K Number
K130709

Validate with FDA (Live)

Date Cleared
2013-10-07

(206 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ViSi Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult (18 years or older) patients. It is indicated for ECG (3 or 5 leadwire), respiration rate, non-invasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and skin temperature in hospital-based facilities; including general medical-surgical floors, intermediate care floors, and emergency departments.

Device Description

The ViSi Mobile Monitoring System is a lightweight, body-worn vital signs monitor featuring a high resolution, full color touch screen display, with visual and audible alarms and alerts. The ViSi Mobile Monitor is designed to continuously non-invasively measure ECG, heart rate, SpO2, blood pressure, pulse rate, respiration rate, and temperature. The ECG, SpO2, and Respiration waveforms are viewable on demand. The ViSi Mobile Monitoring System is capable of one-time and continuous NIBP measurements.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the ViSi Mobile Monitoring System:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (from ISO 81060-2 & AAMI/ANSI SP10)Reported Device Performance (ViSi System)
Intermittent NIBP MAP (Mean Error)≤ ± 5 mmHgMeets criteria
Intermittent NIBP MAP (Standard Deviation)≤ 8 mmHgMeets criteria
cNIBP (Mean Error)≤ ± 5 mmHg (as per ISO 81060-2)Meets criteria
cNIBP (Standard Deviation)≤ 8 mmHg (as per ISO 81060-2)Meets criteria

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the test set. It mentions "clinical performance validation," but provides no numbers.

The data provenance is not specified regarding country of origin. The studies were clinical performance validations, which implies prospective data collection, but this is not explicitly stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

The document does not mention the use of experts to establish ground truth for the test set. Instead, it states that the ViSi's MAP measurement was "validated by comparison to an intra-arterial reference." This indicates that invasive, direct arterial blood pressure measurements were used as the gold standard for comparison.

4. Adjudication Method for the Test Set

No adjudication method is mentioned as the ground truth was established by direct intra-arterial measurements rather than expert review.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No MRMC comparative effectiveness study was done. This device is a vital signs monitor and does not involve "human readers" interpreting images or data that AI would assist with in the typical sense of MRMC studies.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the studies conducted were standalone performance evaluations of the device's ability to accurately measure blood pressure. The performance criteria (mean error and standard deviation) are algorithm-only metrics comparing the device's output to a reference standard.

7. The Type of Ground Truth Used

The ground truth used for both intermittent NIBP MAP and cNIBP was intra-arterial reference measurements. This is considered a gold standard for blood pressure measurement.

8. The Sample Size for the Training Set

The document does not explicitly state the sample size for a training set. This is because the device's performance validation is based on clinical testing against a reference standard, not a machine learning model that requires a distinct training and testing set. The "auto-calibration to NIBP measurements" mentioned for cNIBP implies a form of continuous adaptation, but doesn't define a traditional "training set" in the context of supervised learning.

9. How the Ground Truth for the Training Set Was Established

Since a traditional "training set" in the machine learning sense is not applicable or explicitly mentioned for this device's validation, the method for establishing its ground truth is also not provided. Any internal calibration or learning processes would likely use intra-arterial or NIBP reference measurements as their "truth."

{0}------------------------------------------------

WIRELESS

5 510(K) SUMMARY

Date preparedJune 11, 2013
NameSotera Wireless, Inc.9444 Waples Street, Suite 280San Diego, CA 92121T. 858.373.4841; F. 858.427.4639OCT 07 2013
Contact personEben GordonSenior Director, Regulatory
Trade nameViSi Mobile Monitoring System
Common nameVital signs monitor
Regulation NameCardiac Monitor Including Cardiotachometer and Rate Alarm
Classification number21 CFR 870.2300
Product codeMWI, DRT, DXN, DQA, FLL
Regulatory classII
Predicate devicesViSi Mobile Monitoring System; K112478 (Clearance: 3/27/2012)CNAP Monitor 500i, 500at; K082599 (Clearance: 10/17/2008)
DescriptionThe ViSi Mobile Monitoring System is a lightweight, body-worn vital signsmonitor featuring a high resolution, full color touch screen display, with visualand audible alarms and alerts. The ViSi Mobile Monitor is designed tocontinuously non-invasively measure ECG, heart rate, SpO2, blood pressure,pulse rate, respiration rate, and temperature. The ECG, SpO2, and Respirationwaveforms are viewable on demand. The ViSi Mobile Monitoring System iscapable of one-time and continuous NIBP measurements.
Indications for useThe ViSi Mobile Monitoring System is intended for use by clinicians andmedically qualified personnel for single or multi-parameter vital signs monitoringof adult (18 years or older) patients. It is indicated for ECG (3 or 5 leadwire),respiration rate, heart rate, non-invasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygensaturation of arterial hemoglobin (SpO2), pulse rate, and skin temperature inhospital-based facilities; including general medical-surgical floors, intermediatecare floors, and emergency departments.
Continuous non-invasive blood pressure (cNIBP) testing has not been conductedon ambulatory patients.
Summary ofsubstantial equivalenceThe ViSi System has the following similarities and differences with the indicatedpredicate devices:

{1}------------------------------------------------

  • . Intended use - The ViSi System has the same vital sign monitoring intended use as the ViSi System and CNAP Monitors. The cNIBP measurement of the ViSi System is the same as the CNAP Monitor.
  • . Device Performance - The modified ViSi System with intermittent NIBP MAP and cNIBP measurements have been clinically validated to meet the same performance requirements (mean error ≤ ± 5 mmHg; standard deviation ≤ 8 mmHg) as outlined in the consensus standards for automated non-invasive blood pressure monitors.
  • . Technology - The ViSi System is unchanged from it predicate submission except for the display of intermittent NIBP MAP readings and the continuous non-invasive blood pressure (cNIBP) measurements.

The intermittent NIBP MAP measurement is based on the same technology as is currently in the cleared ViSi System and is equivalent to the CNAP Monitor. The ViSi's MAP measurement has been validated by comparison to an intra-arterial reference and meets the performance requirements of ISO 81060-2.

The ViSi's cNIBP measurement has a different principle of operation from that of the CNAP Monitor predicate device. Both methods are similar in that pressure is determined on a beat-to -beat basis and are joint solutions employing automatic calibration to NIBP measurements.

The question of effectiveness of the cNIBP measurement has been addressed by the clinical performance validation to the requirements of ISO 81060-2, an equivalent consensus standard to which the CNAP Monitor meets (i.e. AAMVANSI SP10).

The safety and effectiveness of the design elements implemented into the ViSi System have been confirmed by their compliance to the prevailing standards. The ViSi System has demonstrated compliance with the following consensus standards: IEC 60601-1-8, IEC 80601-2-30, IEC 62304, ISO 81060-2, and IEC 62366.

General Safety and Effectiveness Concerns - The instructions for use for the ViSi System contains the necessary cautions and warnings to provide for safe and effective use of the device.

The ViSi System has successfully undergone functional testing to demonstrate equivalence to the predicate devices. The following quality assurance measures were applied to the device: Risk analysis, Requirements review, Code inspections, Verification and validation, Bench testing, and Clinical performance testing.

The ViSi System have been tested and found to comply with recognized performance standards for medical devices. The results of all the testing demonstrate that the ViSi System is safe, effective, complies with the appropriate medical device standards, and is substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

October 7, 2013

Sotera Wireless, Inc. Attn: Mr. Eben Gordon Sr. Director, Regulatory Affairs 9444 Waples St., Suite 280 San Diego, CA 92121

Re: K130709

Trade/Device Name: Visi Mobile Monitoring System, Model 92-10010 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (Two) Product Code: MWI, DRT, DXN, DQA and FLL Dated: September 30, 2013 Received: October 1, 2013

Dear Mr. Gordon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 – Mr. Eben Gordon

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.ida.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.

Sincerely yours,

Owen P. Faris -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

4 INDICATIONS FOR USE

510(k) Number (if known): K130709_

ViSi Mobile Monitoring System __ Device Name:

Indications for Use:

The ViSi Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult ( 18 years or older) patients. It is indicated for ECG (3 or 5 leadwire), respiration rate, non-invasive blood pressure (NIBP), continuous noninvasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and skin temperature in hospital-based facilities; including general medical-surgical floors, intermediate care floors, and emergency departments.

Continuous non-invasive blood pressure (cNIBP) testing has not been conducted on ambulatory patients.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over the Counter Use __

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally signed by
Owen P. Faris -S
Date: 2013.10.07
15:13:05 -04'00'

Page _ 1 _ of _ 1_

Page 12

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).