K Number
K140874

Validate with FDA (Live)

Date Cleared
2014-09-02

(151 days)

Product Code
Regulation Number
870.1120
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GH disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

The GH reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

Device Description

Both devices comprise tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb, secured by hook and loop closure, and used to measure the patient's blood pressure. The device tubing is connected to a non-invasive blood pressure measurement system.

AI/ML Overview

This document describes the premarket notification for GH Disposable and GH Reusable Blood Pressure Cuffs. However, it does not contain specific acceptance criteria or an explicit study proving the device meets acceptance criteria in the format requested.

The document states that "Both new devices were designed and manufactured in accordance with the following standards:" and lists several ISO and ANSI/AAMI standards. It concludes that "Based on the comparison of intended use, design, materials and performance, we conclude that the new devices are substantially equivalent to the predicates. The differences between the devices do not raise new questions of safety and effectiveness."

Without the full test report or direct mention of performance metrics in relation to acceptance criteria, I cannot provide the detailed information requested in your table format. The document implies that the device meets the standards cited, which would inherently include performance criteria, but these are not explicitly presented as "acceptance criteria" with "reported device performance."

Therefore, I cannot fulfill all parts of your request. I can, however, extract what information is present regarding testing and standards:

G. Summary of Testing indicates that the devices were designed and manufactured in accordance with the following standards:

  • ANSI/AAMI SPIO, Manual, electronic or automated sphygmomanometers, 2002+AI.:2003+A2:2006+(R)2008: This standard sets requirements for the accuracy and performance of blood pressure measuring devices.
  • ISO 10993-1, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process, 2009: This standard outlines a framework for assessing medical device biocompatibility.
  • ISO 10993-5, Biological evaluation of medical devices- Part 5: Tests for In Vitro cytotoxicity, 2009: This standard specifies test methods for assessing in vitro cytotoxicity of medical devices.
  • ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and delayed-type hypersensitivity, 2002+A1 :2006: This standard describes test methods for assessing skin irritation and sensitization caused by medical devices.

Based on the provided document, the following information is available:

1. A table of acceptance criteria and the reported device performance

  • This information is not explicitly provided in the document. The document states that the devices comply with specific standards (ANSI/AAMI SPIO, ISO 10993 series), implying they meet the performance and safety criteria outlined in those standards. However, the specific acceptance criteria and the measured device performance against those criteria are not detailed.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not provided in the document.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable as this document does not describe a clinical study involving ground truth established by experts for a test set. The testing referenced relates to engineering and biocompatibility standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is a medical device (blood pressure cuff), not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable in the context of diagnostic accuracy. For the device itself, the "ground truth" for compliance testing would be the specifications and requirements defined by the cited standards (e.g., accuracy limits for pressure, chemical composition for biocompatibility tests).

8. The sample size for the training set

  • Not applicable. This is not a machine learning or AI device.

9. How the ground truth for the training set was established

  • Not applicable.

In summary, the provided document is a 510(k) summary indicating substantial equivalence to predicate devices based on design, materials, intended use, and compliance with general medical device standards. It does not provide specific performance data against explicit acceptance criteria in the format requested.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, stacked on top of each other. The profiles are connected by flowing lines that resemble hair or fabric. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 2, 2014

Shanghai GaoHui Rubber and Plastic Company c/o Ms. Rhonda Thompson Alexander Registrar Corp. 144 Research Drive Hampton, VA 23666

Re: K140874

Trade/Device Name: GH Disposable and GH Reusable Blood Pressure Cuffs Regulation Number: 21 CFR 870.1120 Regulatory Class: Class II Product Code: DXQ Dated: July 23, 2014 Received: July 24, 2014

Dear Ms. Alexander:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Rhonda Thompson Alexander

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mitchell Stein

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140874

Device Name GH Reusable Blood Pressure Cuff

The GH blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold or intended for use except as indicated.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K140874

Device Name

GH Disposable Blood Pressure Cuff

Indications for Use (Describe)

The GH Disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold or intended for use except as indicated.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY (21 CFR 807.92)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92

The assigned 510(k) number is:

Premarket Notification [510(k)] Summary

A. General Information
Submitter's Name:Shanghai GaoHui Rubber and Plastic Company
Address:No 381, Hong An Road, Xinnong Zhujing ZhenJinshan District, China 201503
Telephone:+86 21 54360656
Fax Number:+86 21 54357084
Contact Person:Richard Gao, Vice General Manager
Date Prepared:February 12, 2014

B. Device

Trade Name:GH Disposable Blood Pressure CuffGH Reusable Blood Pressure Cuff
Common Name:Blood Pressure Cuff
Product Code:DXQ
Class:Class 2
Regulation Number:21 CFR 870.1120

C. Identification of Legally Marketed Predicate Device

Name:Non-Invasive Blood Pressure Cuff
Manufacture:Caremate Medical Device Co., Ltd.
K Number:K102792
Name:Non-Invasive Blood Pressure Cuff
Manufacture:Unimed Medical Supplies Inc.
K Number:K112544

D. Description of the Device

Both devices comprise tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb, secured by hook and loop closure, and used to measure the patient's blood pressure. The device tubing is connected to a non-invasive blood pressure measurement system.

