K Number
K142740

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-04-06

(194 days)

Product Code
Regulation Number
892.1650
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ziehm Vision RFD 3D system is intended for use in providing both 2D and 3D medical imaging for all adult and pediatric populations, using pulsed and continuous fluoroscopic imaging.

The device provides 2D medical imaging for fluoroscopy, digital subtraction, and acquisition of cine loops during diagnostic interventional and surgical procedures where intra-operative imaging and visualization of complex anatomical structures of both lower and higher contrast density are required. Such procedures may include but are not limited to those of interventional cardiology, heart surgery, hybrid procedures, interventional radiology, interventional angiography, electrophysiology, pediatrics, endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery, neurosurgery, critical care, emergency room procedures, and those procedures visualizing structures of the cervical, thoracic, and lumbar regions of the spine and joint fractures of the upper and lower extremities, and where digital image data is required for Computer-Assisted Surgery procedures.

The device is also intended to provide 3D medical imaging of patients during orthopedic, neurological, intra-operative surgical procedures and where the clinician benefits from 3D visualization of complex anatomical structures, such as but not limited to those of high contrast objects, bones, joints, maxillofacial, cervical, thoracic, and lumbar regions of the spine, pelvis, acetabulum and joint fractures of the upper and lower extremities, and where digital image and C-arm positioning data is required for Computer-Assisted Surgery procedures.

The visualization of such anatomical structures assists the clinical outcome. At the discretion of a physician, the device may be used for other imaging applications. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.

Device Description

The ZIEHM VISION RFD 3D employs X-rays as its imaging technology for visualizing human anatomy in both 2D and 3D imaqing. The X-ray tube in the generator produces X-rays, quided toward the patient under control of the user at the direction of a physician who determines the specific clinical procedure. The images from the system assist the physicians in visualizing the patient's anatomy. This visualization helps to localize regions of pathology and for surgical procedures. The device provides both real-time image capture and post capture visualization and of in vivo surgical procedures and post-surgical outcomes.

The Ziehm Vision RFD 3D mobile fluoroscopy system is a flat panel detector (FD) Computed tomography x-ray system and fluoroscopic X-ray imaging system consisting of two mobile units: a Mobile Stand (C-Arm) and a Monitor Cart/Workstation. The Mobile Stand is comprised of a mono-block high voltage generator, X-ray control, and a C-Profile which is "C" shaped and supports the X-ray generator, and the image receptor Flat Panel Detector (FD).

The device performs both 2D medical imaging and the specialized 4 axes of motorized movement necessary for the 3D imaging. This provides the user/operator the option to use manual or motorized linear and rotational movements of the C- Profile for positioning of the imaging components at various angles and distances with respect to the patient using a control interface, Vision Center, Remote Vision Center or remote Position Control Center.

The motorization of the 4 axes provides the user an alternative for visualizing anatomical structures using a variable iso-centric location. The system working with a variable iso-center allows freely selectable positions of patient anatomy. The variable isocenter and distance control ensures that anatomical structures are safely visualized from different angles without re-adjusting the Carm or moving the patient. The iso-center is not restricted to orbital movements and can hold this iso-center during angulations and vertical travel using the 4 motorized axes. This same motion control provides the bases for 3D views of the patient anatomy. These 3D views are generated by means of an iterative algorithm.

The system uses the images of a scan captured with relation to a predefined scan center to compute the three-dimensional representation of an object. The 3D views are always displayed on the reference screen of the monitor cart. It is possible to display multiplanar reconstructions, orthogonal or freely selectable sections, and different surface reconstructions.

The Distance Control surface detection integrated around the lower edge of the flat panel detects objects, such as patients. When the flat panel approaches an object, the device reduces speed, slowing the motorized movement. The movement stops immediately before entering a defined safety zone.

The mobile stand supports the optional wireless footswitch for optimum positioning for the surgeon by removing the cable on the floor.

