Aspheric

NaturalVue (etafilcon A) Sphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

Toric

NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

Multifocal Spherical

NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

Multifocal Toric

NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.

NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

The lenses are intended for single-use disposable wear.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510k Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

DATEFebruary 13, 2015
APPLICANTVisioneering Technologies, Inc.Stephen Snowdy, CEO4555 Mansell RoadSuite 300Alpharetta, GA 30022770-521-4358 (office phone)404-863-8892 (cell phone)ssnowdy@vtivision.com (email)
OFFICIALCORRESPONDENTPenny Northcutt, RAC, FRAPS, CQA174 Watercolor Way, Suite 103-403Santa Rosa Beach, FL 32459678-428-6978 (phone)866-630-4082 (fax)pennynorthcutt@theregsolutions.com (email)
TRADE NAMENaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic)Contact Lenses
COMMON NAMESoft (hydrophilic) Contact Lenses (daily wear)Soft (hydrophilic) Contact Lenses (disposable)
DEVICECLASSIFICATIONAND PRODUCTCODEClass IILenses, Soft Contact, Daily Wear, per 21 CFR 886.5925,Product Code: LPL
Class IILens, Contact, (Disposable), per 21 CFR 886.5925, ProductCode: MVN
PREDICATEDEVICEK140025 Pegavision Corporation, NaturalVue (etafilcon A)Daily Disposable Soft (Hydrophilic) Contact Lenses

SUBSTANTIAL EQUIVALENCE:

It is Visioneering Technologies, Inc.'s conviction that the data submitted in this 510(k) provides evidence of substantial equivalence for the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses with the same established safety profile and effectiveness of the predicate device K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses.

{5}------------------------------------------------

DESCRIPTION OF THE DEVICE:

The NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are visibility-tinted, containing a UV blocker and are available in a spherical, aspheric, toric, multifocal, and multifocal toric designs. They are manufactured by cast molding method. The device is an ionic hydrogel lens derived from etafilcon A which is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. The copolymer consists of 42% etafilcon A and 58% water by weight when immersed in normal buffered saline solution. The lenses are tinted blue using C.I. Reactive Blue 19 to make them more visible for handling. The lenses contain a UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of the lenses are less than 5% in the UVB range of 280-315nm and less than 30% in the UVA range of 316-380mm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution.

INTENDED USE/INDICATIONS FOR USE:

NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses

Spherical and Aspheric

NaturalVue (etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

Toric

NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

Multifocal

NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eves in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

Multifocal Toric

NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.

NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

The lenses are intended for single-use disposable wear.

{6}------------------------------------------------

TECNOLOGICAL CHARACTERISTICS:

Historical Clinical Use

etafilcon A lenses have been used widely. Their safety and effectiveness have been well established and documented. Their safety and effectiveness can be further exemplified by the following lenses, as well as many others as discussed in the 510(k), cleared by FDA:

  • ACUVUE (etafilcon A) Contact lens, clear and visibility tint with UV blocker, K962804 ● submitted by Vistakon USA
  • ACUVUE (etafilcon A) contact lens, clear and visibility tint with UV blocker, K991134 ● submitted by Vistakon USA
  • Aquamax (etafilcon A) Disposable Soft (Hydrophilic) Contact Lenses, K120028 . submitted by Pegavision Corporation, Taiwan
  • NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, . K140025 submitted by Pegavision Corporation, Taiwan

NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses from the previous FDA cleared K140025 did not require clinical studies as the USAN name and manufacturing processes are the same as the above-mentioned predicate (and reference) devices.

Non-Clinical Study

All tests were conducted in accordance with the May 1994 FDA Guidance Document for Daily Wear Contact Lenses.

The non-clinical performance tests have been performed to demonstrate the safety and effectiveness of NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and establish substantial equivalence to the predicate NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses (K140025). The evidence of substantial equivalence to the predicate lenses is described below.

a) Material Properties Testing

The material properties of NaturalVue (etafilcon A) Soft (Hydrophilic) Contact Lenses were tested as illustrated in the following table.

Proposed DeviceK140025 Predicate
Production MethodCast-MoldedCast-Molded
USAN Nameetafilcon Aetafilcon A
Material ClassificationGroup 4 high water ionicGroup 4 high water ionic
Water Content (%)58%58%
Refractive Index1.40231.4023
Oxygen Permeability(edge corrected) @ 35°C19.3 x 10-11(cm²/sec)(ml O2/ml-mmHg)19.3 x 10-11(cm²/sec)(ml O2/ml-mmHg)
Percent Transmittance
% T at 593nm> 95%> 95%
% T at 380-315nm< 30%< 30%
% T at 315-280nm< 5%< 5%

b) Biocompatibility

The standard cytotoxicity, maximization, and ocular irritation tests were carried out for the predicate K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and therefore apply to NaturalVue (etafilcon A) Daily Disposable Soft

{7}------------------------------------------------

(Hydrophilic) Contact Lenses. Negative responses were recorded for all tests. The validity of blister packaging for these lenses was demonstrated by passing the standard extraction tests.

c) Microbiologv

The steam sterilization process has been validated to deliver a minimum SAL of 10-9, thereby complying with the requirement of FDA Group 4. There is shelf-life stability supporting that these lenses remain sterile through the expiration date claimed for the product.

d) Bacteriostatic Validation

The steam sterilizer was tested for effectiveness by measuring and demonstrating the uniformity of temperature at different locations inside the sterilizer over test period. Tested microorganisms were killed under tested conditions as compared to control.

Lenses remained sterilized and there was no microbial growth for a period of 4 years tested under accelerated conditions. Seal of lens packages remained tight for a period of 4 years as demonstrated by the constant peeling strength tested under accelerated conditions.

e) Leachability

Studies were conducted to determine the leachable materials from the finished lenses. The results show that, at the levels of the detection reported, there are no leachable monomers and additive residues.

CONCLUSION:

Based on the performance and manufacturing verification studies, it can be concluded that the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are equivalent to the predicate device K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses with respect to intended use, principles of operation, and technological characteristics.

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.