K Number
K140831

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-04-15

(13 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Pressure Rated: The MaxZero nulti fuse extension set with needleless connector(s) is for single use only. The extension set may be used for direct injection, intermittent infusion or aspiration. This set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10mL per second.

Non Pressure Rated: The MaxZero multi fise extension set with needleless connector(s) is for single use only. The extension set may be used for direct injection, intermittent infusion, continuous infusion or aspiration.

Device Description

The CareFusion MaxZero Extension Sets with Needleless Connector are intravascular extension sets intended for single patient use, including pediatrics and immunocompromised patients, for direct injection, intermittent infusion continuous infusion or aspiration of drugs, blood and fluids. All MaxZero Extension Sets with Needleless Connector include the previously cleared zero reflux MZ1000 Needleless Connector (K132413) bonded to the extension set tubing. The MZ1000 needleless connector allows thorough and easy disinfection due to a solid, flat smooth surface and eliminates the risk of needlestick injuries. The MaxZero Extension Sets with Needleless Connectors are sterile single patient devices that can be used for seven (7) days and 200 activations. All extension sets included in this submission are not made from natural rubber latex or DEHP.

AI/ML Overview

This is a 510(k) premarket notification for a medical device, not a study evaluating an AI algorithm. Therefore, the questions related to AI/algorithm performance, ground truth, expert adjudication, and sample sizes for training/test sets are not applicable.

The document describes the CareFusion MaxZero Extension Sets with Needleless Connector and aims to demonstrate its substantial equivalence to previously marketed predicate devices. The "acceptance criteria" here refer to the performance standards and tests that the new device must meet to prove its safety and effectiveness, similar to the predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance:

The document broadly states that "All test results met their acceptance criteria and support that the MaxZero Extension Sets with Needleless Connector are appropriately designed for their intended use." However, specific quantitative acceptance criteria or detailed reported performance values for each test are not provided in this summary. The summary only lists the types of tests performed.

Test TypeAcceptance Criteria (Not explicitly quantified in summary)Reported Device Performance (Summary statement)
Microbial ingress and barrier testingEnsure effective barrier against microbesMet acceptance criteria
Hemolysis testingNon-hemolyticMet acceptance criteria
Shelf life performance testingMaintain performance over specified shelf life (7 days, 200 activations)Met acceptance criteria
Harsh Infusates testingCompatibility with harsh infusatesMet acceptance criteria
ISO TestingCompliance with relevant ISO standardsMet acceptance criteria
Priming volume/flow rate testingDemonstrate specified priming volume and flow rateMet acceptance criteria
Biocompatibility testingBiocompatible (per ISO 10993)Met acceptance criteria
Sterilization validation testingEffective sterilizationMet acceptance criteria
Functional/Performance TestingEnsure safe and effective operation for intended useMet acceptance criteria

The document states that a "complete list of functional/performance testing protocols and reports are included in Section 18 of this submission," which is not provided here.

2. Sample size used for the test set and the data provenance:
Not applicable for this type of device submission. This is a medical device, not an AI or data-driven software. The "test set" here refers to physical devices undergoing various laboratory and performance tests. The sample sizes for these physical product tests are not specified in this summary. Data provenance is also not relevant in the context of device testing as described.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This is not an AI algorithm study or a study requiring expert consensus for ground truth. The "ground truth" for this device would be established by objective physical and chemical testing against predetermined performance standards.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable. There is no human interpretation or adjudication described for the device performance tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is a medical device, not an AI system being evaluated with human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a medical device, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this device's performance is established by standardized laboratory testing protocols, national/international standards (e.g., ISO), and engineering specifications designed to ensure the device's safety and effectiveness (e.g., microbial barrier effectiveness, material biocompatibility, pressure ratings).

8. The sample size for the training set:
Not applicable. This is a medical device, not an AI algorithm requiring a training set.

9. How the ground truth for the training set was established:
Not applicable. This is a medical device, not an AI algorithm requiring a training set with ground truth.