The Monitor Cart is a mobile platform that connects to the Mobile Stand by a cable, and which integrates the LCD flat panel display monitors, 2D image processing, Optional 3D image processing, user controls and image recording devices. Interfaces provided for optional peripheral devices such as external monitors, thermal video printers, wireless video display, wireless video server, injector connection and image storage devices (USB, DVD) and DICOM fixed wired and wireless network interfaces.

AI/ML Overview

The provided document is a 510(k) summary for the Ziehm Vision RFD 3D. While it details the device's indications for use, technological characteristics, and comparison to predicate devices, it does not contain information regarding a specific clinical study with acceptance criteria and reported device performance related to diagnostic accuracy or clinical outcomes.

The "Summary of Non-Clinical Test Data" section primarily focuses on:

  • Compliance with internal functional specifications, design controls, and engineering standards.
  • Verification and validation testing.
  • Non-clinical image comparisons with predicate devices (Ziehm Vision RFD K132904 and Ziehm Vision2 FD Vario 3D K073346).
  • Software verification/validation testing aligning with FDA guidance.
  • Compliance with federal performance standards for X-Ray Fluoroscopic equipment (21 CFR 1020.30-32) and relevant voluntary safety standards (IEC standards).

Therefore, I cannot provide the specific information requested in the format of the questions, as the document primarily addresses substantial equivalence based on technological characteristics and compliance with safety and performance standards for an imaging device, rather than a clinical study evaluating diagnostic performance against predefined acceptance criteria for a specific clinical task.

Specifically, the document does not include:

  1. A table of acceptance criteria and reported device performance from a clinical study.
  2. Sample size used for a test set or data provenance for a clinical study.
  3. Number of experts or their qualifications for establishing ground truth in a clinical study.
  4. Adjudication method for a clinical study.
  5. Information about a multi-reader multi-case (MRMC) comparative effectiveness study, or effect sizes for AI assistance.
  6. Results of a standalone algorithm performance without human-in-the-loop.
  7. The type of ground truth used for a clinical study.
  8. Sample size for a training set.
  9. How ground truth for a training set was established.

The document's purpose is to demonstrate substantial equivalence to predicate devices based on technical specifications and non-clinical testing, not to present the results of a clinical performance study with specific acceptance criteria for diagnostic accuracy.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized representation of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and is positioned around the upper half of the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 6, 2015

Ziehm Imaging GmbH % Mr. Richard Westrich Director of Regulatory Affairs and Quality Assurance Ziehm Imaging, Inc. 6280 Hazeltine National Drive ORLANDO FL 32822

Re: K142740

Trade/Device Name: Ziehm Vision RFD 3D Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, OXO, JAA, JAK Dated: March 16, 2015 Received: March 16, 2015

Dear Mr. Westrich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular graphic on the left and the text "ziehmimaging" on the right. The graphic is a stylized representation of a person inside of a circle with lines around it. The text is in a teal color.

Indications for Use

510(k) Number (if known): K142740

ZIEHM VISION RFD 3D Device Name:

Indications for Use:

The Ziehm Vision RFD 3D system is intended for use in providing both 2D and 3D medical imaging for all adult and pediatric populations, using pulsed and continuous fluoroscopic imaging.

The device provides 2D medical imaging for fluoroscopy, digital subtraction, and acquisition of cine loops during diagnostic interventional and surgical procedures where intra-operative imaging and visualization of complex anatomical structures of both lower and higher contrast density are required. Such procedures may include but are not limited to those of interventional cardiology, heart surgery, hybrid procedures, interventional radiology, interventional angiography, electrophysiology, pediatrics, endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery, neurosurgery, critical care, emergency room procedures, and those procedures visualizing structures of the cervical, thoracic, and lumbar regions of the spine and joint fractures of the upper and lower extremities, and where digital image data is required for Computer-Assisted Surgery procedures.

