K Number
K142827

Validate with FDA (Live)

Date Cleared
2015-07-20

(293 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ViSi Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire), respiration rate, noninvasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, basic arrhythmia analysis (Ventricular Tachycardia, Ventricular Fibrillation, Asystole, Atrial Flutter) and alarm in hospital-based facilities; including general medical-surgical floors, intermediate care floors, and emergency departments.

Continuous non-invasive blood pressure (cNIBP) measurements have not been evaluated on patients during ambulation.

The basic arrhythmia analysis feature is intended for use on patients 18 years of age and older. It has not been evaluated on pediativ patients or neonates.

The arrhythmia analysis feature is intended for use by healthcation and treatification and treatment of arrhythmia events. Automated arrhythmia analysis is an adjunct to clinician review of the analysis should precede any therapeutic intervention.

The ViSi Mobile Monitoring System may be used as standalone devices or networked to a ViSi Mobile Remote Viewer through wireless 802.11 communication.

Device Description

The ViSi Mobile Monitoring System is a lightweight, body-worn vital signs monitor featuring a high resolution, full color touch screen display, with visual and audible alarms and alerts. The ViSi Mobile Monitor is designed to continuously non-invasively measure ECG, basic arrhythmias [including ventricular tachycardia, ventricular fibrillation, asystole, and atrial fibrillation/atrial flutter], heart rate, SpO2, blood pressure, pulse rate, respiration rate, and temperature. The ECG, Sp02, and Respiration waveforms are viewable on demand. The ViSi Mobile Monitoring System is capable of one-time and continuous NIBP measurements.

AI/ML Overview

The ViSi Mobile Monitoring System for basic arrhythmia analysis (Ventricular Tachycardia, Ventricular Fibrillation, Asystole, Atrial Fibrillation/Atrial Flutter) includes the following acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (ANSI/AAMI EC57: 2012)Reported Device Performance (ViSi Mobile Monitoring System)
QRS Detection(Not explicitly defined in the provided text, but implied as being comparable to predicate)Comparable to predicate device
Arrhythmia AnalysisBased on ANSI/AAMI EC57: 2012Validated by comparison to AHA, MIT-BIH, CU, and NST databases
False AlarmsMinimized by artifact and unclassified rhythm detectionArtifact detection prevents noisy signals from being detected as beats; Unclassified rhythm for unclassifiable rhythms (>30s triggers technical alarm)

Note: The document specifies that the basic arrhythmia analysis is based on the same technology as the currently cleared Monebo Technologies, Inc Arrhythmia library (K062282), which was adapted for real-time analysis. It also states the device has been validated by comparison to specific databases as prescribed by ANSI/AAMI EC57: 2012.

2. Sample Size Used for the Test Set and Data Provenance

The test set included data from the following databases:

  • AHA (American Heart Association)
  • MIT-BIH Arrhythmia Database
  • CU (Creighton University Arrhythmia Database)
  • NST (Northrop-Swales Transient Database)

The document does not detail the specific sample size (number of recordings or patients) used from each database, nor does it explicitly state the country of origin or whether the data was retrospective or prospective. However, these are established, publicly available retrospective databases commonly used for arrhythmia algorithm validation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The ground truth for the AHA, MIT-BIH, CU, and NST databases was established by expert consensus during the creation of these databases. The specific number and qualifications of experts involved in the original annotation of these databases are not provided in this document but are well-documented within the respective database literature.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method for the test set beyond relying on the established ground truth of the referenced databases (AHA, MIT-BIH, CU, NST). These databases typically feature annotations derived from multiple expert reviews.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is mentioned in the provided text, nor is there any effect size of human readers improving with AI vs without AI assistance. The study described focuses on standalone algorithm performance against established databases.

6. Standalone Performance Study

Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The ViSi Mobile Monitoring System's basic arrhythmia analysis was validated by comparison to the AHA, MIT-BIH, CU, and NST databases as prescribed in ANSI/AAMI EC57: 2012. The QRS detection sensitivity of the ViSi System was reported as comparable to the predicate device.

