K Number
K141625

Validate with FDA (Live)

Date Cleared
2014-07-18

(31 days)

Product Code
Regulation Number
878.4493
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

POLYS YN™ Polyglycolic Acid (PGA) Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in Cardiovascular and Neurological procedures.

Device Description

PGA Sutures are supplied as braided or monofilament, dyed (violet) or undyed and coated or uncoated. The pigment for the violet is D&C Violet #2. Where applicable, the coating is a copolymer of polycarpolactone and calcium stearate. The PGA Suture is available in Size 2 through Size 10-0.

AI/ML Overview

This document describes the 510(k) summary for the PolySynTM (PGA) Surgical Suture. The purpose of the submission is to demonstrate substantial equivalence to legally marketed predicate devices, specifically for additional diameter sizes (USP Size 7-0 through 10-0).

1. Table of Acceptance Criteria and Reported Device Performance:

The primary acceptance criteria for the PolySynTM (PGA) Surgical Suture are conformity to the USP monograph for absorbable sutures and demonstration of substantial equivalence to predicate devices through various performance tests.

Acceptance CriteriaReported Device Performance
Conformity to USP monograph for absorbable sutures"The results of this testing demonstrates that the PolySyn™ suture is substantially equivalent to the predicate devices." (Implies conformity to USP as part of substantial equivalence).
Demonstration of substantial equivalence to predicate device including in vitro post-hydrolysis tensile testing"The results of this testing demonstrates that the PolySyn™ suture is substantially equivalent to the predicate devices."

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the specific sample sizes for each test in the non-clinical laboratory performance testing. It generally refers to "performance testing."

The data provenance is from non-clinical laboratory performance testing. The country of origin of the data is not explicitly stated, but it is implied to be generated by the applicant, Surgical Specialties Corporation, in the United States (given their address). The data is retrospective testing done to support the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

Not applicable. This device is a surgical suture, and the evaluation is based on non-clinical laboratory performance testing against established standards (USP monograph) and predicate device characteristics, not on expert interpretation of medical images or patient data.

4. Adjudication Method for the Test Set:

Not applicable. The evaluation is based on objective laboratory measurements against predefined specifications and comparisons to predicate device data, not on subjective assessments requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

Not applicable. This device is a surgical suture, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study is irrelevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This is a traditional medical device (surgical suture), not an algorithm or AI system.

7. The Type of Ground Truth Used:

The ground truth used for evaluation is based on:

  • USP (United States Pharmacopeia) monograph for absorbable sutures: These are established, scientifically vetted standards for the physical and performance characteristics of surgical sutures.
  • Predicate device characteristics: These are the established performance parameters of the legally marketed devices (K965162 and K022269) to which the new device is compared for substantial equivalence.

8. The Sample Size for the Training Set:

Not applicable. This is a traditional medical device (surgical suture) and does not involve machine learning or AI, thus no "training set."

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As there is no training set for an AI/ML algorithm, there is no ground truth established for it.

{0}------------------------------------------------

JUI. 18 2014

.

Page 1 of 2 K141625

510(k) Summary

ﺗﮯ

Date Prepared:Jul 17, 2014
Company:Surgical Specialties Corporation100 Dennis Dr.Reading, PA 19606
Contact:Paul AmudalaRegulatory Affairs SpecialistPhone: 610-404-3376Fax: 610-404-3924Email: pamudala@surgicalspecialties.com
Device trade name:PolySynTM (PGA) Surgical Suture
Device Common Name:Absorbable Polyglycolic Acid Surgical Suture
Device classification:Suture, Absorbable, Synthetic, Polyglycolic AcidProduct code, GAM21 CFR 878.4493Class II
Legally marketed deviceto which the device issubstantially equivalent:K965162: PolySynTM Surgical Suture (primary)K022269 Coated VICRYLTM (Polyglactin) Suture (reference)
Description of thedevice:PGA Sutures are supplied as braided or monofilament, dyed (violet) or undyed and coated or uncoated. The pigment for the violet is D&C Violet #2. Where applicable, the coating is a copolymer of polycarpolactone and calcium stearate. The PGA Suture is available in Size 2 through Size 10-0.
Indications for Use:PolySynTM (Polyglycolic Acid (PGA)) Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in Cardiovascular and Neurological procedures.
SubstantialEquivalence:The proposed additional diameter sizes, USP Size 7-0 through 10-0, of the PolySynTM suture product line have the same material (K965162), design, intended use and technological characteristics as the predicate device. The only difference between the proposed and predicate (K965162) device is the suture diameter.

{1}------------------------------------------------

K4141625

.

: .

Page 2 of 2

.

Performance tests:Non-clinical laboratory performance testing was conducted toconfirm that the PolySyn™ suture conforms to the USPmonograph for absorbable sutures (as applicable). This testingwas performed in accordance with FDA's Class II SpecialControls Guidance Document: Surgical Sutures, Issued June 3,2003. Additional performance testing was conducted in order todemonstrate substantial equivalence to the predicate deviceincluding in vitro post-hydrolysis tensile testing.
The results of this testing demonstrates that the PolySyn™ sutureis substantially equivalent to the predicate devices.

l

.

.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drue Administration 10903 New Hampshire Avenue Document Contral Center - WO66-G609 Silver Spring, MD 20993-0002

July 18, 2014

Surgical Specialties Corporation Mr. Paul Amudala Regulatory Affairs Specialist 100 Dennis Drive Reading, Pennsylvania 19606

Re: K141625

Trade/Device Name: POLYSYN" (Polyglycolic Acid) Surgical Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: Class II Product Code: GAM Dated: June 18, 2014 Reccived: June 19, 2014

Dear Mr. Amudala:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Paul Amudala

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on fast page.

510(k) Number (if known) K141625

Device Name

POLYS YN™ (Polyglycolic Acid) Surgical Suture

Indications for Use (Describe)

POLYS YN™ Polyglycolic Acid (PGA) Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in Cardiovascular and Neurological procedures.

Type of Use (Select one or both, as applicable)

[ Prescription Use (Part 21 CFR 801 Subpart D)

O Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Peter L. Hud 2014.07.18 07:27

FORM FDA 3881 (1/14)

§ 878.4493 Absorbable poly(glycolide/l-lactide) surgical suture.

(a)
Identification. An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.