K Number
K150041

Validate with FDA (Live)

Date Cleared
2015-06-26

(168 days)

Product Code
Regulation Number
864.5425
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CORA System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The CORA System records the kinetic changes in a venous sample of 3.2% cirrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R, K, Angle, MA, and FLEV.

The CORA System provides specific blood modifiers, in the form of reagents dried-in-place within CORA Cartridges.

Results from the CORA analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.

The indication for CORA System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.

Device Description

The CORA System consists of a four-channel diagnostic analyzer with integrated computer module, system reagents (CK, CRT, CKH, and CFF), and Abnormal Quality Control material and microfluidic test cartridges.

To perform a test, a disposable CORA Cartridge is inserted into the analyzer. Blood or WQC material is added to an entry port on the cartridge and drawn into the cartridge under analyzer control. The amount of the sample drawn into the cartridge is automatically determined by the volume of the blood chambers in the cartridge. Once in the disposable, the sample is metered into as many as four separate analysis channels, depending upon the assays being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows (pumps) within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for visco-elastic changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.

The CORA technology is based on a disposable containing up to four independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring) with a diameter of 2.5mm and a length of 4.5mm. Detection of clotting in the CORA System is performed optically. Under control of the analyzer, approximately 20ul of prepared sample is delivered to the tube, where a meniscus naturally forms at each end of the tube. The tube is positioned so that the lower meniscus partially blocks light traveling from a collimated source toward a photodiode.

During testing, a piezoelectric actuator drives the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The profile has a maximum amplitude of under 10um and contains frequencies from 10-500Hz. Some, but not all, of the measurement cell motion will induce motion in the sample meniscus, which will be detected by the photodiode. The resulting motion of the meniscus is monitored optically and recorded by the analyzer to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, it is possible to determine the frequencies of input motion that caused the greatest deflection of the sample (these are called the resonant frequencies).

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the CORA® (Coagulation Resonance Analysis) System, based on the provided text:

Important Note: The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. It typically does not contain explicit "acceptance criteria" for performance in the same way a clinical trial protocol would. Instead, it demonstrates performance that aligns with a predicate device and adheres to established analytical performance guidelines (like CLSI standards). The "acceptance criteria" as presented below are inferred from the demonstrated performance and the context of regulatory submission for a device of this type.

1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of the submission (510(k) summary), the "acceptance criteria" are implied by the precision limits set and met, and the acceptable agreement with the predicate device.

Test ParameterAcceptance Criteria (Implied / Stated)Reported Device Performance (CV% or Correlation with Predicate)
Analytical Precision (All Reagents)
R parameterCV ≤ 15%All reported CVs for R ≤ 14.8% (Hypo, Normal, Hyper, Anticoagulant Patients)
K parameterCV ≤ 25%All reported CVs for K ≤ 23% (Hypo, Normal, Hyper, Anticoagulant Patients)
Alpha (Angle)CV ≤ 10%Most reported CVs for Angle ≤ 8.7%; a few outliers noted (up to 27.4%) with justification that they are not clinically significant.
MA parameterCV ≤ 10%Most reported CVs for MA ≤ 8.4%; one CFF MA Hypo outcome (11.6%) with justification that the SD is small and not clinically significant.
FLEV parameter(No explicit CV% provided, but expected to be within reasonable limits for clinical use)Most reported CVs for FLEV ≤ 5% (Hypo, Normal, Hyper, Anticoagulant Patients)
Method Comparison (Correlation with TEG 5000)
All parameters (R, K, Alpha, MA, FLEV)Excellent agreement expected, with R-values generally aboveR-values for all parameters (CK, CKH, CRT, CFF) ranged from 0.680 to 0.938, demonstrating strong correlation.
Reader Study
Agreement between CORA and TEG 5000 readingsExcellent agreement expectedDemonstrated "Excellent agreement" between the two devices based on 30 test outcomes per reagent category, per reader.

2. Sample Size Used for the Test Set and Data Provenance

  • Analytical Precision:

    • Normal Donors (original study): Blood draws from 3 donors (1 Hypo, 1 Normal, 1 Hyper) across 5 non-consecutive days, by 2 operators, using 3 reagent lots and 12 analyzers. Each test had 2 replicates. This results in 120 measurements for each parameter per level (Hypo, Normal, Hyper).
    • Spiked Samples (supplemental study): 3 donors for each reagent-parameter and spiking type (hypocoagulable, hypercoagulable). For each donor-reagent-parameter-hypo/hyper type, there were 12 outcomes (combinations of 3 operators, 3 reagent lots, 12 instruments, 2 replicates per instrument).
    • Anticoagulant Patients (supplemental study): 4 Dabigatran patients and 2 Warfarin patients for each reagent-parameter. 12 outcomes for each patient-reagent-parameter.
    • Data Provenance: Not explicitly stated, but performed in Coramed's laboratory. Given it's a US company, it's likely US-derived. This is a retrospective analysis of laboratory test samples.
  • Reference Ranges:

    • Sample Size: Up to 55 normal volunteer subjects from each of three sites, totaling 157 samples. At least 151 valid results for every parameter for all reagents.
    • Data Provenance: Three clinical sites. Subjects chosen to represent demographic populations of these three areas (age, race, gender). This is a prospective study from normal volunteers.
  • Method Comparison:

    • Sample Size: Not explicitly stated as a single number but conducted at three clinical sites on patient samples. Up to 10% contrived samples were added to broaden the comparison range.
    • Data Provenance: Three clinical sites: Mayo Clinic, University of Pittsburgh Medical Center, and Sinai Hospital, Baltimore. Subjects were patients undergoing cardiovascular surgery or cardiology procedures, with blood samples drawn pre- and post-surgery and in the ICU. This is a prospective study from clinical patients, with some contrived samples.
  • Reader Study:

    • Sample Size: 30 test outcomes compared for each reagent test category (e.g., CK, CKH, CRT, CFF).
    • Data Provenance: Three sites.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Analytical Precision, Linearity, Sensitivity & Specificity, Heparin Neutralization: The ground truth for these analytical studies is based on the known concentrations/conditions of the samples (e.g., spiked samples, known interferents) and laboratory measurements. No human "experts" established ground truth in the interpretative sense.

