K Number
K143131

Validate with FDA (Live)

Date Cleared
2015-02-12

(104 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

The gloves are powder free, ambidextrous with beaded-cuff, black colored, single-use disposable devices that come in six sizes (XS, S, M, L, XL, XXL), and supplied in Non-Sterile state.

The gloves are made of Nitrile Butadiene Rubber, designed and manufactured in accordance with ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Its physical and performance characteristics meet all requirements of ASTM D6319-10.

AI/ML Overview

The provided document describes the acceptance criteria and performance of the Kossan International Sdn. Bhd. Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile (K143131). This is a medical device, and the "study" referred to is the non-clinical testing performed to demonstrate substantial equivalence to a predicate device.

Here's the information extracted and organized as requested:


1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Standards Requirements)Reported Device Performance (Results Summary)
DimensionsASTM D6319-10 (Length $\ge$ 230mm, Palm Thickness $\ge$ 0.05mm, Finger Thickness $\ge$ 0.05mm, Specific Width ranges for XS, S, M, L, XL, XXL)Meets Standard Requirements (Explicit values not provided in summary, but stated agreement with standard)
Physical PropertiesASTM D6319-10 (Before Aging: Tensile Strength $\ge$ 14 MPA, Elongation $\ge$ 500%; After Aging: Tensile Strength $\ge$ 14 MPA, Elongation $\ge$ 400%)Meets Standard Requirements (Explicit values not provided in summary, but stated agreement with standard)
Freedom from PinholesASTM D5151-11, ASTM D6319-10 (Pass quality level at G1 AQL 1.5)Tested in accordance with ASTM D5151 test method. Pass quality level at G1 AQL 1.5. Meets Standard Requirements.
Powder Free ResidueASTM D6124-11, ASTM D6319-10 (Values $\le$ 2 mg of residual powder per glove)Result generated values $\le$ 2 mg of residual powder per glove. Meets Standard Requirements.
BiocompatibilityDermal Sensitization (as ISO 10993-10:2010), Primary Skin Irritation Test (as ISO 10993-10:2010)Dermal Sensitization: Magnusson & Kligman Scale is '0'. Device is not a sensitizer. Meets Standard Requirements.Primary Skin Irritation: Primary Irritation Index for Erythema and Edema is '0'. Device is not an irritant. Meets Standard Requirements.

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes (number of gloves or animals/humans) used for each non-clinical test (dimensions, physical properties, pinholes, powder residue, biocompatibility). It only references adherence to ASTM and ISO standards, which would dictate appropriate sample sizes for each test.

  • Data Provenance: The tests were conducted by Kossan International Sdn. Bhd. (Malaysia), making the data provenance likely from Malaysia. The data is non-clinical/laboratory test data, not patient data, so "retrospective or prospective" is not an applicable distinction in this context.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable. The "ground truth" for this device's performance is established by objective measurements against recognized industry standards (ASTM, ISO) in a laboratory setting, not by expert consensus on clinical findings. Therefore, no experts were used to establish a clinical ground truth for the test set in the traditional sense of medical image analysis or diagnostic studies.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where multiple human readers assess data (e.g., images), and discrepancies need to be resolved. For non-clinical, objective laboratory tests of physical properties and biocompatibility, the results are typically determined by measurements and observations according to established protocols, not by expert adjudication.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This document pertains to the regulatory clearance of a physical medical device (examination gloves) based on non-clinical performance and biocompatibility data. It does not involve AI or human readers for diagnostic assessment.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. This document is about a physical medical device, not an algorithm or AI system.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" is based on objective laboratory measurements and tests against predefined industry standards (ASTM D6319-10, ASTM D5151-11, ASTM D6124-11, ISO 10993-10:2010 for biocompatibility). For example:

  • Physical dimensions: measured values compared to specified ranges.
  • Physical properties (tensile strength, elongation): measured values compared to minimum requirements.
  • Freedom from pinholes: water leak test results compared to AQL.
  • Powder free residue: measured weight of residual powder compared to maximum limit.
  • Biocompatibility: animal study results (Magnusson & Kligman Scale, Primary Irritation Index) compared to "non-sensitizer" and "non-irritant" conclusions.

8. The sample size for the training set

This information is not applicable. This is not an AI/machine learning device. No training set was used.


