K Number
K141279

Validate with FDA (Live)

Date Cleared
2014-12-03

(201 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Nitrile films form a barrier to body fluids and bloodborne Pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Nitrile Powder Free Patient Examination Gloves, Blue Color:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 6319-10Meets
Physical PropertiesASTM standard D 6319-10Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 6319-10 and D6124-06(Reapproved 2011)Meets (<2mg/glove)
Biocompatibility: Primary Skin IrritationISO 10993-10: Third Edition 2010-08-01 (not a primary skin irritation)Passes (Not a Primary Skin Irritation)
Biocompatibility: Dermal sensitizationISO 10993-10: Third Edition 2010-08-01 (not a dermal sensitization)Passes (Not a Dermal sensitization)
Dimensions - LengthMeets ASTM D6319-10 (≥230mm min)230mm min for all sizes
Dimensions - WidthMeets ASTM D6319-10 (Small 70-90mm, Medium 85-105mm, Large 100-120mm, Xlarge 110-130mm)Small 83-86mm, Medium 92-98mm, Large 105-109mm, X large 113-119mm
Dimensions - ThicknessMeets ASTM D6319-10 (Finger 0.05mm min., Palm 0.05mm min.)Finger 0.10 mm min., Palm 0.06 mm min.
Physical Properties (specific values)Meets ASTM D D6319-10 (Before aging/after aging: Elongation ≥500%, Tensile Strength ≥ 14MPa)Before aging/after aging: Elongation: 520-580%, Tensile Strength: 22-34 MPa
Freedom from Pinholes (specifics)Meets • 21 CFR 800.20 • ASTM D6319-10 • ASTM D 5151-06 (Reapproved 2011)Meets ASTM D5151-06 (Reapproved 2011) Holes at Inspection Level I AQL2.5
Residual Powder (specifics)Meets ASTM D 6124-06 (Reapproved 2011), below 2mg of residual powderMeets ASTM D 6124-06 (Reapproved 2011), Results generated values below 2mg of residual powder

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for each specific test (e.g., number of gloves tested for dimensions, pinholes, etc.). It generally states that the device "meets" the standards, implying that samples were taken and tested according to the specified ASTM or ISO standards.

The data provenance is from China, as the submitter, Wanghui Glove Co., Ltd., is located in Hebei Province, China. The studies are retrospective in the sense that they are reporting results from tests already conducted to demonstrate compliance with established standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not mention the use of "experts" in the context of establishing ground truth for individual device performance tests. Ground truth for these types of devices is established by conforming to pre-defined, objective, and standardized measurements and test methods outlined in the ASTM and ISO standards themselves. For example, a pinhole test has a clear, objective method for determining the presence or absence of a pinhole (e.g., water leak test).

4. Adjudication Method for the Test Set

No adjudication method is mentioned because the tests are objective and based on established industry standards. There is no subjective human interpretation of results that would require adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size.

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic devices where human interpretation of medical images or data is involved and the AI's impact on reader performance (e.g., radiologists) is being evaluated. This device is a patient examination glove, and its performance is evaluated against physical and chemical standards, not human diagnostic interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, the device's performance was evaluated in a standalone manner. The tests described (dimensions, physical properties, freedom from pinholes, powder residual, biocompatibility) are all performed directly on the glove material or a batch of gloves. There is no "human-in-the-loop" component to these performance assessments; they are objective measurements against set standards.

7. The Type of Ground Truth Used

The ground truth used for this device is based on established industry standards and objective measurements. Specifically:

  • ASTM D6319-10: Standard Specification for Nitrile Examination Gloves for Medical Application
  • ASTM D6124-06 (Reapproved 2011): Standard Test Method for Residual Powder on Medical Gloves
  • 21 CFR 800.20: General requirements for patient examination gloves (specifically referenced for freedom from pinholes)
  • ASTM D5151-06 (Reapproved 2011): Standard Test Method for Detection of Holes in Medical Gloves
  • ISO 10993-10: Third Edition 2010-08-01: Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization

These standards define the acceptable ranges or thresholds for various physical and biological properties.

8. The Sample Size for the Training Set

This information is not applicable and therefore not provided in the document. This device is a physical product (a glove) evaluated against performance standards, not an AI/ML algorithm that requires a "training set" for model development.

9. How the Ground Truth for the Training Set was Established

This information is not applicable as there is no "training set" for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection. The profiles are rendered in a simple, line-art style.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 3, 2014

Wanghui Glove Co., Ltd. C/O Mr. Chu Xiaoan Official Correspondent Rm. 1606, Bldg.1 Jianxiang Yuan No.209 Bei Si Huan Zhong Road Haidian District Beijing, 100083 CHINA

Re: K141279 Trade/Device Name: Nitrile Powder Free Examination Gloves, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient examination glove Regulatory Class: I Product Code: LZA Dated: October 28, 2014 Received: October 30, 2014

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141279

Device Name

Nitrile Powder Free Patient Examination Gloves, Blue Color

Indications for Use (Describe)

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 6 510(k) Summary

510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: K141279 ..

