K Number
K142611

Validate with FDA (Live)

Date Cleared
2014-12-31

(106 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TYRX™ Neuro Absorbable Antibacterial Envelope is intended to hold a vagus nerve stimulator or spinal cord neuromodulator securely in order to create a stable environment when implanted in the body.

TYRX™ Neuro Absorbable Antibacterial Envelope contains the antimicrobial agents rifampin and minocycline which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pulse generator. This device is intended to be used in conjunction with vagus nerve stimulators implanted in the infraclavicular fossa, or spinal cord neuromodulators implanted laterally to the body midline and slightly superior to the gluteal region.

TYRX™ Neuro Absorbable Antibacterial Envelope is intended for single patient, one-time use only.

Device Description

TYRX ™ Neuro Absorbable Antibacterial Envelope is a fully absorbable , dual component sterile device designed to hold a vagus nerve stimulator or spinal cord neuromodulator securely to create a stable environment when implanted in the body. It is constructed of knitted filaments of a commercially available absorbable polymer, Glycoprene II, comprised of glycolide, caprolactone and trimethylene carbonate polymer, and coated with an absorbable polyarylate polymer mixture containing the antimicrobial agents rifampin and minocycline. Rifampin and minocycline have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of an implantable electronic device is to be used in a healthcare facility/hospital by personnel experienced in the procedure of vagus nerve or spinal cord neuromodulator implantation.

AI/ML Overview

This document describes a 510(k) premarket notification for the TYRX™ Neuro Absorbable Antibacterial Envelope. It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets new acceptance criteria through a dedicated study with specific performance metrics.

Therefore, many of the requested elements for a study proving acceptance criteria cannot be extracted directly from this document. The document primarily details comparative information with predicate devices and safety testing.

Here's an analysis of the provided information based on your questions:

1. Table of Acceptance Criteria and Reported Device Performance

This document does not present a table of specific acceptance criteria (e.g., a certain percentage reduction in infection rate, a specific tensile strength at a given time point) that the new device was tested against in its own clinical study. Instead, it relies on demonstrating equivalence to predicate devices, which were previously cleared.

However, it does mention results from in vitro and in vivo studies that support the predicate devices' antimicrobial activity, which is a key performance aspect.

Performance CharacteristicAcceptance Criteria (Implied from Predicate)Reported Device Performance (Inherited/Demonstrated through Equivalence)
Antimicrobial ActivityEffective against common bacteria (MRSA, S. aureus, A. baumannii, S. epidermidis, S. lugdunensis, E. coli)Demonstrated by predicate devices (K132699 and K130943). The subject device uses the identical antimicrobial coating at the same concentrations.
BiocompatibilityBiocompatible, non-pyrogenic"Supplied sterile, biocompatible, and non-pyrogenic." Demonstrated for predicate devices (ISO 10993).
SterilitySterile"TYRX follows the ISO 11137 standard for sterility."
DegradationGlycoprene II mesh degrades into constituent monomers"Bench testing demonstrated that the Glycoprene II mesh degrades into its constituent monomers."
No Chemical InteractionNo chemical or physical interaction between components"Bench testing demonstrated... there is no chemical or physical interaction between the Glycoprene mesh, the polyarylate coating or the antibiotics."
Stable EnvironmentAbility to hold a neuromodulator securely and create a stable environmentIntended Use is identical to the predicate device, implying similar mechanical stability for holding the device.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated for any new clinical test set for the subject device. The document refers to "in vitro studies" and "in vivo efficacy testing" performed on the predicate devices. The sample sizes for these predicate studies are not provided in this document.
  • Data Provenance: The "in vivo efficacy testing" was "previously submitted with the primary predicate, K132699" and "previously submitted with predicate K130943." This implies the data is likely retrospective relative to this 510(k) submission, as it comes from prior submissions. The country of origin is not specified, but given the FDA submission, it would adhere to US regulatory standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided. The document highlights laboratory (in vitro), animal (in vivo), and bench testing, not studies requiring human expert adjudication of clinical outcomes in the same way a diagnostic imaging device might.

