K Number
K143599

Validate with FDA (Live)

Date Cleared
2015-04-17

(120 days)

Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chemtrue® Drug Screen Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Morphine 300, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
BuprenorphineBUPBuprenorphine10
TricyclicAntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMPd-Methamphetamine1000
MorphineMORMorphine300
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ9-THC9 COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Drug Screen Cup Tests panel can consist of any combination of the above listed drug analytes.

The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cutoff concentration levels for these drugs in urine.

The Chemtrue Drug Screen Cup Tests with OPI 2000 are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana, Opiates 2000, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
BuprenorphineBUPBuprenorphine10
Tricyclic
AntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMPd-Methamphetamine1000
Opiates 2000MORMorphine2000
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ9-THC9 COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue Drug Screen Cup Tests with OPI 2000 panel can consist of any combination of the above listed drug analytes.

The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodizzepines, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cutoff concentration levels for these drugs in urine.

Device Description

The Chemtrue® Drug Screen Cup Tests are colloidal gold based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in the Cup format, as indicated by the test name.

AI/ML Overview

The provided document describes the performance characteristics of the Chemtrue® Drug Screen Cup Tests and Chemtrue® Drug Screen Cup Tests with OPI 2000.

Here's an analysis of the acceptance criteria and the studies performed:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" as a pass/fail threshold. Instead, it presents the results of reproducibility, specificity, interference, and method comparison studies. The performance is reported in terms of agreement with GC/MS (reference method) and the percentage of accurate results at various concentrations (negative, 50%, 75%, 125%, 150% of cutoff). For the purpose of this response, "reported device performance" will be directly extracted from the provided tables.

Acceptance Criteria (Implied) and Reported Device Performance for Method Comparison (Accuracy) Study:

The document implies an acceptance criterion for accuracy based on the "Agreement" percentage. For most analytes, the agreement with the GC/MS reference method is 98% or 100%. For discordant results, it is noted that these were confirmed at the drug cutoff level.

AnalyteCutoff Concentration (ng/mL)Agreement with GC/MS (Positive Results)Agreement with GC/MS (Negative Results)Overall Agreement ReportedNotes on Performance
Amphetamine (AMP)100097.6%100%97.6% (for positive agreement)One discordant negative result at 1061 ng/mL
Barbiturates (BAR)300100%100%100%
Benzodiazepines (BZO)300100%98%98% (for negative agreement)One discordant positive result at 253 ng/mL
Cocaine (COC)300100%98%98% (for negative agreement)One discordant positive result at 292 ng/mL
Ecstasy (MDMA)500100%98.3%98.3% (for negative agreement)One discordant positive result at 498 ng/mL
Methamphetamine (MET)1000100%100%100%
Methadone (MTD)300100%100%100%
Morphine 300 (MOR300)300100%100%100%
Opiates 2000 (OPI2000)2000100%98%98% (for negative agreement)One discordant positive result at 1701 ng/mL
Oxycodone (OXY)100100%100%100%
Phencyclidine (PCP)25100%98%98% (for negative agreement)One discordant positive result at 24.6 ng/mL
Marijuana (THC)50100%100%100%

Acceptance Criteria (Implied) and Reported Device Performance for OTC Lay-user Accuracy and Usability Studies:

The implied acceptance criterion here is 100% agreement for all drug concentrations (No Drug Present, 50% of cutoff, 75% of cutoff, 125% of cutoff, 150% of cutoff). For all analytes except THC, 100% agreement was reported across all concentration levels tested. THC showed 90% agreement at 75% of cutoff.