The device is also intended to provide 3D medical imaging of patients during orthopedic, neurological, intra-operative surgical procedures and where the clinician benefits from 3D visualization of complex anatomical structures, such as but not limited to those of high contrast objects, bones, joints, maxillofacial, cervical, thoracic, and lumbar regions of the spine, pelvis, acetabulum and joint fractures of the upper and lower extremities, and where digital image and C-arm positioning data is required for Computer-Assisted Surgery procedures.

The visualization of such anatomical structures assists the clinical outcome. At the discretion of a physician, the device may be used for other imaging applications. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Ziehm Imaging. The logo consists of a circular emblem on the left and the text "ziehmimaging" on the right. The emblem features a stylized human figure within a globe-like structure. The text is in a teal color and appears to be a sans-serif font.

U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center -W066-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002

510 (k) Summary

March 31, 2015

In accordance with 21 CFR §807.92 the following 510(k) summary information is provided:

Submitter Address:Ziehm Imaging GmbHDonaustrasse 3190451 NurembergGermanyPhone: + 49.911.2172-219Fax: +49.911.2172-390
Primary ContactPerson /Agent:Richard L. WestrichDirector of Regulatory Affairs and QualityAssuranceZiehm Imaging, Inc.Phone: (407) 615-8560 Ext 140Fax: (470) 615-8561
Secondary ContactPerson:Stefan FiedlerDirector QM/RAZiehm Imaging GmbHPhone: + 49.911. 2172-219Fax: +49.911.2172-390
Device (Trade Name):Ziehm Vision RFD 3D
Common /UsualNames:Mobile Fluoroscopic C-Arm
Classification:21CFR 892.1650

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular icon with a stylized human figure inside, surrounded by orbital lines. To the right of the icon is the text "ziehmimaging" in a gradient teal color, with the color becoming more saturated from left to right.

RegulationDescription:Image-intensified fluoroscopic x-ray system
Device:Interventional fluoroscopic x-ray system
Product Code:OWB
SubsequentClassification(s)21CFR 892.1650
RegulationDescription:Image-intensified fluoroscopic x-ray system
Device:Image-intensified fluoroscopic x-ray system,mobile
Product Code:OXO
SubsequentClassification(s)21CFR 892.1650
RegulationDescription:Image-intensified fluoroscopic x-ray system
Device:System, x-ray, fluoroscopic, image intensified
Classification ProductCode:JAA
Classification:21CFR 892.1750
RegulationDescription:Computed tomography x-ray system.
Device:System, X-ray, tomography, computed
Classification ProductCode:JAK
Predicate Device:Decision Date:K132904 Ziehm Vision RFD12/05/2013
Classification ProductCode:JAA
Classification:21CFR 892.1650
Subsequent ProductCodes:OWB, OXO

{5}------------------------------------------------

Predicate Device:K073346 Ziehm Vision2 FD Vario 3D
Decision Date:03/12/2008
Classification Product Code:JAA
Classification:21CFR 892.1650
Subsequent Product Codes:OWB, JAK

The ZIEHM VISION RFD 3D employs X-rays as its imaging General Description: technology for visualizing human anatomy in both 2D and 3D imaqing. The X-ray tube in the generator produces X-rays, quided toward the patient under control of the user at the direction of a physician who determines the specific clinical procedure. The images from the system assist the physicians in visualizing the patient's anatomy. This visualization helps to localize regions of pathology and for surgical procedures. The device provides both real-time image capture and post capture visualization and of in vivo surgical procedures and post-surgical outcomes.

The Ziehm Vision RFD 3D mobile fluoroscopy system is a flat panel detector (FD) Computed tomography x-ray system and fluoroscopic X-ray imaging system consisting of two mobile units: a Mobile Stand (C-Arm) and a Monitor Cart/Workstation. The Mobile Stand is comprised of a mono-block high voltage generator, X-ray control, and a C-Profile which is "C" shaped and supports the X-ray generator, and the image receptor Flat Panel Detector (FD).

