K Number
K142598

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-06-12

(270 days)

Product Code
Regulation Number
870.1330
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/ or other any interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease.

Device Description

The OptoWire and OptoMonitor System includes a pressure sensing guidewire (OptoWire), electronic signal processing and display units (OptoMonitor) that process signals received from the OptoWire to display intravascular blood pressure and fractional flow reserve (FFR) values, and various connection cables. The OptoWire and OptoWire cable are sterile, single-use devices; the remaining system components are reusable.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the OptoWire and OptoMonitor System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document details various performance tests (bench and animal) but does not explicitly list specific acceptance criteria values alongside the reported device performance in a single section. Instead, it states that "All internal design criteria were successfully met" and the device "compared favorably to predicate devices tested" for the OptoWire mechanical/functional verification.

However, a detailed comparison table with predicate devices (Volcano Corp. PrimeWire Prestige Plus Pressure Guide Wire and ComboMap Pressure and Flow System) does provide specific performance metrics for the OptoWire and OptoMonitor System that can be inferred as criteria for substantial equivalence.

Inferred Acceptance Criteria & Reported Device Performance (from "Substantial Equivalence Comparison" table):

Feature/ParameterAcceptance Criteria (Inferred from Predicate/Target)OptoWire and OptoMonitor System Performance (Reported)Met Acceptance Criteria?
OptoWire
Guidewire OD0.014" (Predicate: PrimeWire Prestige Plus)0.014"Yes
Guidewire LengthRanges (e.g., 185 cm, 300 cm for predicate)175 cmFavorable Comparison
Guidewire Shaft MaterialStainless Steel; Nitinol (Predicate: PrimeWire)Stainless Steel; NitinolYes
Guidewire CoatingTeflon; GlyDx Hydrophilic (Predicate: PrimeWire)Teflon; SiliconeFavorable Comparison
Guidewire Tip ConfigurationStraight, pre-shaped "J" (Predicate: PrimeWire)StraightFavorable Comparison
Guidewire Tip Length3.0 cm (Predicate: PrimeWire)3.5 cmFavorable Comparison
Radiopaque Tip?Yes (Predicate: PrimeWire)YesYes
Pressure Sensor Location3.0 cm from distal tip (Predicate: PrimeWire)3.5 cm from distal tipFavorable Comparison
Pressure Sensing & Signal TransHard wired strain gauge (Predicate: PrimeWire)Fiberoptic sensor & fiber bundleFavorable Comparison
OptoMonitor
Operating Temperature (Monitor)16°C to 32°C (Predicate: ComboMap)15°C to 30°CYes
Transport Temperature (Monitor)-20°C to 60°C (Predicate: ComboMap)-25°C to 60°CYes
Operating Relative Humidity30% to 75% (Predicate: ComboMap)10% to 85% non-condensingYes
Storage Temperature (Monitor)-20°C to 60°C (Predicate: ComboMap)Room TemperatureFavorable Comparison
Operating Pressure50 to 106 kPa (Predicate: ComboMap)70 to 106 kPaYes
Pressure Range-30 to 300 mmHg (Predicate: PrimeWire/ComboMap)-30 to 300 mmHgYes
Pressure Accuracy+/- 1 mmHg + 1% (-30 to 50 mmHg), +/- 3% (50-300 mmHg) (Predicate: ComboMap)+/- 1 mmHg + 1% (-30 to 50 mmHg), +/- 3% (50-300 mmHg)Yes
Thermal Zero Shift<0.3 mmHg/deg C (Predicate: PrimeWire)<0.3 mmHg/deg CYes
Zero DriftUnknown (Predicate: PrimeWire)<1 mmHg/hMet internal criteria
Electrical IsolationClass 1 (Predicate: ComboMap)Class 2 (double isolation)Favorable Comparison
User InterfaceTouchscreen, Remote Control (Predicate: ComboMap)TouchscreenFavorable Comparison
Auto-zeroingYes (Predicate: ComboMap)YesYes
Real Time CurvesAortic, distal, Pa/Pd trend (Predicate: ComboMap)Aortic, distalFavorable Comparison
Real Time Numerical ValuesMean aortic, mean distal, mean Pd/mean Pa, heartbeat (Predicate: ComboMap)Mean aortic, mean distal, mean Pd/mean PaFavorable Comparison
FFR Capability?Yes (Predicate: PrimeWire/ComboMap)YesYes
FFR Viewing?Yes (Predicate: ComboMap)YesYes
Basis for FFR DeterminationSimultaneous acquisition of 2 pressure valuesSimultaneous acquisition of 2 pressure valuesYes

