K Number
K140526

Validate with FDA (Live)

Date Cleared
2014-05-19

(76 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NxStage System One is indicated for the treatment of acute and chronic renal failure, or fluid overload using hemofiltration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility. The NxStage System One is also indicated for hemodialysis with or without ultrafiltration in the home. All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

Device Description

The NxStage System One is comprised of the NxStage Cycler, an electromechanical control unit; the NxStage Cartridge, a sterile, single-use extracorporeal blood and fluid management circuit (with or without a preattached high permeability filter) that mounts integrally within the NxStage Cvcler. The combined system is designed to deliver hemofiltration, hemodialysis and/or ultrafiltration in an acute or chronic care facility. The NxStage System One is also indicated for hemodialysis with or without ultrafiltration in the home.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the NxStage System One, a hemodialysis system. The submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting a novel clinical study with new acceptance criteria and performance data.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document primarily compares technological characteristics to a predicate device (K122051). While it mentions "predetermined acceptance criteria was met" for performance, it does not explicitly list the acceptance criteria or specific performance values for the new device against those criteria in a table format. Instead, it largely states that the proposed device has "Same" characteristics as the predicate or lists the predicate's specifications as a de facto benchmark.

Therefore, a table that perfectly fits the request cannot be generated from the given information because the specific acceptance criteria values for the current submission's device are not enumerated, nor are detailed "reported device performance" values presented for novel criteria. The comparison is mainly a feature-by-feature match to the predicate.

However, based on the provided "Technological Characteristics Comparison Table," we can infer the "acceptance criteria" were essentially "to match or be equivalent to the predicate device's specifications."

ParameterAcceptance Criteria (Inferred from Predicate)Reported Device Performance (Proposed Device)
Intended UseHemodialysis, Hemofiltration, UltrafiltrationYes
Pumps4 peristaltic pumpsSame
Valves (Clamps)2 solenoid actuated pinch clamps, 8 cam driven pinch clampsSame
Air / Fluid Detectors3 ultrasonic air/fluid detectorsSame
Blood Leak Detector1 optical blood leak detectorSame
Pressure Transducers5 electronic pressure transducersSame
Temperature Sensors1 electronic temperature sensorSame
Blood Flow Rate10-600 ml/minSame
Prescription Fluid / Dialysate Flow0-18000 ml/hr (NX1000-3)0-18000 ml/hr (NX1000-4)
Ultrafiltration Flow0-2400 ml/hrSame
Transmembrane Pressure MonitoringYesSame
Venous Pressure Monitor0 to 400 mmHgSame
Effluent fluid Pressure Monitor0 to 500 mmHgSame
Air Detector SensitivityReduction of detector signal lasting 6 ms minimum (Approximates a 60 micro liter bubble at 400 mmHg venous pressure and 600 ml/min blood flow)Same
Blood Leak Detector Sensitivity15% reduction in detector signal lasting 20 seconds minimum. Signal reduction % based on a 0.45 ml/min leak rate of 32 Hct blood.15% reduction in detector signal lasting 20 seconds minimum. Signal reduction % based on a 0.35 ml/min leak rate of 32 Hct blood.
Effluent Volume Accuracy (sw 4.8+)Greater of 300 ml/ 12 hr or 3% of exchange volume; Therapy Fluid Flow Rate L/hr) Specification greater of > 3 ±100 ml/hr* + 5% UF* or ≤ 3 ± 25 ml/hr*Same
IV Prescription FluidOff-line, sterile- physician prescribed, indicated for infusionSame
DialysateDialysate available as pre-packaged pre-mixed sterile fluids or via the PureFlow SL (K043436 K060296, K090919 & K111174)Same
Compatible Blood Tubing SetNxStage Standard CartridgeSame
Software VersionSoftware version 4.8Software version 4.9

Note: The Blood Leak Detector sensitivity for the proposed device cites 0.35 ml/min leak rate compared to the predicate's 0.45 ml/min. This appears to be a slight improvement in sensitivity, meaning it can detect smaller leaks, which would meet or exceed the predicate's performance.

