K Number
K981798

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-07-13

(53 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LactoSorb®Pop Rivet are indicated for use in soft tissue reattachment in the following shoulder procedures:

    1. instability repairs in the shoulder (Bankart procedures)
    1. SLAP lesion repair
    1. Acromio-clavicular separation repairs
    1. Rotator cuff repairs
    1. Capsular shift or capsulolabral reconstructions
    1. Biceps tenodesis
    1. Deltoid repair
Device Description

The LactoSorb* Pop Rivets are composed of bioresorbable and biocompatible polymers that have been used in surgical procedures for years. LactoSorb* resorbable copolymer is a synthetic polyester derived from lactic and glycolic acids. Polylactic/polyglycolic acid copolymer degrades and resorbs IN VIVO by hydrolysis to lactic and glycolic acids which are then metabolized by the body. In animal studies LactoSorb* has been found to be biocompatible in both soft tissue and bone tissue.

AI/ML Overview

The provided text is a 510(k) summary for a medical device called the LactoSorb® Pop Rivet. It does not contain any information about acceptance criteria or a study proving the device meets those criteria, nor does it discuss AI-related performance metrics.

The summary primarily focuses on:

  • Device Name and Manufacturer: LactoSorb® Pop Rivet by Biomet, Inc.
  • Indications for Use: Soft tissue reattachment in various shoulder procedures (Bankart, SLAP, AC separation, rotator cuff, capsular shift, biceps tenodesis, deltoid repair).
  • Device Composition: Bioresorbable and biocompatible polymers (polylactic/polyglycolic acid copolymer).
  • Mechanism of Degradation: Hydrolyzes to lactic and glycolic acids, which are then metabolized.
  • Biocompatibility: Stated to be biocompatible in soft tissue and bone tissue based on animal studies.
  • Substantial Equivalence: Claim that it does not raise new safety or effectiveness concerns compared to the predicate device.
  • FDA 510(k) Clearance: Letter from the FDA confirming substantial equivalence to a legally marketed predicate device.

Therefore, I cannot provide the requested information regarding acceptance criteria, device performance, study details, or AI-related metrics as this information is not present in the provided document.

{0}------------------------------------------------

JUL 1 3 1998

510(K) SUMMMARY OF SAFETY AND EFF

SUBMITTER: Biomet, Inc. Airport Industrial Park P. O. Box 587 Warsaw IN 46580 -0587

Contact Person: Julie K. Ryan

ﺎ ﻳﻮﺟﺪ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ Product Code: 87HWC .

Device Name: LactoSorb® Pop Rivet

The LactoSorb®Pop Rivet are indicated for use in soft tissue reattachment in the following shoulder procedures:

    1. instability repairs in the shoulder (Bankart procedures)
    1. SLAP lesion repair
    1. Acromio-clavicular separation repairs
    1. Rotator cuff repairs
    1. Capsular shift or capsulolabral reconstructions
    1. Biceps tenodesis
    1. Deltoid repair

ﻢ ﺍﻟﻤﺤﺎﻓﻈﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ

The LactoSorb* Pop Rivets are composed of bioresorbable and biocompatible polymers that have been used in surgical procedures for years. LactoSorb* resorbable copolymer is a synthetic polyester derived from lactic and glycolic acids. Polylactic/polyglycolic acid copolymer degrades and resorbs IN VIVO by hydrolysis to lactic and glycolic acids which are then metabolized by the body. In animal studies LactoSorb* has been found to be biocompatible in both soft tissue and bone tissue.

The use of this device in the shoulder indications listed above does not raise any new types of safety and effectiveness when compared to the predicate device.

00004

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 3 1998

Ms. Julie K. Ryan ·Requlatory Specialist Biomet, Inc. P.O. Box 587 Warsaw, Indiana 46581-0587

Re : K981798 LactoSorb® Pop Rivet Requlatory Class: II Product Codes: JDR, MAI, and HRX Dated: May 20, 1998 May 21, 1998 Received:

Dear Ms. Ryan:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸੈ substantially equivalent determination assumes compliance with ... _ the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Ms. Julie K. Ryan

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Col. M. Whitten, Ph.D.

Ph.D., M.D. M. Witten, Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) NUMBER (IF KNOWN): K981798

LactoSorb* Pop Rivet DEVICE NAME:

INDICATIONS FOR USE:

The LactoSorb* Pop Rivets are indicated for use in soft tissue reattachment in the following procedures:

  • instability repairs in the shoulder (Bankart Procedures) 1 . .
  • SLAP lesion repair 2.
  • 3 . . Acromio-clavicular separation repairs
  • Rotator cuff tear repairsg 4. - 4.
  • Capsular shift or capsuloral reconstructions 5.
  • 6 . Biceps Tenodesis
  • 7 :- Deltood repairs

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X OR Over-The-Counter-Use
(Per 21 CFR 801.109) (Optional Format 1-2-96)

(Division Sign-Off)
Division of General Restorative Devices
510(k) NumberK981798