Note: "Favorable Comparison" indicates that while not identical, the performance is considered acceptable for substantial equivalence based on the technological characteristics and overall safety/effectiveness. "Met internal criteria" refers to tests where no direct predicate comparison was given for a specific metric.

2. Sample Size Used for the Test Set and Data Provenance

The document describes two types of performance testing:

  • Bench Testing: The text states, "Bench testing has been performed to confirm that the OptoWire and OptoMonitor System meets predetermined acceptance criteria and is substantially equivalent to the predicate devices." It lists various tests (Tip Testing, Tensile Test, Fatigue Test, Sensor Sensitivity to Bending, Pressure and Sensor Testing, Age testing, etc.) for both the OptoWire and OptoMonitor components. No specific sample sizes (number of devices/components tested) are provided for these bench tests. The data provenance is implicitly from laboratory/engineering testing.
  • Animal Testing: "Animal studies have been performed to characterize the in-vivo performance of the OptoWire and OptoMonitor System, as compared to the predicate device and other currently marketed devices with the same indications for use." No specific sample size (number of animals) is provided for the animal studies. The data provenance is from animal models.

The document does not specify a country of origin for the data for either bench or animal testing, nor does it specify if the animal study was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The provided document does not mention the use of human experts or expert consensus to establish ground truth for any of the performance tests (bench or animal studies). The ground truth for these tests would likely be based on established engineering principles, reference measurements, and physiological standards measured during the experiments.

4. Adjudication Method for the Test Set

Since no human experts were explicitly mentioned for establishing ground truth, there is no adjudication method described for the test set. Performance was likely assessed against objective, predetermined technical specifications and direct comparisons to predicate devices.

5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study Was Done

A Multi-reader Multi-case (MRMC) comparative effectiveness study was not mentioned in the provided document. The device is for measurement of physiological parameters (pressure, FFR), not for interpretation of images or other data by human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

The provided document describes the performance of the device itself, which includes the OptoWire (pressure sensing guidewire) and OptoMonitor (electronic signal processing and display units). This constitutes a standalone performance evaluation of the system's ability to accurately measure and display intravascular blood pressure and FFR values. The software within the OptoMonitor was validated "in accordance with design specifications and the system operates as intended when used for its intended use."

7. The Type of Ground Truth Used

The ground truth for the performance testing appears to be based on:

  • Established engineering specifications and physical measurements: For bench testing of mechanical properties, sensor accuracy, electrical safety, etc.
  • Physiological standards and comparative measurements: For animal studies, comparing the device's pressure measurements and handling characteristics against predicate devices and other currently marketed devices with the same indications for use in a living system.

There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for substantial equivalence.

8. The Sample Size for the Training Set

The document does not refer to a "training set" in the context of machine learning or AI. The OptoMonitor contains software, which was "developed, documented and validated in accordance with industry standards (IEC 62304) and FDA guidance." This type of software development and validation typically involves testing against requirements and specifications, not training with a distinct dataset in the way a machine learning model would be.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a "training set" in the context of machine learning, this question is not applicable. The software validation involved demonstrating that the system software operates in accordance with design specifications and performs as intended.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 12, 2015

Opsens. Inc. % Pamela Papineau President Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer, Massachusetts 01432

Re: K142598

Trade/Device Name: OptoWire and OptoMonitor System Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX, DXO Dated: May 13, 2015 Received: May 14, 2015

Dear Pamela Papineau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

A. Mitchell Stein

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142598

Device Name OptoWire and OptoMonitor System

Indications for Use (Describe)

To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/ or other any interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 – 510(k) Summary

