K Number
K141138

Validate with FDA (Live)

Date Cleared
2014-05-30

(28 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SURFLASH® Safety I.V. Catheter is inserted into the patient's vascular system for short term use (<30 days) to withdraw blood samples, administer fluid intravenously, or monitor blood pressure by attaching a monitoring line. The needle shaft cover and tip shield feature aids in the prevention of needle stick injuries. These cathers may patient population with consideration given to adequacy of vasular anatony and appropriateness for the solution being infused and duration of therapy. The 18 to 22 gauge catheters are suitable for use with power injectors rated for a maximum of 325 psi.

Device Description

The SURFLASH® Safety I.V. Catheter is manually operated and contains a passive safety feature that automatically activates as the needle is withdrawn from the catheter.
The SURFLASH® Safety I.V. Catheter is a device consisting of catheter assembly (catheter, caulking pin, and catheter hub), needle assembly (needle, needle housing, transparent flash chamber (needle connecting part) with filter, and needle protector) and a passive needle-shielding mechanism (shutter and needle shaft cover consisting of inner cylinder, outer cylinder, and junction cylinder).
The devices are an over-the needle, peripheral catheter made of a slender, flexible, radio-opaque plastic with a hub that is inserted into the patient's vascular system for short term (<30 days) use to withdraw blood samples, administer fluids intravenously, or monitor blood pressure by attaching a monitoring line.
The grooved cannula, which allows the clinician to visualize flashback detection through the groove, indicates that there is confirmation of vessel entry. The whole length of the cannula, including the sharp end of the cannula, is covered by the needle shielding mechanism as the needle is withdrawn from catheter's hub to aid in the prevention of needle stick injuries.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the SURFLASH® Safety I.V. Catheter, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The submission declares that the device "complies with the acceptance criteria established based on the predicate" and that "No deviations from ISO standards were identified in the testing to standards." However, specific numerical acceptance criteria (e.g., minimum force for needle breaking shutter) are not explicitly stated in the provided text. The table below lists the performance tests conducted and confirms that the device conformed to the relevant standards or internal requirements.

Performance TestAcceptance Criteria (Implicitly: Conformance to relevant ISO standards or internal requirements)Reported Device Performance
Force to needle breaking shutter (Puncture resistance of needle shield)Required by ISO 23908Conformed to ISO 23908
Tensile strength of needle shaft and needle housing (Break strength of safety mechanism)Required by ISO 23908Conformed to ISO 23908
Force to release shutter and reactive force when the shutter opens to cover the cannula tip (Reaction force generated by the activation mechanism)Required by internal standardConformed to internal standard
Force to detach needle shaft cover from catheter hub (Force to activate safety feature)Required by ISO 23908Conformed to ISO 23908
Simulated maximum pressureRequired by internal standardConformed to internal standard
Burst pressure (whole device)Required by ISO 10555-1Conformed to ISO 10555-1
Collapse (under negative pressure)Required by ISO 10555-1Conformed to ISO 10555-1
Catheter to catheter hub tensile strength (Force at break of catheter / hub)Required by ISO 10555-1Conformed to ISO 10555-1
Strength of union between transparent flash chamber (needle connecting part) and needleRequired by ISO 10555-5Conformed to ISO 10555-5
Needle attachment to catheter (Initial sliding friction)Required by internal standardConformed to internal standard
Detection of flashback at catheter tipRequired by internal standardConformed to internal standard
Detection of flashback at transparent flash chamber (needle connecting part)Required by ISO 10555-5Conformed to ISO 10555-5
Flow rateRequired by ISO 10555-5Conformed to ISO 10555-5
Vent fittingRequired by ISO 10555-5Conformed to ISO 10555-5
Drop testRequired by internal standardConformed to internal standard
Conical fittings of catheter hubRequired by ISO 10555-1Conformed to ISO 10555-1
Pressure monitoringRequired by internal standardConformed to internal standard
Simulated use studyRequired by ISO 23908All safety features activated effectively, no de-activations after discarding. All participants could read IFU and activate safety feature without further explanation.
SurfaceRequired by ISO 10555-1Conformed to ISO 10555-1
Corrosion resistanceRequired by ISO 10555-1Conformed to ISO 10555-1
Radio-detectabilityRequired by ISO 10555-5Conformed to ISO 10555-5
Catheter unitRequired by ISO 10555-5Conformed to ISO 10555-5
MaterialRequired by ISO 10555-5Conformed to ISO 10555-5
Needle pointRequired by ISO 10555-5Conformed to ISO 10555-5
Catheter body tensile strength (Force at break)Required by ISO 10555-1Conformed to ISO 10555-1
Catheter StiffnessRequired by internal standardConformed to internal standard
Catheter ElongationRequired by internal standardConformed to internal standard
Biocompatibility TestsAdherence to ISO 10993-1 and FDA G95-1 guidelinesMaterials are biocompatible throughout shelf life. Biocompatibility established under K122544, no re-testing needed.
CytotoxicityN/ASatisfactory
SensitizationN/ASatisfactory
Intracutaneous reactivityN/ASatisfactory
Systemic toxicity (acute)N/ASatisfactory
PyrogenN/ASatisfactory
GenotoxicityN/ASatisfactory
Implantation / Subchronic toxicityN/ASatisfactory
HemolysisN/ASatisfactory
Physicochemical CharacterizationN/ASatisfactory
SterilizationSterility Assurance Level (SAL) of 10^-6Achieved SAL of 10^-6 using Ethylene Oxide
Risk AnalysisConformity to ISO 14971:2007Determined no new issues of safety or effectiveness.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not explicitly state the numerical sample size for each performance test. It mentions that "Performance testing was conducted to ensure the safety and effectiveness... verify conformity to the applicable part of ISO standards, and demonstrate substantial equivalence to the predicate device." For the "Simulated Use Study," it refers to "study participants" without specifying their number.
  • Data Provenance: The testing was non-clinical performance and biocompatibility testing. The data originates from internal testing conducted by Terumo Corporation's Kofu Factory in Japan, as indicated by the manufacturer's location and the context of the submission. The text implicitly suggests retrospective data collection as these are standardized tests assessing device properties.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This section is not applicable as the studies were non-clinical performance and biocompatibility tests, not studies requiring expert ground truth in a clinical setting (e.g., image interpretation). The "ground truth" for these tests consisted of established ISO standards and internal specifications.
  • For the Simulated Use Study, "healthcare workers who use these types of Safety I.V. Catheters" were involved, but their number and specific qualifications (e.g., years of experience) are not provided.