| Analyte | Cutoff (ng/mL) | Agreement (No Drug present) | Agreement (50% cutoff) | Agreement (75% cutoff) | Agreement (125% cutoff) | Agreement (150% cutoff) |
|---|---|---|---|---|---|
| AMP | 1000 | 100% | 100% | 100% | 100% | 100% |
| BAR | 300 | 100% | 100% | 100% | 100% | 100% |
| BZO | 300 | 100% | 100% | 100% | 100% | 100% |
| BUP | 10 | 100% | 100% | 100% | 100% | 100% |
| COC | 300 | 100% | 100% | 100% | 100% | 100% |
| MDMA | 500 | 100% | 100% | 100% | 100% | 100% |
| MET | 1000 | 100% | 100% | 100% | 100% | 100% |
| MTD | 300 | 100% | 100% | 100% | 100% | 100% |
| MOR | 300 | 100% | 100% | 100% | 100% | 100% |
| OPI | 2000 | 100% | 100% | 100% | 100% | 100% |
| OXY | 100 | 100% | 100% | 100% | 100% | 100% |
| PCP | 25 | 100% | 100% | 100% | 100% | 100% |
| TCA | 1000 | 100% | 100% | 100% | 100% | 100% |
| THC | 50 | 100% | 100% | 90% | 100% | 100% |


2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Method Comparison Studies (Accuracy Study):

    • Sample Size: "On average of 85 clinical specimens for each drug test and a total of 586 samples were tested."
    • Data Provenance: The document states "blind-labeled clinical specimen correlation study". It does not specify the country of origin of the data or explicitly state if it was retrospective or prospective, but the use of "clinical specimens" suggests these were real-world samples.
  • Reproducibility (Precision) Studies:

    • Sample Size: For each analyte and concentration level, 30 devices were tested (3 operators * 3 lots * 10 replicates for each level of Negative, 50%, 75%, Cutoff, 125%, 150% of the cutoff). The tables show n=210 for Negative concentration and n=30 for each of the other concentration levels (50%, 75%, Cutoff, 125%, 150% of cutoff) over 10 non-consecutive days.
    • Data Provenance: The study used "GC/MS confirmed drug spiked urine controls." This indicates artificially prepared samples by spiking drugs into drug-free urine, not clinical specimens. The location or type of institution where this was performed is not specified.
  • OTC Lay-user Accuracy and Usability Studies:

    • Sample Size: 100 intended lay-users participated.
    • Data Provenance: "GC/MS confirmed urine samples" were used. These samples were artificially prepared by "spiking drugs into drug-free urine pool" and aliquoted into individual blind-labeled containers. The study was conducted at "three (3) intended user sites," implying it was conducted in a real-world or simulated real-world setting for OTC use, but the origin of the drug-free urine or spiking drugs is not detailed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The ground truth for all performance studies (reproducibility, method comparison, and OTC lay-user accuracy) was established using Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) as the preferred confirmatory methods.

GC/MS and LC/MS are analytical chemistry techniques that provide definitive identification and quantification of substances. Therefore, the "ground truth" is based on the results of these highly accurate chemical methods, rather than human expert interpretation of images or clinical findings. The expertise implicitly lies in the technicians and analytical instruments performing the GC/MS/LC/MS analysis. No specific "number of experts" or their "qualifications" like radiologists with X years of experience are applicable here, as it's a lab-based chemical assay validation.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The concept of a "2+1" or "3+1" adjudication method is typically used in studies where multiple human readers interpret results and a consensus or tie-breaker is needed, especially in imaging studies.

In this context, the adjudication method is "none" for human readers. Any "adjudication" between the device's results and the ground truth was done by comparing the device's qualitative positive/negative result against the quantitative GC/MS/LC/MS cut-off value. If there was a discrepancy (discordant result), the GC/MS result was considered the definitive truth. The "blind-labeled clinical specimen correlation study" implies that the device results were compared after the GC/MS truth was known, or at least independently, to determine agreement.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done.

This type of study is relevant for evaluating human performance improvements with AI assistance, typically in diagnostic imaging where human readers interpret and classify cases. The Chemtrue® Drug Screen Cup Tests are in vitro diagnostic devices (qualitative immunoassays) for drug detection in urine, interpreted visually by a single user. There is no "AI assistance" involved in the interpretation process described in this document, nor is there a cohort of human readers making independent diagnostic judgments that would then be compared.


6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, a standalone performance study was done in essence. The "Reproducibility (Precision) Studies," "Specificity Study," and "Interference" studies examine the intrinsic performance of the device (immunoassay) itself under controlled conditions and against known reference methods (GC/MS). The tables showing detection rates at various concentrations and cross-reactivity are direct measures of the device's standalone performance.

