Intended Use

ProGrip™ Self-Gripping Polyester Mesh is indicated for inguinal and incisional hernias repair.
Parietex™ Plug and Patch System is indicated for the reinforcement of soft tissues during repair of groin hernia defects by open approach.
ProGrip™ Laparoscopic Self-Fixating Mesh is indicated for the reinforcement of soft tissues during repair of inguinal hernia defects by laparoscopic approach.

Device Description

ProGrip™ Self-Gripping Polyester Mesh: The mesh and the overlapping flaps of the pre-cut versions are made of knitted monofilament polyester and have polylactic acid monofilament resorbable pins on one of the sides. These pins facilitate placing, positioning and fixation of the overlapping flap and the mesh to the surrounding tissue. A colored yarn marker on the medial edge of the pre-cut mesh helps orientation. The monofilament polylactic acid pins are bioresorbable and contribute to the fixation of the mesh to surrounding tissue during at least 8 weeks. The polylactic acid pins degrade and resorb in vivo by hydrolysis and are metabolized by the body into CO and H2O.
Parietex™ Plug and Patch System: The Parietex™ Plug and Patch System is a kit composed of: a pre-cut patch made from monofilament polyester; a semi-resorbable disk which is a bi-component made of polyester monofilament and polylactic acid.
ProGrip™ Laparoscopic Self-Fixating Mesh: The mesh is made of knitted monofilament polyester with monofilament polylactic acid resorbable grips on one side and a resorbable film made of collagen from porcine origin and glycerol, on the other side. The grips allow positioning and fixation of the mesh to the surrounding tissue, while the collagen film facilitates mesh handling and deployment. The mesh presents a green band (polyester dyed with D&C green no. 6) to facilitate their orientation. The monofilament polylactic acid grips are bioresorbable and provide the fixation of the mesh to surrounding tissue for at least 8 weeks. The polylactic acid grips degrade and resorb in vivo by hydrolysis and are metabolized by the body into CO2 and H2O.

AI/ML Overview

This medical device submission (K142900) is for a change in the raw material formulation of several surgical mesh products. The purpose is to demonstrate that the new formulation of polyester yarn does not change the substantial equivalence of the modified devices to their existing, legally marketed predicate devices. Therefore, the "acceptance criteria" and "device performance" are not presented in terms of clinical accuracy metrics, but rather in terms of demonstrating equivalent performance characteristics through bench testing and biocompatibility.

Here's an analysis based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Demonstrated Equivalence)Reported Device Performance (Subject Device)
Indication for UseNo changes in indications for use compared to predicate devices.No changes to the stated indications for use for ProGrip™ Self-Gripping Polyester Mesh, Parietex™ Plug and Patch System, and ProGrip™ Laparoscopic Self-Fixating Mesh. (Pages 7, 3, 4, 5)
Raw MaterialsThe new polyester formulation maintains material properties comparable to the existing polyester, such that overall device performance is maintained. (Implied)The new polyester formulation maintains technological characteristics equivalent to the existing polyester (same yarn supplier). (Page 8)
Performance CharacteristicsDemonstrates equivalent mechanical, physical, and functional performance to predicate devices, as assessed by relevant bench tests.Bench testing conducted in accordance with "FDA's Guidance for the Preparation of a Premarket Notification Application for Surgical Mesh issued March 2, 1999" (Page 8). This testing demonstrated equivalent performance characteristics.
BiocompatibilityDemonstrates biocompatibility for a permanent implant as per ISO 10993-1.Biocompatibility studies conducted on the proposed polyester-based devices in accordance with ISO 10993-1 (a recognized FDA standard). (Page 8)
Stability/Shelf LifeMaintains equivalent stability and shelf life compared to predicate devices.Stability studies conducted, and proposed devices' shelf life has been demonstrated. (Page 8)
DesignNo changes in device design that would impact safety or effectiveness.No changes to the design of the devices. (Page 8 - implied by statement "Summary comparing the technological characteristics... in terms of ... Design")

2. Sample Size Used for the Test Set and the Data Provenance

This submission is for a material change in a surgical mesh, not a diagnostic AI device. Therefore, there is no "test set" in the traditional sense of patient data or clinical images. The "test sets" here refer to:

  • Bench Test Samples: The document doesn't specify the exact number of mesh samples or tests conducted for the bench testing. This information would typically be detailed in the full 510(k) submission, but not in the summary.
  • Biocompatibility Test Samples: Similarly, the document does not specify the number of animals or in-vitro tests performed for biocompatibility.
  • Data Provenance: The data provenance is from bench testing and preclinical studies conducted by the manufacturer (Sofradim Production/Covidien) to evaluate the new polyester formulation. These are not "retrospective or prospective" in the clinical trial sense, but rather laboratory-based evaluations. The country of origin for the manufacturing entity is France (Sofradim Production) and the contact person is in the US (Connecticut).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This question is not applicable to this type of 510(k) submission. There is no "ground truth" derived from expert review of patient data for an AI or diagnostic tool. The "ground truth" here is established by the scientific and engineering principles of material science, biocompatibility testing (toxicology and pathology experts for animal studies if applicable), and mechanical testing, following recognized standards (e.g., FDA guidance for surgical mesh, ISO 10993-1).