4. Adjudication Method for the Test Set

  • Not applicable. This was non-clinical testing against established standards and internal requirements, not a study requiring adjudication of expert interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • Not applicable. This submission is for a physical medical device (I.V. catheter) and does not involve AI or human readers for diagnostic interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. This submission is for a physical medical device and does not involve an algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" for the non-clinical performance tests was rooted in established international standards (ISO standards), FDA guidance documents (e.g., for sharps injury prevention features), and internal specifications derived from the predicate device. For the biocompatibility tests, it was adherence to ISO 10993-1 and FDA G95-1 guidelines.

8. The Sample Size for the Training Set

  • Not applicable. This is not a machine learning or AI device that would require a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. This is not a machine learning or AI device that would require a training set.

{0}------------------------------------------------

MAY 3 0 2014

510(k) SUMMARY

A. SUBMITTER INFORMATION (807.92(a)(1))

Prepared by: Phillip Michael Lester · Regulatory Affairs Specialist Terumo Medical Corporation Tel. (410) 392-7321 Fax (410) 398-6079

Prepared for: Owner/Operator Terumo Corporation 44-1, 2-Chome, Hatagaya Shibuya-Ku, Tokyo, Japan 151-0072 Registration Number: 8010026

Manufacturer and Sterilization Facility

Kofu Factory of Terumo Corporation 1727-1, Tsuijiarai, Showa-Cho Nakakoma-Gun, Yamanashi, Japan 409-3853 Registration Number: 9681835

Contact Person: Phillip Michael Lester Regulatory Affairs Specialist Terumo Medical Corporation 950 Elkton Boulevard Elkton, MD 21921 Tel. (410) 392-7321 Fax (410) 398-6079 E-mail: philmichael.lester@terumomedical.com

Date prepared: May 1, 2014

{1}------------------------------------------------

B. DEVICE NAME (807.92(a)(2))

Proprietary Name:SURFLASH® Safety I.V. Catheter
Common Name:Intravascular Catheter
Classification Name:Catheter, Intravascular, Therapeutic, Short-Term Less Than30 Days
Regulation Number:21 CFR 880.5200
Regulatory Class:II
Review Panel:General Hospital
Product Code:FOZ

C. PREDICATE DEVICE (807.92(a)(3))

The legally marketed device to which substantial equivalence is claimed is the current device, SURFLASH® Safety I.V. Catheter (K122544), manufactured by Kofu Factory of Terumo Corporation.

D. REASON FOR 510(K) SUBMISSION

This Special 510(k) is being submitted due to a design modification to the previously cleared device to extend the range of product offering to include 14G and 16G catheters and a new length of 51 mm (2").

