K Number
K143037

Validate with FDA (Live)

Date Cleared
2015-01-20

(90 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SnapShot Fixation System is indicated for use in soft tissue reattachment procedures in the following shoulder procedures:

Bankart repair, SLAP repair, acromio-clavicular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, and deltoid repair.

The SnapShot Fixation System is also indicated for supplementary fixation when used in conjunction with a primary fixation device in surgical procedures requiring graft fixation.

Device Description

The SnapShot Fixation System consists of a SnapShot implant and deployment gun. The SnapShot implant is a PEEK rivet comprised of two components, a cannulated body and a bullet. The SnapShot deployment gun comes packaged pre-loaded with the SnapShot implant and is designed to ease insertion of the SnapShot implant into the bone hole. A reusable deployment gun and reusable drill are also available and can be used with the implant reloads.

AI/ML Overview

The provided text describes a medical device, the SnapShot Fixation System, and its 510(k) premarket notification for FDA clearance. However, it does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and a study proving those criteria are met. This document is a summary of the submission for clearance, not the full study report.

Specifically, the document states:

  • "Non-clinical laboratory testing was performed to verify the fixation strength of the SnapShot Fixation System in mechanical pullout testing as compared to the predicate LactoSorb Pop Rivet (K981798) for specific indications for use."
  • "The average pullout strength of the SnapShot Fixation System implants was statistically equivalent to or greater than that of the Lactosorb Pop Rivet."
  • "When testing supplemental fixation using a resorbable interference screw for primary fixation, the average pullout strength of the construct utilizing the SnapShot Fixation System for supplementary fixation was statistically greater than the construct with no supplemental fixation."
  • "Clinical Tests: None provided as a basis for substantial equivalence."

Based on the information provided, here's what can be extracted and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria for Non-Clinical Testing (Implied from the text):

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Fixation Strength (Pullout) vs. Predicate"statistically equivalent to or greater than" the Lactosorb Pop Rivet (K981798)"The average pullout strength of the SnapShot Fixation System implants was statistically equivalent to or greater than that of the Lactosorb Pop Rivet."
Supplemental Fixation Strength"statistically greater than" the construct with no supplemental fixation (when used with a primary fixation device like a resorbable interference screw)"the average pullout strength of the construct utilizing the SnapShot Fixation System for supplementary fixation was statistically greater than the construct with no supplemental fixation." (when the primary fixation was a resorbable interference screw)

Note: The specific numerical values (e.g., in Newtons) for acceptance criteria and performance are not provided in this document. The document only states the comparative outcome ("statistically equivalent to or greater than," "statistically greater than").


2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified. The document states "Non-clinical laboratory testing was performed," but does not give the number of samples or specimens tested.
  • Data Provenance: The study was a "Non-clinical laboratory testing" (mechanical testing). The country of origin for the data is not explicitly stated, but given the sponsor (Biomet Inc., Warsaw, IN) and the FDA submission, it's highly likely to be conducted in the USA or by labs commissioned by a US company. It is by definition retrospective in relation to the submission date, as the tests were completed before the 510(k) was filed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This was a non-clinical mechanical performance study. "Ground truth" in the clinical sense (e.g., expert consensus on medical images or patient outcomes) is not relevant to this type of testing. The "ground truth" for mechanical testing is established by internationally recognized standards for mechanical testing of medical devices, which are not mentioned in this summary but would have been followed.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As this was a non-clinical mechanical performance study, there was no need for expert adjudication of results in the way it's done for clinical or imaging studies. Mechanical test results are typically objective measurements.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical fixation system (screws/rivets) for soft tissue reattachment, not an AI or imaging diagnostic device. Therefore, MRMC studies and AI assistance are not relevant to its evaluation.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not a software algorithm.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For non-clinical mechanical testing, the "ground truth" is defined by the physical measurements obtained from calibrated testing equipment under controlled conditions, following established mechanical testing standards (e.g., ASTM, ISO standards for orthopedic implants). These standards themselves can be considered the "ground truth" methodology.


8. The sample size for the training set

Not applicable. This describes a non-clinical mechanical performance study for a physical device, not an AI or machine learning model that would require a 'training set'.