The "Method Comparison Studies" (Accuracy study, section H.5) also evaluate the device's performance against the GC/MS reference method. While "three operators performed the testing," their role was to visually interpret the test lines, which is a direct readout of the device's chemical reaction. The accuracy is reported as the device's performance against the gold standard, not the operators' modified performance.

The "OTC Lay-user Accuracy and Usability Studies" (section H.6) assess "lay-user accuracy," which means how accurately un-trained individuals could read and interpret the device standalone results. This is still a standalone performance assessment of the device, albeit under user-influenced conditions.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The primary type of ground truth used for performance evaluation was Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). These are highly accurate and definitive analytical chemistry methods for identifying and quantifying substances in a sample.

  • For reproducibility studies, GC/MS confirmed drug spiked urine controls were used.
  • For method comparison studies, the device was compared to the GC/MS Reference Method using blind-labeled clinical specimens.
  • For OTC lay-user studies, GC/MS confirmed urine samples (spiked into drug-free urine pool) were used.

8. The sample size for the training set

The document describes the validation studies for the device but does not specify or mention a "training set" in the context of machine learning or AI algorithm development. The device is a lateral flow immunoassay, which relies on chemical reactions and visual interpretation, not on trained algorithms that require a separate training dataset.


9. How the ground truth for the training set was established

As there is no mention of a "training set" for an AI or machine learning algorithm in this document (the device is a qualitativeunoassay), this question is not applicable. The ground truth for the device's validation was established using GC/MS/LC/MS as detailed above.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle with three heads, which are represented by curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 17, 2015

CHEMTRON BIOTECH, INC. JANE ZHANG DIRECTOR OF QA/RA 9245 BROWN DEER ROAD SAN DIEGO CA 92121

Re: K143599

Trade/Device Name: Chemtrue® Drug Screen Cup Tests Chemtrue® Drug Screen Cup Tests With OPI 2000 Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, LFG, DKZ, DIO, LDJ, DNK, LAF, LCM, JXM, DIS, DJC, DJR Dated: April 9, 2015 Received: April 13, 2015

Dear Ms. Jane Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Stayce Beck -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143599

Device Name

Chemtrue Drug Screen Cup Tests

Indications for Use (Describe)

The Chemtrue® Drug Screen Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Morphine 300, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
BuprenorphineBUPBuprenorphine10
TricyclicAntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMPd-Methamphetamine1000
MorphineMORMorphine300
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ9-THC9 COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Drug Screen Cup Tests panel can consist of any combination of the above listed drug analytes.

The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cutoff concentration levels for these drugs in urine.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K143599

Device Name

Chemtrue Drug Screen Cup Tests with OPI 2000

Indications for Use (Describe)

The Chemtrue Drug Screen Cup Tests with OPI 2000 are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana, Opiates 2000, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
BuprenorphineBUPBuprenorphine10
Tricyclic
AntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMPd-Methamphetamine1000
Opiates 2000MORMorphine2000
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ9-THC9 COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue Drug Screen Cup Tests with OPI 2000 panel can consist of any combination of the above listed drug analytes.

The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodizzepines, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cutoff concentration levels for these drugs in urine.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

510(k) Summary

AS REQUIRED BY 21 CFR 807.92(c)

A. SUBMITTER: Chemtron Biotech, Inc. 9245 Brown Deer Road, Suite B, San Diego, CA 92121. TEL: 858-450-0044; FAX: 858-450-0046

Contact Person: Jane Zhang, Director of OA/RA Official FDA Correspondent 9245 Brown Deer Road, Suite B San Diego, CA 92121 Office: (858) 450-0044; FAX: (858) 450-0046 Email: jane@uschemtronbio.com

Date Prepared: April 15, 2015

B. DEVICE

Trade or Proprietary Name: Chemtrue® Drug Screen Cup Tests Chemtrue® Drug Screen Cup Tests with OPI 2000

Common Name: Multi-Drug Urine Test Panel Regulatory Class: Class II

Regulatory Information:

Drug of AbuseProductCodePanelRegulation Section
AmphetamineDKZToxicology 9121CFR 862.3100, Amphetamine Test System
CocaineDIOToxicology 9121 CFR 862.3250, Cocaine and metabolites Test System
MethamphetamineLAFToxicology 9121 CFR 862.3610, Methamphetamine Test System
MorphineDNKToxicology 9121 CFR 862.3640, Morphine Test System
OpiatesDJGToxicology 9121 CFR 862.3650 Opiates Test System
PhencyclidineLCMToxicology 91Unclassified, Enzyme immunoassay Phencyclidine
Marijuana (THC)LDJToxicology 9121 CFR 862.3870, Cannabinoids Test System
BenzodiazepinesJXMToxicology 9121 CFR 862.3170, Benzodiazepines Test System
BarbituratesDISToxicology 9121 CFR 862.3150, Barbiturates Test System
Ecstasy (MDMA)DJCToxicology 9121 CFR 862.3610, Methamphetamine Test System
MethadoneDJRToxicology 9121 CFR 862.3620, Methadone Test System
OxycodoneDJGToxicology 9121 CFR 862.3650, Opiate Test System

In addition, previously FDA cleared Chemtrue® BUP/TCA Drug Screen Cup Tests (K142396/CR140334):

Drug of AbuseProduct CodePanelRegulation Section
Buprenorphine (BUP)DIGToxicology 9121CFR 862.3650
Tricyclic Antidepressants (TCA)LFGToxicology 9121 CFR 862.3910

C. PREDICATE DEVICE

{7}------------------------------------------------

C-1.K061718

INNOVACON® Integrated E-Z Cup, submitter: INNOVACON Laboratories, Inc.

  • C-2. K060896
    OnSite CupKit™, submitter: Varian, Inc.

D. INDICATIONS FOR USE:

Device Name: Chemtrue® Drug Screen Cup Tests

Chemtrue® Drug Screen Cup Tests with OPI 2000

Indications for Use:

Chemtrue® Drug Screen Cup Tests:

The Chemtrue® Drug Screen Cup Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana, Morphine 300, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
BuprenorphineBUPBuprenorphine10
Tricyclic AntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP /METd-Methamphetamine1000
MorphineMORMorphine300
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ⁹-THC-9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Drug Screen Cup Tests panel can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

{8}------------------------------------------------

Chemtrue® Drug Screen Cup Tests with OPI 2000:

The Chemtrue® Drug Screen Cup Tests with OPI 2000 are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana, Opiates 2000, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
BuprenorphineBUPBuprenorphine10
Tricyclic AntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP/METd-Methamphetamine1000
Opiates 2000OPIMorphine2000
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ9-THC-9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Drug Screen Cup Tests with OPI 2000 panel can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

E. DEVICE DESCRIPTION

The Chemtrue® Drug Screen Cup Tests are colloidal gold based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in the Cup format, as indicated by the test name.

F. SUBSTANTIAL EQUIVALENCE INFORMATION:

Comparison with the predicate devices is outlined below:

{9}------------------------------------------------

Similarities and Differences
ItemCandidate DevicesPredicate (K061718 and K060896)
Intended UseSameFor qualitative detection of drugs of abuse inhuman urine
ResultsSameQualitative
MethodologySameLateral flow, competitive bindingimmunoassay based on the principle of antigenand antibody immunochemistry.
Storage2 - 30°C until the expiration dateK061718: 2 - 30°C until the expiration date;K060896: 15 – 30°C.
Intended UsersPrescription and Over-the-Counter(OTC) usersK061718:Prescription users, including point-of-care;K060896:Prescription users
Analytes and CutOffs (ng/mL)Buprenorphine – sameBarbiturates – sameTricyclic Antidepressants – sameOpiates 2000 – sameMDMA – sameMethadone – sameOxycodone - samePropoxyphene – not included in thedeviceAmphetamine – 1000Benzodiazepines – 300Cocaine - 300Methamphetamine – 1000Morphine - 300Phencyclidine – sameMarijuana – sameK061718Buprenorphine – 10Barbiturates - 300Tricyclic Antidepressants – 1000Opiates 2000 – 2000MDMA - 500Methadone - 300Oxycodone - 100Propoxyphene - 300K061718 and K060896Amphetamine – 1000 / 300Benzodiazepines – 300 / 200Cocaine - 300 / 150Methamphetamine – 1000 / 500 / 300Morphine - 300 / 2000Phencyclidine – 25Marijuana – 50
FormatCup onlyK061718-Cup and Dipcard; K060896-Cup

