K Number
K141427

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-07-24

(420 days)

Product Code
Regulation Number
864.5700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. For in vitro diagnostic use. For professional use only.

Device Description

The VerifyNow System is a turbidimetric-based optical detection system that measures plateletinduced aggregation. The system consists of an instrument, a disposable test device and quality control materials. Quality control measures include an instrument based electronic quality control (EOC). two levels of wet quality controls (WOC), internal quality controls, and shipping controls. The instrument controls all assay sequencing, temperature, reagent-sample mixing and performs self-diagnostics. The degree of platelet function is determined and the result is displayed.

The VerifyNow PRUTest device contains three lyophilized reagent pellets in separate reaction chambers within the test device: 1) ADP pellet consisting of a preparation of Fibrinogen and BSA coated beads, adenosine-5-diphosphate (ADP), prostaglandin E1 (PGE1), dye, buffer, and a preservative; 2) TRAP pellet (Internal Control) consisting of a preparation of iso-TRAP (Thrombin Receptor Activating Peptide), Fibrinogen and BSA coated beads, buffer, dye, and a preservative; and 3) No-Agonist Pellet (NAP) consisting of a preparation of BSA coated beads, dye, buffer, and a preservative.

AI/ML Overview

The provided text describes the performance studies for the VerifyNow PRUTest device, demonstrating its substantial equivalence to a predicate device (VerifyNow P2Y12 Assay).

Here's an analysis of the acceptance criteria and the studies that prove the device meets these criteria:


1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly list a single table of "acceptance criteria" for all tests, but rather describes the criteria within the context of each study. Below is a compilation of the criteria and performance for the key studies described:

Study/ParameterAcceptance CriteriaReported Device Performance
Method ComparisonSlope not statistically significantly different from 1.0; Intercept not statistically significantly different from 0; Correlation (r) greater than 0.9.P2Y12 Rep 1 vs. PRUTest Rep 1: Slope = 1.01 (95% CI: 0.97–1.05), p-value = 0.56; Intercept = -0.77 (-8.00-6.50); Correlation R = 0.98. P2Y12 Avg. vs. PRUTest Avg.: Slope = 1.04 (95% CI: 1.00-1.07), p-value = 0.07; Intercept = -4.57 (-11.2-2.05); Correlation R = 0.98. Performance met criteria.
Precision (WQC Level 1)PRU result ≤ 30 (indicates negative control/no aggregation).All WQC Level 1 results fell within the acceptable range of ≤ 30 PRU. At the nine-month time point, the range of PRU values generated with WOC Level 1 did not exceed 30 PRU.
Precision (WQC Level 2)Within-run, Between-run, Between-day, Between-lot, and Total %CVs for WQC Level 2 (specific values are not given as general acceptance criteria, but implied by the successful analysis against CLSI EP05-A2). For stability, Percent recovery ≥ 90 to ≤ 110% of baseline PRU.Between-lot: Total SD = 22.2, Total %CV = 7.4. Between instrument: Total SD = 24.2, Total %CV = 8.0. For stability, percent recovery of WQC Level 2 passed acceptance criterion (≥ 90 to ≤ 110%) at nine months.
Precision (Whole Blood - Multiple Lot)Individual CVs for Within-run, Between-run, Between-day, Between-lot components each < 10% (for most categories, N/A for Between-day for specific donors).Most individual donor results showed component %CVs < 10%, though some Between-day values were higher (e.g., Donor 1 Between-day %CV 15.0). All specific donor data are provided in Table H. However, the overall conclusion is not explicitly stated if all individual criteria were met for every donor.
Precision (Whole Blood - Multiple Instrument)Individual CVs for Within-run, Between-run, Between-day, Between-lot components each < 10% (for most categories, N/A for Between-day for specific donors).Similar to multiple lot, most individual donor results showed component %CVs < 10%, with some exceptions (e.g., Donor 7 Between-run %CV 17.7 and Between-day %CV 57.1). All specific donor data are provided in Table I.
Open Pouch Device Stability (Whole Blood)Percent recovery of VerifyNow PRUTest values ≤ 10% deviation from baseline.Met acceptance criteria after exposure to 61-73% RH and 21-24°C temperatures for 10 hours.
Open Pouch Device Stability (WQC Level 2)PRU result remained within the allowable range (230–362).Remained within the allowable range for 10 hours under the same conditions.
Sample Wait TimeNot explicitly stated, but the study determined an appropriate value.Determined to be 10 minutes (minimum storage time after venipuncture).
Sample StabilityNot explicitly stated, but the study determined an appropriate value.Determined to be four hours (blood sample must be assayed prior to four hours from draw time).
Interfering SubstancesPercent recovery which deviated less than 10% from baseline PRUTest values.Each interferent tested showed percent recovery deviating less than 10% from baseline PRUTest values. None of the substances significantly interfered.
Reagent Stability (WQC Level 1)PRU result < 30.At the nine-month time point, the range of PRU values generated did not exceed 30 PRU. All test intervals met criteria.
Reagent Stability (WQC Level 2)Percent recovery ≥ 90 to ≤ 110% of the baseline PRU result.At the nine-month time point, percent recovery passed criteria. All lots fell within ± 10% of the control condition.
Reagent Stability (Donor Whole Blood)Percent recovery ≥ 90 to ≤ 110% of the baseline PRU result.At the nine-month time point, percent recovery passed criteria. Aggregated lot data fell within ± 10% of the control condition.
Reagent Stability (Inhibited Whole Blood)Percent recovery ≥ 90 to ≤ 110% of the baseline PRU result.At the nine-month time point, percent recovery passed criteria. Aggregated lot data fell within ± 10% of the control condition.