4. Adjudication Method for the Test Set

Not applicable for the types of tests described (in vitro, in vivo efficacy in animals, bench testing).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This device is a medical implant (envelope for neuromodulators) containing antibiotics, not an AI-powered diagnostic or assistive tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable, as this is not an algorithm. However, the in vitro and in vivo animal efficacy studies can be considered standalone performance assessments of the antimicrobial properties and safety profile.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

  • For antimicrobial activity (in vitro): Direct observation of bacterial inhibition/killing.
  • For in vivo efficacy (animal studies): Likely based on direct measurement of infection rates or bacterial counts at the surgical site in animal models, comparing treated groups to control groups. This would be considered "outcomes data" in an animal model.
  • For biocompatibility: Standardized toxicological and immunological endpoints as per ISO 10993.
  • For degradation and interaction: Bench testing and chemical analysis.

8. The Sample Size for the Training Set

Not applicable. This device does not involve machine learning or AI that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. This device does not involve machine learning or AI that requires a "training set."

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 31, 2014

Medtronic Tvrx. Inc. % Regina Novak Senior Regulatory Specialist 1 Deer Park Drive Suite G Monmouth Junction, New Jersey 08852

Re: K142611

Trade/Device Name: Tyrx Neuro Absorbable Antibacterial Envelope Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: November 25, 2014 Received: November 26, 2014

Dear Ms. Novak,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (and for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Felipe Aquel -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement

510(k) number (if known):K142611
------------------------------------

Device Name: TYRX™ Neuro Absorbable Antibacterial Envelope

TYRX™ Neuro Absorbable Antibacterial Envelope is intended to hold a vagus nerve stimulator or spinal cord neuromodulator securely in order to create a stable environment when implanted in the body.

TYRX™ Neuro Absorbable Antibacterial Envelope contains the antimicrobial agents rifampin and minocycline which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pulse generator. This device is intended to be used in conjunction with vagus nerve stimulators implanted in the infraclavicular fossa, or spinal cord neuromodulators implanted laterally to the body midline and slightly superior to the gluteal region.

TYRX™ Neuro Absorbable Antibacterial Envelope is intended for single patient, one-time use only.

Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

{3}------------------------------------------------

510(k) Summary

In accordance with 21 CFR 807.92, Medtronic TYRX, Inc. provides this summary of the safety and effectiveness information available for TYRX™ Neuro Absorbable Antibacterial Envelope , as well as the substantial equivalence decision making process used for the TYRX™ Neuro Absorbable Antibacterial Envelope subject device.

Sponsor/Applicant Name and Address:Medtronic TYRX Inc.1 Deer Park Drive Suite GMonmouth Junction, N.J. 08852
Establishment Registration Number:3005619263
Sponsor Contact Information:Regina NovakSenior Regulatory Affairs SpecialistPhone: 732-964-1492Fax: 732-246-8677Email:regina.novak@medtronic.comAlternate Contact:Susan Olinger, JDDirector, Regulatory AffairsPhone: 732-964-1110Fax: 763-246-8677Email:susan.olinger@medtronic.com
Date of preparation of 510(k) SummarySeptember 15, 2014
New Device Trade/Proprietary Name:TYRX TM Neuro AbsorbableAntibacterial Envelope
Device Common Name:Surgical Mesh
Regulatory Classification:Class II
PROCODE:FTL

{4}------------------------------------------------

Predicate Device Name and 510(k) Number

AIGISRx N Antibacterial Envelope (K132699) (Primary Predicate)

AIGISRx Resorbable Antibacterial Envelope (K130943)

Device Description:

TYRX ™ Neuro Absorbable Antibacterial Envelope is a fully absorbable , dual component sterile device designed to hold a vagus nerve stimulator or spinal cord neuromodulator securely to create a stable environment when implanted in the body. It is constructed of knitted filaments of a commercially available absorbable polymer, Glycoprene II, comprised of glycolide, caprolactone and trimethylene carbonate polymer, and coated with an absorbable polyarylate polymer mixture containing the antimicrobial agents rifampin and minocycline. Rifampin and minocycline have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of an implantable electronic device is to be used in a healthcare facility/hospital by personnel experienced in the procedure of vagus nerve or spinal cord neuromodulator implantation.

Indications for Use

TYRX TM Neuro Absorbable Antibacterial Envelope is intended to hold a vagus nerve stimulator or spinal cord neuromodulator securely in order to create a stable environment when implanted in the body.

TYRX ™ Neuro Absorbable Antibacterial Envelope contains the antimicrobial agents rifampin and minocycline which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pulse generator. This device is intended to be used in conjunction with vagus nerve stimulators implanted in the infraclavicular fossa, or spinal cord neuromodulators implanted laterally to the body midline and slightly superior to the gluteal region.

