K Number
K991134

Validate with FDA (Live)

Date Cleared
1999-05-06

(34 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or nonaphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism.

The ACUVUE BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism.

The ACUVUE TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

The ACUVUE TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00D of astigmatism or less.

ACUVUE UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/ planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

Device Description

The device description does not change from that cleared under K962804.

AI/ML Overview

The provided text describes a 510(k) submission for ACUVUE (etafilcon A) soft (hydrophilic) contact lenses. This submission is for a labeling modification to add an additional Indications Statement. It is not an AI/ML device, and therefore, many of the requested criteria regarding AI/ML device performance and studies are not applicable.

Here's an analysis of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not provide specific acceptance criteria or reported device performance for an AI/ML device. Instead, it refers to the existing clearance of the same lens material (etafilcon A) and its associated performance as sufficient due to the nature of the 510(k) being a labeling modification.

Acceptance Criteria (Not explicitly stated for AI/ML)Reported Device Performance (Historical for etafilcon A contact lenses)
UV Blocking (UVA)82%
UV Blocking (UVB)97%
Clinical SafetySupported by previously submitted data (N18-033/S31 and K962804)
Non-Clinical SafetySupported by previously submitted data (N18-033/S31 and K962804)

2. Sample Size Used for the Test Set and Data Provenance

Not applicable as this is not an AI/ML device requiring a test set for algorithm performance evaluation. The submission relies on existing clinical and non-clinical data for the lens material itself from previous submissions (N18-033/S31 and K962804). The provenance of that historical data is not detailed in this document.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable as this is not an AI/ML device using a test set with ground truth established by experts.

4. Adjudication Method for the Test Set

Not applicable as this is not an AI/ML device using a test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This is not an AI/ML device, so such a study to evaluate human reader improvement with AI assistance is not relevant.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone performance study was not done. This is not an AI/ML device.

7. The Type of Ground Truth Used

The concept of "ground truth" as typically applied to AI/ML device performance (e.g., expert consensus, pathology, outcomes data) is not directly applicable here. The demonstration of safety and effectiveness for a contact lens material (etafilcon A) generally relies on:

  • Clinical safety and efficacy data: Data from human trials assessing visual acuity, comfort, ocular health, and adverse events.
  • Non-clinical data: Laboratory testing for material properties, microbiology, toxicology, chemistry, shelf-life, and leachability.

The submission states that this data was sufficient for the predicate devices and supports the current device due to no changes in material or manufacturing process.

8. The Sample Size for the Training Set

Not applicable as this is not an AI/ML device requiring a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable as this is not an AI/ML device requiring a training set with established ground truth.

Conclusion:

This 510(k) submission is for a labeling modification for an existing contact lens product. The "device" in question is a contact lens, not an AI/ML diagnostic or therapeutic system. Therefore, most of the requested information pertaining to AI/ML device acceptance criteria, studies, and ground truth establishment is not relevant to this specific document. The submission relies on the previously established safety and effectiveness of the etafilcon A material clear in prior 510(k)s and PMAs.

{0}------------------------------------------------

Summary of 510(k) Submission

Name andaddress ofsubmitterVistakon, Johnson & Johnson Vision Products, Inc.4500 Salisbury Road, Suite 300Jacksonville, Florida 32216Contact: Martine D. MartinoPhone: (904) 443-1808Date Prepared: April 1, 1999
Identification ofdeviceThe trade name is ACUVUE (etafilcon A) soft (hydrophilic) contactlenses, clear and visibility tint, with UV blocker, for daily wear The common or usual name is Soft (hydrophilic) Contact Lens (dailywear). The FDA Classification is Class II.
PredicatedevicesThe predicate device is ACUVUE (etafilcon A) soft (hydrophilic) contactlenses, clear and visibility tint, with UV blocker, for daily wear covered underK962804.
Description ofdeviceThe device description does not change from that cleared under K962804.
Continued on next page
Intended useThe reason for the 510(k) is to add the following Indications Statement (see bold text ) for ACUVUE (etafilcon A) soft (hydrophilic) contact lenses, clear and visibility tint, with UV blocker, for daily wear. The revised Indications Statement is included with this 510(k):
The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism.
The ACUVUE BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism.
The ACUVUE TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.
The ACUVUE TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00D of astigmatism or less.
ACUVUE UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/ planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

{1}------------------------------------------------

Summary of 510(k) Submission, Continued

Continued on next page

{2}------------------------------------------------

Summary of 510(k) Submission, Continued

CharacteristicsThe characteristics do not change. They are the same as previously submittedin N18-033/S31 and K962804.
Non-clinicalstudiesNon-clinical studies (microbiology, toxicology, chemistry, shelf-life, andleachability) on the lens material were not conducted because the lensmaterial, etafilcon A, does not change. Non-clinical studies were conducted to support the following modifiedActions in the labeling. The average UV blocking for ACUVUE Contact Lenses with UVBlocker for UVA is 82% and for UVB is 97%.
Clinical studiesThis 510(k) describes a labeling modification: an additional IndicationsStatement. There is no change in lens material, the manufacturing process,nor the parameters and properties, therefore, the clinical data previouslysubmitted in N18-033/S31 and K962804 supports the clinical safety of thesubject device.
Conclusionsdrawn fromstudiesAdditional studies were not conducted, therefore, the conclusions drawn fromstudies previously submitted in N18-033/S31 and K962804 support the non-clinical and clinical safety of the subject device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized representation of a human figure, consisting of three overlapping profiles facing to the right, with flowing lines suggesting movement or connection.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 6 1999

Vistakon™ Johnson & Johnson Vision Care Products, Inc. C/O Ms. Martine D. Martino P.O. Box 10157 Jacksonville, FL 32247

Re: K991134

Trade Name: ACUVUE (etafilcon A) soft (hydrophilic) contact lens, clear and with visibility tint, with UV blocker, for daily wear Regulatory Class: II Product Code: LPL Dated: April 1, 1999 Received: April 2, 1999

Dear Ms. Martino:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Ms. Martine D. Martino

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A Ralph L. Rosenthal, M.D.

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications Statement

K991134 510(k) Number:

ACUVUE (etafilcon A) Soft (hydrophilic) Contact Lens Clear and Visibility Device Name: Tint with UV Blocker

The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or nonaphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism.

The ACUVUE BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism.

The ACUVUE TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

The ACUVUE TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00D of astigmatism or less.

ACUVUE UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/ planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)

Karin Waibmtor
(Division Sign-Off)
Division of Ophthalmic Devices
Prescription Use ✓ 510(k) Number K991134 OR Over the Coumer

Continued on next page

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.