2. Sample Size Used for the Test Set and Data Provenance:

  • Method Comparison: 119 samples. The origin is not specified, but the study involved "three investigational intended use sites." It is implied to be prospective, collected for the purpose of the study.
  • Reference Range Study: 152 evaluable healthy donors. Provenance not specified, but likely collected prospectively for the study.
  • Expected PRUTest values (Treatment-naïve ACS patients): 84 patients. Provenance not specified, likely collected prospectively for the study.
  • Expected PRUTest values (Clopidogrel-treated ACS patients): 71 patients. Provenance not specified, likely collected prospectively for the study.
  • Precision Studies (WQC Level 2): 240 tests (for both Between-lot and Between-instrument). Likely collected prospectively for the study.
  • Precision Studies (Whole Blood): 9 donors (each sampled multiple times, e.g., 30 to 120 reps per donor across days/runs/lots/instruments). Provenance not specified, likely collected prospectively for the study.
  • Open Pouch Device Stability: Whole blood samples (n=3) and WQC Level 2. Likely collected prospectively for the study.
  • Interfering Substances: Samples from 11 different donors. Likely collected prospectively for the study.
  • Reagent Stability (WQC Level 1, 2): Multiple lots (3 each of WQC levels and reagent devices), tested at time zero, 3, 6, and 9 months (number of individual tests not specified but for WQC Level 1 at 9 months, N=108). Likely collected prospectively for the study.
  • Reagent Stability (Donor Whole Blood): Whole blood samples from normal donors (specific N not specified, but tested across 3 reagent lots at 0, 3, 6, 9 months). Table N specifically shows data for 2 donors, at 9-month time point (N=3 for each lot/donor combination, total of 18 samples per donor for the specific comparison). Likely collected prospectively for the study.
  • Reagent Stability (Inhibited Whole Blood): 2 donors (Donors 399 and 1077). Similar breakdown as above, tested across 3 reagent lots at 9 months. Likely collected prospectively for the study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

The document describes in vitro diagnostic (IVD) device performance. For this type of device, ground truth is typically established by comparative measurements against a recognized reference method or by established laboratory-based assays (like the predicate device or clinical reference ranges) rather than expert consensus on image interpretation or clinical diagnosis. Therefore, "experts" in the context of establishing ground truth as typically understood for imaging AI models (e.g., radiologists) are not applicable here. The ground truth relies on the inherent biochemical mechanism of the device (platelet aggregation) and comparison to established methods.

4. Adjudication Method for the Test Set:

Not applicable in the conventional sense for image or clinical diagnosis adjudication. The "ground truth" for the analytical performance studies (method comparison, precision, stability, interference) is derived from quantitative measurements and statistical analysis, often against the predicate device or a control condition.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is an automated platelet aggregation system, an IVD. It is not an AI-assisted diagnostic tool that aids human "readers" (like radiologists or pathologists) in interpreting complex data. The device itself performs the measurement and outputs a quantitative result (PRU value).

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Yes, the studies describe the standalone performance of the VerifyNow PRUTest system. The system "automatically dispenses... blood... with no blood handling required by the user." The instrument "controls all assay sequencing, temperature, reagent-sample mixing and performs self-diagnostics. The degree of platelet function is determined and the result is displayed." This indicates it is an automated, standalone system.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc):

The ground truth used primarily consists of:

  • Comparative Analytical Performance: Measurements generated by the legally marketed predicate device (VerifyNow P2Y12 Assay) and comparison against it.
  • Established Reference Materials/Controls: Wet Quality Controls (WQC) with known characteristics.
  • Clinically Relevant Ranges: Reference ranges and expected values established statistically from healthy donors and patient populations (treatment-naïve, treated with clopidogrel).
  • Statistical Analysis: Adherence to CLSI guidelines (EP28-A3c, EP05-A2, EP07-A2) for establishing reference intervals, evaluating precision, and testing interference.