  • Reference Ranges: Ground truth is established by obtaining samples from "normal volunteer subjects" and excluding outliers based on standard criteria. There is no mention of "experts" establishing a ground truth for individual samples.

  • Method Comparison: Ground truth is the results obtained from the predicate TEG 5000 device.

  • Reader Study:

    • Number of Experts: Nine readers (three experienced TEG 5000 doctors at each of the three sites).
    • Qualifications: "Experienced TEG 5000 doctors." Specific years of experience or board certification are not detailed.

4. Adjudication Method for the Test Set

  • Analytical Precision, Linearity, Sensitivity & Specificity, Heparin Neutralization, Reference Ranges, Method Comparison: No adjudication method is described as these are quantitative analytical studies or comparisons to a predicate device. Ground truth is either inherent in the sample preparation or derived from the predicate device's measurements.

  • Reader Study: No explicit adjudication method (e.g., 2+1, 3+1) is described. The study assessed "agreement between the two devices" based on the doctors' readings of both device outputs. It implies that the TEG 5000 reading was considered the reference for comparison, but not necessarily an adjudicated ground truth in the sense of multiple experts independently labeling cases and resolving discrepancies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? Yes, a "Reader Study" was conducted, which has elements of an MRMC study in that it involved multiple readers and multiple cases.

  • Effect Size of Human Readers Improve with AI vs. Without AI Assistance: This information is not applicable and not provided in the document. The CORA System is a diagnostic instrument that provides "semi-quantitative indications of the hemostasis state," similar to the predicate TEG 5000. It does not incorporate AI nor does it aim to "assist" human readers in interpreting complex images or deriving diagnoses. It measures parameters directly; the "Reader Study" was about comparing the readings of the outputs from the CORA System to the predicate TEG 5000, not about human performance with/without AI assistance.

6. Standalone Performance (Algorithm only without human-in-the-loop performance)

  • Was a standalone performance study done? Yes, the entire set of analytical performance studies (Analytical Precision, Interference, Linearity, Sensitivity and Specificity, Heparin Neutralization) and the Method Comparison study effectively represent standalone (algorithm only) performance. The CORA System generates numerical outputs (R, K, Angle, MA, FLEV) directly. The human "reader" in the "Reader Study" is interpreting the output of the device, not interacting with an algorithm to refine a diagnosis. Therefore, the core function of the device is standalone.

7. Type of Ground Truth Used

  • Analytical Precision, Interference, Linearity, Sensitivity and Specificity, Heparin Neutralization: Ground truth was established by known sample properties (e.g., spiked concentrations of substances, known degrees of hemodilution) or through controlled laboratory measurements.
  • Reference Ranges: Ground truth was established by samples from "normal volunteer subjects", with outliers removed using standard statistical criteria.
  • Method Comparison: Ground truth was the predicate device (TEG 5000) measurements. This is a common approach in 510(k) submissions to demonstrate equivalence.
  • Reader Study: Ground truth was readings from the predicate device (TEG 5000) outcomes by experienced clinicians, to which the CORA System's outcomes were compared. It's essentially a comparison of the device outputs' interpretability.

8. Sample Size for the Training Set

The document does not explicitly describe a "training set" for an AI/algorithm in the conventional sense. The CORA System is a measurement device that senses physical changes and calculates parameters. The development of its algorithms (how it analyzes frequency data to provide parameters) would have involved extensive engineering and potentially iterative testing, but this is distinct from the "training set" concept often associated with machine learning.

The analytical studies (precision, linearity, etc.) and method comparison effectively validate the final algorithms/calculations after their development.

9. How the Ground Truth for the Training Set Was Established

As noted above, the concept of a "training set" and its "ground truth" in the context of an AI/ML algorithm isn't directly applicable here. The device's parameters are derived from physical measurements and established formulas related to viscoelastic properties of blood. The underlying physics and physiology are the "ground truth" upon which the measurement technique is built. Any internal calibration or algorithm development would have relied on known samples or precisely measured physical responses, but these are not explicitly detailed as a "training set" in the submission.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles three human profiles facing to the right, with lines extending above them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 26, 2015

Coramed Technologies, LLC Mr. Norman Brunner Director of RA/QA 6225 W. Howard Street Niles, IL 60714

Re: K150041

Trade/Device Name: CORA® (Coagulation Resonance Analysis) System Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: II Product Code: JPA Dated: May 21, 2015 Received: May 22, 2015

Dear Mr. Brunner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Leonthena R. Carrington -S

Leonthena R. Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150041

Device Name CORA Hemostasis System

Indications for Use (Describe)

The CORA System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a blood sample. The CORA System records the kinetic changes in a venous sample of 3.2% cirrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R, K, Angle, MA, and FLEV.

The CORA System provides specific blood modifiers, in the form of reagents dried-in-place within CORA Cartridges.

Results from the CORA analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.

The indication for CORA System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K150041

Device Name CORA Citrated Multichannel Cartridge

Indications for Use (Describe)

The CORA System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The Citrated Multichannel Cartridge, to be used with the CORA System instrument, contains four independent assays (CK, CRT, CKH and CFF), described below.

The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the CORA System. Clotting characteristics are described by the functional parameters Clotting Time (R), Speed of Clot Formation (K and Alpha angle) and Maximum Clot Strength (MA).