4. Adjudication Method for the Test Set

Not applicable. There is no human adjudication process involved in determining the "truth" for bench testing or biocompatibility studies of a material change. The results are objective measurements against established criteria/standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (especially those involving human interpretation of images) to measure human performance with and without AI assistance. This submission is for a surgical mesh material change, which does not involve human readers interpreting cases.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

No, a standalone performance study was not done. This concept applies to AI algorithms performing a diagnostic task independently. This submission is for a surgical mesh material change.

7. The Type of Ground Truth Used

The "ground truth" used is based on:

  • Objective physical and mechanical measurements from bench testing of the mesh material (e.g., tensile strength, pore size, burst strength – implied by "performance characteristics" and "FDA's Guidance for the Preparation of a Premarket Notification Application for Surgical Mesh").
  • Biocompatibility test results against established criteria for cytotoxicity, sensitization, irritation, acute systemic toxicity, subchronic toxicity, genotoxicity, implantation, and hemocompatibility, as defined by ISO 10993-1 for permanent implants.
  • Stability data demonstrating the integrity of the material over time.

These are not "expert consensus," "pathology" (in the clinical sense for a patient diagnosis), or "outcomes data" from clinical trials.

8. The Sample Size for the Training Set

Not applicable. There is no "training set" as this is not an AI/machine learning device. The material change is evaluated through direct testing against established scientific and regulatory standards.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for an AI model.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three faces overlapping to create a sense of depth and community.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 28, 2014

Sofradim Production % Ms. Mary Mellows Surgical Devices, a global business unit of Covidien 60 Middletown Avenue North Haven, Conneticut 06473

Re: K142900

Trade/Device Name: ProGrip™ Self-Gripping Polyester Mesh Parietex™ Plug and Patch System ProGrip™ Laparoscopic Self-Fixating Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: October 3, 2014 Received: October 6, 2014

Dear Ms. Mellows:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142900

Device Name Parietex TM Plug and Patch System

Indications for Use (Describe)

Parietex In Plug and Patch System is indicated for the reinforcement of soft tissues during repair of groin hernia defects by open approach

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K142900

Device Name

ProGrip™ Laparoscopic Self-Fixating Mesh

Indications for Use (Describe)

ProGrip™ Laparoscopic Self-Fixating Mesh is indicated for the reinforcement of soft tissues during repair of inguinal hernia defects by laparoscopic approach

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K142900

Device Name

ProGrip™ Self-Gripping Polyester Mesh

Indications for Use (Describe)

ProGrip™ Self-Gripping Polyester Mesh is indicated for inguinal and incisional hernias repair.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary

Submitter InformationName:Sofradim Production (subsidiary of Covidien LLC)
Address:116, avenue du formans01600 Trevoux, France
Phone number+33 (0)4 74 08 90 00
Fax number:+33 (0) 4 74 08 90 02
Establishment Registration:9615742
Name of contact person:Mary MellowsSenior Regulatory Specialist60 Middletown AvenueNorth Haven, CT 06473
Phone:203-492-5284
Date prepared:October 3, 2014
Name of device:
Trade or proprietary name:ProGrip™ Self-Gripping Polyester MeshParietex™ Plug and Patch SystemProGrip™ Laparoscopic Self-Fixating Mesh
Common or usual name:Surgical Mesh
Classification name:Mesh, Surgical, Polymeric
Classification panel:General and Plastic Surgery (79)
Regulation:21 CFR 878.3300
Product Code:FTI

{6}------------------------------------------------

Legally marketed devices to which equivalence is claimed:

ParietexTM Progrip™ Mesh (K103682) ParietexTM Pluq and Patch (K101519) Progrip™ Laparoscopic Self-Fixating Mesh (K123479)

Reason for 510(k) Submission:

The purpose of this 510(k) is to notify the Agency of the addition of another formulation of raw material, polyester, from the yarn supplier who manufactures the monofilament yarns used in the proposed devices. Eventually the current polvester will no longer be available by the supplier.