8. The Sample Size for the Training Set:

The document does not specify a "training set" in the context of an algorithm that learns from data in the way a machine learning model does. This device is described as a turbidimetric-based optical detection system, implying a fixed algorithm and chemical reagent system rather than a machine learning algorithm requiring a distinct training phase with a labeled dataset. Its development would involve optimization of reagents, optical detection parameters, and algorithms to convert optical signals to PRU values, which is a different paradigm than typical AI/ML training.

9. How the Ground Truth for the Training Set Was Established:

As there is no "training set" in the AI/ML context, this question is not directly applicable. The "ground truth" for the device's design and calibration would have been established during its development through a combination of:

  • Biochemical principles of platelet aggregation.
  • Empirical optimization of reagent concentrations and measurement parameters.
  • Comparison to gold standard methods or other validated platelet function assays during the research and development phase to ensure the system accurately reflects platelet P2Y12 receptor blockade.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a stylized eagle with three human profiles incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 30, 2015

ACCUMETRICS, INC. FRANK LADUCA CHIEF SCIENTIFIC OFFICER 3985 SORRENTO VALLEY BLVD. SUITE B SAN DIEGO. CA92121

Re: K141427

Trade/Device Name: VerifyNow PRUTest Regulation Number: 21 CFR 684.5700 Regulation Name: Automated platelet aggregation system Regulatory Class: Class II Product Code: JOZ Dated: July 23, 2015 Received: July 24, 2015

Dear Dr. LaDuca:

This letter corrects our substantially equivalent letter of July 24, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If vou desire specific advice for your device on our labeling regulation (21 CFR Part Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Leonthena R. Carrington -S

Leonthena R. Carrington, MS. MBA, MT (ASCP) Director Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141427

Device Name VerifyNow PRUTest

Indications for Use (Describe)

The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. For in vitro diagnostic use. For professional use only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) summary is submitted in accordance with 21 CFR §807.92.

Owner:Accumetrics, Inc.3985 Sorrento Valley Blvd.San Diego, CA 92121 USAContact: Frank M. LaDuca, PhD; Chief Scientific OfficerTelephone: (858) 263-2450Fax: (858) 875-0603
Prepared:22 July 2015
Trade name:Accumetrics, Inc. VerifyNow PRUTest ™
Common name:Platelet reactivity test
Classification name:Automated platelet aggregation system(21 CFR §864.5700; Product Code: JOZ)
Predicate device:VerifyNow P2Y12 Assay, Accumetrics, Inc., K051231

Device description:

The VerifyNow System is a turbidimetric-based optical detection system that measures plateletinduced aggregation. The system consists of an instrument, a disposable test device and quality control materials. Quality control measures include an instrument based electronic quality control (EOC). two levels of wet quality controls (WOC), internal quality controls, and shipping controls. The instrument controls all assay sequencing, temperature, reagent-sample mixing and performs self-diagnostics. The degree of platelet function is determined and the result is displayed.

The VerifyNow PRUTest device contains three lyophilized reagent pellets in separate reaction chambers within the test device: 1) ADP pellet consisting of a preparation of Fibrinogen and BSA coated beads, adenosine-5-diphosphate (ADP), prostaglandin E1 (PGE1), dye, buffer, and a preservative; 2) TRAP pellet (Internal Control) consisting of a preparation of iso-TRAP (Thrombin Receptor Activating Peptide), Fibrinogen and BSA coated beads, buffer, dye, and a preservative; and 3) No-Agonist Pellet (NAP) consisting of a preparation of BSA coated beads, dye, buffer, and a preservative.

Test principle:

The patient sample is citrate-anticoagulated whole blood, which is automatically dispensed from the blood collection tube into the test device by the instrument, with no blood handling required by the user. Fibrinogen-coated microparticles are used in the VerifyNow PRUTest device to bind activated platelet GP IIb/IIIa receptors. ADP is incorporated into the assay to activate platelets. and the reagent is formulated to specifically measure P2Y12-mediated platelet aggregation.

{4}------------------------------------------------

When the activated platelets are exposed to the fibrinogen-coated microparticles, aggregation occurs in proportion to the number of activated platelet receptors. The VerifyNow PRUTest reports results in P2Y12 Reaction Units (PRU).