The device performs both 2D medical imaging and the specialized 4 axes of motorized movement necessary for the 3D imaging. This provides the user/operator the option to use manual or motorized linear and rotational movements of the C- Profile for positioning of the imaging components at various angles and distances with respect to the patient using a control interface, Vision Center, Remote Vision Center or remote Position Control Center.

The motorization of the 4 axes provides the user an alternative for visualizing anatomical structures using a variable iso-centric location. The system working with a variable iso-center allows freely selectable positions of patient anatomy. The variable isocenter and distance control ensures that anatomical structures are safely visualized from different angles without re-adjusting the Carm or moving the patient. The iso-center is not restricted to orbital movements and can hold this iso-center during angulations and vertical travel using the 4 motorized axes. This same motion control provides the bases for 3D views of the patient anatomy. These 3D views are generated by means of an iterative algorithm.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular graphic with a stylized human figure inside, next to the text "ziehmimaging". The text is in a teal color and appears to be a sans-serif font.

The system uses the images of a scan captured with relation to a predefined scan center to compute the three-dimensional representation of an object. The 3D views are always displayed on the reference screen of the monitor cart. It is possible to display multiplanar reconstructions, orthogonal or freely selectable sections, and different surface reconstructions.

The Distance Control surface detection integrated around the lower edge of the flat panel detects objects, such as patients. When the flat panel approaches an object, the device reduces speed, slowing the motorized movement. The movement stops immediately before entering a defined safety zone.

The mobile stand supports the optional wireless footswitch for optimum positioning for the surgeon by removing the cable on the floor.

The Monitor Cart is a mobile platform that connects to the Mobile Stand by a cable, and which integrates the LCD flat panel display monitors, 2D image processing, Optional 3D image processing, user controls and image recording devices. Interfaces provided for optional peripheral devices such as external monitors, thermal video printers, wireless video display, wireless video server, injector connection and image storage devices (USB, DVD) and DICOM fixed wired and wireless network interfaces.

  • The Ziehm Vision RFD 3D system is intended for use in providing Indications for Use: both 2D and 3D medical imaging for all adult and pediatric populations, using pulsed and continuous fluoroscopic imaging.
    The device provides 2D medical imaging for fluoroscopy, digital subtraction, and acquisition of cine loops during diagnostic interventional and surgical procedures where intra-operative imaging and visualization of complex anatomical structures of both lower and higher contrast density are required. Such procedures may include but are not limited to those of interventional cardiology, heart surgery, hybrid procedures, interventional radiology, pediatrics, endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery, neurology, neurosurgery, critical care, emergency room procedures, and those procedures visualizing structures of the cervical, thoracic, and lumbar regions of the spine and joint fractures of the upper and lower extremities, and where digital image data is required for Computer-Assisted Surgery procedures.

The device is also intended to provide 3D medical imaging of patients during orthopedic, neurological, intra-operative surgical procedures and where the clinician benefits from 3D visualization

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for ziehmimaging. The logo consists of a circular graphic on the left and the company name "ziehmimaging" on the right. The text is in a teal color and appears to be a sans-serif font.

of complex anatomical structures, such as but not limited to those of high contrast objects, bones, joints, maxillofacial, cervical, thoracic, and lumbar regions of the spine, pelvis, acetabulum and joint fractures of the upper and lower extremities, and where digital image and C-arm positioning data is required for Computer-Assisted Surgery procedures.

The visualization of such anatomical structures assists the clinician in the clinical outcome. At the discretion of a physician, the device may be used for other imaging applications. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.

Summary of The Ziehm Vision RFD 3D is based on the Ziehm Vision RFD Technological K132904 including the major system control, x-ray generator, Characteristics: Pulsed fluoroscopy, image detector, 4 Axes motorization, software, with imaging workstation upgraded to integrate the 3D image capture and control functions from our Ziehm Vision2 FD Vario 3D K073346. The subject device employs the same fundamental scientific technologies, as the predicate devices for system controls, full DICOM compatibility, and intuitive graphical user interface. The technological comparison chart in Section 12, of this premarket submission reveals that the proposed device shares the technological characteristics performed by the predicate devices.