{0}------------------------------------------------

CAREFUSION

K140831

510(k) Summary

APR 1 5 2014

Submitter Information

A. Company Name:CareFusion
B. Company Address:10020 Pacific Mesa Blvd.San Diego, CA 92121
C. Company Phone:(858) 617-3042
D. Company Fax:(858) 617-5982
E. Contact Person:Larry Walker
F. Date Summary Prepared:March 20, 2014

Device Identification

A. Trade NameMaxZero Extension Sets with Needleless Connector
B. Common Name:IV Administration Sets
C. Classification:IV Administration Set, Needleless Connector,Closed Access, 21 CFR 880.5440, (Product code FPA)

Legally Marketed Predicate Device for Substantial Equivalence

Predicate DeviceManufacturer510(k) #Date Cleared
MZ1000 Needleless ConnectorCareFusionK132413August 29, 2013
Medegen Pressure RatedExtension SetsCareFusion (formerlyMedegen)K083472December 9, 2008
Medegen IntravascularAdministration Set andExtension SetCareFusion (formerlyMedegen)K051499June 22, 2005

Rational for Substantial Equivalence

The information provided in the premarket notification demonstrates that the subject CareFusion MaxZero Extension Set with Needleless Connector is substantially equivalent to the legally marketed predicated devices. The subject MaxZero Extension Sets with Needleless Connector and the predicate Medegen Extension Sets are intended to be used for the delivery and/or aspiration of fluids to/from an IV catheter in a hospital environment. The subject and the predicate devices are similar in physical properties, materials, and configuration. Each device includes connectors that allow for needleless access to the IV line during IV therapy eliminating the risk of needle injury. The subject device incorporates the predicate MZ1000 Needleless Connector bonded directly to IV tubing. Components of the subject devices are made of materials that are substantial equivalent to the predicate devices listed above and this submission includes comprehensive biocompatibility testing for all device materials included in this submission.

{1}------------------------------------------------

Device Description

The CareFusion MaxZero Extension Sets with Needleless Connector are intravascular extension sets intended for single patient use, including pediatrics and immunocompromised patients, for direct injection, intermittent infusion continuous infusion or aspiration of drugs, blood and fluids. All MaxZero Extension Sets with Needleless Connector include the previously cleared zero reflux MZ1000 Needleless Connector (K132413) bonded to the extension set tubing. The MZ1000 needleless connector allows thorough and easy disinfection due to a solid, flat smooth surface and eliminates the risk of needlestick injuries. The MaxZero Extension Sets with Needleless Connectors are sterile single patient devices that can be used for seven (7) days and 200 activations. All extension sets included in this submission are not made from natural rubber latex or DEHP.

ModelNumberDescriptionTubingIDTubingOD
MZ5301Pressure rated extension set, MaxZero connector, slide clamp, spin maleluer lock [≈ 7" (18 cm); ≈ 0.40 ml]0.042"0.079"
MZ5302Pressure rated extension set, MaxZero connector, slide clamp, spin maleluer lock [≈ 7" (18 cm); ≈ 0.40 ml]0.042"0.079"
MZ5303Pressure rated extension set, MaxZero connector, slide clamp, spin maleluer lock [≈ 7" (18 cm); ≈ 0.50 ml]0.060"0.144"
MZ5304Pressure rated extension set, MaxZero connector, slide clamp, spin maleluer lock [≈ 7" (18 cm); ≈ 0.50 ml]0.060"0.144"
MZ5305Pressure rated extension set, MaxZero connector, spin make luer lock [≈7" (18 cm); ≈ 0.50 ml]0.060"0.144"
MZ5306Pressure rated extension set, minibore tubing, MaxZero connector, spinmake luer lock. [≈ 7" (18 cm); ≈ 0.40 ml]0.042"0.079"
MZ5307Bi-fuse pressure rated extension set, minibore tubing, (2) MaxZeroconnectors, (2) side clamps, spin male luer lock. [≈ 7" (18 cm); ≈ 0.80ml]0.042"0.079"
MZ5308Bi-fuse pressure rated extension set, (2) MaxZero connectors, (2) slideclamps, spin make luer lock. [≈ 6" (15 cm); ≈ 0.90 ml]0.060"0.144"
MZ9284Bi-fuse extension set, (2) MaxZero connectors, (2) check valves (2) slideclamps, spin make luer lock [≈ 7" (18 cm); ≈ 0.90 ml]0.042"0.079"
MZ9285Bi-fuse extension set, (2) MaxZero connectors, (2) slide clamps, spinmake luer lock [≈ 7" (18 cm); ≈ 0.90 ml]0.60"0.144"

The following model numbers are subject to this submission:

Intended Use .

Pressure Rated: The MaxZero multi fuse extension set with needleless connector(s) is for single use only. The extension set may be used for direct injection, intermittent infusion, continuous infusion or aspiration. This set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10mL per second.

Non Pressure Rated: The MaxZero multi fuse extension set with needleless connector(s) is for single use only. The extension set may be used for direct injection, intermittent infusion, continuous infusion or aspiration.