Intended use:

The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. For in vitro diagnostic use. For professional use only.

Indications for use:

Same as the intended use.

Summary of the technological characteristics of the VerifyNow PRUTest (subject device) compared to VerifyNow P2Y12 Assay (predicate device):

The technological characteristics of the Accumetrics Inc. VerifyNow PRUTest are compared with the VerifyNow P2Y12 Assay (K051231) in the table below. This comparison demonstrates the substantial equivalence of this device to the predicate device. The devices are similar in intended use, principle of operation and specimen type. Both devices have similar analytical performance. The modifications from the predicate device are:

Reagents:

  • ADP pellet: The reagent pellet was reformulated to contain 80% Fibrinogen/20% BSA coated beads from the previous 100% Fibrinogen coated beads.
  • TRAP pellet: This reagent was moved from channel 4 to channel 3. The formulation . was unchanged. This channel is used as an internal assay control.
  • NAP pellet: The No-Agonist pellet was introduced to channel 4. The purpose of this ● channel is to prevent a software error and provide no further input into the algorithm for generating the PRU result.

Controls (WQC):

  • Level 1 (B-Carbon Sol Filler): This control was reformulated to contain BSA instead of ● Fish Skin Gelatin.
  • . Level 2: The GPRP peptide conjugate length was modified from 2 x 10° MW AD to 500 MW AD.

Test Device/Instrumentation:

  • . The humidity sensor was removed from the test device.
  • Port cover added to instrument. ●
  • The instrument reports results for the ADP pellet measurement in PRU units.

{5}------------------------------------------------

ItemAccumetrics, Inc.VerifyNow PRUTest(Subject Device)Accumetrics, Inc.VerifyNow P2Y12 Assay (K051231)(Predicate Device)
Similarities
Principle ofOperationFibrinogen bead / platelet aggregationwith optical detection.Same.
Specimen TypeCitrate-anticoagulated whole blood.Same.
Testing SitePoint of care or laboratory.Same.
Assay ResultsP2Y12 Reaction Units (PRU).Same.
ControlsPre-packaged, two-level Wet QualityControls (WQC); Internal control ineach assay device.Same.
Test Procedure1. Add cartridge to instrument.2. Attach blood collection tube.No other operator intervention.Same.
CalibrationFactory.Same.
Time to result3 minutes.Same.
Analytical ClaimsHematocrit range = 33-52%Same.
Modifications
Intended UseThe Accumetrics, Inc. VerifyNowPRUTest is a whole blood test used inthe laboratory or point of care settingto measure the level of platelet P2Y12receptor blockade. For in vitrodiagnostic use. For professional useonly.The VerifyNow P2Y12 Assay is awhole blood assay used in thelaboratory or point of care setting tomeasure the level of platelet P2Y 12receptor blockade.
ReagentsADP Pellet contains 80%Fibrinogen/20% BSA coated beadsADP Pellet contains 100% Fibrinogencoated beads
TRAP Pellet found in channel 3.TRAP Pellet found in channel 4.
NAP Pellet introduced into channel 4.No NAP Pellet
Level 1 (B-Carbon Sol Filler) containsBSALevel 1 (F-Carbon Filler) contains FishSkin Gelatin
Level 2 GPRP Peptide Conjugatelength truncated to 500,000 kDaLevel 2 GPRP
Test DeviceNo humidity sensorContains humidity sensor
Analytical ClaimsStability = 9 months (ongoing real-time study in process)Stability = 18 months

Table A. Comparison of Predicate to Subject Device

Reference range:

A reference range study (Healthy Donor PRU Reference Range) was conducted to measure platelet function in a healthy donor population using VerifyNow PRUTest. A total of 152 evaluable healthy donors were recruited and their blood samples tested with the VerifyNow PRUTest. The data were evaluated as recommended in EP28-A3c "Defining, Establishing and

{6}------------------------------------------------

Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline – Third Edition. " The reference range determined from this study, expressed as the central 95% Confidence Interval of the mean, is 182-335 PRU. The data are summarized in the table below.

Table B. Reference Range

NMeanSD95% confidence interval (CI)
Lower limit (95% CI)Upper limit (95% CI)
15226642182 (116–197)335 (324–354)

Expected PRUTest values:

Treatment-naïve patients in the intended use population

Whole blood samples from 84 patients with acute coronary syndrome (ACS) that were not receiving a P2Y12 receptor inhibiting drug were tested with VerifyNow PRUTest. The data were evaluated as recommended in EP28-A3c "Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory: Approved Guideline – Third Edition." The range of values expressed as the central 95% Confidence Interval of the mean is 180-376 PRU. The data are summarized in the table below.

