K Number
K142817

Validate with FDA (Live)

Date Cleared
2015-04-23

(206 days)

Product Code
Regulation Number
870.1200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IntraNovo 25 Microcatheter is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremities, and all coronary vessels. The Microcatheter is also intended for drug infusion in intra-arterial therapy and infusion of embolic materials for hemostasis. The Microcatheter should not be used in cerebral vessels.

Device Description

The IntraNovo 25 Microcatheter (the "IntraNovo 25" or the "Microcatheter") is an infusion catheter intended for intravascular use. The Microcatheter consists of a single lumen. The shaft consists of a lubricous inner liner made from polytetraflouroethylene (PTFE) with a stainless steel coil over the inner liner. The outer liner consists of different lengths of colored polyether block amide with varying durometers hardness. A lubricious hydrophilic coating covers the distal end of the IntraNovo 25 Microcatheter. Depending on the model, one or two radiopaque marker bands are placed at the distal end of the Microcatheter and above the stainless steel coil. A polycarbonate hub is attached to the proximal end of the Microcatheter with a strain relief placed over the distal end of the hub. The device shaft has an outer diameter of 2.9 French size on the proximal end and 2.7 French size on the distal end with an inner diameter of 0.025" throughout the shaft. The device is available in lengths of 110 cm, 130 cm, and 150 cm. The following accessories are included with the Microcatheter: a stainless steel steam shaping mandrel to allow for manual shaping of the distal tip and a 3.0 mL syringe. The device is supplied sterile and is intended for single use.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called the "IntraNovo 25 Microcatheter." It outlines the device's characteristics, intended use, and a comparison to a predicate device, along with a summary of testing performed to demonstrate substantial equivalence.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in a separate section for each test, but rather implies that individual tests passed based on internal test methods or referenced ISO/ASTM standards. The "Test Results" column for each test simply states "Pass."

Here's a table summarizing the tests and reported performance:

Test #Test NameApplicable Standard or Internal Test MethodReported Device Performance (Acceptance Criteria)
1Guidewire & Guide Catheter CompatibilityInternal Test MethodT=0 Pass, T=2 Pass (Implies compatibility was demonstrated)
2In-Vitro Track ForceInternal Test MethodT=0 Pass, T=2 Pass (Implies adequate track force)
3Durability of Hydrophilic CoatingInternal Test MethodT=0 Pass, T=2 Pass (Implies coating remained durable)
4Lubricity of Hydrophilic CoatingInternal Test MethodT=0 Pass, T=2 Pass (Implies coating retained lubricity)
5Tip Shape RetentionInternal Test MethodT=0 Pass, T=2 Pass (Implies tip shape was retained)
6Static Burst PressureISO 10555T=0 Pass, T=2 Pass (Implies device withstood static burst pressure as per standard)
7Dimensional & Physical AttributesISO 10555T=0 Pass, T=2 Pass (Implies dimensions and physical characteristics met standard)
8Corrosion ResistanceISO 10555T=0 Pass, T=2 Pass (Implies resistance to corrosion as per standard)
9Dynamic Burst PressureInternal Test MethodT=0 Pass, T=2 Pass (Implies device withstood dynamic burst pressure)
10Air LeakInternal Test MethodT=0 Pass, T=2 Pass (Implies no air leakage)
11Liquid LeakageInternal Test MethodT=0 Pass, T=2 Pass (Implies no liquid leakage)
12Tensile StrengthISO 10555T=0 Pass, T=2 Pass (Implies adequate tensile strength as per standard)
13Flow RateInternal Test MethodT=0 Pass, T=2 Pass (Implies satisfactory flow rate)
14Kink ResistanceInternal Test MethodT=0 Pass, T=2 Pass (Implies good kink resistance)
15RadiopacityASTM-F640-12T=0 Pass, T=2 Pass (Implies radiopacity met standard)
16Torque to FailureInternal Test MethodT=0 Pass, T=2 Pass (Implies adequate torque resistance)
17Catheter StiffnessInternal Test MethodT=0 Pass, T=2 Pass (Implies appropriate catheter stiffness)
18Packaging IntegrityASTM F-88-09, ASTM-1929-98T=0 Pass, T=2 Pass (Implies packaging integrity as per standards)
19Therapeutic AgentsInternal Test MethodT=0 Pass, T=2 Pass (Implies compatibility with therapeutic agents)
20Female Luer Hub VerificationISO 594T=0 Pass, T=2 Pass (Implies luer hub met standard)
21Shipping and Transportation SimulationISTA 3PAT=0 Pass, T=2 Pass (Implies device integrity after shipping simulation)
22Coating IntegrityInternal Test MethodT=0 Pass, T=2 Pass (Implies coating integrity maintained)
23Torque ResponseInternal Test MethodT=0 Pass, T=2 Pass (Implies satisfactory torque response)
24Particulate EvaluationUSP <788>T=0 Pass, T=2 Pass (Implies particulate levels met USP standards)
Biocompatibility Tests (e.g., Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Pyrogenicity, Hemolysis, Coagulation, Complement Activation, In vivo Thrombogenicity)ISO 10993, ASTM, USP methodsPass (Implies meeting biocompatibility requirements)
Sterilization Process Validation (Ethylene Oxide Gas)-10^-6 (Pass) (Implies sterilization achieved required SAL)

