K Number
K132413

Validate with FDA (Live)

Device Name
MZ1000
Manufacturer
Date Cleared
2013-08-29

(27 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MZ1000 is a sterile single patient use connector for needleless access to the IV line and/or IV catheter during IV therapy. The MZ1000 can be used for direct injection, intermittent infusion, continuous infusion or aspiration.

Device Description

The MZ1000 is a zero reflux needleless connector intended for single patient use, including pediatrics and immunocompromised patients, for direct injection, intermittent infusion, continuous infusion or aspiration of drugs, blood and fluids which using a vascular access device. The MZ1000 is a closed, luer activated device that eliminates the risk of needlestick injuries. The MZ1000 does not require a specific clamping sequence or technique in order to be used safely. The flush volume for the MZ1000 is five milliliters (5ml). The MZ1000 can be used for seven (7) days and 200 activations.

AI/ML Overview

The MZ1000 is a zero reflux needleless connector intended for single patient use, including pediatrics and immunocompromised patients, for direct injection, intermittent infusion, continuous infusion or aspiration of drugs, blood, and fluids using a vascular access device. The MZ1000 is a closed, luer-activated device that eliminates needlestick injuries and does not require a specific clamping sequence or technique for safe use.

1. Acceptance Criteria and Reported Device Performance

Performance CriteriaAcceptance CriteriaReported Device Performance
Zero refluxThe device must demonstrate zero reflux during use.Met acceptance criteria
Low flush volume (5ml)The device must have a flush volume of 5ml.Met acceptance criteria (flush volume is 5ml)
7 days & 200 activationsThe device must be able to be used for 7 days and withstand 200 activations.Met acceptance criteria
Needleless connectorThe device must function as a needleless connector.Met acceptance criteria
No interstitial or dead space internal to the connectorThe device must not have any internal interstitial or dead space.Met acceptance criteria
Non-hemolyticThe device must be non-hemolytic.Met acceptance criteria
Flow rate of >6L/hourThe device must achieve a flow rate greater than 6 Liters per hour.Met acceptance criteria
325 psi & maximum flow rate of 10ml/secondsThe device must withstand 325 psi and a maximum flow rate of 10ml/second.Met acceptance criteria
Disinfect with 70% IPAThe device must be able to be disinfected with 70% Isopropyl Alcohol (IPA).Met acceptance criteria
Cleanse with 2% chlorhexidine gluconate or iodineThe device must be able to be cleansed with 2% chlorhexidine gluconate or iodine.Met acceptance criteria
Functional, microbial ingress, sterility, biocompatibility, and other performance criteriaAll functional, microbial ingress, sterility, biocompatibility, and other performance criteria outlined in the testing protocol must be met.Met all acceptance criteria

2. Sample Size and Data Provenance

  • Sample size for the test set: Not explicitly stated. The document mentions "non-clinical data (bench testing)" that included functional, microbial ingress, sterility, and other performance criteria.
  • Data provenance: Bench testing. The country of origin of the data is not specified but is implicitly from the manufacturer, CareFusion, based in San Diego, CA, USA. The data is prospective, as it was generated through specific testing for the submission.

3. Number of Experts and Qualifications for Ground Truth

Not applicable. This submission does not involve expert-based ground truth for image or clinical data. The assessment is based on objective performance specifications and laboratory testing of the medical device.

4. Adjudication Method for the Test Set

Not applicable. As the testing involves objective performance criteria for a medical device (e.g., flow rate, pressure, microbial ingress), there is no adjudication method in the context of human expert review. The results are determined by the outcome of bench tests against predefined specifications.

5. Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. The MZ1000 is a medical device (needleless connector), not an AI algorithm for diagnostic interpretation. Therefore, a multi-reader, multi-case comparative effectiveness study of human readers with and without AI assistance is not relevant to this submission.

6. Standalone Performance Study

A standalone performance study (i.e., algorithm only without human-in-the-loop performance) was done for the MZ1000. The "Performance Data" section explicitly states: "The MZ1000 (subject device) met all of the acceptance criteria for all functional, microbial Ingress, sterility, biocompatibility and other performance criteria which verify it to be substantially equivalent to the predicate devices." This refers to the device performing according to its specifications in bench tests, without human intervention or interpretation of results beyond verification of test outcomes.

7. Type of Ground Truth Used

The ground truth used is based on objective scientific, engineering, and microbiological testing standards and specifications. This includes:

  • Functional performance criteria: e.g., flow rate, pressure resistance, flush volume, number of activations.
  • Microbial ingress testing: to assess the barrier integrity against microorganisms.
  • Sterility testing: to confirm the absence of viable microorganisms.
  • Biocompatibility testing: to ensure the device is safe for contact with biological systems.