E. Intended Use

The GH disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for

{5}------------------------------------------------

single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

The GH reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

FeaturesShanghai GaoHui Rubber andPlastic CompanyCaremate Medical Device Co.,Ltd.
GH Disposable Blood Pressure CuffNon-invasive Blood Pressure Cuff
Intended useThe GH disposable blood pressurecuff is an accessory used inconjunction with noninvasive bloodpressure measurement systems.The cuff is non-sterile and forsingle-patient use. It is available inneonatal, pediatric and adult sizes.The cuff is not designed, sold, orintended for use except asindicated.Indirect measurement of bloodpressure
Patient PopulationsAdults/PediatricsAdults/Pediatrics
Tube ConfigurationOne or two tubeOne or two tube
Size (range in cm)Conform to AHA bladder sizesrecommendationsConform to AHA bladder sizesrecommendations
Neonate 1 (3-6)Neonate 1 (3.3-5.6)
Neonate 2 (4-7)Neonate 2 (4.2-7.1)
Neonate 3 (5-11)Neonate 3 (5-10.5)
Neonate 4 (7-12)Neonate 4 (6.9-11.7)
Neonate 5 (9-15)Neonate 5 (8.9-15)
Neonate 6 (8-12)Neonate 6 (7.7-10.5)
Neonate 7 (10-14)Neonate 7 (9.8-13.3)
Neonate 8 (12-17)Neonate 8 (12.4-16.8)
Infant 9(16-22)Infant 9(15.8-21.3)
Child 10(20-27)Child 10(20-27)
Small adult 11(26-35)Small adult 11(25.3-34.3)
Adult 12 (32-43)Adult 12 (32.1-43.4)
Thigh 13 (46-66)Thigh 13 (46-66)
SterilityNot supplied sterileNot supplied sterile
Pressure limits0-300mmHg0-300mmHg
SterilityNon-sterileNon-sterile
BiocompatibilityComply with ISO 10993biocompatibility evaluationComply with ISO 10993biocompatibility evaluation

F. Comparison to the Predicate Device

{6}------------------------------------------------

The GH Disposable Blood Pressure Cuff has the same Intended Use, basic construction, and technology specification as the predicated device. Both devices are wrapped the patient's arm or leg and secured by a hook and loop fastener commonly called Velcro. Both devices are available in the same size and range and are intended for the same patient populations. The materials of both devices are all conformed to ISO 10993. We extend the length of the cuffs in order to accommodate special groups, such as overweight subjects. Based on the performance testing in this submission, the slight difference on the range of these blood pressure cuffs does not raise any safety or effectiveness issue.

FeaturesShanghai GaoHui Rubber and Plastic CompanyGH Reusable Blood Pressure CuffUnimed Medical Supplies Inc.K112544Non-invasive Blood Pressure -Cuff
Intended UseIndirect measurement of blood pressureIndirect measurement of blood pressure
Indications for useThe GH Reusable Blood Pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.The Unimed Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in pediatric and adult sizes. The cuff is not designed, sold or intended for use except as indicated.
Patient PopulationsAdults/PediatricsAdults/Pediatrics
Tube ConfigurationOne or two tubeOne or two tube
Size (range in cm)Conform to AHA bladder sizes recommendationsNeonate 1 (3-6)Neonate 2 (4-7)Neonate 3 (5-11)Neonate 4 (7-12)Neonate 5 (9-15)Neonate 6 (8-12)Neonate 7 (10-14)Neonate 8 (12-17)Infant 9(16-22)Child 10(20-27)Small adult 11(26-35)Adult 12 (32-43)Thigh 13 (46-66)Conform to AHA bladder sizes recommendationsNeonatal (6-11)Infant (10-19)Pediatric(16-26)Small adult(20-28)Adult (25-35)Large Adult (33-47)Thigh (46-66)

{7}------------------------------------------------

Repeated inflation10,000 inflations10,000 inflations
3,000 hook and loop closures3,000 hook and loop closures
Pressure limits0-300mmHg0-300mmHg
SterilityNon-sterileNon-sterile
BiocompatibilityComply with ISO 10993Comply with ISO 10993
biocompatibility evaluationbiocompatibility evaluation
Material of cuffCuff : woven nylon.light and soft, Nontoxic, anti bacterialCuff: Pu Synthetic Leatherlight and soft, Nontoxic, anti bacterial
Material of tubePVC : stable ; corrosion resistance ;heat-resistancePVC : stable ; corrosion resistance ;heat-resistance
Material of BladderTPU : stable ; corrosion resistance ;TPU : stable ; corrosion resistance ;
Material ofhook&loop fastenerNylon: stable,corrosion resistance ;heat-resistance , anti pollution,fastcolorsNylon:stable,corrosion resistance ;heat-resistance , anti pollution,fastcolors

G. Summary of Testing

Both new devices were designed and manufactured in accordance with the following standards:

  • . ANSI/AAMI SPIO, Manual, electronic or automated sphygmomanometers, 2002+AI.:2003+A2:2006+(R)2008
  • ISO 10993-1, Biological evaluation of medical devices -- Part 1: Evaluation and testing . within a risk management process, 2009
  • . ISO 10993-5, Biological evaluation of medical devices- PartS5: Tests for In Vitro cytotoxicity, 2009
  • . ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and delayed-type hypersensitivity, 2002+A1 :2006

H. Conclusion of Substantial Equivalence

Based on the comparison of intended use, design, materials and performance, we conclude that the new devices are substantially equivalent to the predicates. The differences between the devices do not raise new questions of safety and effectiveness.

§ 870.1120 Blood pressure cuff.

(a)
Identification. A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.(b)
Classification. Class II (performance standards).