K Number
K142943

Validate with FDA (Live)

Date Cleared
2015-08-03

(297 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The metallic locking bone plate and screw are intents with age above 21 as indicated fixation of fractures, including ulna, radius, humerus, femur, tibia, clavicle, and fibula for the following indications:

  • ostcotomies, mal-unions, and non-unions;
  • single, segmental, and comminuted fractures;
  • normal bone density and osteopenic bone.
Device Description

The proposed device, metallic Locking Bone Plate and Screw consists of bone plate and screw. The bone plates are used for internal fixation of bone fracture, including ulna, radius, humerus, femur and tibia. Implant is manufactured with Ti6Al4V Eli and titanium which conforms to ASTM F 136 and ASTM F67.
The proposed devices are provided un-sterilized. They shall be sterilized prior to use by healthcare provider. The proposed devices shall never be reused.

AI/ML Overview

This document is a 510(k) premarket notification approval for a medical device: "Metallic Locking Bone Plate and Screw". It primarily focuses on demonstrating substantial equivalence to predicate devices based on non-clinical performance data and material biocompatibility, rather than providing details on a study proving the device meets acceptance criteria regarding clinical outcomes or diagnostic performance.

Therefore, many of the requested elements are not applicable to the information contained in this FDA letter and summary. However, I can extract the relevant information based on the document's content.


Device Name: Metallic Locking Bone Plate and Screw

Purpose of the Submission: To demonstrate substantial equivalence to legally marketed predicate devices.

1. Table of acceptance criteria and the reported device performance

Acceptance Criteria (indirectly inferred from testing standards)Reported Device Performance (as demonstrated by testing)
Mechanical performance meeting established standards for metallic bone plates (e.g., bend testing, screw properties)Complies with ASTM F382 - 99(2008) Standard Specification and Test Method for Metallic Bone Plates (bend testing).
Complies with ASTM F543 - 13 Standard Specification and Test Methods for Metallic Medical Bone Screws.
Biocompatibility of materialsMade of pure titanium and Ti-6Al-4V, complying with ASTM F1472 and ASTM F67. Materials are "well-characterized" and have been "employed successfully in human implant applications in contact with soft tissue and bone for over a decade."
Sterilization requirementsProvided un-sterilized; "shall be sterilized prior to use by healthcare provider."
Single-use requirement"The proposed devices shall never be reused."

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in the context of clinical or diagnostic performance data or sample sizes for mechanical testing. The performance data provided is based on non-clinical bench testing and material specifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for mechanical performance is established by recognized engineering standards (ASTM). Biocompatibility is based on material science and established historical use, not expert consensus on a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical or diagnostic test set requiring expert adjudication was conducted.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a passive implantable device, not an AI-assisted diagnostic or therapeutic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an algorithm-based device.

7. The type of ground truth used

The ground truth for this device's performance relies on:

  • Engineering Standards: Bench testing results compared against established ASTM standards for metallic bone plates and screws.
  • Material Specifications and Historical Use: Biocompatibility is established by the use of materials (pure titanium and Ti-6Al-4V) that conform to ASTM standards (F1472, F67) and have a long history of successful implantation in humans.

8. The sample size for the training set

Not applicable. No "training set" in the context of machine learning or clinical trials was used. The device design and performance are based on engineering principles and material science, informed by prior knowledge and testing of similar devices and materials.

9. How the ground truth for the training set was established

Not applicable. There was no training set. The "ground truth" for the device's design and materials is derived from established engineering and material science principles, as well as the performance of predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure, with flowing lines suggesting movement or flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

August 3, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Canwell Medical Company, Ltd % Mr. Mike Gu Osmunda Medical Device Consulting Company, Ltd. 7th Floor Jingui Business Building 982 Cogyuan Road, Baiyun District 510420 Guangzhou China

Re: K142943

Trade/Device Name: Metallic Locking Bone Plate and Screw Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: July 3, 2015 Received: July 6, 2015

Dear Mr. Gu,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labcling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

Page 2 - Mr. Mike Gu

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K142943

Device Name Metallic Locking Bone Plate and Screw

Indications for Use (Describe)

The metallic locking bone plate and screw are intents with age above 21 as indicated fixation of fractures, including ulna, radius, humerus, femur, tibia, clavicle, and fibula for the following indications:

  • ostcotomies, mal-unions, and non-unions;

  • single, segmental, and comminuted fractures;

  • normal bone density and osteopenic bone.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Traditional 510(k) Submission_Metalic bone plate system

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided: l. SUBMITTER

.

