(84 days)
The Xia® Growth Rod Conversion Set is indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The Xia® Growth Rod Conversion Set may be used with any cleared Xia® 4.5 Spinal System rod construct. The Xia® Growth Rod Conversion Set is not intended for use in conjunction with staples.
The Xia® Growth Rod Conversion Set consists of connectors designed to convert a traditional fusion construct into a non-fusion growth enabling construct that can be surgically lengthened on a periodic basis as the patient grows. The Xia® Growth Rod Conversion Set components are manufactured from titanium alloy and are designed to interact with constructs consisting of hooks, screws, connectors, and 4.5mm diameter rods. The Xia® Growth Rod Conversion Set is intended for use only with Xia® 4.5 Spinal System fusion constructs cleared for pediatric use.
The provided text describes the Xia® Growth Rod Conversion Set, a medical device. However, it does not contain any information about acceptance criteria or a study proving the device meets acceptance criteria in the context of diagnostic performance (e.g., accuracy, sensitivity, specificity for an AI-powered device).
This document is a 510(k) premarket notification for a medical device (Xia® Growth Rod Conversion Set), which primarily focuses on demonstrating substantial equivalence to a predicate device.
Specifically, the "Summary of the Performance Data" section (Page 4, Table "510(k) Summary") states:
"Engineering analysis demonstrated that introduction of the Growth Rod Conversion Set does not adversely affect performance of the Xia® 4.5 Spinal System and does not represent a new, worst case scenario. No additional performance data was provided."
This indicates that the "study" conducted was an engineering analysis to ensure the new component does not negatively impact the performance of an existing, cleared system, rather than a clinical study evaluating diagnostic accuracy or a specific set of performance metrics against acceptance criteria.
Therefore, I cannot provide the requested information as it is not present in the provided text. The document is about a different type of medical device review process (510(k) for a physical implantable device) that relies on substantial equivalence and engineering analysis, not diagnostic performance metrics or human reader studies.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 29, 2022
Stryker Spine Garry T. Hayeck, Ph.D. Senior Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401
Re: K142114
Trade/Device Name: Xia® Growth Rod Conversion Set Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: PGM
Dear Dr. Hayeck:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 27, 2014. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation number, 21 CFR 888.3070.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Ronald Jean, OHT6: Office of Orthopedic Devices, (301)796-5650, Ronald.Jean@fda.hhs.gov
Sincerely,
Ronald P. Jean -S
Ronald P. Jean, Ph.D. Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol of medicine, with three intertwined snakes and a staff. The logo is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2014
Stryker Spine Garry T. Hayeck, Ph.D. Senior Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401
Re: K142114
Trade/Device Name: Xia® Growth Rod Conversion Set Regulatory Class: Unclassified Product Code: PGM Dated: July 31, 2014 Received: August 4, 2014
Dear Dr. Hayeck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K142114
Device Name Xia® Growth Rod Conversion Set
Indications for Use (Describe)
The Xia® Growth Rod Conversion Set is indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, earlyonset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The Xia® Growth Rod Conversion Set may be used with any cleared Xia® 4.5 Spinal System rod construct. The Xia® Growth Rod Conversion Set is not intended for use in conjunction with staples.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) Summary: Stryker Spine Xia® Growth Rod Conversion Set | |
|---|---|
| Submitter | Stryker Spine2 Pearl CourtAllendale, NJ 07401 |
| Contact Person | Garry T. Hayeck, Ph.D.Senior Regulatory Affairs SpecialistPhone: 201-760-8043Fax: 201-760-8406E-mail: garry.hayeck@stryker.com |
| Date Prepared | September 29, 2014 |
| Trade Name | Xia® Growth Rod Conversion Set |
| Common Name | Growing Rod System |
| Proposed Class | Unclassified |
| Product Code | PGM |
| Predicate andReference Devices | The Xia® Growth Rod Conversion Set was shown to be substantiallyequivalent to the primary predicate device listed below:Medtronic Sofamor Danek, CD HORIZON® Growth Rod ConversionSet, K133904The following reference device was also cited in support of thissubmission:Stryker Spine, Xia® 4.5 Spinal System, K121342 |
| Device Description | The Xia® Growth Rod Conversion Set consists of connectors designedto convert a traditional fusion construct into a non-fusion growthenabling construct that can be surgically lengthened on a periodicbasis as the patient grows.The Xia® Growth Rod Conversion Set components are manufacturedfrom titanium alloy and are designed to interact with constructsconsisting of hooks, screws, connectors, and 4.5mm diameter rods.The Xia® Growth Rod Conversion Set is intended for use only with Xia®4.5 Spinal System fusion constructs cleared for pediatric use. |
| Indications for Use | The Xia® Growth Rod Conversion Set is indicated in patients withpotential for additional spinal growth under 10 years of age whorequire surgical treatment to obtain and maintain correction ofsevere, progressive, life-threatening, early-onset spinal deformitiesassociated with thoracic insufficiency, including early-onset scoliosis.The Xia® Growth Rod Conversion Set may be used with any clearedXia® 4.5 Spinal System rod construct. The Xia® Growth Rod ConversionSet is not intended for use in conjunction with staples. |
| Summary ofTechnologicalCharacteristics | Characteristics of the Xia® Growth Rod Conversion Set aresubstantially equivalent to those of the above mentioned CDHORIZON® Growth Rod Conversion Set based on material, geometry,intended use, and fundamental scientific technology, |
| Summary of thePerformance Data | Engineering analysis demonstrated that introduction of the GrowthRod Conversion Set does not adversely affect performance of theXia® 4.5 Spinal System and does not represent a new, worst casescenario. No additional performance data was provided. |
| Conclusion | This system is as safe and effective as the previously cleared CDHORIZON® Growth Rod Conversion Set listed above due to theirshared technological and material characteristics, and principles ofoperation. |
{5}------------------------------------------------
N/A