Table C. Expected PRU values in treatment-naïve patients

NMean(PRU)SD(PRU)95% confidence interval (CI)
8427448Lower limit (95% CI)180 (160–200)Upper limit (95% CI)376 (358–395)

ACS patients receiving treatment with clopidogrel

Whole blood samples from 71 patients with ACS receiving dual treatment with aspirin and clopidogrel were tested with VerifyNow PRUTest. The data were evaluated as recommended in EP28-A3c "Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory: Approved Guideline - Third Edition. " The range of values expressed as the central 95% Confidence Interval of the mean is 6-300 PRU. The data are shown in the table below.

Table D. Expected PRU values in patients receiving clopidogrel treatment

NMean(PRU)SD(PRU)95% confidence interval (CI)
71156736 (0-34)300 (269-329)

{7}------------------------------------------------

Performance studies

Method Comparison

A method comparison study (VerifyNow P2Y12 and PRUTest Method Comparison Study) was performed using the VerifyNow P2Y12 Assay System (predicate device) and the VerifyNow PRUTest System (subject device). A total of three investigational intended use sites were used to gather data for this investigation. The VerifyNow test analysis was performed at each site by intended users who were trained by the manufacturer. Test device and instrument configurations are described below.

Results:

The PRU results from both the VerifyNow P2Y12 (predicate device) and VerifyNow PRUTest (subject device) demonstrate comparable performance, with a slope not statistically significantly different from 1.0 and an intercept not statistically significantly different from 0, and a correlation (r) greater than 0.9 as shown in the table below.

RegressionSlopeSlopeInterceptCorrelation
ComparisonNMethod(95% CI)p-value(95% CI)Coefficient R
P2Y12 Rep 1 vs.119Ordinary1.01-0.770.98
PRUTest Rep 1Least Squares(0.97—1.05)0.56(-8.00-6.50)
P2Y12 Avg. vs.119Ordinary1.04-4.57
PRUTest Avg.Deming(1.00-1.07)0.07(-11.2-2.05)0.98

Table E. Method Comparison

CLSI Precision Studies

Complex precision for the VerifyNow PRUTest was assessed using Wet Quality Controls (WOC). Levels 1 and 2, and intermediate precision was assessed using whole blood samples from healthy donors and from ACS patients receiving clopidogrel with or without aspirin by testing three different lots of test devices and three instruments. Data were analyzed in accordance with Clinical Laboratory Standards Institute (CLSI) guidance EP05-A2, "Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Second Edition."

Testing over several days with samples from the same patient is expected to yield variable results between days because platelet reactivity in response to orally ingested medications known to vary based upon time between dosages, as well as the time between dosage and blood collection. A summary of the precision verification studies performed is shown in Table F.

{8}------------------------------------------------

Table F. Description of precision studies

TypeDescription
Complex precision using WQC Levels 1and 2Measured intermediate precision with WQCLevel 2 samples over multiple days.
Complex precision between instrumentsusing WQC Level 2Measured Level 2 WQC between-instrumentprecision over multiple days using a single lot ofPRUTest devices.
Intermediate precision using whole bloodsamples from donors and ACS patientsMeasured intermediate precision and instrument-to-instrument precision with whole bloodsamples over multiple days using subjects ondual antiplatelet therapy, healthy subjects dosedin vivo with P2Y12 clopidogrel / P2Y12inhibitor, and normal healthy donors
Intermediate precision betweeninstruments using whole blood samplesfrom donors and ACS patientsMeasured whole blood between-instrumentprecision over multiple days using a single lot ofPRUTest devices.

Complex precision using WQC Levels 1 and 2:

WQC Level 1 and Level 2 samples were tested over 20 nonconsecutive days. WQC Level 1 is a negative control (blank) from which no aggregation is expected. All WQC Level 1 results fell within the acceptable range of ≤ 30 PRU and data are not shown. WQC Level 2 results are detailed below. Analysis of variance (ANOVA) was used to estimate the components of variance in accordance with CLSI guidance EP05-A2.

Table G. Reproducibility using WQC Level 2

Between-lot

NMeanPRUWithin-runBetween-runBetween-dayBetween-lotTotal
SD%CVSD%CVSD%CVSD%CVSD%CV
240300.513.64.57.72.58.32.813.54.522.27.4

Between instrument

NMeanPRUWithin-runBetween-runBetween-dayBetween-lotTotal
SD%CVSD%CVSD%CVSD%CVSD%CV
240302.722.67.50.00.08.52.80.00.024.28.0

Intermediate precision using whole blood samples

Whole blood samples from nine donors were drawn once or twice a day, run in duplicate, using three device lots on three instruments over a minimum of five non-consecutive days. Analysis of variance (ANOVA) was used to estimate the components of variance within individual donors in accordance with CLSI guidance EP05-A2. The results of these precision measurements are

{9}------------------------------------------------

detailed in the following two tables. Table H evaluates precision across multiple PRUTest device lots using whole blood samples. Table I evaluates precision across multiple VerifyNow instruments using whole blood samples.