E. DEVICE DESCRIPTION (807.92(a)(4))

(SUMMARY OF TECHNOLOGICAL CHARACTERISTICS)

Principle of Operation/Technology

The SURFLASH® Safety I.V. Catheter is manually operated and contains a passive safety feature that automatically activates as the needle is withdrawn from the catheter.

Design/Construction

The SURFLASH® Safety I.V. Catheter is a device consisting of catheter assembly (catheter, caulking pin, and catheter hub), needle assembly (needle, needle housing, transparent flash chamber (needle connecting part) with filter, and needle protector) and a passive needle-shielding mechanism (shutter and needle shaft cover consisting of inner cylinder, outer cylinder, and junction cylinder).

The devices are an over-the needle, peripheral catheter made of a slender, flexible,

{2}------------------------------------------------

radio-opaque plastic with a hub that is inserted into the patient's vascular system for short term (<30 days) use to withdraw blood samples, administer fluids intravenously, or monitor blood pressure by attaching a monitoring line.

The grooved cannula, which allows the clinician to visualize flashback detection through the groove, indicates that there is confirmation of vessel entry. The whole length of the cannula, including the sharp end of the cannula, is covered by the needle shielding mechanism as the needle is withdrawn from catheter's hub to aid in the prevention of needle stick injuries.

Material

The stainless steel cannula is placed in the catheter to maintain rigidity and is withdrawn after the catheter is placed in the vascular system. The catheter that is advanced into the vessel is made of polyurethane.

Specifications

The modified SURFLASH® Safety I.V. Catheter is available in 3 sizes, with combinations of 14 and 16 gauge catheter diameters and 32 and 51 mm (1 1/2' and 2'') lengths.

F. INTENDED USE (807.92(a)(5))

The SURFLASH® Safety I.V. Catheter is inserted into the patient's vascular system for short term use (<30 days) to withdraw blood samples, administer fluid intravenously, or monitor blood pressure by attaching a monitoring line. The needle shaft cover and tip shield feature aids in the prevention of needle stick injuries. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy. The 18 to 22 gauge catheters are suitable for use with power injectors rated for a maximum of 325 psi.

Note: This intended use is identical to the predicate device, K122544.

G. TECHNOLOGICAL CHARACTERISTIC COMPARISON (807.92(a)(6))

The SURFLASH® Safety I.V. Catheter, subject of this Special 510(k), has the same technological characteristics as the SURFLASH® Safety I.V. Catheter (K122544) in

{3}------------------------------------------------

terms of operation principle, design/construction, material and specifications.

A comparison of the technological characteristics is summarized on the table below.

Comparison of the technological characteristics to predicate device
Trade NameModifiedSURFLASH® Safety I.V. Catheter(subject of this 510(k))CurrentSURFLASH® Safety I.V. Catheter(K122544)
ManufacturerTerumo CorporationTerumo Corporation
OperationprincipleConduit as fluid pathway.Operated manually.Conduit as fluid pathway.Operated manually.
Design /ConstructionIdentical to the current K122544a. Catheter assembly (Catheter*, caulking pin* and catheter hub*)b. Needle assembly (Needle*, Needle housing and Transparent flash chamber (needle connecting part))c. Needle shield assembly (Shutter, Inner cylinder, Outer cylinder, Junction cylinder)d. Filter (Filter cap not available)e. Needle protector* Contacting patient body directly or indirectlya. Catheter assembly (Catheter*, caulking pin* and catheter hub*)b. Needle assembly (Needle*, Needle housing and Transparent flash chamber (needle connecting part))c. Needle shield assembly (Shutter, Inner cylinder, Outer cylinder, Junction cylinder)d. Filter (Filter cap not available)e. Needle protector* Contacting patient body directly or indirectly
Needle Shape /MaterialIdentical to the current K122544Notched cannula. Ditch on the cannula surface from the heel of the bevel to rear of the catheter tip.Stainless steelNotched cannula. Ditch on the cannula surface from the heel of the bevel to rear of the catheter tip.Stainless steel
FlashbackIdentical to the current K122544When catheter and needle are properly placed inside vessel, flashback can be confirmed after blood flows through ditch on the cannula surface.When catheter and needle are properly placed inside vessel, flashback can be confirmed after blood flows through ditch on the cannula surface.
Cathetermaterial.Identical to the current K122544PolyurethaneBiocompatibility of material confirmed.PolyurethaneBiocompatibility of material confirmed.
RadiopaquemediumIdentical to the current K122544Yes (Barium sulfate)Yes (Barium sulfate)
Comparison of the technological characteristics to predicate device
Trade NameModifiedSURFLASH® Safety I.V. Catheter(subject of this 510(k))CurrentSURFLASH® Safety I.V. Catheter(K122544)
Safetymechanism* Passive needle shielding* A needle shield passively covers theinner needle when the needle iswithdrawn from the catheter.* Whole length cannula including cannulatip is shielded.* Passive needle shielding* A needle shield passively covers theinner needle when the needle iswithdrawn from the catheter.* Whole length cannula including cannulatip is shielded.
Color code ofcatheter hubIdentical to the current K122544In accordance with ISO 10555-5In accordance with ISO 10555-5
PackageIdentical to the current K122544Sterility barrier: Blister package,Cast Polypropylene(CPP)/Polyethylene(PE)two layer laminated film and paperUnit boxCardboardSterility barrier: Blister packageCast Polypropylene(CPP)/Polyethylene(PE)two layer laminated film and paperUnit boxCardboard
SterilizationmethodIdentical to the current K122544Ethylene oxideEthylene oxide
Range of SizesIdentical to the current K122540014 and 16G, 1 ¼"-2"18-24G, ¾" - 1 ¼"