General Information

Owner's Name:Opsens, Inc.
Address:2014 rue Cyrille Duquet, #125Quebec, QC G1N 4N6 Canada
Telephone:418-682-9996
Fax Number:418-682-9939
Contact Person:Pamela Papineau, RAC (US, EU, CAN)Delphi Medical Device Consulting, Inc.
Address:5 Whitcomb AvenueAyer, MA 01432
Telephone Number:(978) 772-3552
Fax Number:(978) 796-5460
Subject Device:
Trade Name:OptoWire and OptoMonitor System
Common Name:Intravascular Pressure Monitoring System
Product Code:DXQ / DXO
FDA Regulation:21 CFR 870.1330 - Catheter Guide Wire21 CFR 870.2870 – Catheter Tip Pressure Transducer
Device Classification:Class II
Predicate Devices:
Product Name:PrimeWire Prestige Plus Pressure Guide Wire (Volcano Corp.)
Common Name:Pressure Guide Wire
Product Code:DQX
FDA Regulation:21 CFR 870.1330 – Catheter Guide Wire
Device Classification:Class II
Premarket Notification:K111395
Product Name:ComboMap Pressure and Flow System (Volcano Corp.)

ComboMap Pressure and Flow System (Volcano Corp.) Intravascular Pressure Monitoring System OBJ 21 CFR 870.1200 – Diagnostic Intravascular Catheter Class II K041134

Indications for Use:

Premarket Notification:

Common Name:

FDA Regulation: Device Classification:

Product Code:

To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or other any interventional procedures. Blood pressure

{4}------------------------------------------------

measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease.

Device Description:

The OptoWire and OptoMonitor System includes a pressure sensing guidewire (OptoWire), electronic signal processing and display units (OptoMonitor) that process signals received from the OptoWire to display intravascular blood pressure and fractional flow reserve (FFR) values, and various connection cables. The OptoWire and OptoWire cable are sterile, single-use devices; the remaining system components are reusable.

Substantial Equivalence:

The OptoWire and OptoMonitor System is substantially equivalent to the Volcano Corp. PrimeWire Prestige Plus Pressure Sensing Guidewire (K111395) and the Volcano Corp. ComboMap Pressure and Flow System (K041134). Substantial equivalence is based on indications for use, technological characteristics, device materials, performance characteristics; physical and functional test results, and conformity with standards. A substantial equivalence summary table is provided at the end of this 510(k) Summary.

Biocompatibility Testing:

Testing has been performed on the OptoWire device in accordance with the ISO 10993 family of standards to confirm that all patient-contacting materials are biocompatible. This testing included cytotoxicity, intracutaneous toxicity, acute systemic toxicity, sensitization, hemocompatibility and pyrogenicity.

Sterilization, Packaging and Shelf Llife:

The OptoWire is sterilized to a 106 SAL using an ethylene oxide process that has been validated in accordance with ISO 11135-1. The remaining system components are non-sterile. Testing has been performed to confirm that all packaging (sterile barrier and outer packaging) meets predetermined acceptance criteria with respect to sterile package seal strength and seal integrity, and that the packaging adequately protects all system components throughout transportation and storage. The sterile product shelf life has been validated through package and product testing of unaged devices as well as devices that have been held at an elevated temperature to simulate the labelled shelf life. Samples will also be tested after real time aging.

Software:

The OptoMonitor contains software that has been developed, documented and validated in accordance with industry standards (IEC 62304 - Medical device software - Software life cycle processes) and FDA guidance (General Principles of Software Validation; Final Guidance for FDA and Industry). Complete software design, risk and validation documents have been provided in accordance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Software validation has demonstrated that the system software operates in accordance with design specifications and the system operates as intended when used for its intended use.

{5}------------------------------------------------

Electrical Safety and EMC Testing:

The OptoMonitor has been tested in accordance with IEC 60601-1 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance) to demonstrate electrical safety; testing in accordance with IEC 60601-1-2 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests) has confirmed electromagnetic compatibility.