Important Note: The document specifies that "The IntraNovo 25 Microcatheter met the predetermined acceptance criteria." However, it does not explicitly list these criteria for each test (e.g., "burst pressure must be > X psi"). The "Pass" indicates that these internal criteria were met.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify the exact sample size (number of devices) used for each bench test. It mentions that testing was performed on "non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2)," suggesting a comparison between fresh and aged devices. However, the number of devices in each group is not provided.
  • Data Provenance: The data is generated from bench testing performed by Texas Medical Technologies, Inc. (the manufacturer). This is not human clinical data, so concepts like "country of origin of the data" and "retrospective or prospective" do not directly apply in the same way they would for clinical studies. It's pre-market, in-vitro/bench data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

N/A. This is bench testing of a physical device, not a diagnostic or prognostic algorithm that requires ground truth establishment by human experts reading medical images or clinical data. The "ground truth" for these tests is defined by the physical or chemical properties measured against a standard or an internal specification.

4. Adjudication Method for the Test Set

N/A. As this is bench testing, there is no adjudication by human experts in the sense of reviewing interpretations or diagnoses. The "adjudication" is inherent in whether the device's physical performance (e.g., burst pressure, flow rate, and biocompatibility) quantitatively meets the predetermined acceptance criteria for each test.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

N/A. This document is for a microcatheter, a physical medical device, not an AI software algorithm or diagnostic tool that would typically undergo an MRMC study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

N/A. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the performance evaluation of the IntraNovo 25 Microcatheter in this context is established by:

  • Bench Test Specifications: Predetermined performance targets and limits derived from engineering specifications, industry best practices, and regulatory standards (ISO, ASTM, USP).
  • Referenced Standards: Adherence to established international and national standards (e.g., ISO 10555 for catheters, ISO 10993 for biocompatibility, ASTM for materials, USP for particulates). The acceptance criteria for many of these tests are defined within these standards.

8. The Sample Size for the Training Set

N/A. This is a physical medical device, not a machine learning model, so there is no "training set."

9. How the Ground Truth for the Training Set was Established

N/A. This is a physical medical device, not a machine learning model, so there is no "training set" or ground truth establishment for such a set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The image is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 23, 2015

Texas Medical Technologies, Inc. % Viviana Gonzalez Consultant Pomamed Consulting LLC Po Box 9818 San Juan, PR 00908-0818

Re: K142817

Trade/Device Name: IntraNovo 25 Microcatheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DOO Dated: March 9, 2015 Received: March 18, 2015

Dear Viviana Gonzalez,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman -S

Bram D. Zuckerman, ivi.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142817

Device Name

IntraNovo 25 Microcatheter

Indications for Use (Describe)

The IntraNovo 25 Microcatheter is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremities, and all coronary vessels. The Microcatheter is also intended for drug infusion in intra-arterial therapy and infusion of embolic materials for hemostasis. The Microcatheter should not be used in cerebral vessels.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1 – Company Information & Contact Person

Company Name:Texas Medical Technologies Inc.
Company Address:9005 Montana Ave. Ste. AEl Paso, Texas 79925
Telephone:(915) 774-4321
Fax:(915) 774-4323
Contact Person:Cesar Rios, Quality Assurance & Regulatory Manager
Date Prepared:4/22/2015

2 - Device Name & Classification

Proprietary Name:IntraNovo 25
Common Name:Microcatheter
Classification Name:Diagnostic Intravascular Catheter
Regulation Number:21 CFR 870.1200
Product Code:DQO
Device Class:II

3 - Predicate Device

Legally Marketed Substantially Equivalent Predicate Device

Proprietary Name:PROGREAT
Company Name:Terumo Medical Corporation
Common Name:Catheter
Classification Name:Diagnostic Intravascular Catheter
Regulation Number:21 CFR 870.1200
Product Code:DQO
Device Class:II
510(k) Number:K033583