G. TEST PRINCIPLE

The Chemtrue® Drug Screen Tests are rapid lateral flow immunoassays in which chemically modified drugs (drug-protein conjugates) compete with drugs that may be present in urine. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the anti-drug antibody-colloidal gold conjugate pads are placed at the forward end of the membrane. If target drugs are present in the urine specimen below its cut-off concentration, the solution of the colored antibody-colloidal gold conjugates moves along with the sample solution by capillary action across the membrane to the immobilized drug-protein conjugate zone on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band indicates a

{10}------------------------------------------------

negative result. If the target drug level exceeds its cut-off concentration, the drug/metabolite antigen competes with drug-protein conjugates on the test band region for the limited antibody on the colored drug antibody-colloidal gold conjugate pad. The drug will saturate the limited antibody binding sites and the colored antibody-colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.

A band should form in the control region (C) of the devices regardless of the presence of drug in the sample to indicate that the test has been performed properly.

Monoclonal anti-drug antibodies are used on the BUP/AMP/COC/MET/MOR/OPI2000/PCP/THC/ BAR/MDMA/MTD/OXY Test devices which are derived from mouse. The polyclonal anti-drug antibodies are used on TCA/BZO Test devices which are derived from sheep/mouse.

H. PERFORMANCE CHARACTERISTICS

Performance data is only provided for AMP, COC, MAMP, OPI2000, MOR300, PCP, THC, BZO, BAR, MDMA, MTD, and OXY, as the new analytes. BUP/TCA analytes of the candidate devices were previously cleared under K142396/CR140334.

1. Reproducibility (Precision) Studies:

The precision study was conducted by three (3) Operators with three (3) lots in replicates of 10 devices/lot at each concentration level of Negative, 50%, 75%, cut-off, 125% and 150% of the cutoff which are GC/MS confirmed drug spiked urine controls. The study was conducted over a ten (10) nonconsecutive days. The samples were blind coded according to a random table and randomly distributed to three operators by the project Manager. The data is analyzed and summarized in the tables below:

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301614
125% of cutoff30300
150% of cutoff30300
Table 1a. AMP Cup Test: Cutoff: 1000 ng/mL
--------------------------------------------------

{11}------------------------------------------------

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301614
125% of cutoff30300
150% of cutoff30300

Table 1b. BAR Cup Test: Cutoff: 300 ng/mL

Table 1c. BZO Cup Test: Cutoff: 300 ng/mL

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301515
125% of cutoff30300
150% of cutoff30300

Table 1d. COC Cup Test: Cutoff: 300 ng/mL

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301812
125% of cutoff30300
150% of cutoff30300

Table 1e. MDMA Cup Test: Cutoff: 500 ng/mL

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301713
125% of cutoff30300
150% of cutoff30300

{12}------------------------------------------------

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301614
125% of cutoff30300
150% of cutoff30300

Table 1f. MET Cup Test: Cutoff: 1000 ng/mL

Table 1g. MTD Cup Test: Cutoff: 300 ng/mL

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301317
125% of cutoff30300
150% of cutoff30300

Table 1h. MOR 300 Cup Test: Cutoff: 300 ng/mL

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301614
125% of cutoff30300
150% of cutoff30300

Table 11. OPI 2000 Cup Test: Cutoff: 2000 ng/mL

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301812
125% of cutoff30300
150% of cutoff30300

{13}------------------------------------------------

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301614
125% of cutoff30300
150% of cutoff30300

Table 1j. OXY Cup Test: Cutoff: 100 ng/mL

Table 1k. PCP Cup Test: Cutoff: 25 ng/mL

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301812
125% of cutoff30300
150% of cutoff30300

Table 1L. THC Cup Test: Cutoff: 50 ng/mL

Concentration LevelnTOTAL
+-
Negative2100210
50% of cutoff30030
75% of cutoff30030
Cutoff301515
125% of cutoff30300
150% of cutoff30300
  1. Specificity Study: These studies were conducted by adding various drugs, drug metabolites, and other structurally-similar compounds likely to be present in the actual urine specimen.