{2}------------------------------------------------

Substantial Equivalence Table

DeviceCareFusion MaxZeroExtension Sets withNeedleless ConnectorCareFusion (formerlyMedegen) Pressure RatedExtension SetsCareFusion (formerlyMedegen) IntravascularAdministration Set andExtension Set
510(k) #TBD Subject deviceK083472K051499
Fluid contactingmaterialsTubing- PVCLuer- Co-polyesterNeedleless Connector -Polycarbonate/siliconeMale Spin lock - ABSthermoplasticBifurcated Connecter-PVCTubing- PVCLuer- Co-polyesterCheck Valves- ABSthermoplasticNeedleless Connector -Polycarbonate/siliconeMale Spin lock - ABSthermoplasticBifurcated Connecter- ABSTubing- PVCLuer- Co-polyesterCheck Valves- ABSthermoplasticNeedleless Connector -Polycarbonate/siliconeMale Spin lock - ABSthermoplasticBifurcated Connecter-ABS
NeedlelessConnectorCareFusion MZ1000(K132413)CareFusion MaxPlus Tru-Swab Connector(K072542)CareFusion NAC PlusNeedleless Connector(K011193)
Functional UseDirect Injection,intermittent infusion,continuous infusion,aspirationDirect Injection,intermittent infusion,continuous infusion,aspirationDirect Injection,intermittent infusion,continuous infusion,aspiration
PackagingTyvek/film pouchTyvek/film pouchTyvek/film pouch
Sterilization MethodE-BeamE-BeamIrradiation
Usable Life7 days 200 activations3 days 96 activationsPer CDC guidelines

Technical Characteristics

Technological Characteristics
Zero Reflux Needleless Connector
Designed to prevent microbial ingress
Needleless connector can be disinfected with 3 sec scrub with 70% IPA
Maximum clinical use of 7 days 200 activations (single patient use)
Non-hemolytic
Can be used with low power injectors with maximum of 325 psi & flow rate of10ml/second (pressure rated sets)
Not made with DEHP
Safe for use in MRI environment
Not made with natural latex rubber
Sets can be used with harsh infusates

Clinical Data

There is no clinical data included in this submission.

Non-Clinical Data

CareFusion performed design verification performance testing to verify, demonstrate and support the claim of substantial equivalence to the predicate devices. All test results met their acceptance criteria and support that the MaxZero Extension Sets with Needleless Connector are appropriately designed for their intended use.

Testing performed included:

  • . Microbial ingress and barrier testing

{3}------------------------------------------------

  • . Hemolysis testing
  • . Shelf life performance testing
  • . Harsh Infusates testing
  • . ISO Testing
  • . Priming volume/flow rate testing
  • . Biocompatibility testing to applicable section of ISO 10993
  • . Sterilization validation testing

A complete list of functional/performance testing protocols and reports are included in Section 18 of this submission.

Conclusion

The results of the non-clinical testing exhibited that no new issues of safety and efficacy are raised with the proposed introduction of the MaxZero Extension Sets with Needleless Connector. The device met the acceptance criteria for all functional, microbial ingress, sterility, biocompatibility and other performance criteria, which verify it to be substantially equivalent to the predicate devices. The conclusion drawn from the performance testing demonstrate that the MaxZero Extension Set with Needleless Connector is as safe as effective and performs at least as safe and effectively as the legally marketed predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 15, 2014

CareFusion C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo MN 55313

Re: K140831

Trade/Device Name: MaxZero Extension Sets with Needleless Connector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: March 27, 2014 Received: April 4, 2014

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/8 description: The image contains a logo with the text "Tejashri Purohit-Sheth, M.D." and "Clinical Deputy Director" stacked on top of each other. Below that is the word "DAGRID". The word "FOR" is in the bottom right corner of the image.

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K140831

Device Name

MaxZero Extension Set with Needleless Connector

Indications for Use (Describe)

Pressure Rated: The MaxZero nulti fuse extension set with needleless connector(s) is for single use only. The extension set may be used for direct injection, intermittent infusion or aspiration. This set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10mL per second.

Non Pressure Rated: The MaxZero multi fise extension set with needleless connector(s) is for single use only. The extension set may be used for direct injection, intermittent infusion, continuous infusion or aspiration.

Type of Use (Select one or both, as applicable)

[ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

THE THE FORIEDA USE ONLY AND THE ONLY AND THE ORIENT 1. . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/6/Picture/14 description: The image shows the letters FDA in a stylized, outlined font. The letters are arranged horizontally, with the F on the left, D in the middle, and A on the right. Each letter is composed of multiple parallel lines, giving it a layered or 3D effect. The image has a slightly distressed or textured appearance, as if it were printed on a rough surface.

Digitally signed by Richard C. Chapman Date: 2014.04.15 11:15:24 -04'00'

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of Information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.