The CRT assay monitors the hemostasis process via both the intrinsic pathways in 3.2% citrated whole blood specimens on the CORA System. Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA). The CRT MA parameter is equivalent to the CK MA parameter but the final MA value is reached more quickly using the CRT assay.

The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the CORA System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests. The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the CORA System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameters Maximum Clot Strength (MA) and the Estimated Functional Fibrinogen Level (FLEV).

Results from the CORA analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for CORA System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K150041

Device Name

CORA Hemostasis System - Abnormal Wet Quality Control (WQC) Material

Indications for Use (Describe)

The CORA System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample.

The Abnormal Wet Quality Control Material is to be used for monitoring the accuracy and precision of tests carried out on the CORA System.

Results from the CORA analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.

The indication for CORA System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Coramed Technologies, LLC 6225 W. Howard Street, Niles, Illinois 60714 847-647-8800 | 847-510-0502 FAX

CORA® System 510(k) Summary

APPLICANT INFORMATION

Name of Manufacturer: Coramed Technologies, LLC

Manufacturer Street Address: 6225 W. Howard St.

City, State, Zip: Niles, IL 60714

Phone Number: (847) 647-8800

FAX Number: (847) 510-0502

Contact person for all communications: Norman E. Brunner

Email for contact person: nbrunner@coramedtech.com

Date that Summary was prepared: June 23, 2015

DEVICE INFORMATION

Trade name (proprietary name): CORA® (Coagulation Resonance Analysis) System and the following assays and reagents

  • Citrated Assays and Reagents
    • . CK (Citrated Kaolin) (Kaolin + CaCl2, for citrated blood) reagent
    • CRT (Citrated RapidTEG) (RapidTEG® + CaCl2, for citrated blood) reagent ●
    • CKH (Citrated Kaolin with Heparinase) (Kaolin + CaCl> + Heparinase, for citrated blood) . reagent
    • CFF (Citrated Functional Fibrinogen) (Tissue Factor (TF) + CaCl2 + ReoPro®, for citrated ● blood) reagent

Common name (usual name): Whole Blood Hemostasis System

Classification Name: 21CFR 864.5425 Multipurpose System for in vitro Coagulation Studies

PREDICATE DEVICE

  • Thrombelastograph® Coagulation Analyzer (TEG®) 5000 Series, K002177, Product O Code JPA (System, Multipurpose, for In Vitro Coagulation Studies), Haemoscope Corporation

{6}------------------------------------------------

DESCRIPTION OF THE DEVICE

System Description

The CORA System consists of a four-channel diagnostic analyzer with integrated computer module, system reagents (CK, CRT, CKH, and CFF), and Abnormal Quality Control material and microfluidic test cartridges. See below for a description of system reagents. Reagents are dried-in-place within the cartridges during manufacturing. Abnormal Quality Control material is lyophilized and can be reconstituted with water as needed for WOC testing with reagent cartridges.

To perform a test, a disposable CORA Cartridge is inserted into the analyzer. Blood or WQC material is added to an entry port on the cartridge and drawn into the cartridge under analyzer control. The amount of the sample drawn into the cartridge is automatically determined by the volume of the blood chambers in the cartridge. Once in the disposable, the sample is metered into as many as four separate analysis channels, depending upon the assays being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows (pumps) within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for visco-elastic changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.

The CORA Measurement Technique

The CORA technology is based on a disposable containing up to four independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring) with a diameter of 2.5mm and a length of 4.5mm. Detection of clotting in the CORA System is performed optically. Under control of the analyzer, approximately 20ul of prepared sample is delivered to the tube, where a meniscus naturally forms at each end of the tube. The tube is positioned so that the lower meniscus partially blocks light traveling from a collimated source toward a photodiode.

During testing, a piezoelectric actuator drives the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The profile has a maximum amplitude of under 10um and contains frequencies from 10-500Hz. Some, but not all, of the measurement cell motion will induce motion in the sample meniscus, which will be detected by the photodiode. The resulting motion of the meniscus is monitored optically and recorded by the analyzer to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, it is possible to determine the frequencies of input motion that caused the greatest deflection of the sample (these are called the resonant frequencies).

Resonance is the tendency of a material or structure to oscillate with greater amplitude at some frequencies than others. The exact frequencies at which resonance occurs will depend on the stiffness and mass of the sample. Stiffness, in turn, is a function of a material's modulus of elasticity and the boundary conditions to which the material is exposed, such as the geometry and materials of a test cell. By holding these boundary conditions and sample mass constant from run to run, the CORA System allows direct comparison of elasticity between samples.

In a typical test, blood that has been delivered to the measurement cell will not clot for several minutes. During this time the sample has no inherent stiffness except that provided by surface tension, and since this remains constant the measured resonant frequencies will not change. Once clotting begins, however, the elastic modulus and thus the resonant frequencies increase rapidly. During fibrinolysis, the process is reversed, with elastic modulus and resonant frequencies decreasing. In tests where clotting does not occur, the resonant frequency of the sample will not change. During coagulation, however, a clot will bind to the test tube (ring) and the resonant frequency will rise with increasing firmness of the clot. The

{7}------------------------------------------------

CORA Analyzer collects meniscus motion data, tracks changing resonant frequencies and analyzes the frequency data to provide semi-quantitative parameters describing the clot. Results are presented in a format similar to the TEG 5000.