The following structurally-related compounds were tested for cross-reactivity and found to be positive if the levels were greater than the following listed concentrations:

{14}------------------------------------------------

Amphetamine related compounds:

SubstancesConc. (ng/mL)% Cross Reactivity
d-Amphetamine1,000100
d,l-Amphetamine2,50040
l-Amphetamine>100,000< 1
d-Methamphetamine>100,000< 1
l-Methamphetamine>100,000< 1
(d,l)-MDMA [(d,l)-3,4-Methylenedioxymethamphetamine]>100,000< 1
Ephedrine>100,000< 1
Pseudoephedrine>100,000< 1
(d,l)3,4-Methylenedioxyamphetamine (MDA)3,00033.3
Phentermine5,00020
MDEA>100,000< 1
d,l-Methamphetamine>100,000< 1
Phenylephrine>100,000< 1

Barbiturates related compounds:

SubstancesConcentration (ng/mL)% Cross Reactivity
Secobarbital300100
Pentobarbital300100
Alphenal50060
Amobarbital80037.5
Aprobarbital50060
Barbital10,0003
Butabarbital50060
Butalbital3,00010
Cyclopentobarbital75040
Phenobarbital2,00015

Benzodiazepines related compounds:

SubstancesConcentration (ng/mL)% Cross Reactivity
Oxazepam300100
Alprazolam300100
α-Hydroxyalprazolam100300
Bromazepam50060
Chlordiazepoxide2,50012
Clobazam200150
Clonazepam10,0003
Clorazepate35085.7
Desalkylflurazepam65462
Diazepam200150
Estazolam50060
Flunitrazepam37580

{15}------------------------------------------------

Flurazepam90333
Lorazepam60050
Lormetazepam7,5004
Midazolam90033.3
Nitrazepam200150
Nordiazepam150200
Temazepam35085.7
Triazolam1,00030

Cocaine related compounds:

SubstancesConcentration (ng/mL)% Cross Reactivity
Benzoylecgonine300100
Cocaine50060
Cocaethylene200001.5

MDMA (Methylenedioxymethamphetamine) related compounds:

SubstancesConc. (ng/mL)% Cross Reactivity
d,1-(3,4)-Methylenedioxymethamphetamine (MDMA)500100
3,4-Methylenedioxyamphetamine (MDA)15,0003.3
3,4-Methylenedioxyethylamphentamine (MDEA)1,00050
d-Methamphetamine100,0000.5
d-Amphetamine100,0000.5
l-Methamphetamine>100,000< 0.5
Ephedrine>100,000< 0.5
Pseudoephedrine>100,000< 0.5
d,1- Amphetamine>100,000<0.5
l-Amphetamine>100,000< 0.5
Phentermine>100,000<0.5
d,1- Methamphetamine>100,000<0.5
Phenylephrine>100,000<0.5

{16}------------------------------------------------

Methamphetamine related compounds:

SubstancesConc. (ng/mL)% Cross Reactivity
d-Methamphetamine1,000100
d,l-Methamphetamine5,00020
d-Amphetamine10,00010
l- Amphetamine> 100,000< 1
Ephedrine> 100,000< 1
(R)-(-)-Phenylephrine10,00010
Pseudoephedrine> 100,000< 1
d,l-MDMA (3,4- Methylenedioxymethamphetamine)5,00020
d,l-MDEA (Methylenedioxyethylamphetamine)100,0001
d,l-MDA (3,4- Methylenedioxyamphetamine)>100,000< 1
l-Methamphetamine>100,000< 1
d,l- Amphetamine>100,000< 1
Phentermine>100,000< 1

Methadone related compounds:

SubstancesConcentration (ng/mL)% Cross Reactivity
Methadone300100
Doxylamine100,0000.3
EDDP>100,000< 0.3
Pheniramine>100,000< 0.3

Morphine 300 related compounds:

SubstancesConcentration (ng/mL)% Cross Reactivity
Morphine300100
Codeine300100
6-Acetylmorphine50060
Diacetyl morphin (Heroin)2,00015
Hydrocodone50,0000.6
Hydromorphone5,0006
Oxycodone50,0000.6
Oxymorphone>100,000< 0.3
Procaine>100,000< 0.3