4 - Device Description

The IntraNovo 25 Microcatheter (the "IntraNovo 25" or the "Microcatheter") is an infusion catheter intended for intravascular use. The Microcatheter consists of a single lumen. The shaft consists of a lubricous inner liner made from polytetraflouroethylene (PTFE) with a stainless steel coil over the inner liner. The outer liner consists of different lengths of colored polyether block amide with varying durometers hardness. A lubricious hydrophilic coating covers the distal end of the IntraNovo 25 Microcatheter. Depending on the model, one or two radiopaque marker bands are placed at the distal end of the Microcatheter and

{4}------------------------------------------------

above the stainless steel coil. A polycarbonate hub is attached to the proximal end of the Microcatheter with a strain relief placed over the distal end of the hub.

The device shaft has an outer diameter of 2.9 French size on the proximal end and 2.7 French size on the distal end with an inner diameter of 0.025" throughout the shaft. The device is available in lengths of 110 cm, 130 cm, and 150 cm.

The following accessories are included with the Microcatheter: a stainless steel steam shaping mandrel to allow for manual shaping of the distal tip and a 3.0 mL svringe.

The device is supplied sterile and is intended for single use.

The following table lists the models and sizes available for the IntraNovo 25 Microcatheter.

Design &DevelopmentModel NumberCommercialModel NumberShaftLength(cm)Outer DiameterInnerDiameterMarker BandConfigurationTip Shape
SMC-25110-SMC-2711-2SN1102.7 Fr.2.9 Fr.0.025"2 Marker BandsStraight(shapeable)
SMC-25130-SMC-2713-2SN1302.7 Fr.2.9 Fr.0.025"2 Marker BandsStraight(shapeable)
SMC-25150-SMC-2715-2SN1502.7 Fr.2.9 Fr.0.025"2 Marker BandsStraight(shapeable)
SMC-25110-1SMC-2711-1SN1102.7 Fr.2.9 Fr.0.025"1 Marker BandStraight(shapeable)
SMC-25130-1SMC-2713-1SN1302.7 Fr.2.9 Fr.0.025"1 Marker BandStraight(shapeable)
SMC-25150-1SMC-2715-1SN1502.7 Fr.2.9 Fr.0.025"1 Marker BandStraight(shapeable)

Table 4.1. IntraNovo 25 Microcatheter Models and Sizes

5 - Indications for Use

The IntraNovo 25 Microcatheter is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremities, and all coronary vessels. The Microcatheter is also intended for drug infusion in intra-arterial therapy and infusion of embolic materials for hemostasis. The Microcatheter should not be used in cerebral vessels.

  • 6 Summary of Technological Characteristics Comparison
    Based on a comparison of the indications for use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging, it is determined that the IntraNovo 25 Microcatheter is substantially equivalent to the predicate device. Table 6.1 below provides a comparison of the IntraNovo 25 Microcatheter and the predicate.

{5}------------------------------------------------

Technical Characteristics /Principle of OperationIntraNovo 25MicrocatheterPROGREAT Catheter(Predicate)SubstantiallyEquivalent?
Length100cm - 150cm100cm - 150cmYes
Distal: 2.7 Fr.Proximal 2.9 Fr.Distal: 2.7 Fr.Proximal: 2.9 Fr.Yes
Outer DiameterYes
Inner Diameter0.025"0.025"Yes
Maximum Pressure5,171 kPa (750 psi)5,171 kPa (750 psi)Yes
Distal CurveStraight (shapeable)Straight (shapeable)Yes
Inner Liner MaterialPolytetraflouroethylene(PTFE)*Polytetraflouroethylene(PTFE)Yes
Coil Reinforcement MaterialStainless Steel*TungstenYes
Radiopaque MarkerPlatinum/Iridium*Tungsten coil and/orPlatinum marker bandYes
Outer Shaft MaterialPolyether Block Amide*UnknownYes
Luer MaterialPolycarbonate*UnknownYes
Luer ConnectorFemale Luer Connector*Female Luer ConnectorYes
Hydrophilic CoatedYes*YesYes
Anatomical Site UsePeripheral, CoronaryPeripheral, CoronaryYes
Supplied AccessoriesShaping Mandrel, InjectionSyringeShaping Mandrel,Injection SyringeYes
Delivery to SiteOver-the-wireOver-the-wireYes
Guidewire CompatibilityMaximum 0.021"Maximum 0.021"Yes
PackagingPolyethylene Hoop andTyvek PouchPolyethylene Hoop andTyvek PouchYes
SterilizationEtO GasEtO GasYes

Table 6.1 Comparison of the IntraNovo 25 Microcatheter and the Predicate Device.