Device description:

ProGrip™ Self-Gripping Polyester Mesh

The mesh and the overlapping flaps of the pre-cut versions are made of knitted monofilament polyester and have polylactic acid monofilament resorbable pins on one of the sides. These pins facilitate placing, positioning and fixation of the overlapping flap and the mesh to the surrounding tissue. A colored yarn marker on the medial edge of the pre-cut mesh helps orientation.

The monofilament polylactic acid pins are bioresorbable and contribute to the fixation of the mesh to surrounding tissue during at least 8 weeks. The polylactic acid pins degrade and resorb in vivo by hydrolysis and are metabolized by the body into CO and H2O.

Parietex™ Plug and Patch System

The Parietex™ Plug and Patch System is a kit composed of:

  • . a pre-cut patch made from monofilament polyester
  • . a semi-resorbable disk which is a bi-component made of polyester monofilament and polylactic acid.

ProGrip™ Laparoscopic Self-Fixating Mesh

The mesh is made of knitted monofilament polyester with monofilament polylactic acid resorbable grips on one side and a resorbable film made of collagen from porcine origin and glycerol, on the other side. The grips allow positioning and fixation of the mesh to the surrounding tissue, while the collagen film facilitates mesh handling and deployment. The mesh presents a green band (polyester dyed with D&C green no. 6) to facilitate their orientation.

The monofilament polylactic acid grips are bioresorbable and provide the fixation of the mesh to surrounding tissue for at least 8 weeks. The polylactic acid grips degrade and resorb in vivo by hydrolysis and are metabolized by the body into CO2 and H2O.

Intended use of the device:

ProGrip™ Self-Gripping Polyester Mesh

ProGrip™ Self-Gripping Polyester Mesh is intended for the reinforcement of tissue during surqical repair.

No changes to the intended use have been made in this submission.

{7}------------------------------------------------

Parietex™ Pluq and Patch System

Parietex™ Pluq and Patch System is intended for the reinforcement of soft tissues during surqical repair.

No changes to the intended use have been made in this submission.

ProGrip™ Laparoscopic Self-Fixating Mesh

ProGrip™ Laparoscopic Self-Fixating Mesh is intended for the reinforcement of tissue during surqical repair.

No changes to the intended use have been made in this submission.

Indications for use:

ProGrip™ Self-Gripping Polvester Mesh

ProGrip™ Self-Gripping Polyester Mesh is indicated for inguinal and incisional hernias repair.

No changes to the indication for use have been made in this submission.

Parietex™ Plug and Patch System

Parietex™ Plug and Patch System is indicated for the reinforcement of soft tissues during repair of groin hernia defects by open approach.

No changes to the indication for use have been made in this submission.

ProGrip™ Laparoscopic Self-Fixating Mesh

ProGrip™ Laparoscopic Self-Fixating Mesh is indicated for the reinforcement of soft tissues during repair of inguinal hernia defects by laparoscopic approach. No changes to the indication for use have been made in this submission.

Summary comparing the technoloqical characteristics of the subject and predicate devices:

The proposed ProGrip™ Self-Gripping Polyester Mesh, Parietex™ Plug and Patch System and ProGrip™ Laparoscopic Self-Fixating Mesh manufactured with another formulation of raw material, polyester, from the same yarn supplier as the current polyester are equivalent to predicate Parietex™ Progrip™ Mesh (K103682), Parietex™ Plug and Patch (K101519) and Progrip™ Laparoscopic Self-Fixating Mesh (K123479) in terms of the following technological characteristics:

  • -Indication
  • Raw materials -
  • -Performance characteristics
  • -Biocompatibility
  • Stability -
  • -Design

Performance data: This change consists of the addition of another formulation of raw material, polyester, for the subject devices.

{8}------------------------------------------------

Bench testing has been conducted in accordance with FDA's Guidance for the Preparation of a Premarket Notification Application for Surgical Mesh issued March 2, 1999 to evaluate the performance characteristics of proposed devices.

Stability Studies have been conducted and the proposed devices shelf life has been demonstrated.

Biocompatibility studies have been conducted on the proposed polyester based devices in accordance with ISO 10993-1 for a permanent implant, a recognized standard by FDA (#2-156).

  • Conclusion: Bench testing and preclinical tests results demonstrate that proposed devices are substantially equivalent to the predicates Parietex™ Progrip™ Mesh (K103682), Parietex™ Plug and Patch (K101519) and Progrip™ Laparoscopic Self-Fixating Mesh (K123479).

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.