The following table compares the dominant performance data of the subject device with the predicate devices to substantiate equivalence of the subject device and predicate devices.

Features/TechnologySubject DeviceZiehm Vision RFD3D (K142740)Predicate DeviceZiehm Vision RFD(K132904)Predicate DeviceZiehm Vision2 FDVario 3D (K073346)
Mobile fluoroscopic c-armYesYesYes
Product CodesOWB, OXO, JAA,JAKOWB, JAA, OXOJAA, OWB, JAK
User interface touchcontrol panelYes, same designYes, same designYes, same design
X-ray generator and tubehousing assemblymonoblock technologyYes, samemonoblock designrotating anodeYes, samemonoblock designrotating anodemonoblock designFixed anode tube
KV Range 40-120 kVYesYes40-110 kVOpt 120 kV
Max power output25 kWOptional 20 kW or7.5 kW25 kWOptional 20 kW or7.5 kW2.02 kW
Pulsed fluoroscopy7.5 kW 1.5-75mA7.5 kW 1.5-75mA1.5-20mA

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular icon with a stylized human figure inside, surrounded by orbiting lines. To the right of the icon is the company name, "ziehmimaging", written in a teal-colored, sans-serif font.

20 kW 1.5-200mA25 kW 1.5-250mA20 kW 1.5-200mA25 kW 1.5-250mA
X-ray tubeRotating anode0.3, 0.6 focal spotRotating anode0.3, 0.6 focal spotStationary anode0.6 focal spot
Collimator/beam limiterasymmetrical shuttersYes, same designYes, same designYes, same design
Virtual collimationYes, same designYes, same designYes, same design
X-ray detector (FPD)Yes, same designYes, same designYes, same design
Amorphous silicon20 cm x 20 cm or30 cm x 30 cm20 cm x 20 cm or30 cm x 30 cm20 cm x 20 cm
FPD Matrix size20 cm x 20 cm1024x1024 pixelsor30 cm x 30 cm1536 x 1536 pixels20 cm x 20 cm1024x1024 pixelsor30 cm x 30 cm1536 x 1536 pixels20 cm x 20 cm1024x1024 pixels
FPD's same paneltechnology.Yes, same designYes, same designYes, same design
AERC Dose control systemYes, same designkV/mA curve typesUse of all threeYes, same designkV/mA curve typesUse of only twoYes, same designkV/mA curve typesUse of only two
Image data port fornavigation systemYes, same designExport of imagedata andparametersYes, same designExport of imagedata andparametersYes, same designExport of image dataand parameters
Manual and motorizedaxes of movementYes, both manualand motorized 4axes of movementYes, both manualand motorized 4axes of movementYes both manual andmotorized 3 axes ofmovement
Image Post Processing 2DYes, same designYes, same designYes, same design
3D Imaging WorkstationWith Post ProcessingYes, uses same 3DImage WorkstationNo 3D ImageCaptureYes, uses same 3DImage Workstation
3D ReconstructionAlgorithmYesNo 3D WorkstationYes
DICOM functionalityincluding RSDRYes, same designYes, same designYes, same design
Monitors Dual 19" TFTFlat Screen DisplaysYes, same designYes, same designYes, same design
Vision CenterGraphical userinterface touch panelYes, same designYes, same designYes, same design

Summary of Non-Clinical Test Data:

Ziehm Vision RFD 3D is based on the direct modifications to a cleared predicate device Ziehm Vision RFD (K132904); The Ziehm Vision RFD 3D design changes were completed in accordance with Ziehm Imaging GmbH Quality Management System Design Controls and Engineering, standards compliance, and Verification and Validation testing were successfully conducted. Tests were performed on the Ziehm Vision RFD 3D which demonstrated that the device is safe and effective, performs comparably to the

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular emblem on the left and the company name "ziehmimaging" on the right. The text is in a teal color and appears to be a sans-serif font.