7. Type of Ground Truth Used

The type of ground truth used was expert consensus based on the established annotations within the AHA, MIT-BIH, CU, and NST databases.

8. Sample Size for the Training Set

The document does not provide specific details on the sample size used for the training set. It states that the basic arrhythmia analysis is based on the same technology as a previously cleared device (Monebo Technologies, Inc Arrhythmia library, K062282). This implies that a substantial training set was likely used during the development of the original algorithm, and possibly for adaptive training, but no specific numbers are given for the ViSi Mobile's training.

9. How the Ground Truth for the Training Set Was Established

Similarly, the document does not explicitly state how the ground truth for the training set was established. Given that the algorithm is based on a previously cleared library, it is highly probable that the training data and its ground truth were established through similar expert-annotated ECG datasets as those used for validation (e.g., subsets or similar types of data to the AHA, MIT-BIH databases).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a bird or abstract human figures, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 20, 2015

Sotera Wireless, Inc. Eben Gordon Regulatory Consultant 10020 Huennekens Street San Diego, California 92121

Re: K142827

Trade/Device Name: ViSi Mobile Monitoring System, ViSi Mobile Chest Sensor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including St-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DSL, DRT, DXN, DQA, FLL Dated: June 10, 2015 Received: June 11, 2015

Dear Eben Gordon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Arthel Stein

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K142827

Device Name ViSi Mobile Monitoring System

Indications for Use (Describe)

The ViSi Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire), respiration rate, noninvasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, basic arrhythmia analysis (Ventricular Tachycardia, Ventricular Fibrillation, Asystole, Atrial Flutter) and alarm in hospital-based facilities; including general medical-surgical floors, intermediate care floors, and emergency departments.

Continuous non-invasive blood pressure (cNIBP) measurements have not been evaluated on patients during ambulation.

The basic arrhythmia analysis feature is intended for use on patients 18 years of age and older. It has not been evaluated on pediativ patients or neonates.

The arrhythmia analysis feature is intended for use by healthcation and treatification and treatment of arrhythmia events. Automated arrhythmia analysis is an adjunct to clinician review of the analysis should precede any therapeutic intervention.

The ViSi Mobile Monitoring System may be used as standalone devices or networked to a ViSi Mobile Remote Viewer through wireless 802.11 communication.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a logo for Sotera Wireless. The logo features the word "Sotera" in a bold, black font. Above the word "Sotera" is a blue heart with three curved lines emanating from it. Below the word "Sotera" is the word "WIRELESS" in a smaller, blue font.

510(K) SUMMARY

Date preparedJuly 7, 2015
NameSotera Wireless, Inc.10020 Huennekens StreetSan Diego, CA 92121T. 858.427.4620; F. 858.999.2487
Trade nameViSi Mobile Monitoring System
Common nameVital signs monitor
Regulation NameArrhythmia detector and alarm (including ST-segment measurement and alarm)
Classification number21 CFR 870.1025
Product codeMHX, DSI, DRT, DXN, DQA, FLL
Regulatory classII
Predicate devicesAutomated ECG Analysis and Interpretation Analysis Software Library, K062282(Clearance: 03/22/2007)Acuity Central Monitoring Station, K052160 (Clearance: 12/16/2005)ViSi Mobile Monitoring System; K122036 (Clearance: 8/15/2012)ViSi Mobile Monitoring System; K130709 (Clearance: 10/7/2013)
DescriptionThe ViSi Mobile Monitoring System is a lightweight, body-worn vital signsmonitor featuring a high resolution, full color touch screen display, with visualand audible alarms and alerts. The ViSi Mobile Monitor is designed tocontinuously non-invasively measure ECG, basic arrhythmias [includingventricular tachycardia, ventricular fibrillation, asystole, and atrialfibrillation/atrial flutter], heart rate, SpO2, blood pressure, pulse rate, respirationrate, and temperature. The ECG, Sp02, and Respiration waveforms are viewableon demand. The ViSi Mobile Monitoring System is capable of one-time andcontinuous NIBP measurements.
510(k) NumberK142827
Indications for useThe ViSi Mobile Monitoring System is intended for use by clinicians andmedically qualified personnel for single or multi-parameter vital signs monitoringof adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire),respiration rate, heart rate, noninvasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygensaturation of arterial hemoglobin (SpO2), pulse rate, skin temperature, basicarrhythmia analysis (Ventricular Tachycardia, Ventricular Fibrillation, Asystole,Atrial Fibrillation/Atrial Flutter) and alarm in hospital-based facilities; includinggeneral medical surgical floors, intermediate care floors, and emergency