Canwell Medical Co., Ltd.

No. 466 South Xianhua Street, High-tech Industrail Zone, Jinhua, Zhejiang, 321016 P. R. of China

Primary Contact Person: Mike Gu

Regulatory Affairs Manager

OSMUNDA Medical Device Consulting Co., Ltd

Tel: (+86) 20-6232 1333

Fax: (+86) 20-8633 0253

Secondary Contact Person: Chen Hui

Regulatory Affairs Manager

Canwell Medical Co., Ltd.

Tel: +86-(0)579-8911-9235

Fax: +86-(0)579-8223-9558

September 30, 2014 Date Prepared:

DEVICE II.

Name of Device:Metallic Locking Bone Plate and Screw
Common/Usual Name:Bone Fixation Plate and Bone Fixation Screw
Classification Names:Plate, Fixation, Bone (21 CFR 888.3030)
Screw, Fixation, Bone (21 CFR 888.3040)
Regulation Class:II
Product Code:HRS and HWC

!!!. PREDICATE DEVICE

{5}------------------------------------------------

Traditional 510(k) Submission_Metalic bone plate system

Locking Bone Plates and Screws' predicate devices: K073159, K073432, K130108 and K130009;

This predicate has not been subject to a design-related recall.

No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The proposed device, metallic Locking Bone Plate and Screw consists of bone plate and screw. The bone plates are used for internal fixation of bone fracture, including ulna, radius, humerus, femur and tibia. Implant is manufactured with Ti6Al4V Eli and titanium which conforms to ASTM F 136 and ASTM F67.

The proposed devices are provided un-sterilized. They shall be sterilized prior to use by healthcare provider. The proposed devices shall never be reused.

V. INDICATIONS FOR USE

The metallic locking bone plate and screw are intended for adult patients with age above 21 as indicated fixation of fractures, including ulna, radius, humerus, femur, tibia, clavicle, and fibula for the following indications:

  • osteotomies, mal-unions, and non-unions
  • single, segmental, and comminuted fractures
  • normal bone density and osteopenic bone

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Metallic Locking Bone Plate and screw employ the same plate and screw technology as its predicate devices K073159, K073432, K130108 and K130009.

Metallic Locking Bone Plate and Screw is a structure formed by locking plate (which has at least one taper-thread hole) and locking screw (which has a head of taper-thread). Wherein the metal locking plate is a fracture fixation device with threaded holes, When the screw is revolved tightened in the plate, the plate system becomes an angle-fixed device. Locking plate can have both locking and non-locking screw holes fitting with various types of screws (also known as locking compression plate). In practice, positioning a plurality of screws in the locking plate maintains fracture anatomical reduction, which restores physiological function after bone healing.

2 / 3

{6}------------------------------------------------

Traditional 510(k) Submission Metalic bone plate system

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination:

  • . ASTM F382 - 99(2008) Standard Specification and Test Method for Metallic Bone Plates , bend testing;
  • . ASTM F543 - 13 Standard Specification and Test Methods for Metallic Medical Bone Screws metallic Bone Plates.

Biocompatibility testing:

Metallic Locking Bone Plate and Screw is made of pure titanium and Ti-6Al-4V, complying with the ASTM F1472 and ASTM F67; the materials have been employed successfully in human implant applications in contact with soft tissue and bone for over a decade. Due to the wellcharacterized level of biological response exhibited by this alloy, it has been used as a control material in Practice F 981

There is no Ultrahigh Molecular Weight Polyethylene (UHMWPE) Used in Orthopedic Devices

Animal and clinical study

The subject of this premarket submission, metallic Locking Bone Plate and Screw, does not require clinical studies to support substantial equivalence.

VIII. CONCLUSIONS

The non-clinical data support the safety of the device and the performance testing report demonstrate that the metallic Locking Bone Plate and Screw should perform as intended in the specified use conditions. Canwell Medical Co., Ltd. considers the metallic locking bone plate and screw do not raise any new issues of safety or effectiveness.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.