{17}------------------------------------------------

SubstancesConcentration (ng/mL)% Cross Reactivity
Morphine2000100
Codeine2000100
6-Acetylmorphine1500133.3
Diacetyl morphin (Heroin)2000100
Ethylmorphine1500133.3
Hydrocodone50,0004
Hydromorphone50,0004
Norcodeine100,0002
Normorphine100,0002
Oxycodone100,0002
Oxymorphone100,0002
Paracetamol (or Acetaminophen)100,0002
Thebaine100,0002

Opiates 2000 related compounds:

Oxycodone related compounds:

SubstancesConcentration (ng/mL)% Cross Reactivity
Oxycodone100100
Codeine100,0000.1
Hydrocodone100,0000.1
Oxymorphone100100

PCP related compounds:

SubstancesConcentration (ng/mL)% Cross Reactivity
Phencyclidine25100
Pheniramine>100,0000.025

THC related compounds:

SubstancesConcentration (ng/mL)% Cross Reactivity
11-nor-Δ9-THC-9-COOH50100
11-nor-Δ8-THC-9-COOH30167
Δ9-Tetrahydrocannabinol12,0000.4
Cannabidol>100,0000.05
Cannabinol>100,0000.05

3. Interference:

  • 3-1. Over one hundred of potential interferents were tested and found not to cross-react when tested at concentrations of 100 µg/mL at ±25% of the drug cut-off concentrations.
    Table 3. The following compounds do not interfere with the tests:

{18}------------------------------------------------

Endogenous Compounds:

AlbuminCreatinineRiboflavin
BilirubinGlucoseSodium Chloride
CholesterolHemoglobinUric Acid

Un-structurally related compound:

Un-structurally related compound:
Acetaminophen5, 5-DiphenylhydantoinOctopamine
AcetoneDopamineOxalic Acid
Acetylsalicylic Acid1-Erythromycin,Papaverine
AmoxicillinEstradiolPenicillin-G
AmpicillinEstronePerphenazine
R-(-)-ApomorphineEthanolPhenelzine
L-Ascorbic AcidFenofibratePhenylethylamine
AspirinFentanylPrednisone
AspartameFotemustinePromazine
AtropineFurosemidePromethazine
BaclofenGemfibrozild-Propoxyphene
BenzocaineGuaiacolglyceryl etherd,l-Propranolol
Benzoic AcidGentisic acidPyridoxine
CarisoprodolHydralazinePyrilamine
ChloramphenicolHydrocortisonePyrogallol
Chlordiazepoxide3-HydroxytyramineQuinidine
(+)-Chlorpheniramine(+/-)-IsoproterenolQuinine
ChlorpromazineKetamineQuinolinic Acid
ClofibrateMeprobamateRanitidine
ClonidineMethapyrileneSalicylic Acid
CortisoneMethylphenidateSulfamethazine
(-)-CotinineNalidixic AcidSulindac
Creatine HydrateNaloxoneTetracycline
CyclobenzaprineNaltrexoneTetrahydrozoline
Cyclodextrin-r(+)-NaproxenThiamine
CyproheptadineNiacinamideThioridazine
DeoxycorticosteroneNicotinic AcidTramadol
DextromethorphanNifedipineTrifluoperazine
Diclofenac19-NorethindroneTryptamine
DiflunisalNorpropoxypheneTyramine
4-Dimethyl-aminoantipyrineNoscapineZomepirac sodium salt
Diphenhydramine

Additional interference study: In addition to the cross-reactivity and interference studies presented in this submission, all the drug tests were tested with each of the 14 drugs at 150% and 50% of the drug cut-off urine samples. The results confirmed that there is no interference or cross-reactivity among these drug tests.

{19}------------------------------------------------

Effect of Urine pH and Specific Gravity Studies: The testing results demonstrate that the urine pH ranges from 2.0 to 9.0 at ±50% of the drug cut-off concentrations do not affect the test performance. The specific gravity (SG) ranges of 1.001, 1.010, 1.015, 1.020, 1.025 and 1.030 at ±50% of the drug cut-off concentrations do not affect the test results.