{4}------------------------------------------------

H. NON CLINICAL TESTS (807.92(b)(1))

Performance

Performance testing was conducted to ensure the safety and effectiveness of the SURFLASH® Safety I.V. Catheter throughout the shelf life, verify conformity to the applicable part of ISO standards, and demonstrate substantial equivalence to the predicate device. Additionally, performance testing other than to the ISO Standards was performed on the device to verify the modified device against the current. The device complies with the acceptance criteria established based on the predicate.

No new issues of safety and effectiveness were raised with the testing performed. Performance testing demonstrates that the SURFLASH® Safety I.V. Catheter conforms to the ISO standards, is substantially equivalent to the predicate device, and is acceptable for clinical use throughout the shelf life.

The following performance tests were performed on the device:

{5}------------------------------------------------

.

Performance testTestingTesting by internalstandard
ISO 23908ISO 10555-1ISO 10555-5
Force to needle breaking shutter(Puncture resistance of needle shield)XNot requiredNot requiredRequired by ISO
Tensile strength of needle shaft andneedle housing (Break strength of safetymechanism)XNot requiredNot requiredRequired by ISO
Force to release shutter and reactiveforce when the shutter opens to coverthe cannula tip (Reaction forcegenerated by the activation mechanism)Not requiredNot requiredNot requiredX
Force to detach needle shaft cover fromcatheter hub (Force to activate safetyfeature)XNot requiredNot requiredRequired by ISO
Simulated maximum pressureNot requiredNot requiredNot requiredX
Burst pressure ( whole device)Not requiredXNot requiredRequired by ISO
Collapse (under negative pressure)Not requiredXNot requiredRequired by ISO
Catheter to catheter hub tensile strength(Force at break of catheter / hub)Not requiredXNot requiredRequired by ISO
Strength of union between transparentflash chamber (needle connecting part)and needleNot requiredNot requiredXRequired by ISO
Needle attachment to catheter (Initialsliding friction)Not requiredNot requiredNot requiredX
Detection of flashback at catheter tipNot requiredNot requiredNot requiredX
Detection of flashback at transparentflash chamber (needle connecting part)Not requiredNot requiredXRequired by ISO
Flow rateNot requiredNot requiredXRequired by ISO
Vent fittingNot requiredNot requiredXRequired by ISO
Drop testNot requiredNot requiredNot requiredX
Conical fittings of catheter hubNot requiredXNot requiredRequired by ISO
Pressure monitoringNot requiredNot requiredNot requiredX
Simulated use studyXNot requiredNot requiredRequired by ISO
SurfaceNot requiredXNot requiredRequired by ISO
Corrosion resistanceNot requiredXNot requiredRequired by ISO
Radio-detectatbilityNot requiredNot requiredXRequired by ISO
Catheter unitNot requiredNot requiredXRequired by ISO
MaterialNot requiredNot requiredXRequired by ISO
Needle pointNot requiredNot requiredXRequired by ISO
Catheter body tensile strength (Force atbreak)Not requiredXNot requiredRequired by ISO
Catheter StiffnessNot requiredNot requiredNot requiredX
Catheter ElongationNot requiredNot requiredNot requiredX

.

X: Confirmed the conformance to the standard.

{6}------------------------------------------------

No deviations from ISO standards were identified in the testing to standards.