Both the TEG 5000 and CORA System monitor the elastic modulus of clots. Method Comparison testing has been performed, vielding data from approximately 300 patients from three sites (Mayo Clinic, University of Pittsburgh Medical Center, and Sinai Hospital, Baltimore). These data include all applicable parameters from the four citrated tests (CK, CKH, CRT and CFF). The TEG 5000 measures the changes of the linkage forces of the clot between the cup and pin over time, while the CORA System measures the changes of the resonant frequency of the clot over time. Despite the difference in the way the two analyzers measure the changes in the modulus of elasticity of the clot over time, they measure the same physical phenomenon and produce the same result when converted from their specific units of measurement (forces for the TEG5000 and frequency for the CORA System) to millimeters. The fact that these results are similar across all reagents used in both technologies is demonstration that the two technologies measure the same phenomenon, the changes in elastic modulus.

CORAParameterDefinitionParameter Relation toHemostasis
RR is the time from the start of the test until initial fibrinformation. This represents the enzymatic portion ofcoagulation.Normal / reduced / increasedspeed of coagulation initiation
KK is the time after R needed to reach a certain level ofclot strength. This represents clot kinetics.Normal / reduced / increasedspeed of coagulation amplificationand propagation
Alpha α(angle)Alpha is the angle representing the rate increase in theclot strength and the rapidity of fibrin build-up and cross-linking.Normal / reduced / increasedspeed of coagulation amplificationand propagation
MAMA, or Maximum Amplitude, represents the maximumfirmness of the clot during the test.Normal / reduced / increasedclot elasticity/strength
FLEVFLEV is an estimate of the fibrinogen level in the bloodsample.Estimated Functional FibrinogenLevel

The following definitions apply to calculated parameters in the CORA System:

Citrated Assays

The CK assay is a semi-quantitative in vitro diagnostic assay for monitoring the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the CORA System. Clotting characteristics are described by the functional parameters Clotting Time (R), Speed of Clot Formation (K and Alpha angle), and Maximum Clot Strength (MA). K and Alpha are complementary parameters and should be used in conjunction with Clotting Time (R) and Maximum Clot Strength (MA).

The CRT assay is a semi-quantitative in vitro diagnostic assay for monitoring the hemostasis process via both the intrinsic and extrinsic pathway in 3.2% citrated whole blood specimens on the CORA System. Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA). The CRT MA parameter is equivalent to the CK MA parameter but the final MA value is reached more quickly using the CRT assay.

The CFF assay is a semi-quantitative in vitro diagnostic assay for monitoring hemostasis of 3.2% citrated whole blood specimens in the CORA System after blocking platelet contributions to clot strength. CFF is used in conjunction with CK and clotting characteristics are described by the functional parameter Maximum Clot Strength (MA) and the Estimated Functional Fibrinogen Level (FLEV).

{8}------------------------------------------------

The CKH assay is a semi-quantitative in vitro diagnostic assay used in conjunction with the CORA System to monitor the effects of heparin in the blood stream of a patient. CKH is used in conjunction with CK and heparin influence is determined by comparing Clotting Times (R) between the two tests.

Wet Quality Control (WQC)

Abnormal Wet Quality Control (WOC) material consists of bovine plasma with buffers and stabilizers, and is used instead of hypocoagulable blood, in coniunction with the Citrated Multichannel system cartridge. When used with the Citrated Multichannel cartridge, this material provides consistent and repeatable values for CORA parameters to confirm reagent quality and viability.

Abnormal WQC Intended Use

The Abnormal WOC (Wet Quality Control) Material is intended to be used for monitoring the accuracy and precision of tests carried out on the CORA System.

INTENDED USE AND INDICATIONS FOR USE

The CORA System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The CORA System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R. K. Angle, MA, and FLEV.

The CORA System provides specific blood modifiers, in the form of reagents dried-in-place within CORA Cartridges.

Results from the CORA analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history. the clinical picture and. if necessary, further hemostasis tests.

The indication for CORA System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARING THE CORA SYSTEM TO THE TEG 5000 PREDICATE DEVICE

ItemTEG® 5000 PredicateCORA® System
Analyzer
Technological PurposeMonitoring the response of a clot tolow levels of applied strainMonitoring the response of a clot tolow levels of applied strain
What is measuredChanges in clot elasticity over timeChanges in clot elasticity over time
Initial Warm Up Time5 min5 min
Time to Complete a TestVaries with assaySame as TEG 5000
Assays and Reagents
Intrinsic Contact ActivationReagent (CK)Kaolin and CaCl2Kaolin and CaCl2, same materials asTEG 5000

Table of Similarities

{9}------------------------------------------------

ItemTEG® 5000 PredicateCORA® System
Intrinsic Contact ActivationReagent with Heparinasefor heparin reversal (CKH)Kaolin and CaCl2 with HeparinaseKaolin, Heparinase and CaCl2, samematerials as TEG 5000
Citrated RapidTEG (CRT)(Tissue Factor and KaolinActivation)Tissue Factor (TF), Kaolin andCaCl2Tissue Factor (TF), Kaolin andCaCl2, same materials as TEG5000
Citrated FunctionalFibrinogen(CFF) / Platelet-Blocked(Tissue Factor Activation)ReoPro®,Tissue Factor and CaCl2,same materials as CORAReoPro®, Tissue Factor and CaCl2,same materials as TEG 5000
Calcium Chloride Reagent(CC) (for re-calcification)CaCl2 reagentCaCl2 same materials as TEG 5000
Quality Control Material(WQC Abnormal Control)TEG coagulation control – Level IITEG coagulation control –Abnormal, same materials as TEG5000