IDRx1DaysRunsRepsMean PRUSDWithin-run%CVSDBetween-run%CVSDBetween-day%CVSDBetween-lot%CV
Acceptance criteria:<10<10N/A2<10
125130644.77.4------9.615.00.030.03
2010211824410.44.20.00.021.99.00.00.0
3152601626.33.95.13.117.710.90.00.0
41523019012.06.3------15.17.90.00.0
5010212028911.23.97.72.727.09.30.00.0
6251302218.94.0------17.07.70.00.0
7182961239.67.818.715.157.646.70.00.0
8110211421613.06.07.33.416.57.60.00.0
92513016310.96.7------13.38.20.00.0

Table H: Precision of whole blood samples - multiple lot experiment

Table I: Precision of whole blood samples - multiple instrument experiment

IDRx1DaysRunsRepsMean PRUSDWithin-run %CVSDBetween-run %CVSDBetween-day %CVSDBetween-lot %CV
Acceptance criteria:<10N/A2<10
125130677.310.8------9.313.80.030.03
20828925313.65.40.00.020.48.19.33.7
3152601609.35.84.12.518.111.33.62.3
441000---------------------------
509210129010.73.75.41.921.47.411.43.9
6251302297.93.5------14.46.38.13.5
7152601158.87.620.417.765.957.10.00.0
81525423114.76.412.15.39.24.07.93.4
92513016011.27.0------11.67.30.00.0

1 0 = healthy, no treatment; 1 = dual therapy with clopidogrel plus aspirin; 2 = healthy, given clopidogrel Key: 75 mg/day: 2 not applicable as acceptance criteria for this parameter is undefined in the text; 3 between-lot SD of 0 and %CV of 0 can be observed when the contribution of one source of variability is very small relative to other sources of variability, or if all of the sample elements are the same; " Donor 4 had insufficient blood volume to allow for inclusion in the multiple instrument experiment

Open Pouch Device Stability

Extended Open Pouch Stability at Elevated Humidity was conducted to evaluate the stability of the reagents in the PRUTest test device when exposed to environmental humidity. The results of the study are detailed below.

{10}------------------------------------------------

TimeunpouchedPRU whole bloodPRU WQC Level 2
Mean (n=3)Pct. recoveryAllowable range 230–362
0274100.0286
10277101.1254
14Attention 32*N/A3
18Attention 32*N/A7
22Attention 32*N/A18
26Attention 32*N/A21

Table J. Open Pouch Device Stability

  • Attention 32 is generated because the initial DC value is too high, and the VerifyNow system recognizes that either the blood sample and reagent is too clear, or the reagent pellet is not performing as intended

Percent recovery of VerifyNow PRUTest values in whole blood met the acceptance criteria after exposure to 61-73% RH and 21-24°C temperatures for 10 hours. PRU result with WQC Level 2 remained within the allowable range for 10 hours under the same conditions. These findings support a claim of 10 hour open-pouch stability.

Sample Wait Time and Sample Stability

A sample stability and sample wait time study was conducted to determine the appropriate values. Sample wait time was determined to be 10 minutes, meaning that a blood sample must be stored for a minimum of 10 min after collection by venipuncture. Sample stability was determined to be four hours, meaning that the blood sample must be assayed prior to four hours from the time of blood draw.

Interfering substances

An "interference screen," was performed which followed the recommendations outlined in Clinical Laboratory Standards Institute (CLSI) guidance EP07-A2, "Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition." Compounds were tested at a worstcase concentration - at least three-fold greater than therapeutics concentrations. Samples from 11 different donors were assayed in singlicate across eight VerifyNow PRUTest instruments with one device lot of VerifyNow PRUTest using a randomized scheme. Interference testing was performed with six donors using native blood samples and blood samples spiked with 2-MeSAMP, a P2Y12 receptor inhibitor.