9. How the ground truth for the training set was established

Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 20, 2015

Biomet Manufacturing Corporation Mr. Adam Cargill Regulatory Affairs Specialist 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581

Re: K143037

Trade/Device Name: SnapShot Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDR Dated: October 20, 2014 Received: October 22, 2014

Dear Mr. Cargill:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

Page 2 - Mr. Adam Cargill

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143037

Device Name SnapShot Fixation System

Indications for Use (Describe)

The SnapShot Fixation System is indicated for use in soft tissue reattachment procedures in the following shoulder procedures:

Bankart repair. SLAP repair, acromio-clavicular separation, rotator cuff repair, capsulo labral reconstruction, biceps tenodesis, and deltoid repair.

The SnapShot Fixation System is also indicated for supplementary fixation when used in conjunction with a primary fixation device in surgical procedures requiring graft fixation.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "BIOMET" in a simple, sans-serif font. The letters are all capitalized and appear to be a dark gray color. The background is plain white, which makes the text stand out. The font is clean and modern.

510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the SnapShot Fixation System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.

Sponsor:Biomet Inc.56 East Bell DrivePO Box 587Warsaw, IN 46581Office: 574-267-6639Main Fax: 574-267-8137Establishment Registration Number: 1825034
Contact:Adam CargillRegulatory Affairs Specialist, Sports Medicine
Date:January 16, 2015
Subject Device:Trade Name: SnapShot Fixation SystemCommon Name: Soft Tissue Fixation DeviceClassification Name: JDR - Staple, Fixation, Bone (21 CFR 888.3030)

Legally marketed devices to which substantial equivalence is claimed:

Biomet Lactosorb Pop Rivet (K981798) Biomet PEEK Knotless Anchor (K070389) – reference for materials

Device Description

The SnapShot Fixation System consists of a SnapShot implant and deployment gun. The SnapShot implant is a PEEK rivet comprised of two components, a cannulated body and a bullet. The SnapShot deployment gun comes packaged pre-loaded with the SnapShot implant and is designed to ease insertion of the SnapShot implant into the bone hole. A reusable deployment gun and reusable drill are also available and can be used with the implant reloads.

Intended Use and Indications for Use

The SnapShot Fixation System is indicated for use in soft tissue reattachment procedures in the following shoulder procedures:

Bankart repair, SLAP repair, acromio-clavicular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, and deltoid repair.

{4}------------------------------------------------

BIOMET

The SnapShot Fixation System is also indicated for supplementary fixation when used in conjunction with a primary fixation device in surgical procedures requiring graft fixation.

Summary of Technological Characteristics

The technological characteristics (materials, design, sizing, and indications) of the SnapShot Fixation System are similar or identical to the predicate device or other previously cleared devices.

The rationale for substantial equivalence is based on consideration of the following characteristics:

  • . Intended Use: The proposed SnapShot Fixation System has the same intended use as the predicate Lactosorb Pop Rivet (K981798) for soft tissue to bone fixation.
  • Indications for Use: The proposed SnapShot Fixation System has similar indications for use as the predicate Lactosorb Pop Rivet (K981798), for soft tissue to bone fixation, with the addition of providing supplementary fixation in surgical procedures requiring graft fixation.
  • Materials: The proposed SnapShot Fixation System utilizes the same implant material, polyetheretherketone (PEEK), as the PEEK Knotless Anchor (K070389).
  • . Design Features: The proposed SnapShot Fixation System has the same design features and is dimensionally similar to the predicate Lactosorb Pop Rivet (K981798).

Summary of Performance Data (Nonclinical and/or Clinical)

  • Non-Clinical Tests
    • о Non-clinical laboratory testing was performed to verify the fixation strength of the SnapShot Fixation System in mechanical pullout testing as compared to the predicate LactoSorb Pop Rivet (K981798) for specific indications for use. The average pullout strength of the SnapShot Fixation System implants was statistically equivalent to or greater than that of the Lactosorb Pop Rivet. When testing supplemental fixation using a resorbable interference screw for primary fixation, the average pullout strength of the construct utilizing the SnapShot Fixation System for supplementary fixation was statistically greater than the construct with no supplemental fixation.
  • Clinical Tests
    • None provided as a basis for substantial equivalence. o

Substantial Equivalence Conclusion

The proposed SnapShot Fixation System has similar intended use, technological characteristics, and mechanical performance as the Lactosorb Pop Rivet (K981798). The performance testing data identified no new risks and substantial equivalence to the legally marketed predicate device.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.