K Number
K143644

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-09-14

(266 days)

Product Code
Regulation Number
878.4300
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Premium Surgiclip™ (All models) The Premium Surgiclip™ clip applier has application in many types of surgical procedures to occlude vessels and other tubular structures and for vagotomy, sympathectomy and radiographic markings

Endo Clip™ M, ML, L The Endo Clip™ clip applier has application in endoscopic procedures to achieve occlusion of vessels and other tubular structures, and for radiographic markings

Endo Clip™ 5mm The Endo Clip™ clip applier has applications in endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings

Endo Clip™ II The Endo Clip™ II clip applier has application in many types of endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings

AcuClip™ The instrument is primarily indicated for patients undergoing laparoscopic surqical procedures involving dissection and occlusion of blood vessels, ducts and other tubular structures. It can also be used to mark anatomical structures during surgery, and for radiographic marking.

Device Description

The Premium Surgiclip™ clip applier consists of an applier shaft with attached handles and integrated cartridge containing 15 or 20 titanium clips. The clip applier jaw is placed around a vessel or other tubular structure. As the handles of the applier are brought together, the clip is closed around the vessel or structure. As the handles are released, a new clip is automatically loaded into the clip applier jaw.

Models available are:

Premium Surqiclip™ M – 11.5 and Premium Surgiclip™ II M-11.5 - medium sized clips with an 11.5 inch shaft

Premium Surgiclip™ M - 9.75 and Premium Surgiclip™ II M-9.75 - medium sized clips with an 9.75 inch shaft

Premium Surgiclip™ S-9.0 – small sized clips with an 9.0 inch shaft

Premium Surgiclip™ L-13.0 - large sized clips with an 13.0 inch shaft

The Endo Clip™ clip applier contains 20 titanium clips in the ML and M sizes, and 15 titanium clips in the L size. They are designed for introduction and use through all appropriately sized Covidien™ trocar sleeves, or larger sized trocar sleeves with the use of a converter.

The Endo Clip™ 5mm clip applier contains 12 titanium clips. The applier is designed for introduction and use through an appropriately sized Covidien™ trocar sleeve, or larger with the use of a converter. The overall length of the shaft is approximately 28 cm (11").

The Endo Clip™ II clip applier contains 20 titanium clips. The appliers are designed for introduction and use through all appropriately sized Covidien™ trocar sleeves, or larger sized trocar sleeves with the use of a converter.

The AcuClip™ right angle clip applier is loaded for delivery of twenty (20) 8 mm titanium ligating clips. It is designed for introduction and use through all appropriate sized Covidien™ trocar sleeves or larger trocar sleeves with the use of a converter. The distal portion of the AcuClip™ right angle clip applier forms a hook that is placed around the vessel or other tissue structure to be occluded. The clips are advanced into the jaws by squeezing the handle approximately half-way. The clips are then closed by fully depressing the handle.

AI/ML Overview

This document is a 510(k) premarket notification for surgical clips and clip appliers. It is not a study that proves a device meets acceptance criteria but rather a submission to demonstrate substantial equivalence to previously cleared predicate devices. The document explicitly states that "In-vitro and In-vivo performance testing were not conducted as a requirement of the labeling change to add a contraindication."

Therefore, I cannot provide the requested information about acceptance criteria, device performance, sample sizes, ground truth establishment, or multi-reader multi-case studies because no new performance studies were conducted or reported in this document.

This submission focuses on a labeling change (specifically, the addition of a contraindication) for existing devices already marketed. The FDA's review in this case is to determine if the device, with the proposed labeling change, is still substantially equivalent to its predicates, not to re-evaluate its fundamental performance against criteria through new studies.

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 14, 2015

Covidien Mr. Michael Koczocik Product Specialist, Regulatory Affairs 60 Middletown Ave North Haven, Connecticut 06473

Re: K143644

Trade/Device Name: Premium Surgiclip™; Endo Clip™; AcuClip™ Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: August 4, 2015 Received: August 7, 2015

Dear Mr. Koczocik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv vours.

Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the logo for Covidien. The logo consists of a blue square made up of smaller squares of varying shades of blue on the left. To the right of the square is the word "COVIDIEN" in a dark blue sans-serif font.

Form Approved: OMB No. 0910-0120

to occlude vessels and other

Food and Drug AdministrationExpiration Date: January 31, 2017
Indications for UseSee PRA Statement below.
510(k) Number (if known)K143644
Device NamePremium Surgiclip
Indications for Use (Describe)The Premium Surgiclip TM clip applier has application in many types of surgical procedures to occlude vessels and other tubular structures and for vagotomy, sympathectomy and radiographic markings

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Servaces (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square made up of smaller squares on the left and the word "COVIDIEN" in blue on the right. The word "COVIDIEN" is in a sans-serif font.

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

Indications for Use

510(k) Number (if known) K143644

Device Name Endo Clip™M

Indications for Use (Describe)

Endo Clip™M ML, L - The Endo Clip™ clip applier has application in endoscopic procedures to achieve occlusion of vessels and other tubular structures, and for radiographic markings.

Endo Clip™ 5mm - The Endo Clip™ clip applications in endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings.

Endo Clip™ II - The Endo Clip™ II clip applier has application in many types of endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the word "COVIDIEN" in blue font. To the left of the word is a blue square with a white rectangle in the middle. The square is divided into four smaller squares, with the top left and bottom right squares being a darker shade of blue than the top right and bottom left squares. The logo is simple and modern, and the word "COVIDIEN" is easy to read.

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

Indications for Use

510(k) Number (if known)

K143644

Device Name AcuClip™

Indications for Use (Describe)

The instrument is primarily indicated for patients undergoing laparoscopic surgical procedures involving dissection and occlusion of blood vessels, ducts and other tubular structures. It can also be used to mark anatomical structures during surgery, and for radiographic marking.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{5}------------------------------------------------

510(k) Summary

This 510(k) summary of data used to demonstrate substantial equivalence is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR § 807.98

NAME:COVIDIEN
ADDRESS:60 Middletown AvenueNorth Haven, Connecticut 06473 USA
CONTACT PERSON:Michael KoczocikProduct Specialist, Regulatory Affairs
PHONE NUMBER:(203) 492-6312
FAX NUMBER:(203) 492-5029
DATE PREPARRED:March 13, 2015
TRADE/PROPRIETARY NAME:Premium Surgiclip™Endo Clip™AcuClip™
COMMON/USUAL NAME:Surgical clip appliers with implantable clips
CLASSIFICATION NAME:Implantable Clip per 21 CFR § 878.4300
PRODUCT CODE:FZP
CLASSIFICATION PANEL NAME:General and Plastic Surgery
FDA PANEL NUMBER:79
DEVICE CLASS:Pursuant to 21 CFR § 878.4300 implantable clipsare Class II devices
PREDICATE DEVICE(S):Premium Surgiclip™ III (K853650), (K142869)Endo Clip™ (K061288), (K883018)AcuClip™ (K920599)

{6}------------------------------------------------

INTENDED USE:

Premium Surgiclip™ (All models) The Premium Surgiclip™ clip applier has application in many types of surgical procedures to occlude vessels and other tubular structures and for vagotomy, sympathectomy and radiographic markings

Endo Clip™ M, ML, L The Endo Clip™ clip applier has application in endoscopic procedures to achieve occlusion of vessels and other tubular structures, and for radiographic markings

Endo Clip™ 5mm The Endo Clip™ clip applier has applications in endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings

Endo Clip™ II The Endo Clip™ II clip applier has application in many types of endoscopic procedures to achieve occlusion of vessels and other tubular structures and for radiographic markings

AcuClip™ The instrument is primarily indicated for patients undergoing laparoscopic surqical procedures involving dissection and occlusion of blood vessels, ducts and other tubular structures. It can also be used to mark anatomical structures during surgery, and for radiographic marking.