2-MeSAMP not added2-MeSAMP added
MeanPRUMeanPRUPct.MeanPRUMeanPRUPct.
Interferent / Test concentrationControlTestRecoveryControlTestRecovery
A3P5PS2100 µM274.3265.096.696.897.0100.3
Acetaminophen1.32 mM274.3263.796.296.899 1102.5
Betamethasone64 uM274.3270.598.696.895.198.3
Caffeine306 uM274.3264.896.596.899.5102.8

Table K. Interfering Substance

{11}------------------------------------------------

Captopril23 μΜ274.3262.995.996.8100.0103.4
Catechin86 μΜ274.3273.199.696.899.0102.3
Celecoxib8.5 µg/mL318.3319.198.0129.9128.098.6
Cilostazol60 μΜ328.8326.699.4139.6141.8101.5
Cimetidine79 μΜ328.8315.395.9139.6140.9100.9
DMSOb0.11%318.3322.198.9129.9131.9101.5
Dipyridamole20 μΜ328.8319.997.3139.6146.8105.1
Diltiazem15 μΜ328.8321.197.7139.6139.9100.2
Ethanol87 mM328.8321.197.3139.6141.5101.3
Fish oil32 mg/mL328.8317.196.5139.6142.0101.7
Glucosamine HCl9.4 μΜ201.4205.8102.2139.4137.298.4
Heparin, LMWc1833 U/L201.4208.6103.6139.4141.8101.7
Hydrochlorothiazide20 μΜ201.4204.3101.4139.4138.599.4
Ibuprofen2.4 mM201.4200.099.3139.4137.498.6
Insulin3 ng/mL201.4208.9103.7139.4134.996.8
Lidocaine51 μΜ201.4206.8102.7139.4137.599.6
Nitroglycerin0.1 µg/mL249.5241.096.6129.8132.1101.8
Norfluoxitine7.27 μΜ249.5241.496.7129.8132.4102.0
Norverapamil4.5 μΜ249.5243.397.5129.8133.8103.1
Omeprazole4.5 μΜ269.1265.698.8142.0143.9101.3
Oxypurinol99 μΜ249.5242.097.0129.8131.3101.2
Pravastatin56 μΜ249.5249.199.9129.8129.9100.1
Propranolol7.7 μΜ249.5242.897.3129.8135.6104.5
Salicylic acid4.3 mM249.5241.196.6129.8128.999.3
Streptokinase400 U/mL325.6317.697.5129.9132.6102.1
Theophylline220 μΜ325.6318.397.8129.9131.4101.2
L-Thyroxine32 nM201.4198.598.6139.4138.499.3
α-Tocopherol58 μΜ318.3311.196.5129.9133.1102.5
Triglycerides37 mM318.3315.997.0129.9129.399.5
Warfarin32 μΜ318.3319.198.0129.9129.499.6

ª Putative thiamine-phosphate pyrophosphorylase

b DMSO = dimethyl sulfoxide c LMW = low molecular weight

Each interferent spiked into whole blood samples at the above concentrations showed percent recovery which deviated less than 10% from baseline PRUTest values. Therefore, none of these substances significantly interfere with VerifyNow PRUTest results.

Stability

Wet Quality Controls (WQC)

Device stability of the VerifyNow PRUTest reagent device has been evaluated over nine months. Three lots of WQC level 1, three lots of WQC Level 2 and three lots of reagent devices are stored at -70°C and at 20±5 °C. Each of the three lots of WQC levels 1 and 2 were tested with the three lots of reagent devices at time zero, 3, 6, and 9 months under both storage conditions.

{12}------------------------------------------------

WQC Level 1

WQC Level 1 simulates a patient sample with highly inhibited platelets and does not produce a significant aggregation profile. The CVs for Level 1 WOC are not calculated because the nominal PRU value is zero. The acceptance criterion for stability of WOC Level 1 was a PRU result < 30. At the nine-month time point, the range of PRU values generated with WOC Level 1 did not exceed 30 PRU.

WQC Level 2

The acceptance criterion for stability for WQC Level 2 was percent recovery ≥ 90 to ≤ 110% of the baseline PRU result. At the nine-month time point, percent recovery of WQC Level 2 passed acceptance criterion for stability when compared to time zero.

Isochronous reagent stability study with donor whole blood samples

Each of three reagent lots was tested at time zero, 3, 6, and 9 months using whole blood samples from normal donors. The acceptance criterion for stability for whole blood samples was percent recovery ≥ 90 to ≤ 110% of the baseline PRU result. At the nine-month time point, percent recovery of PRUTest results in whole blood passed acceptance criterion for stability when compared to time zero. This further supports the nine month stability claim for VerifyNow™ PRUTest reagent storage at 20±5 ℃.

Conclusion

The results of this study indicate that the PRUTest device is stable for up to nine months when stored at ambient temperature (20°C ± 5°C).