Simulated Use Study

An additional simulated use study was conducted in accordance with FDA's Guidance for Industry and Staff, Medical Devices with Sharps Injury Prevention Features, issued on August 9, 2005 and ISO 23908: 2011, Sharps injury protection-Requirements and test methods-Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling. The objectives of this study were designed to confirm that study participants are able to read and follow the Instructions for Use to safely and effectively operate the safety feature of the modified device. The objectives were:

    1. Verify that proper activation of the sharps injury prevention feature of the device can be accomplished in simulated clinical environments by healthcare workers who use these types of Safety I.V. Catheters.
    1. Determine if the Instructions for Use are adequate for proper activation of the sharps injury prevention feature.

All safety features activated effectively and there were no de-activations after discarding. All participants were able to read and activate the catheter safety feature without further explanation or training.

Conclusion: The safety feature of the device operates safely and effectively as intended. Users with various levels of experience can read the IFU and operate the device safely and effectively.

Biocompatibility

In accordance with ISO 10993-1, the SURFLASH® Safety I.V. Catheter is classified as:

  • Catheter and lubricant: Externally Communicating Device, Circulating Blood, . Prolonged Exposure (24 hours to 30 days).
  • All other materials contacting with patient's body: Externally Communicating . Device, Blood Path Indirect, Prolonged Exposure (24 hours to 30 days).

As mentioned in the below table, the device's materials contacting with the patient's

{7}------------------------------------------------

body were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO - 10993, Biological Evaluation of Medical Devices part 1: Evaluation and Testing.

Screening tests were performed on accelerated aged whole devices to show that the biocompatibility is maintained throughout the shelf life of the product. Results of the testing demonstrate that the materials contacting with patient's body are biocompatible throughout the shelf life of the product.

Test
Cytotoxicity
Sensitization
Intracutaneous reactivity
Systemic toxicity (acute)
Pyrogen
Genotoxicity
Implantation / Subchronic toxicity
Hemolysis
Physicochemical
Characterization

Biocompatibility of the device and materials was established under K122544 for the current SURFLASH® Safety I.V. Catheter, manufactured by Kofu Factory of Terumo Corporation. None of the data raises any new issues of safety and effectiveness.

No changes have been made since 510(k) clearance that would require re-testing; therefore, biocompatibility established under K122544 is still applicable.

Sterilization

The sterility of the device is assured using a sterilization method validated in accordance with Method C: Half-cycle method in Annex B of EN ISO11135-1: 2007 Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. The SURFLASH® Safety I.V. Catheter is sterilized to provide a Sterility Assurance Level (SAL) of 10-0.

{8}------------------------------------------------

Risk Analysis

A Product Risk Analysis was conducted in accordance with ISO 14971: 2007, taking into account the modifications to the previous device, and it was determined that there were no new issues of safety or effectiveness.

I. CLINICAL TESTS (807.92(b)(2))

This 510(k) does not include data from clinical tests.

J. CONCLUSION (807.92(b)(3))

In summary, the SURFLASH® Safety I.V. Catheter, subject of this Special 510(k), is substantially equivalent in its intended use, technology / principal of operation, materials, and performance to the current device:

SURFLASH® Safety I.V. Catheter (K122544), manufactured by Kofu Factory of Terumo Corporation, Japan.

There is no significant difference that raises any new issues of safety and effectiveness.

{9}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609

Silver Spring, MD 20993-0002

Public Health Service

May 30. 2014

Terumo Medical Corporation Mr. Phillip Lester Regulatory Affairs Specialist 950 Elkton Boulevard Elkton, MD 21921

Re: K141138

Trade/Device Name: SURFLASH® Safety I.V. Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: May 1, 2014 Received: May 2, 2014

Dear Mr. Lester:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warrantics. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{10}------------------------------------------------

Page 2 - Mr. Lester

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

MarySTBunner-S

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K141138

Device Name

SURFLASH® Safety I.V. Catheter

Indications for Use (Describe)

The SURFLASH® Safety I.V. Catheter is inserted into the patient's vascular system for short term use (<30 days) to withdraw blood samples, administer fluid intravenously, or monitor blood pressure by attaching a monitoring line. The needle shaft cover and tip shield feature aids in the prevention of needle stick injuries. These cathers may patient population with consideration given to adequacy of vasular anatony and appropriateness for the solution being infused and duration of therapy. The 18 to 22 gauge catheters are suitable for use with power injectors rated for a maximum of 325 psi.

Type of Use (Select one or both, as applicable)

[ Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017 See PRA Statement on last page.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FOR FOR FDA USE ONLY -. . . . . . ﭘﮍ ﮨﮯ ﮐﮧ ﻣ Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

Digitally signed by Richard C. Chapman -S Date: 2014.05.30 09:43:53 -04'00'

{12}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).