Premarket Notification [510(k)] Summary

1.0 Submitter:

Submitter's name :Wanghui Glove Co., Ltd.
Submitter's address :Cuizhuang, Pachigang Town, LuannanCounty, Hebei Province, 063502, China
Phone number :(86) 315-4169201
Fax number :(86) 315-4430333
Name of contact person:Zhang Liang
Date of preparation :2014-10-28

2.0 Name of the Device

Device Name:Nitrile Powder Free Patient ExaminationGloves, Blue Color
Proprietary/Trade name:Nitrile Powder Free Patient ExaminationGloves, Blue Color
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LZA

3.0 Predicate device

Device Name:Nitrile Powder Free Patient Examination Gloves,Blue Color
Company name:Tangshan Zhonghong Pulin Plastic Co.,Ltd.
510(K) Number:K120970

4.0 Device Description:

4.1 How the device functions:

Nitrile films form a barrier to body fluids and bloodborne Pathogens

4.2 Scientific concepts that form the basis for the device

The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

{4}------------------------------------------------

4.3 Physical and performance characteristics such as design, materials and physical properties:

Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

5.0 Device Intended Use (Indication for use):

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6.0 Summary of the Technological Characteristics of the Device:

The Nitrile Powder Free Patient Examination Gloves, Blue Color, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 6319-10.Meets
Physical PropertiesASTM standard D 6319-10.Meets
Freedomfrompinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 6319-10 andD6124-06(Reapproved 2011).Meets<2mg/glove
BiocompatibilityPrimary Skin Irritation in rabbitsISO 10993-10: Third Edition2010-08-01.Dermal sensitization in theguinea pig ISO 10993-10: ThirdEdition 2010-08-01.PassesNot a Primary SkinIrritationPassesNot a Dermalsensitization

7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:

Nitrile Powder Free Patient Examination Gloves, Blue Color, meet requirements per ASTM D6319-10.per ASTM D6124-06(Reapproved 2011), per 21 CFR 800.20 and ISO 10993-10: Third Edition 2010-08-01.

The performance test data of the non clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

Substantial Equivalent Based on Assessment of Clinical Performance Data: 8.0

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

Features &DescriptionPredicate DeviceSubject DeviceResultComparison
CompanyTangshan Zhonghong PulinPlastic Co., Ltd.Wanghui Glove Co., Ltd.--
510(K) NumberK120970K141279

9.0 Substantial Equivalence Comparison:

{5}------------------------------------------------

Product nameNitrile Powder Free Patient Examination Gloves, Blue ColorNitrile Powder Free Patient Examination Gloves, Blue Colorsame
Product CodeLZALZAsame
SizeSmall/ Medium/ Large/X largeSmall/ Medium/ Large/X largesame
Intend for useNitrile Powder Free Patient Examination Gloves, Blue Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.Nitrile Powder Free Patient Examination Gloves, Blue Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.Substantially equivalent
Device Description and SpecificationsMeets ASTM D6319-10Meets ASTM D6319-10Substantially equivalent
Dimensions --LengthMeets ASTM D6319-10≥230mm min230mm min for all sizesSubstantially equivalent
Dimensions -- WidthMeets ASTM D6319-10Small 70-90 mmMedium 85-105mmLarge 100-120mmXlarge 110-130 mmSmall 83-86 mmMedium 92-98mmLarge 105-109mmX large 113-119 mmSubstantially equivalent
Dimensions --ThicknessMeets ASTM D6319-10Finger 0.05mm min.Palm 0.05mm min.Thickness (mm) min.Finger 0.10 mm min.Palm 0.06 mm min.Substantially equivalent
Physical PropertiesMeets ASTM D D6319-10Before aging/after agingElongation ≥500%Tensile Strength≥ 14MPaBefore aging/after agingElongation :520-580%Tensile Strength:22-34 MPaSubstantially equivalent
Freedom from PinholesMeets• 21 CFR 800.20• ASTM D6319-10• ASTM D 5151-06(Reapproved 2011)Meets ASTM D5151-06(Reapproved 2011)Holes at Inspection Level I AQL2.5Substantially equivalent
Residual PowderMeets ASTM D 6124-06(Reapproved 2011)below 2mg of residual powderMeets ASTM D 6124-06(Reapproved 2011)Results generated values below 2mg of residual powderSubstantially equivalent
Materials used to fabricate the devicesNitrileNitrileSubstantially equivalent
Dusting or Donning Powder:PUPUSubstantially equivalent
Dusting or Donning Powder: namePUSurface Coating AgentSubstantially equivalent
Compare performanceMeets ASTM D5151-06Meets ASTM D5151-06Substantially equivalent
data supportingsubstantialequivalence(Reapproved 2011)ASTM D6319-10ASTM D6124-06(Reapproved 2011)(Reapproved 2011)ASTM D6319-10ASTM D6124-06(Reapproved 2011)
Single PatientUseSingle Patient UseSingle Patient UseSubstantiallyequivalent
BiocompatibilityUnder the conditions of thisstudy, the test article was anon-irritant and non-sensitizer.SKIN IRRITATION DERMALand SENSITIZATIONSTUDIES Meets ISO10993-10:2002/Amd.1:2006Under the conditions of thisstudy, the test article was anon-irritant and non-sensitizer.SKIN IRRITATIONDERMAL andSENSITIZATION STUDIESMeets ISO 10993-10: ThirdEdition 2010-08-01.Substantiallyequivalent
Labeling for thelegally marketeddevice to whichsubstantialequivalence isclaimed.-Powder Free-Patient Examination Glove-Single Use Only- Manufactured For:- Lot-Blue color- Non sterile-Powder Free-Patient Examination Glove-Single Use Only- Manufactured For:- Lot-Blue color- Non sterileSubstantiallyequivalent

{6}------------------------------------------------

Substantial Equivalence Comparison: 10.0

It can be concluded that the Nitrile Powder Free Patient Examination Gloves, Blue Color meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims.

It can be concluded that the Nitrile Powder Free Patient Examination Gloves, Blue Color is as safe, as effective, and performs as well as the predicate device, Nitrile Powder Free Patient Examination Gloves, Blue Color, Tangshan Zhonghong Pulin Plastic Co.,Ltd.. K120970.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.