Table L. WQC Level 1 Results: 3, 6, and 9 Month Stability

20 ± 5°C Condition
Time Zero3 Months6 Months9 Months
NPRU RangeNPRU RangeNPRU RangeNPRU Range
3241 - 51081 - 31080 - 41080 - 3

At all test intervals, all Level 1 WQC results met the acceptance criteria of ≤30.

{13}------------------------------------------------

3 Months6 Months9 Months
NMeanSDNMeanSD%Differencefrom -70°CConditionNMeanSDNMeanSD%Differencefrom -70°CConditionNMeanSDNMeanSD%Differencefrom70°CCondition
WQCPRUTestLot 1629611.662896.0-2.1629712.162918.2-2.062864.9628713.50.3
Lot 1PRUTestLot 262819.0627614.2-2.0628017.9628810.43.2628513.062737.2-4.2
PRUTestLot 362713.8626511.8-2.0626511.9628110.36.2627618.0627013.2-2.1
WQCPRUTestLot 162857.762937.02.8628722.8629110.31.162879.262779.3-3.5
Lot 2PRUTestLot 262786.562707.8-3.0629211.662808.4-4.0628212.0626811.1-4.9
PRUTestLot 362727.962627.2-3.6628011.8626610.0-4.962748.6626311.8-4.0
WQCPRUTestLot 1627611.262885.54.3629122.6628516.1-2.2629511.562854.6-3.4
Lot 3PRUTestLot 262719.862666.9-1.7629212.6628620.6-2.0627812.0628413.42.0
PRUTestLot 3626911.962639.8-2.262678.7627023.31.462644.862725.63.0
Aggregated5427811.85427514.4-1.15428318.05428215.5-0.55428113.45427512.5-1.9
Table M. WQC Level 2 Results (Comparison to Control [-70°C]): 3, 6, and 9 months
--------------------------------------------------------------------------------------

The table above shows Level 2 WQC Results from the three, six, and nine month time point results using control PRUTest devices stored at ≤ -70 ℃ compared to those stored at 20 ± 5° C. All lots fell within ± 10% of the control condition (≤ -70 °C)

{14}------------------------------------------------

9 Months
-70 C Condition20 ± 5°C Condition
NPRU MeanSDNPRU MeanSD% Difference From -70 C Condition
Donor 1PRUTest Lot 1325412.1323523.4-7.6
Donor 1PRUTest Lot 2326214.6323922.0-8.9
Donor 1PRUTest Lot 332564.2327465.07.2
Donor 1Aggregated925710.4924940.8-3.2
Donor 2PRUTest Lot 131845.831956.96.2
Donor 2PRUTest Lot 231969.931934.2-1.5
Donor 2PRUTest Lot 3318410.0321345.715.7
Donor 2Aggregated91889.6920025.26.6

Table N. Stability at nine-month time point in whole blood compared to -70°C Condition

Whole blood PRU results are shown above Table . Donors were tested at the nine month time point with PRUTest devices stored at 20 ± 5°C and compared to devices stored at ≤ -70°C. For all three PRUTest device lots and both donors the aggregated lot data at (20 ± 5°C) fell within ± 10% of the control condition (≤ -70 ℃).

Table O. Stability at nine-month time point in whole blood inhibited with 2-MeSAMP compared to newly manufactured devices

RecentlyManufactured Lot20 ± 5°C Condition
NPRUMeanSDNPRUMeanSD% Difference FromRecentlyManufactured Lot
Donor399RecentlyManufacturedLot (Control)514013.8514011.9-0.4
512513.0-10.7
51369.0-2.7
Aggregated1513412.4-4.6
Donor1077RecentlyManufacturedLot (Control)513020.3514010.37.4
513027.30.3
514416.82.4
Aggregated1513818.96.1

{15}------------------------------------------------

Inhibited whole blood PRU results are shown above Table. Donors were tested at the nine month time point with PRUTest devices stored at 20 ± 5℃ and compared to newly manufactured devices. For all three PRUTest device lots and both donors the aggregated lot data at (20 ± 5°C) fell within ± 10% of the control condition.

Conclusions drawn from the nonclinical and clinical performance studies:

The results of the bench and clinical performance testing demonstrate that the VerifyNow PRUTest (subject device) is substantially equivalent in performance to the predicate VerifyNow P2Y12 Assay (predicate device).

§ 864.5700 Automated platelet aggregation system.

(a)
Identification. An automated platelet aggregation system is a device used to determine changes in platelet shape and platelet aggregation following the addition of an aggregating reagent to a platelet-rich plasma.(b)
Classification. Class II (performance standards).