departments.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Sotera Wireless. The logo consists of the word "Sotera" in a bold, black font, with the word "WIRELESS" in a smaller, light blue font underneath. Above the word "Sotera" is a light blue graphic that resembles a heart with radio waves emanating from it.

Continuous non-invasive blood pressure (cNIBP) measurements have not been evaluated on patients during ambulation.

The basic arrhythmia analysis feature is intended for use on patients 18 years of age and older. It has not been evaluated on pediatric patients or neonates.

The arrhythmia analysis feature is intended for use by healthcare professionals trained in the identification and treatment of arrhythmia events. Automated arrhythmia analysis is an adjunct to clinical assessment; clinician review of the analysis should precede any therapeutic intervention.

The ViSi Mobile Monitoring System may be used as standalone devices or networked to a ViSi Mobile Remote Viewer through wireless 802.11 communication.

Summary of The basic arrhythmia analysis is based on the same technology as the currently substantial equivalence cleared Monebo Technologies, Inc Arrhythmia library (K062282), which has been adapted for real time analysis in the ViSi System. The ViSi Monitoring System's basic arrhythmia analysis has been validated by comparison to the AHA, MIT-BIH, CU, and NST databases as prescribed in ANSI/AAMI EC57: 2012. The QRS detection sensitivity of the ViSi System was comparable to the predicate device.

The safety and effectiveness of the design elements implemented into the ViSi System have been confirmed by compliance to Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm (October 28, 2003).

The ViSi System's basic arrhythmia analysis and alarms incorporates two additional rhythm classifications, artifact and unclassified, to minimize false alarms. Artifact detection prevents noisy signals from being detected as beats, thus excluding those signals from rhythm detection. In this way, false positive ventricular fibrillation and asystole events are minimized. If the artifact condition persists for >30 seconds, an equipment technical alarm is triggered. If the algorithm cannot classify the current rhythm, the rhythm is labeled as unclassified. If the condition persists for >30 seconds, an equipment technical alarm is triggered. These differences do not raise new questions of either safety or effectiveness.

Non-Clinical Testing

The ViSi System has successfully undergone database and functional testing to demonstrate equivalence to the predicate devices. Verification and validation testing was identified to support the safety and effective of the ViSi System and were relied upon in the determination of substantial equivalence. These nonclinical tests included:

  • . Performance against databases for arrhythmia analysis as specified in ANSI/AAMI EC57
  • Electrical safety testing per AAMI/ANSI ES60601-1:2005 ●
  • . Electromagnetic compatibility testing per IEC 60601-1-2 3rd Edition including Radiated Emissions, Electrostatic Discharge, Radio Frequency

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Sotera Wireless. The logo features the word "Sotera" in a bold, sans-serif font, with a light blue line underneath. Above the word "Sotera" is a light blue heart-shaped icon with three curved lines emanating from it, resembling radio waves. Below the line and the word "Sotera" is the word "WIRELESS" in a smaller, sans-serif font.

Electromagnetic Field Amplitude Modulated, Radio Frequency Common Mode, and Power Frequency Magnetic Field

  • Alarm system testing per AAMI/ANSI/IEC 60601-1-8:2006
  • . ECG performance according to AAMI/ANSI/IEC 60601-2-27:2011

The ViSi System has been tested and found to comply with the Special Controls Guidance, meet ANSI/AAMI EC57 performance requirements, and comply with recognized consensus standards for medical devices providing adequate support of substantial equivalence to the predicates devices. The results of all the testing demonstrate that the ViSi System is as safe, as effective, and performs as well as or better than the predicate device.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.