Performance Testing - Bench:

Bench testing has been performed to confirm that the OptoWire and OptoMonitor System meets predetermined acceptance criteria and is substantially equivalent to the predicate devices. The following is a summary of the performance (bench) testing presented in this 510(k):

ComponentTestingPurposeMet AcceptanceCriteria?
OptoWire
Mechanical /FunctionalVerification• Tip Testing• WireFlexibility/Support• Tensile Test• Torque Tests• Turn to Failure• Fatigue Test• Coating Testing• Compatibility Test• Surface Inspection• Corrosion Test• Dimensional testing• Sensor Sensitivity toBending• Pressure and SensorTesting• Optical Contrast• Optical Sensor Testing• Ambient PressureRange• Signal Loss• Connection/Disconnection Testing• Age testing• FunctionalityandPerformance• Safety• UsabilityYes. All internaldesign criteria weresuccessfully met.Subject devicecompared favorablyto predicate devicestested.
OptoMonitor
• SCUHardware• DUHardware• HCUHardware• System• visual & design• specs & environment• single fault conditions• electrical performance• Age Testing• Transportation Testing• Cleaning• FunctionalityandPerformance• Reliability• Safety• Usability• EnvironmentalYes. All internaldesign criteria weresuccessfully met.

Performance Testing - Animal:

Animal studies have been performed to characterize the in-vivo performance of the OptoWire and OptoMonitor System, as compared to the predicate device and other currently marketed devices with the same indications for use. Animal testing included assessment of the system handling characteristics, compatibility with other devices, and pressure measurement capabilities.

Conclusion:

The OptoWire and OptoMonitor System has been demonstrated to be substantially equivalent to the predicate devices.

