K Number
K130009

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-03-22

(79 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stryker VariAx 2 Compression Plating System is intended for internal fixation of bones in adult patients.

The Stryker VariAx 2 Compression Plating System is indicated for internal fixation of fractures in the Radius, Ulna, Humerus, Clavicle, Distal tibia and Fibula for the following indications:

  • osteotomies, mal-unions, and non-unions
  • single, segmental, and comminuted fractures
  • normal bone density and osteopenic bone
Device Description

The VariAx 2 Compression Plating System is an internal fixation device that consists of various plates used with compatible screws to fit different types of fractures in the radius, ulna, humerus, clavicle, distal tibia and fibula. The subject components will be available sterile and non-sterile. The plates will be available in sizes ranging from 30-246mm in length.

AI/ML Overview

The provided document is a 510(k) summary for the VariAx 2 Compression Plating System, which is a medical device for internal fixation of bone fractures. This type of submission focuses on demonstrating "substantial equivalence" to a predicate device already on the market, rather than conducting new clinical studies with detailed acceptance criteria and performance metrics for the new device as would be the case for a novel AI/software medical device.

Therefore, many of the specific questions regarding acceptance criteria, study design, expert involvement, and ground truth, as typically applied to AI/ML device evaluations, are not directly applicable or available in this document.

Here's an attempt to address your questions based on the provided text, highlighting where information is not present or not relevant to this type of device submission:

Acceptance Criteria and Device Performance (Based on provided text)

CriterionReported Device Performance
Substantial Equivalence to Predicate DeviceDemonstrated through non-clinical laboratory testing
Dynamic Fatigue Testing PerformanceNot explicitly detailed, but part of non-clinical testing to demonstrate substantial equivalence.
Intended UseInternal fixation of bones in adult patients.
Indications for UseInternal fixation of fractures in Radius, Ulna, Humerus, Clavicle, Distal tibia, and Fibula for osteotomies, mal-unions, non-unions, single, segmental, and comminuted fractures, and in normal and osteopenic bone.

Study Details (Based on provided text)

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable. The submission relies on non-clinical laboratory testing (dynamic fatigue testing) to demonstrate substantial equivalence, not a clinical test set with patient data.
    • Data Provenance: Not applicable for a clinical test set. The non-clinical testing was performed for the VariAx 2 Compression Plating System components.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device underwent non-clinical testing, not a clinical study involving expert-established ground truth on a test set.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No clinical test set.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a traditional medical device (bone plating system), not an AI/ML powered device. Therefore, no MRMC study or AI-related effectiveness was assessed.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is a hardware medical device; there is no algorithm or software for standalone performance evaluation in the context of AI.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable in the sense of clinical ground truth for AI/ML devices. The "ground truth" for this submission would be engineering standards and requirements for mechanical performance and materials, verified through non-clinical testing.
  7. The sample size for the training set:

    • Not applicable. There is no training set for this traditional medical device.
  8. How the ground truth for the training set was established:

    • Not applicable. There is no training set.

{0}------------------------------------------------

K130009 Page 1/2 ·

510(k) Summary

MAR 2 2 2013

Proprietary Name: VariAx 2 Compression Plating System Common Name: Bone plates Bone Screws Classification Name and Reference: Single/multiple component metallic bone fixation appliances and accessories 21 CFR §888.3030 Smooth or threaded metallic bone fixation fastener 21 CFR §888.3040 Regulatory Class: Class II 87 HRS: Plate, Fixation, Bone Product Codes: 87 HWC : Screw, Fixation, Bone Stryker Trauma AG Sponsor: Bohnackerweg 1 CH-2545 Selzach Switzerland Contact Person: Estela Celi Regulatory Affairs Specialist

Date Prepared:

December 29, 2012

estela.celi@stryker.com

325 Corporate Drive Mahwah, NJ 07430

Phone: (201) 831-6461 Fax: (201) 831-3461

Description

This Traditional 510(k) submission is being supplied to the U.S. FDA to provide authorization to market the new VariAx 2 Compression Plating System. The VariAx 2 Compression Plating System is an internal fixation device that consists of various plates used with compatible screws to fit different types of fractures in the radius, ulna, humerus, clavicle, distal tibia and fibula. The subject components will be available sterile and non-sterile. The plates will be available in sizes ranging from 30-246mm in length.

{1}------------------------------------------------

Intended Use

The Stryker VariAx 2 Compression Plating System is intended for internal fixation of bones in adult patients.

Indications

The Stryker VariAx 2 Compression Plating System is indicated for internal fixation of fractures in the Radius, Ulna, Humerus, Clavicle, Distal tibia and Fibula for the following indications:

  • osteotomies, mal-unions, and non-unions t
  • single, segmental, and comminuted fractures .
  • . normal bone density and osteopenic bone

Summary of Technologies

Device comparison showed that the proposed device is substantially equivalent in intended use, materials and performance characteristics to the following predicate device:

  • . K000684 Small Fragment Dynamic Compression Locking (DCL) System
  • K082807- SYNTHES (USA) 3.5 AND 4.5 MM LOCKING COMPRESSION PLATE SYSTEM WITH EXPANDED INDICATIONS

Non-Clinical Testing

Non-clinical laboratory testing was performed for the VariAx 2 Compression Plating System components to determine substantial equivalence. Testing demonstrated that the VariAx 2 Compression Plating System is substantially equivalent to the predicate device currently cleared for marketing.

The following testing was performed

Dynamic Fatigue Testing .

Clinical Testing

Clinical testing was not required for this submission.

Conclusion

The VariAx 2 Compression Plating System is substantially equivalent to the predicate device identified in this premarket notification.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 22, 2013

Stryker Trauma AG % Ms. Estela Celi Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430

Re: K130009

Trade/Device Name: VariAx 2 Compression Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 18, 2013 Received: January 24, 2013

Dear Ms. Celi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Estela Celi

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Erin DKeith

Mark N. Melkerson -Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): _K130009

Device Name: VariAx 2 Compression Plating System

Indications for Use:

The Stryker VariAx 2 Compression Plating System is indicated for internal fixation of fractures in the Radius, Ulna, Humerus, Clavicle, Distal tibia and Fibula for the following indications:

  • osteotomies, mal-unions, and non-unions .
  • single, segmental, and comminuted fractures .
  • normal bone density and osteopenic bone · ●

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabet 标签Frank -S

Division of Orthopedic Devices

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.