Table of Differences

ItemTEG® 5000 PredicateCORA® System
AnalyzerThrombelastography analyzer,separate computer and softwareFully integrated Thrombelastographyanalyzer
Measuring TechniqueDirect-contact measurement of shearelasticity of a coagulating sampleNon-contact measurement of shearelasticity of a coagulating sample
Measuring ChannelsTwo, each independent andinterchangeable (can be used withany approved reagent)Four, each independent andinterchangeable (can be used withany approved reagent)
Signal TransducerElectromechanical detection (rotaryvariable inductive transformer) ofrotary motion of a pin suspended inthe sampleOptical detection (siliconphotodiode) of the motion of a freesurface of the sample
Temperature Control20° to 40°C20° to 50°C
Sample Volume (perchannel)360-380 μl63 μl
Total Reaction Volume(single channel)360-380 μl20 μl
Mains Supply Voltage120V, 60Hz and 220V, 50Hz modelavailable100-240V, 50-60Hz (internationalpower supply)
Analyzer Input Voltage24 volts AC, 30 watts max12 volts DC, 60 watts max
EnvironmentLevel and vibration free position, nosolar radiationOperating temperature:10° to 35 °CStorage Temperature:-30° to +50 °C (analyzer only)Relative humidity 20-80% (non-condensing)Stable and level surface.Operating Temperature10° to 32°CStorage Temperature:-20° to 50°C (analyzer only)Relative humidity 20 to 80% (non-condensing)

{10}------------------------------------------------

ItemTEG® 5000 PredicateCORA® System
Sample PreparationPerformed by the operator usingpipettes to reconstitute reagents andmix reagents with the samplePerformed under analyzer controlwithin the disposable cartridge
PipettingManual accurate pipettes(10, 20, 50, 100, 340, 360, 500,1000μl)Unmetered transfer pipette orsyringe; blood sample is added untilit fills to a level above the linemarked on the blood intake well ofthe cartridge
ConsumablesCups & Pins (acrylic plastic)Carrier (acrylic plastic) withmicrofluidics laminate and test rings(acrylic plastic)

SUMMARY OF NON-CLINICAL PERFORMANCE DATA

Analytical Precision

Testing was performed in Coramed's laboratory for precision, using CLSI EP5-A2 as guidance. Acceptance criteria for all reagents are: CV ≤ 15% for the R parameter, CV ≤ 25% for the K parameter, and CV ≤ 10% for the Alpha and MA parameters. Three types of donor citrated whole blood (CWB) were used in this precision testing:

  • Hypo (donors with natural coagulation levels of R parameter near the upper limit of the reference range and MA parameter near the lower limit of the reference range);
  • Normal (donors with natural coagulation levels of R and MA parameters near the center of the ● reference ranges):
  • Hyper (donors with natural coagulation levels of R parameter near the lower limit of the ● reference range and MA parameter near the upper limit of the reference range

Testing was performed with blood draws from three donors (one Hypo, one Normal, and one Hyper) on each of five days (non-consecutive). Testing was performed by two operators using three reagent lots and twelve analyzers, two replicates. The structure of this precision test is shown below.

Sample Type (Hypo, Normal or Hyper)
Day 1 (Total of 5 days)
OperatorOperator 1Operator 2
Reagent lot123123
Analyzer123456789101112
Replicates1,21,21,21,21,21,21,21,21,21,21,21,2

Structure of Precision Testing

Precision test estimates by test, parameter and donor sample test level are shown in the table on the following page. All acceptance criteria were met.

{11}------------------------------------------------

TestParameterLevelnMeanReagent LotOperator*Analyzer (within Operator, Reagent Lot)Day (within Analyzer, Operator, Reagent Lot)RepeatabilityTotalTotal without Day (Within Day)
SD%CVSD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
CFFFLEVHypo120269.82.360.9%00.0%00.0%5.322.0%5.872.2%8.273.1%6.332.3%
CFFFLEVNormal120376.800.0%2.250.6%00.0%5.491.5%3.981.1%7.151.9%4.571.2%
CFFFLEVHyper120597.400.0%00.0%00.0%49.388.3%9.621.6%50.318.4%9.621.6%
CFFMAHypo12014.80.110.8%00.0%00.0%0.292.0%0.332.2%0.453.0%0.352.3%
CFFMANormal12020.600.0%0.130.6%00.0%0.31.5%0.221.1%0.391.9%0.251.2%
CFFMAHyper12032.700.0%00.0%00.0%2.718.3%0.531.6%2.768.4%0.531.6%
CKHRHypo1208.200.0%00.0%0.182.2%0.364.3%0.769.2%0.8610.0%0.789.5%
CKHRNormal1206.000.0%0.050.9%0.091.6%0.142.3%0.427.1%0.467.7%0.447.3%
CKHRHyper1204.900.0%0.051.1%00.0%0.285.7%0.510.0%0.5812.0%0.510.0%
CKAngleHypo12068.700.0%1.091.6%00.0%1.612.3%3.014.4%3.595.2%3.24.7%
CKAngleNormal12072.100.0%00.0%00.0%1.141.6%1.732.4%2.072.9%1.732.4%
CKAngleHyper12076.700.0%00.0%00.0%1.842.4%1.772.3%2.563.3%1.772.3%
CKKHypo1201.80.031.9%0.063.6%00.0%0.137.2%0.1810.0%0.2313.0%0.1911.0%
CKKNormal1201.400.0%00.0%00.0%0.096.7%0.1612.0%0.1913.0%0.1612.0%
CKKHyper1200.900.0%00.0%00.0%0.1212.0%0.1820.0%0.2223.0%0.1820.0%
CKMAHypo12052.40.330.6%0.551.0%0.260.5%0.831.6%0.771.5%1.332.5%1.032.0%
CKMANormal12059.400.0%0.50.8%0.480.8%0.571.0%0.861.4%1.242.1%1.11.9%
CKMAHyper12068.000.0%0.20.3%00.0%0.91.3%0.520.8%1.061.6%0.560.8%
CKRHypo1208.900.0%0.293.3%00.0%0.647.1%0.859.5%1.112.0%0.910.0%
CKRNormal1206.500.0%00.0%00.0%0.284.3%0.497.6%0.568.7%0.497.6%
CKRHyper1205.20.132.5%00.0%00.0%0.315.8%0.6613.0%0.7414.0%0.6713.0%
CRTMAHypo12052.40.390.7%0.220.4%00.0%0.921.8%0.591.1%1.182.3%0.741.4%
CRTMANormal12062.700.0%0.150.2%0.140.2%0.370.6%0.310.5%0.520.8%0.370.6%
CRTMAHyper12069.400.0%00.0%00.0%0.911.3%0.20.3%0.931.3%0.20.3%