{6}------------------------------------------------

OptoWire and OptoMonitor System(current submission; Opsens, Inc.)PrimeWire Prestige PlusPressure Guide Wire(K111395; Volcano Corp.)ComboMap Pressure and Flow System(K041134; Volcano Corp.)
Device Common/Usual NameIntravascular Pressure Guide Wireand Monitoring SystemIntravascular Pressure Guide WireIntravascular Pressure Monitoring System
Device ClassClass IIClass IIClass II
Product Code / RegulationDQX / 21 CFR 870.1330DXO / 21 CFR 870.2870DQX / 21 CFR 870.1330OBJ / 21 CFR 870.2870
Regulation NamePressure Guide WireDiagnostic Intravascular MonitorPressure Guide WireDiagnostic Intravascular Monitor
Indications for UseTo measure pressure in blood vessels,including both coronary and peripheralvessels, during diagnostic angiographyand/or any interventional procedures.Blood pressure measurements providehemodynamic information, such asfractional flow reserve, for the diagnosisand treatment of blood vessel disease.To measure pressure in blood vessels,including both coronary and peripheralvessels, during diagnostic angiographyand/or any interventional procedures.Blood pressure measurements providehemodynamic information for the diagnosisand treatment of blood vessel disease.Intended for use with signal processingsystem such as Volcano ComboMap.A multi-mode system intended for use inall blood vessels, including coronary andperipheral arteries, to measureintravascular blood pressure and/or bloodflow velocities during diagnosticangiography and/or interventionalprocedures.Intended for use with pressure sensingguidewire such as Volcano PrimeWire.
System ComponentsSterile, disposable guidewireReusable signal processor / monitorEmbedded softwareConnecting cablesSterile, disposable guidewireReusable signal processor / monitorEmbedded softwareConnecting cables
System CapabilitiesMeasurement of intravascular bloodpressure including FFR.Measurement of intravascular bloodpressure and flow including FFR(when used with pressure/flow system)Measurement of intravascular bloodpressure and flow including FFR(when used with pressure wire)
Prescription UseRx OnlyRx OnlyRx Only
Pressure Sensing & SignalTransmission TechnologyFiberoptic sensor & fiber bundle embeddedin guidewireHard wired strain gauge embedded inguidewireHard wired strain gauge embedded incompatible guidewire
OptoWire and OptoMonitor System(current submission; Opsens, Inc.)PrimeWire Prestige PlusPressure Guide Wire(K111395; Volcano Corp.)ComboMap Pressure and Flow System(K041134; Volcano Corp.)
Sterile, Single Use PatientContact Component?Yes – OptoWireYes – PrimeWire Prestige PlusNo – computer-controlled pressure andflow monitoring instrument intended foruse with compatible pressure guidewire
FFR Capability?YesYesYes
FFR Viewing?YesN/AYes
Basis for FFR DeterminationSimultaneous acquisition of 2 pressurevalues: distal pressure from sensorembedded in OptoWire; aortic pressurefrom external pressure transducerSimultaneous acquisition of 2 pressurevalues: distal pressure from sensorembedded in PrimeWire; aortic pressurefrom external pressure transducerSimultaneous acquisition of 2 pressurevalues: distal pressure from sensorembedded in compatible guidewire; aorticpressure from external pressure transducer
Operating Temperature(Monitor)15°C to 30°CN/A16°C to 32°C
Transport Temperature(Monitor)-25°C to 60°CN/A-20°C to 60°C
Operating Relative Humidity(Monitor)10% to 85% non-condensingN/A30% to 75%
Storage Temperature(Monitor)Room TemperatureN/A-20°C to 60°C
Operating Pressure70 to 106 kPaN/A50 to 106 kPa
Pressure Range-30 to 300 mmHg-30 to 300 mmHg-30 to 300 mmHg
Pressure Accuracy+/- 1 mmHg plus +/- 1% of reading(pressure range -30 to 50 mmHg) or +/- 3%of reading (pressure range 50 to 300mmHg)Unknown+/- 1 mmHg plus +/- 1% of reading(pressure range -30 to 50 mmHg) or +/-3% of reading (pressure range 50 to 300mmHg)
Thermal Zero Shift<0.3 mmHg/deg C<0.3 mmHg/deg CN/A
Zero Drift<1 mmHg/hUnknownN/A
Electrical IsolationClass 2 (double isolation)N/AClass 1
User InterfaceTouchscreenN/ATouchscreen, Remote Control
Auto-zeroingYesN/AYes
OptoWire and OptoMonitor System(current submission; Opsens, Inc.)PrimeWire Prestige PlusPressure Guide Wire(K111395; Volcano Corp.)ComboMap Pressure and Flow System(K041134; Volcano Corp.)
Real Time CurvesAortic instantaneous pressure, aortic mean pressure, distal instantaneous pressure,distal mean pressureN/AAortic instantaneous pressure, aortic mean pressure, distal instantaneous pressure,distal mean pressure, Pa/Pd trend curve
Real Time Numerical ValuesMean aortic pressure, mean distal pressure,mean Pd/mean PaN/AMean aortic pressure, mean distal pressure,mean Pd/mean Pa, heartbeat
Minimum Pd/Pa Cursor(Detection of FFR Locus)YesN/AYes
Recording ValuesInstantaneous Pa, Pd and Pd/Pa; mean Pa;mean Pd; mean Pd/mean PaN/AInstantaneous Pa, Pd and Pd/Pa; mean Pa;mean Pd; mean Pd/mean Pa
Display MonitorLCDN/ALCD
Aortic InputHigh Level (100 mmHg/V)N/AHigh Level (100 mmHg/V)Low Level (5uV/V/mmHg)
Aortic OutputNoN/ALow Level (5uV/V/mmHg)
Distal InputOptoWire (optical)N/ACompatible pressure wire (PrimeWire)(electrical)
Distal OutputLow Level (5uV/V/mmHg)N/ALow Level (5uV/V/mmHg)
Guidewire OD0.014"0.014"N/A
Guidewire Length175 cm185 cm, 300 cmN/A
Guidewire Shaft MaterialStainless Steel; NitinolStainless Steel; SSN/A
Guidewire CoatingTeflon; SiliconeTeflon; GlyDx Hydrophilic coatingN/A
Guidewire Tip ConfigurationStraightStraight, pre-shaped "J"N/A
Guidewire Tip Length3.5 cm3.0 cmN/A
Radiopaque Tip?YesYesN/A
Pressure Sensor Location3.5 cm from distal tip3.0 cm from distal tipN/A

Substantial Equivalence Comparison

{7}------------------------------------------------

{8}------------------------------------------------

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.