{7}------------------------------------------------

Device Descriptions

Reorder Codes: Premium Surgiclip™

134031 - Premium Surgiclip™ M-11.5 134044 - Premium Surgiclip™ M-9.75 134046 - Premium Surgiclip™ S-9.0 134048 - Premium Surgiclip™ L-13.0 134051 - Premium Surgiclip™ II M-9.75 134053 - Premium Surgiclip™ II M-11.5

Description of Premium Surgiclip™ Devices

The Premium Surgiclip™ clip applier consists of an applier shaft with attached handles and integrated cartridge containing 15 or 20 titanium clips. The clip applier jaw is placed around a vessel or other tubular structure. As the handles of the applier are brought together, the clip is closed around the vessel or structure. As the handles are released, a new clip is automatically loaded into the clip applier jaw.

Models available are:

Premium Surqiclip™ M – 11.5 and Premium Surgiclip™ II M-11.5 - medium sized clips with an 11.5 inch shaft

Premium Surgiclip™ M - 9.75 and Premium Surgiclip™ II M-9.75 - medium sized clips with an 9.75 inch shaft

Premium Surgiclip™ S-9.0 – small sized clips with an 9.0 inch shaft

Premium Surgiclip™ L-13.0 - large sized clips with an 13.0 inch shaft

Reorder Codes: Endo Clip™

176615 – Endo Clip™ ML 176619 – Endo Clip™ M 176625 — Endo Clip™ L

The Endo Clip™ clip applier contains 20 titanium clips in the ML and M sizes, and 15 titanium clips in the L size. They are designed for introduction and use through all appropriately sized Covidien™ trocar sleeves, or larger sized trocar sleeves with the use of a converter.

{8}------------------------------------------------

176620 – Endo Clip™ 5mm

The Endo Clip™ 5mm clip applier contains 12 titanium clips. The applier is designed for introduction and use through an appropriately sized Covidien™ trocar sleeve, or larger with the use of a converter. The overall length of the shaft is approximately 28 cm (11").

176657 – Endo Clip™ II

The Endo Clip™ II clip applier contains 20 titanium clips. The appliers are designed for introduction and use through all appropriately sized Covidien™ trocar sleeves, or larger sized trocar sleeves with the use of a converter.

Reorder Codes: AcuClip™

OMS-A8 - AcuClip™

The AcuClip™ right angle clip applier is loaded for delivery of twenty (20) 8 mm titanium ligating clips. It is designed for introduction and use through all appropriate sized Covidien™ trocar sleeves or larger trocar sleeves with the use of a converter. The distal portion of the AcuClip™ right angle clip applier forms a hook that is placed around the vessel or other tissue structure to be occluded. The clips are advanced into the jaws by squeezing the handle approximately half-way. The clips are then closed by fully depressing the handle.

SUMMARY COMPARING THE TECHNOLOGICAL CHARACTERISTICS OF THE PROPOSED AND PREDICATE DEVICE(S)

MATERIALS:

PERFORMANCE DATA:

The clip appliers are designed to occlude vessels and other tissue and tubular structures. The technological characteristics of the proposed devices remain the same as the 510(k) cleared predicate devices. (Premium Surgiclip™ III K853650, K142869),(EndoClip™ K883018 & EndoClip™ III K061288), & (Acuclip K920599)

All components of the Premium Surgiclip™ Endo Clip™, and AcuClip™ are comprised of materials which are in accordance with ISO 10993-1

In-vitro and In-vivo performance testing were not conducted as a requirement of the labeling change to add a contraindication.

{9}------------------------------------------------

CONCLUSION:

The addition of the contraindication does not alter the performance of the devices.The Premium Surgiclip™, Endo Clip™, and AcuClip™ are
determined to be substantially equivalent to predicate devices (Premium Surgiclip™ III) K142869),(EndoClip K061288), & (Acuclip K920599)

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.