These are established benchmarks for medical device performance.

8. Sample Size for the Training Set

Not applicable. The MZ1000 is a physical medical device, not an AI or machine learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As mentioned, the MZ1000 is a physical medical device and does not utilize a training set in the context of AI or machine learning.

{0}------------------------------------------------

MZ1000

510(k) Summary

CareFusion

10020 Pacific Mesa Blvd.

San Dlego, CA 92121 (858) 617-5925

(858) 617-5977

April 22, 2013

Michelle J. Badal

SUBMITTER INFORMATION

  • Company Name: A.
  • Company Address: B.
  • Company Phone: C. Company Fax:
  • Contact Person: D.
  • Date Summary Prepared: പ്

DEVICE IDENTIFICATION

  • Trade Name: MZ1000 A.
  • IV Administration Set Common Name: в.
  • Classification: IV Administration Set, Needleless Connector, Closed ੁੱ Access, (21 CFR 880.5440, Product Code FPA)

PREDICATE DEVICES

The MZ1000 is of comparable type and is substantially equivalent to the following predicate devices:

Predicate DeviceManufacturer510(k)No.Date Cleared
MaxPlus Tru-SwabPositive DisplacementConnectorCareFusion (FormerlyMedegen)Қ072542Sept 25, 2007
NeutronICU Medical Inc.K100434August 9,2010
Neutral Luer ActivatedDeviceBaxter HealthcareCorporationK120443May 22, 2012

Confidential

{1}------------------------------------------------

DEVICE DESCRIPTION

The MZ1000 is a zero reflux needleless connector intended for single patient use, including pediatrics and immunocompromised patlents, for direct injection, intermlitent infusion, continuous infusion or aspiration of drugs, blood and fluids which using a vascular access device. The MZ1000 is a closed, luer activated device that eliminates the risk of needlestick injurles. The MZ1000 does not require a specific clamping sequence or technique in order to be used safely. The flush volume for the MZ1000 is five millillters (5ml). The MZ1000 can be used for seven (7) days and 200 activations.

INTENDED USE

The MZ1000 Is a sterile single patient use connector for needleless access to the IV line and/or IV catheter during IV therapy. The MZ1000 can be used for direct injection, intermittent infusion, continuous infusion or aspiration.

TECHNOLOGICAL CHARACTERISTICS

Technological Characteristics
Zero reflux
Low flush volume - 5ml
7 days & 200 activations
Needleless Connector
No interstitial or dead space internal to the connector
Non-hemolytic
Flow rate of >6L/hour
325 psi & maximum flow rate of 10ml/seconds
Disinfect with 70% IPA
Device can be cleansed with 2% chlorhexidine gluconate or
iodine

PERFORMANCE DATA

The MZ1000 (subject device) met all of the acceptance criteria for all functional, microbial Ingress, sterility, blocompatiblity and other performance criteria which verify it to be substantially equivalent to the predicate devices. Results of the testing demonstrate that there are no new issues of safety and efficacy that are raised with the MZ1000.

SUBSTANTIAL EQUIVALENCE

The results of the testing demonstrate that there are no new issues of safety and efficacy that are raised with the introduction of the MZ1000. The results also demonstrate substantial equivalence to the predicate devices, and that they are safe and effective for their intended use.

Confidentlal

028

{2}------------------------------------------------

CLINICAL AND NON-CLINICAL DATA:

There is no clinical data Included in this submission. Non-clinical data (bench testing) is included in the MZ1000 Performance Testion in the submission as simulated use environmental testing. The testing Included functional, microbial ingress, sterility, and other performance criterla to verify the safe and effectlye use of the device. The results of the testing demonstrates that there are no differences between the predicate and the subject device that raise any new Issues of safety or effectiveness

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission of protecting the health of all Americans and providing essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 29, 2013

CareFusion C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25" Street, NW BUFFALO MN 55313

Re: K132413

Trade/Device Name: Intravascular Administration Set, Needleless Connector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: July 31, 2013 Received: August 14, 2013

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to eonine.co proc to this 20, 1977, the accordance with the provisions of the Federal Food, Drug, de nees that have over require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, mererere, mans of the Act include requirements for annual registration, listing of general oblic of overving practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability admontion. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be may be subject to trainiens, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours,

Kwame O. Ulmer -S

Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Carefusion

Indications for Use

510(k) Number (if known): K132413

Device Name: MZ1000

Indications For Use:

The MZ1000 is a sterile single patient use connector for needleless access to the IV line and/or IV catheter during IV therapy. The MZ1000 can be used for direction, intermittent infusion, continuous infusion or aspiration

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Richard C. Chapman 2013.08.28 15:47:16 -04'00'

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.