9. How the ground truth for the training set was established

This information is not applicable. This is not an AI/machine learning device. No training set was used.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three profiles overlapping to create a sense of depth and connection. The profiles are rendered in black and white. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 12, 2015

Kossan International Sdn. Bhd. Ms. Cho Sow Fong Manager Regulatory Affairs Wisma Kossan, Lot 782, Jalan Sungai Putus Off Batu 3 3/4, Jalan Kapar Klang, Selangor Darul Ehsan MALAYSIA 42100

Re: K143131

Trade/Device Name: Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile. Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: January 6, 2015 Received: January 8, 2015

Dear Ms. Fong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Fong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143131

Device Name

Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

[× Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for KOSSAN International. The word "KOSSAN" is in large, red, bold letters. Above the "K" is a gray graphic of three curved lines. Below the word "KOSSAN" is the word "INTERNATIONAL" in smaller, gray letters.

510(k) Premarket Notification K143131

FDA 510(k), Premarket Notification: 510(k) Summary of Safety and Effectiveness Information

Date Prepared: February 10, 2015

  • 1.0 Submitter:
    Kossan International Sdn. Bhd. Wisma Kossan, Lot 782, Jalan Sungai Putus, Off Batu 3 34, Jalan Kapar, 42100 Klang, Selangor, Malaysia

Telephone No.: +603 3291 0516 Fax No.: +603 3291 0542

2.0 Contact Person:

Contact:Ms Cho Sow Fong
Telephone No.:+603 3291 0516
Fax No. :+603 3291 0542

3.0 Name of Device:

Trade Name(s): Powder Free Nitrile Patient Examination GloveBlack Colored, Non-Sterile.
Common Name: Powder-Free Nitrile Patient Examination Glove
Classification Name: Patient Examination Glove
Regulation Number: 21 CFR 880.6250
Classification Number: Class I
Product Code: 80 LZA

4.0 Identification of the Legally Marketed Device:

Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile, Class I Patient Examination Gloves, Nitrile - 80 LZA, meets all of the requirements of ASTM D 6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application.

Predicate Device 1: K121529, Black Powder Free Nitrile Patient Exam Glove.

There are no different technological characteristics compared to the Predicate Devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for KOSSAN International. The word "KOSSAN" is in large, red, bold letters. Above the "K" is a gray design of three curved lines. Below the word "KOSSAN" is the word "INTERNATIONAL" in smaller, gray letters.

510(k) Premarket Notification K143131

5.0 Description of Device:

Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

The gloves are powder free, ambidextrous with beaded-cuff, black colored, single-use disposable devices that come in six sizes (XS, S, M, L, XL, XXL), and supplied in Non-Sterile state.

The gloves are made of Nitrile Butadiene Rubber, designed and manufactured in accordance with ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Its physical and performance characteristics meet all requirements of ASTM D6319-10.

6.0 Indications for Use:

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

7.0 Summary of the Technological Characteristics of the Device:

Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile processes the following technological characteristic (as compared to ASTM or equivalent standards):

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for KOSSAN International. The logo features a stylized letter 'K' with a red color scheme, accompanied by the text 'KOSSAN' in bold, red letters. Below the main text, the word 'INTERNATIONAL' is written in a smaller, gray font.

510(k) Premarket Notification K143131

CharacteristicStandardsRequirementsResults SummaryConclusions
DimensionsASTM D6319-10Length $\ge$ 230mmPalm Thickness $\ge$ 0.05mmFinger Thickness $\ge$ 0.05mmWidthX-Small 70-80mmSmall 80-90mmMedium 90-100mmLarge 101-111mmX-Large 111-121mmXX-Large 121-131mmMeetsStandardRequirements
PhysicalPropertiesASTM D6319-10Before Aging After Aging Tensile Strength $\ge$ 14 MPA $\ge$ 14 MPA Elongation $\ge$ 500% $\ge$ 400%MeetsStandardRequirements
Freedom frompinholesASTM D5151-11ASTM D6319-10Tested in accordance with ASTM D5151 testmethod. Pass quality level at G1 AQL 1.5MeetsStandardRequirements
Powder FreeResidueASTM D6124-11ASTM D6319-10Result generated values $\le$ 2 mg of residual powderper glove.MeetsStandardRequirements
BiocompatibilityDermal Sensitization(as ISO 10993-10:2010)Primary Skin IrritationTest (as ISO 10993-10:2010)Magnusson &Kligman Scale is '0'.Under the conditions of the study, the device is not asensitizer.Primary Irritation Index for Erythema and Edemais '0'.Under the conditions of the study, the device is notan irritant.MeetsStandardRequirements

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for KOSSAN International. The word "KOSSAN" is in large, red, bold letters. Above the word is a gray graphic of three curved lines. Below the word "KOSSAN" is the word "INTERNATIONAL" in smaller, gray letters.