*Operator = Operator + operator-by-reagent lot interaction

{12}------------------------------------------------

In addition, Coramed's Citrated Reagent Precision Protocol Rev. 02 was written to supplement precision data originally provided to FDA in 510(k) K150041. This protocol specifies testing to be performed outside of reference range values but within the Analytical Measurement Range (AMR) limits. This report presents data and statistics resulting from the execution of this protocol.

The following pages contain tables of precision statistics. Table 1 shows standard deviations (StDev) and coefficients of variance (CV) for data using blood samples from normal donors but spiked to achieve hypocoagulable (hypo) and hypercoagulable (hyper) results. Three donors for each reagent-parameter and spiking type were tested using the following structure taken from the protocol. As defined in this structure, there are 12 outcomes for each donor-reagent-parameter-hypo/hyper type. resulting in 48 sets of statistics. Lot-to-lot, Instrument-to-Instrument, Operator, Replicate (Repeatability) and Total Precision statistics are given.

SampleHypocoagulable
Operator123
Reagent Lot231213
Instrument123456789101112
ReplicatesABABABABABAB
SampleHypercoagulable
Operator123
Reagent Lot231213
Instrument123456789101112
ReplicatesABABABABABAB

The focus of this study was on achieving values for the primary coagulation parameters R, MA and FLEV outside the reference ranges but within AMR limits, and taking the resulting K and Alpha (angle) values as is. Total Precision %CV values were all within pre-established limits, except for one CFF MA hypo outcome (11.6% for a limit of 10%) and 2 outcomes for Angle. The K and Angle parameters indicate the kinetics of the clotting process, and are not clinically significant as the main parameters representing coagulation, platelet and fibrinogen activities and abnormalities. The CFF MA CV% is higher because the MA value is small (mean of 6.5 mm) and although the Standard Deviation is very small (0.8mm) the resulting CV% is just over the limit. When the SD is so small there is no clinical significance for the differences making this CV% acceptable.

{13}------------------------------------------------

Table 2 shows standard deviations (StDev) and coefficients of variance (CV) for data using blood samples from patients taking anticoagulants Dabigatran and Warfarin. Four Dabigatran patients and two Warfarin patients for each reagent-parameter were tested using the following structure taken from the protocol. As defined in this structure, there are 12 outcomes for each patient-reagent-parameter, resulting in 48 sets of statistics. Lot-to-lot, Instrument, Operator-to-Operator, Replicate (Repeatability) and Total Precision statistics are given.

SampleHypocoagulable
Operator123
Reagent Lot231213
Instrument123456
ReplicatesABABAB

Results for this testing were generally within limits, except for a few minor exceptions. As the data shows, all the MAs, for CK, CRT, and CFF, are within the specified expected CV% across the board while for the R parameter, there are no significant differences between lots, operators and devices and the only CVs (for CK and CKH - R parameter) somewhat exceeding 15% come from repeatability (16.9% for MH-DB003 for CK and 15.9% for MH-DB004 for CKH). Even though the, K and Angle are not parameters that are considered clinically significant the raw data indicates that the single angle CV% that is above the limit was due to an outlier while the K parameter has high CVs due to the fact that elongated coagulation results in higher CVs for that parameter.