  • Denotes a patient-contacting material.

7 - Testing Summary

The following bench tests were performed to evaluate the design elements and performance characteristics of the IntraNovo 25 Microcatheter and to demonstrate substantial equivalence to the predicate device. The IntraNovo 25 Microcatheter met the predetermined acceptance criteria. Testing was performed on non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2). Tests results show that the IntraNovo 25 Microcatheter is substantially equivalent to the predicate device.

{6}------------------------------------------------

7.1- Bench Testing Table

Table 7.1 below provides a summary of the bench testing performed on the IntraNovo 25 Microcatheter.

Test #Test NameApplicable Standard or Internal Test MethodTest Results (T=0) and (T=2)
1Guidewire & Guide Catheter CompatibilityInternal Test MethodT=0 PassT=2 Pass
2In-Vitro Track ForceInternal Test MethodT=0 PassT=2 Pass
3Durability of Hydrophilic CoatingInternal Test MethodT=0 PassT=2 Pass
4Lubricity of Hydrophilic CoatingInternal Test MethodT=0 PassT=2 Pass
5Tip Shape RetentionInternal Test MethodT=0 PassT=2 Pass
6Static Burst PressureISO 10555T=0 PassT=2 Pass
7Dimensional & Physical AttributesISO 10555T=0 PassT=2 Pass
8Corrosion ResistanceISO 10555T=0 PassT=2 Pass
9Dynamic Burst PressureInternal Test MethodT=0 PassT=2 Pass
10Air LeakInternal Test MethodT=0 PassT=2 Pass
11Liquid LeakageInternal Test MethodT=0 PassT=2 Pass
12Tensile StrengthISO 10555T=0 PassT=2 Pass
13Flow RateInternal Test MethodT=0 PassT=2 Pass
14Kink ResistanceInternal Test MethodT=0 PassT=2 Pass
15RadiopacityASTM-F640-12T=0 PassT=2 Pass
16Torque to FailureInternal Test MethodT=0 PassT=2 Pass
17Catheter StiffnessInternal Test MethodT=0 PassT=2 Pass
18Packaging IntegrityASTM F-88-09ASTM-1929-98T=0 PassT=2 Pass
19Therapeutic AgentsInternal Test MethodT=0 PassT=2 Pass
20Female Luer Hub VerificationISO 594T=0 PassT=2 Pass
21Shipping and Transportation SimulationISTA 3PAT=0 PassT=2 Pass
22Coating IntegrityInternal Test MethodT=0 PassT=2 Pass
23Torque ResponseInternal Test MethodT=0 PassT=2 Pass
24Particulate EvaluationUSP <788>T=0 PassT=2 Pass

Table 7.1. Bench Testing Performed on the IntraNovo 25 Microcatheter.

{7}------------------------------------------------

7.2 – Biocompatibility

The IntraNovo 25 Microcatheter is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (≤24 hours). Biocompatibility testing was performed in accordance with ISO 10993 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process" (2009). Table 7.2 below describes the testing performed to determine biocompatibility.

Test NameTest DescriptionTest Results
Cytotoxicity - TestCytotoxicity - MEM Elution Test (ISO 10993-5:2009)Pass
Cytotoxicity - TestCytotoxicity-MTT Quantitative Evaluation (ISO 10993-5:2009)Pass
Sensitization - TestMaximization Test for Delayed-Type Hypersensitivity Test (ISO 10993-10:2010)Pass
Irritation - TestIntracutaneous (Intradermal) Reactivity Test (ISO 10993-10:2010)Pass
Systemic Toxicity - TestAcute Systemic Toxicity Test (ISO 10993-11:2006)Pass
Pyrogenicity TestPyrogen Test in Rabbits (ISO10993-11/USP 37 <151>, rev. 5/2014)Pass
Hemolysis - TestHemolysis, Direct and Extraction Method (ASTM 756-08)Pass
CoagulationProthrombin Time AssayPass
Complement Activation - TestComplement ActivationPass
In vivo Thrombogenicity - TestThromboresistance in DogsPass

Table 7.2. Summary of Biocompatibility Testing for the IntraNovo 25 Microcatheter

8 - Sterilization Testing Summary

Validation Sterilization ProcessSterility Assurance Level (SAL)Validation Result
Ethylene Oxide Gas10-6Pass

9 - Conclusion

The IntraNovo 25 Microcatheter is substantially equivalent in intended use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging to the predicate device. Differences between the devices do not raise any new issues of safety or effectiveness.

§ 870.1200 Diagnostic intravascular catheter.

(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).