510(k) Premarket Notification K143131

8.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data

Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile have been tested against the applicable ASTM standards listed above, and meet the requirements set forth in those standards.

There is no difference between the Subject Device and the Predicate Device with respect to performance standard and technological characteristics.

9.0 Substantial Equivalent Based on Assessment of Clinical Performance Data

Clinical data was not needed.

10.0 Conclusion

The Substantial Equivalent Comparison Table below outlines the similarity, and/or differences between the subject device and the predicate device for the substantial equivalent determination.

As such, this device is substantially equivalent to predicate device.

It can be concluded that the Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile, is substantially equivalent to the predicate device identified in this 510(k) summary.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Kossan International SDN. BHD. The logo features the word "KOSSAN" in red, with the word "INTERNATIONAL" in gray underneath. The company's registration number, (273178-M), is included to the right of the company name.

Substantial Equivalent Comparison Table

CharacteristicsSubject DevicePredicate DeviceComments
IdentificationK143131Powder Free Nitrile Patient ExaminationGlove, Black Colored, Non-SterileK121529Black Powder Free Nitrile Exam GloveN/A
Device Classification Name/Regulation NumberPatient Examination Glove/21 CFR Part 880.6250Patient Examination Glove/21 CFR Part 880.6250Substantially Equivalent
Product Code80 LZA80 LZASubstantially Equivalent
Intended UseIntended for medical and dental purposesthat is worn on the examiner's hand toprevent contamination between patient andexaminer.Intended for medical and dental purposesthat is worn on the examiner's hand toprevent contamination between patient andexaminer.Substantially Equivalent
Indications for UseA patient examination glove is a disposabledevice intended for medical purposes that isworn on the examiner's hand to preventcontaminationbetweenpatientandexaminer.The Examination Glove is a disposabledevice intended for medical purposes that isworn on the examiner's hand or finger toprevent contamination between patient andexaminer.Substantially Equivalent
MaterialsNitrileNitrileSubstantially Equivalent
ColorBlackBlackSubstantially Equivalent
DesignAmbidextrous, in different sizes per ASTMD6319 dimension requirement.Ambidextrous, in different sizes per ASTMD6319 dimension requirement.Substantially Equivalent
PerformanceI. SterilityII. Freedom from holesIII. DimensionIV. Physical PropertiesV. Powder Free ResidueNot Applicable (Non-Sterile)Passes at AQL 1.5Meets ASTM D6319Meets ASTM D6319Meets ≤ 2 mg/gloveNot Applicable (Non-Sterile)Passes at AQL 1.5Meets ASTM D6319Meets ASTM D6319Meets ≤ 2 mg/gloveSubstantially EquivalentSubstantially EquivalentSubstantially EquivalentSubstantially EquivalentSubstantially Equivalent

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Kossan International SDN. BHD. The logo features the word "KOSSAN" in red, with the word "INTERNATIONAL" in gray underneath. The company's registration number, (273178-M), is also included in the logo. The logo is simple and professional, and it is likely used on the company's website, marketing materials, and other official documents.

CharacteristicsSubject DevicePredicate DeviceComments
Single UseYesYesSubstantially Equivalent
Biocompatibility TestUnder the conditions of the study, thedevice is not a sensitizer or an irritant.Passesi. Primary Skin Irritation Test -Gloves are non-irritatingii. Dermal Sensitization Test -Gloves do not display any potential forsensitizationSubstantially Equivalent
PackagingPacked in Dispenser BoxesPacked in Dispenser BoxesSubstantially Equivalent
Labeling Features- Non-sterile- Powder Free- Examination Gloves- Ambidextrous, by Size- Single Use Only- Device Color: Black- Manufactured for:- Lot Number:- Quantity by Weight- Made in Malaysia- Non-sterile- Powder Free- Examination Gloves- Ambidextrous, by Size- Single Use Only- Device Color: Black- Manufactured for:- Lot Number:- Quantity by Weight- Made in MalaysiaSubstantially Equivalent

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.