TYRX "1" Neuro Absorbable Antibacterial Envelope is intended for single-patient, one-time use only.

{5}------------------------------------------------

The Indications for Use statement for TYRX Neuro Absorbable Antibacterial Envelope is identical to the primary predicate. Both devices have the same intended use which is to create a stable environment for a vagus nerve stimulator or spinal cord neuromodulator when implanted.

Comparison of Technological Characteristics with the Predicate Device

TYRX TM Neuro Absorbable Antibacterial Envelope (subject device) is identical in Indication for Use to its predicate, AIGIS N (K132699). Both devices are intended to hold a vagus nerve stimulator or spinal cord neuromodulator securely in order to create a stable environment when implanted in the body. The difference is that TYRX TM Neuro Absorbable Antibacterial Envelope is fully absorbable containing the absorbable Glycoprene II substrate mesh, whereas AIGIS N is partially absorbable; containing the nonabsorbable polypropylene polymer substrate mesh. Both AIGIS N and TYRX Neuro Absorbable Antibacterial Envelope are coated with the identical absorbable polyarylate polymer coating containing the antibiotics rifampin and minocycline in concentrations of 102ug/cm2 .

TYRX ™ Neuro Absorbable Antibacterial Envelope has the same substrate material found in the FDA cleared AIGISRx R (K130943)(predicate for mesh substrate only). Both devices are fully absorbable. The fully absorbable substrate mesh is constructed of knitted filaments of a commercially available absorbable polymer. Glycoprene II, comprised of glycolide, caprolactone and trimethylene carbonate polymer. There is no difference in the manufacturing processes for the subject device. The difference is that the predicate device (K 132699) substrate mesh is composed of polypropylene and the TYRX "" Neuro Absorbable Antibacterial Envelope (subject device) has Glycoprene II as the substrate mesh.

Biocompatability Results

TYRX™ Neuro Absorbable Antibacterial Envelope is supplied sterile, biocompatible, and nonpyrogenic. TYRX follows the ISO 11137 standard for sterility. Bench testing demonstrated that the Glycoprene II mesh degrades into its constituent monomers, and there is no chemical or physical interaction between the Glycoprene mesh, the polyarylate coating or the antibiotics.

Biocompatability testing of the predicate devices in accordance with ISO Standard 10993 demonstrated the biocompatibility and safety of the subject device.

{6}------------------------------------------------

Animal Studies

In in vitro studies, predicates K132699 and K130943 demonstrated antimicrobial activity against Methicillin Resistant Staphylococcus aureus (MRSA), Staphylococcus aureus, Acinetobacter baumannii, Staphylococcus epidermidis; K130943 also demonstrated antimicrobial activity against Staphylococcus lugdunensis and Escherichia coli.

In vivo efficacy testing, previously submitted with the primary predicate, K132699, demonstrated effectiveness in reducing infections. The bacteria tested were Staphylococcus aureus, Acinetobacter baumannii, Staphylococcus epidermidis and Escherichia coli. In vivo efficacy testing, previously submitted with predicate K130943, demonstrated effectiveness in reducing infections with these additional bacteria, methicillin-resistant Staphylococcu aureus (MRSA), and Staphylococcus lugdunensis. It should be noted that the in vivo and in vitro activity of the TYRX "" Antibacterial Envelope antimicrobials is variable against nonepidermidis strains of coagulase-negative Staphylococci.

CONCLUSION:

In summary, the TYRX Neuro Absorbable Antibacterial Envelope is identical in terms of physical construct to the currently marketed AIGIS N (K132699) except that for the subject device, a bioabsorable, multi filament surgical mesh (Glycoprene II) will replace the monofilament polypropylene surgical mesh substrate in the non absorbable predicate device. The polymer coating containing the antibiotics rifampin and minocycline in concentrations of 102µg/cm2 applied to the substrate will remain unchanged. The device is sterile and nonpyrogenic.

Based on the 510(k) summaries and the information provided, we conclude that TYRX ™ Neuro Absorbable Antibacterial Envelope is safe and effective for its intended use, and is substantially equivalent to the predicate device AIGIS N (K132699) from the perspective of Indication for Use and substantially equivalent to the predicate device (AIGIS R, K130943) from the perspective of technological characteristics and shelf life (1 month).

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.