  • Stability Study: To establish and support the shelf life and expiration date, stability studies were 4. conducted under accelerated temperature (at 60°C and 40°C), and real time (25°C±2°C) with three (3) lots of each device format. The stability study results support two (2) years shelf-life of the products at (2 to 30℃). The real time stability study is still on-going.
    1. Method Comparison Studies:

Chemtrue® Drug Screen Cup Tests were compared to the GC/MS Reference Method. The accuracy of the Chemtrue® Test devices were evaluated against the confirmed GC/MS values in this blindlabeled clinical specimen correlation study (On average of 85 clinical specimens for each drug test and a total of 586 samples were tested). Three operators performed the testing. Each blind labeled sample was randomly distributed to each operator by the Clinical Research Cooperator. The results are summarized in the tables below:

Concentrations by GC/MS (ng/mL)
Chemtrue®Drug Screen Cup(-)(+)
Negative(<50% of the C/O)Near cutoffnegative(50% of the C/O to thecutoff)Near cutoffpositive(Cutoff to 150% of theC/O)Positive(≥150% of thecutoff)%Agreement
AMP(+)00162597.6%
(-)672510100%
BAR(+)001228100%
(-)441100100%
BZO(+)011525100%
(-)4090098%
COC(+)011525100%
(-)4090098%
MDMA (+)011228100%
(-)42140098.3%
MET(+)001525100%
(-)401000100%
MTD(+)001426100%
(-)401000100%
MOR300(+)001674100%
(-)401000100%
OPI2000 (+)011427100%
(-)4090098%
  • Table 5a. Summary from method comparison (Accuracy) study of Chemtrue® Drug Screen Cup Test results versus GC/MS

{20}------------------------------------------------

OXY(+)00931100%
(-)401000100%
PCP(+)011624100%
(-)40100098%
THC(+)001525100%
(-)401100100%

Table 5b. DISCORDANT RESULTS: There are six (6) discordant results in total of all the 12-drug tests:

Cutoff Value (ng/mL)Analyte assay(POS/NEG)Drug/Metabolite GC/MS value (ng/mL)
Drug AnalyteGC/MS Value (ng/mL)
Amphetamine 1000-Amphetamine1,061
Cocaine 300+Benzoylecgonine292
MDMA 500+MDMA498
PCP 25+Phencyclidine24.6
BZO 300+Oxazepam253
OPI 2000+Morphine and Codeine1,701

The discordant results were confirmed at the drug cutoff level with the GC/MS concentrations.

  1. OTC Lay-user Accuracy and Usability Studies:

This study demonstrates OTC accuracy and usability with AMP/BAR/BZO/COC/MDMA/MET/ MTD/MOR/OPI2000/OXY/PCP/THC and previously cleared BUP/TCA Tests (K142396). One hundred (100) intended lay-users participated in this OTC accuracy and usability study from three (3) intended user sites with GC/MS confirmed urine samples at the following concentrations: negative, 50%, 75%, 125% and 150% of the cutoff by spiking drugs into drug-free urine pool. Each sample was aliquot into an individual blind-labeled container. Each lay-user was provided with a package insert in English only, one blind labeled sample and one test device. The results are summarized below:

DrugAnalyteCut Off(ng/mL)ResultsDrug Concentrations (Per GC/MS Values)
No Drug present50% of thecutoff75% of thecutoff125% of thecutoff150% of thecutoff
AMP1000# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
BAR300# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
BZO300# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
# of positive0001010
BUP10# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
COC300# of positive000
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
MDMA500# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
MET1000# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
MTD300# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
MOR300# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
OPI2000# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
OXY100# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
PCP25# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
TCA1000# of positive0001010
# of Negative60101000
Total6010101010
Agreement100%100%100%100%100%
THC50# of positive0011010
# of Negative6010900
Total6010101010
Agreement100%100%90%100%100%

Table 6a. Chemtrue® Drug Screen Cup Test vs GC/MS Value Analysis

{21}------------------------------------------------

{22}------------------------------------------------

Lay-users were also given surveys on the ease of understanding the package insert instructions. ≥ 98% of the lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on the package insert and the score revealed a reading grade level of less than 7.

I. CONCLUSION:

Based on the test principle and performance characteristics of the proposed device, it is concluded that the candidate devices are substantially equivalent to the predicate device.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).