TestParam/SpikingPatientIDNMeanLotOperatorInstrumentRepetitionTotal
StDevCVStDevCVStDevCVStDevCVStDevCV
CKRhyperD101122.100000.14.50.13.10.15.5
CKRhyperDI08122.7000l0.13.50.12.40.14.3
CKRhyperDI69122.20.12.8000.13.10.12.50.14.6
CKRhypoD1411216.1O01.16.91.380.74.31.710.8
CKRhypoDI 691215.30.85.3001.610.21.17.1213.2
CKRhypoDI081212.10.1l0.10.90.97.20.75.61.19.3
CKKhyperD101120.5000003.904.105.7
CKKhyperD108120.403.803.807.206.5010.7
CKKhyperDI69120.5000008.208.20.111.6
CKKhypoD1411230.25.20.2б0.39.90.28.20.414.6
CKKhypoDI69123.10.516.4000:514.80.3100.722.7
CKKhypoDI08122.2000.16.40.3120.16.20.314.6
CKAnglehyperD10112800.81.11.51.87.296.98.710.112.6
СКAnglehyperDI081285.10.10.2000.10.10.10.10.20.2
CKAnglehyperDI691284.30.10.1000.10.10.10.10.20.2
CKAnglehypoD1411254.700008.415.45.49.91018.3
CKAnglehypoDI 691242.712.41.94.51.84.31.63.73.17.2
CKAnglehypoDI081266.734.4004.87.23.356.39.5
CKMAhyperDI081271.6000.50.70.30.40.30.40:50.8
CKMAhyperDI691270000.10.10.20.20.10.20.20.3
CKMAhyperD1211271.100000.91.30.60.81.11.5
CKMAhypoD14112420.20.6001.53:50.92.11.74.1
CKMAhypoDI691246.600000.81.80.51.1l2.1
СКMAhypoDI081245.4000.81.81.22.70.921.73.7
СКНRhyperD1011220.14.80002.40.12.80.1ર્ડ
CKHRhyperDI08122.600000.12.6O1.60.13.1
CKHRhyperDI69122.200000ાં .50101.8
CKHRhypoD1411215.70016.40.74.50.64.11.38.2
CKHRhypoDI 691214.5000.85.20.85.70.53.21.17.9
CKHRhypoDI081212.30.65000.86.30.54.41.18.8
CRTMAhyperDI081270.80.20.3000.10.20.10.20.20.3
CRTMAhyperD1211272.50.20.3000.40.50.40.50:50.8
CRTMAhyperD14512730.20.2000.20.30.20.20:30.4
CRTMAhypoD1411241.300000.71.80.51.20.92.2
CRTMAhypoDI 691244.4000.30.70.61.30.40.90.81.7
CRTMAhypoDI081245.5000.40.912.10.51.11.12.5
CFFMAhyperD1011249.90.10.1000.50.90.30.50.51.1
CFFMAhyperD1411244.300000.51.20.30.60.61.3
CFFFMAhyperD145124700000.71.40.5l0.81.7
CFFMAhypoD1011210.20.21.5000.330.21.70.43.6
CFFFMAhypoD125129.5000.21.80.440.32.70.55
CFFMAhypoDI69126.5000.710.10.34.90.4б0.811.6
CFFFLEVhyperD10112910.71.20.1008.60.950.5101.1
CFFFLEVhyperD1411280900009.61.25.20.610.91.3
CFFFFLEVhyperD14512858.1000011.91.48.5l14.71.7
CFFFLEVhypoD10112186.42.91:5005.533.11.76.83.6
CFFFLEVhypoDI2512173.4003.11.8744.62.68.75
CFFFLEVhypoDI6912157.51.81.21.10.721.31.71.13.22.1
Spiked Precision - Table 1
TestParamPatientIDNMeanLotOperatorInstrumentRepetitionTotal
StDevCVStDevCVStDevCVStDevCVStDevGV
CKRMH-DB001121300000.43.21.28.91.29.4
CKRMH-DB0031214.40000002.416.92.416.9
CKRMH-DB0041211.80.540.54.3001.613.91.714.8
CKRMH-WR001129.90.770.32.70.32.60.54.90.88.5
CKRMH-WR0021210.1110.2000.33.21101.413.6
CKKMH-DB001121.30000000.218.20.218.2
CKKMH-DB002122.400000.415.4142144.5
CKKMH-DB003121.30000000.319.80.319.8
CKKMH-DB004120.90.111.400000.3330.334.4
CKKMH-WR001122.20000000.28.20.28.2
CKKMH-WR0021220000000.419.10.419.1
CKAngleMH-DB0011272.50000003.653.65
CKAngleMH-DB0021255.5000010.318.611.620.815.227.4
CKAngleMH-DB0031272.70000003.24.33.24.3
CKAngleMH-DB0041279.50000003.44.33.44.3
CKAngleMH-WR00112650000001.11.71.11.7
CKAngleMH-WR0021267.1001.11.60034.43.14.6
CKMAMH-DB0011260.20000000.91.50.91.5
CKMAMH-DB0021264.5000.20.4000.610.71
CKMAMH-DB0031264.1000.40.7000.50.80.71
CKMAMH-DB0041271.90.10.10.50.6000.50.70.70.9
CKMAMH-WR0011260.60000000.91.40.91.4
CKMAMH-WR0021259.7000011.60.81.31.22
CKHRMH-DB0011212.20.86.800000.75.818.2
CKHRMH-DB0021219.3001.68.4002.211.42.613.5
CKHRMH-DB0031213.5000.54001.410.61.511.1
CKHRMH-DB0041211.30000001.815.91.815.9
CKHRMH-WR001128.4000.10.8000.89.70.89.8
CKHRMH-WR002129.10.55.100000.77.40.88.6
CRTMAMH-DB0011264.90.10.1000.10.10.10.20.10.2
CRTMAMH-DB0021267.70.10.100000.20.20.20.2
CRTMAMH-DB00312660000000.10.20.10.2
CRTMAMH-DB0041273.80.10.1000.10.10.20.20.20.3
CRTMAMH-WR00112640.10.100000.20.30.20.3
CRTMAMH-WR0021264.70000000.30.40.30.4
CFFMAMH-DB0011220.70.10.4000.31.30.10.60.31.4
CFFMAMH-DB0021223.70.31.30.31.4000.20.70.41.6
CFFMAMH-DB0031223.40.2100000.31.50.41.7
CFFMAMH-DB0041251.91.12.200000.40.81.12
CFFMAMH-WR0011220.70000000.41.70.41.7
CFFMAMH-WR0021224.70.10.40.31.2000.31.30.41.7
CFFFLEVMH-DB00112377.31.40.4004.81.32.40.65.31.4
CFFFLEVMH-DB00212432.25.81.36.21.40030.76.81.6
CFFFLEVMH-DB00312426.54.5100006.21.57.31.7
CFFFLEVMH-DB00412947.120.92.200007.50.819.32
CFFFLEVMH-WR00112377.40000006.51.76.51.7
CFFFLEVMH-WR002124501.60.45.21.20061.37.71.7

{14}------------------------------------------------

MH-DB002

СК

R

12

21.3

0

0

0

0

1

4.7

2.1

2.3

11

10.1

{15}------------------------------------------------

Patient-Derived Hypo Samples (Dabigatran and Warfarin) – Table 2

Interference

Testing was performed in Coramed's laboratory for interference, using CLSI EPT-A2 as guidance.