2. Sample Size Used for the Test Set and Data Provenance:

The document mentions "Performance, verification and validation testing was conducted" and that "Results of this testing have documented that the proposed NxStage System One is substantially equivalent." However, it does not provide any specific sample sizes for test sets or details about data provenance (e.g., country of origin, retrospective/prospective). The study appears to be entirely bench testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This information is not provided in the document. Given that the submission is focused on demonstrating substantial equivalence through bench testing, it is unlikely that "experts" were used in the way one might for clinical ground truth establishment (e.g., radiologists interpreting images).

4. Adjudication Method for the Test Set:

This information is not provided. Again, this is a bench testing submission, so adjudication methods typical for clinical studies would not apply.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study was mentioned or conducted. The submission is not about human reader performance with or without AI assistance but rather about the technical specifications and safety of a medical device (hemodialysis system).

6. Standalone (Algorithm Only) Performance Study:

Since this is a physical medical device (hemodialysis system) with software components, not an AI algorithm for interpretation, the concept of a "standalone" or "algorithm-only" performance study in an AI context does not apply in the traditional sense. The device's performance relies on the integrated electromechanical control unit and cartridge. The software functionality itself was presumably validated as part of the overall device's performance testing (e.g., software version 4.9 compared to 4.8).

7. Type of Ground Truth Used:

The "ground truth" for this submission appears to be engineering specifications and performance standards for the hemodialysis system components and overall function, as established and validated during the design and testing phases. This is evidenced by the detailed comparison of parameters like flow rates, pressure monitor ranges, and detector sensitivities.

8. Sample Size for the Training Set:

This information is not provided. This device is not an AI/ML algorithm that is "trained" on a dataset in the conventional sense. Its functionality is based on pre-programmed parameters and electromechanical design, not machine learning model training.

9. How the Ground Truth for the Training Set Was Established:

As this is not an AI/ML device that undergoes "training," this question is not applicable. The "ground truth" for its design and operation would be based on established medical and engineering principles, regulatory requirements, and the performance of the predicate device.

{0}------------------------------------------------

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content contained in this 510(k) summary has been provided in conformance with 21 CFR §807.92

A. Date Prepared:February 28, 2014
B. Submitter's Information:
Name:NxStage Medical, Inc.

Address:

IC.

350 Merrimack Street Lawrence, MA 01843

FDA Establishment Owner/Operator Number: Contact Person:

9045797

Mary Lou Stroumbos Director, Regulatory Affairs

Phone: Fax:

Manufacturer:

NxStage Medical, Inc. 350 Merrimack Street Lawrence, MA 01843

FDA Establishment Registration Number:

Sterilization Site:

3003464075

(978) 687-4872

(978) 687-4750

Steris Isomedix (NxStage Cartridge Express) 1000 S. Sarah Place Ontario, CA 91761

{1}------------------------------------------------

  • C. Device Name:
    Trade/Proprietary Name: Common/Usual Name: Classification Name: Regulation Number: Product Code: Device Classification: Device Panel:

NxStage System

Hemodialysis System High Permeability Hemodialysis System 876.5860 78 KDI Class II Gastroenterology/Urology

D. Substantial Equivalence:

The NxStage System One has the same intended use and utilizes the same fundamental technology as the predicate NxStage System One. The NxStage System One has been compared to the legally marketed predicate devices as cleared through K122051 (April 23, 2013) and was found to be substantially equivalent.

E. Device Description/Indications for Use:

The NxStage System One is comprised of the NxStage Cycler, an electromechanical control unit; the NxStage Cartridge, a sterile, single-use extracorporeal blood and fluid management circuit (with or without a preattached high permeability filter) that mounts integrally within the NxStage Cvcler. The combined system is designed to deliver hemofiltration, hemodialysis and/or ultrafiltration in an acute or chronic care facility. The NxStage System One is also indicated for hemodialysis with or without ultrafiltration in the home.

{2}------------------------------------------------

NxStage Medical, Inc. NxStage System One 510(k) Premarket Notification

Indications for use:

The NxStage System One is indicated for the treatment of acute and chronic renal failure or fluid overload using hemofiltration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility.

The NxStage System One is also indicated for hemodialysis with or without ultrafiltration in the home

All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

F. Technological Characteristics:

The proposed device has the same technological characteristics and is similar in design and configuration as compared to the predicate device. The · proposed device is designed with similar software, components and features also used in the predicate device.