predicate devices, and is substantially equivalent to the predicate devices. Tests included verification/validation testing to internal functional specifications (including software) and non-clinical image comparisons involving flat panel display images taken with the new device and the predicate devices Ziehm Vision RFD K132904 and Ziehm Vision2 FD Vario 3D K073346. Documentation provided demonstrates compliance of the modified device Ziehm Vision RFD 3D to all FDA requirements stated in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, including results of verification/validation tests to software requirements and software risk hazards. Performance testing confirmed that the Ziehm Vision RFD 3D complies with 21 CFR 1020.30-32 Federal Performance Standards for X-Ray Fluoroscopic equipment and with relevant voluntary safety standards for Electrical Safety and Electromagnetic Compatibility testing, specifically IEC standards listed in the table below. Together, the verification/validation activities successfully confirmed device requirements has been fulfilled, and that system functionality is consistent with the user needs, intended uses, and the Ziehm Vision RFD 3D device correctly performs as designed, and raises no new questions reqarding either safety or effectiveness. Therefore, when compared to the predicate devices the Ziehm Vision RFD 3D supports a determination of substantial equivalence to the predicate devices.

Compliance Standards:

FDA/CDRH From 3626(5/11)A Guide for the Submission of Initial Reports on Diagnostic X-RaySystems and Their Major Components.
21 CFR 1020.30-32Federal Performance Standard for Diagnostic X-ray Systems.
MDD 93/42/EECAnnex II of the Medical Devices Directive (MDD) 93/42/EEC
AAMI/ANSI ES60601-1Medical Electrical Equipment -- Part 1: General Requirements forBasic Safety and Essential Performance (IEC 60601-1:2005, mod),Date: 2010, Conformance Standard #19-5
IEC 60601-1:2005Medical Electrical Equipment, General Requirements for Safety.Edition 3.0, Date: 2005-12-15.No Recognized Conformance Standard but general basis ofAAMI/ANSI ES60601-1.
IEC 60601-1-2: 2007Medical Electrical Equipment, General Requirements for Safety,Electromagnetic Compatibility. Edition 3.0, Date: 2007-03-30,Conformance Standard #19-1

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for ziehmimaging. The logo consists of a circular graphic with a stylized human figure inside, surrounded by orbital lines. To the right of the graphic is the text "ziehmimaging" in a teal, sans-serif font. The text appears to be slightly gradient, with the left side being lighter than the right.

  • IEC 60601-1-3: 2008 Medical Electrical Equipment, Radiation Protection in Diagnostic Xray Equipment. Edition 2.0, Date: 2008-01-22, Conformance Standard #12-210
  • Medical electrical equipment, Part 2-43: Particular requirements IEC 60601-2-43: 2010 for basic safety and essential performance of X-ray equipment for interventional procedures. Edition 2.0, Date: 2010-03-25, Conformance Standard #12-202
  • Medical electrical equipment, Part 2-54: Particular requirements IEC 60601-2-54: 2009 for the basic safety and essential performance of X-ray equipment for radiography and radioscopy. Edition 1.0, Date: 2009-06-29, Conformance Standard #12-274
    • Safety of laser products, Equipment Safety, requirements, and IEC 60825-1: 2007 user quide Edition 2.0, Date: 2007-03-30 Conformance Standard #12-273
      • ISO 14971: 2007 Application of risk management to medical devices. Edition 2.0, Date: 2007-03-01, Conformance Standard #5-40
        • Ziehm Imaging GmbH considers the Ziehm Vision RFD 3D to be as Conclusion: safe, as effective, and performs substantially equivalent to the predicate device Ziehm Vision RFD (K132904) and Ziehm Vision2 FD Vario 3D (K073346) in accordance with its labeling.

End of 510(k) Summary

Richard L. Westrich Official Agent and Correspondent for Ziehm Imaging GmbH. Ziehm Imaging Inc, 6280 Hazeltine National Drive Orlando, FI. 32822

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.