{16}------------------------------------------------

For CK, potential interferents tested were Absence of a Discard Tube, Short Draw, Hemolysis, Hemodilution and epsilon aminocaprioc acid (EACA). Only Hemolysis and Hemodilution were found to be interferents.

For CRT, potential interferents tested were Absence of a Discard Tube, Short Draw, Hemolysis, Hemodilution and epsilon aminocaprioc acid (EACA). Short Draw, Hemolysis and Hemodilution were found to be interferents.

For CKH, Protamine was tested for interference. Protamine was found to be an interferent for CKH at concentrations above 0.062 mg/ml.

For CFF, potential interferents tested were Na Heparin and Hemodilution. Heparin was found to be an interferent for CFF above Heparin concentrations of 1 IU/ml, and Hemodilution was found to be an interferent above hemodilution levels of 30%.

Linearity

Testing was performed in Coramed's laboratory for linearity, using CLSI EP6-A as guidance.

CFF was tested and found to be linear with varying concentrations of Fibrinogen (RiaSTAP) added. These concentrations ranged from 0 mg/dl to 565.5 mg/dl, over an FLEV range of 156 mg/dl to 1145 mg/dl. This corresponds to an MA range of 8.5 mm to 62.7 mm.

Sensitivity and Specificity

CK was tested for Sensitivity and Specificity with respect to Na Heparing R parameters of five blood samples with different clinically relevant concentrations of Heparin to five samples without heparin. Both Sensitivity and Specificity were found to be 100%.

Heparin Neutralization

Heparin Neutralization of CKH was tested using blood samples spiked with Unfractionated Heparin (UFH) and Low Molecular Weight Heparin (LMWH) and comparing these results to data obtained from testing CK with un-spiked blood samples. CKH was found to completely neutralize UFH levels of 5 IU/ml blood and LMWH levels of 0.013 mg/ml blood.

SUMMARY OF CLINICAL PERFORMANCE DATA

Testing was performed at three clinical sites for Reference Ranges and Method Comparison.

{17}------------------------------------------------

Reference Ranges

Reference Ranges for the CORA assays were estimated using the CLSI C28-A3c Guideline on three reference sample groups. Blood samples from up to 55 normal volunteer subjects were taken at each of the three sites, for a total of 157 samples. There were at least 151 valid results for every parameter for all reagents. Samples were excluded on a per-parameter basis using standard outlier criteria. An alternative method would have been to exclude all parameters for any tests where one or more parameter results were determined to be outliers. However, it was decided to retain all valid parameter results in calculating reference ranges because a larger number of samples adds statistical credibility. Also, it does not appear that using the alternative approach would have any significant effect on the reference ranges. Subjects were chosen representing demographic populations of the three areas, regarding age, race and gender. These reference ranges are shown below.

Reagent↓NR(min)NK(min)NAngle(deg)NMA(mm)NFLEV(mg/dl)
CK1574.6-9.11570.8-2.115563 - 7815152 - 69
CRT15252 - 70
CKH1554.3-8.3
CFF15115 - 32152278-581

Reference Ranges for CKH are shown but not usually required because the only purpose is to neutralize the heparin in a patient's blood, regardless of the patient's hemostasis condition. Comparing results from the CK and CKH tests for the R parameter can assist in determining whether or not there is heparin in the blood sample.

Reference Ranges for Wet Quality Control (WQC) Abnormal material when used in CORA assay cartridges were estimated. Using three lots of Abnormal Quality Control and three citrated reagent lots, a total of over 135 test results were obtained.

Reagent ↓& Abnormal WQC(AWQC)R(min)K(min)Angle(deg)MA(mm)FLEV(mg/dl)
CK - AWQC0.8 - 1.50.6 - 0.875 - 8332 - 47
CRT - AWQC32 - 46
CKH – AWQC0.8 - 1.5
CFF - AWQC30 - 60563-873

Method Comparison

Method Comparison studies were conducted at three clinical sites on patient samples following CLSI EP09-A3 Guideline. The subjects enrolled were patients undergoing cardiovascular surgery or cardiology procedures, with blood samples drawn pre- and post-surgery and in the ICU. In order to broaden the range of comparison, up to 10% contrived samples were added. Summary statistics are presented below.

{18}------------------------------------------------

Method Comparison Statistics Summary – All Clinical Sites
ParameterRIntercept95% CISlope95% CI
CK - R (min)0.868-0.325-0.425-0.2251.0691.0341.103
CK - K (min)0.740-0.097-0.2920.0980.9060.7691.042
CK – Alpha (deg)0.6807.305-13.16027.7690.9400.6401.240
CK - MA (mm)0.9242.3510.1664.5360.9480.9120.983
CKH – R (min)0.812-0.329-0.393-0.2651.0571.0261.089
CRT CK – ΜΑ (mm)0.929-0.126-1.9091.6560.9780.9481.007
CFF – MA (mm)0.938-0.774-1.8360.2891.1261.0721.180
CFF - FLEV (mg/dl)0.928-12.495-36.81011.8191.1271.0591.194

In addition, a Reader Study was conducted to assess the comparison of CORA System and TEG 5000 results based on reading printed test outcomes. The study was performed at three sites, using three experienced TEG 5000 doctors as readers at each site, for a total of nine readers. 30 test outcomes were compared for each of the reagent test categories, for a total of 270 reading comparisons per reader. Excellent agreement between the two devices was demonstrated.

In both the Method Comparison Study and Reader Study, the CORA RapidTEG (CRT) MA parameter was compared to the TEG 5000 Kaolin (CK) MA parameter, to demonstrate that the CRT MA parameter is equivalent to the CK MA parameter but the final MA value is reached more quickly using the CRT assay.

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTING

The data and information provided in this submission support a substantial equivalence determination for the CORA System and the TEG 5000 predicate device.

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.