Table 1Device Technological Characteristics Comparison Table
ParameterProposed DeviceNxStage System OnePredicate DeviceNxStage System One(K122051)
Intended UseHemodialysisYesYes
HemofiltrationUltrafiltrationYesYesYesYes
Technology / Components:
PumpsSame4 peristaltic pumps
Valves (clamps)Same2 solenoid actuated pinch clamps8 cam driven pinch clamps
Air / fluid detectorsSame3 ultrasonic air/ fluid detectors
Blood leak detectorSame1 optical blood leak detector
Pressure transducersSame5 electronic pressure transducers
Temperature sensorsSame1 electronic temperature sensor

{3}------------------------------------------------

.

NxStage Medical, Inc.

Table 1Device Technological Characteristics Comparison Table
ParameterProposed DeviceNxStage System OnePredicate DeviceNxStage System One(K122051)
Flow Rates:BloodPrescription Fluid /DialysateFlowUltrafiltrationSame0-18000 ml/hr (NX1000-4)Same10-600 ml/min0-18000 ml/hr (NX1000-3)0-2400 ml/hr
Transmembrane PressureMonitoring SpecificationSameYes
Venous Pressure MonitorSame0 to 400 mmHg
Effluent fluid Pressure MonitorSame0 to 500 mmHg
Air DetectorSameReduction of detector signal lasting 6 msminimum (Approximates a 60 micro literbubble at 400 mmHg venous pressure and600 ml/min blood flow)
Blood Leak Detector15% reduction in detector signallasting 20 seconds minimum.Signal reduction % based on a 0.35ml/min leak rate of 32 Hct blood.15% reduction in detector signal lasting 20seconds minimum. Signal reduction %based on a 0.45 ml/min leak rate of 32 Hctblood.
Effluent Volume AccuracySameGreater of 300 ml/ 12 hr or 3% of exchangevolume (For software versions 4.7 andbelow)For software versions 4.8 and higher:TherapyFluid FlowRate L/hr) Specification greater of > 3 ±100 ml/hr*+ 5% UF*or ≤ 3 ± 25 ml/hr* *Representing 95/90 tolerance intervalestablished under controlled laboratorytesting conditions.
IV Prescription FluidSameOff-line, sterile- physician prescribed,indicated for infusion
DialysateSameDialysate available as pre-packaged pre-mixed sterile fluids or via the PureFlow SL(K043436 K060296, K090919 & K111174)
Compatible Blood Tubing SetSameNxStage Standard Cartridge
SoftwareSoftware version 4.9Software version 4.8

{4}------------------------------------------------

NxStage Medical, Inc. NxStage System One 510(k) Premarket Notification

G. Summary of Non-Clinical Test/Performance Testing - Bench

NxStage believes that the information and data provided in this submission clearly describes the proposed device and demonstrates that the device is adequately designed for the labeled indications for use. Performance, verification and validation testing was conducted to characterize performance of the proposed device and the predetermined acceptance criteria was met. Results of this testing have documented that the proposed NxStage System One is substantially equivalent to the predicate devices and is suitable for the labeled indications for use.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract shapes that resemble a human figure.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 19, 2014

NxStage Medical, Inc. Mary Lou Stroumbos Director, Regulatory Affairs 350 Merrimack Street Lawrence, MA 01843

Re: K140526

Trade/Device Name: NxStage® System One™ Regulation Number: 21 CFR& 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: March 4, 2014 Received: March 5, 2014

Dear Mary Lou Stroumbos,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{6}------------------------------------------------

Page 2 - Mary Lou Stroumbos

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/6/Picture/6 description: The image shows the text "Benjamin Fisher -S". The words are written in a simple, bold font. The text is arranged horizontally, with "Benjamin" on the left, followed by "Fisher" and then "-S" on the right.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): K140526

NxStage® System One TM Device Name:

The NxStage System One is indicated for the Indications for Use: treatment of acute and chronic renal failure, or fluid overload using hemofitration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility.

The NxStage System One is also indicated for hemodialysis with our without ultrafiltration in the home.

All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Benjamin Fisher -S
2014.05.19 13:21:21 -04'00'

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”