K Number
K122051

Validate with FDA (Live)

Date Cleared
2013-04-23

(285 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NxStage System One is indicated for the treatment of acute and chronic renal failure, or fluid overload using hemofiltration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility. The System is also indicated for hemodialysis with or without ultrafiltration in the home.

The NxStage System One is also indicated for Therapeutic Plasma Exchange in a clinical environment.

All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

Device Description

The NxStage System One is comprised of the NxStage Cycler, an electromechanical control unit and the NxStage Cartridge, a sterile, singleuse extracorporeal blood and fluid management circuit (with or without a preattached high permeability filter) that mounts integrally within the NxStage Cycler. The combined system is designed to deliver hemofiltration. hemodialysis and/or ultrafiltration in an acute or chronic care facility and is also indicated for hemodialysis with or without ultrafiltration in the home. The NxStage System One is also indicated for Therapeutic Plasma Exchange in a clinical environment.

AI/ML Overview

This submission describes a Special 510(k) Device Modification for the NxStage System One, a hemodialysis system. The modification involves increasing the therapy fluid pump rate from 200 ml/min (12 L/hr) to 300 ml/min (18 L/hr).

1. Table of Acceptance Criteria and Reported Device Performance:

The provided document states that "All predetermined acceptance criteria were met" for the performance, verification, safety, and software testing. However, the specific acceptance criteria and detailed device performance metrics are not explicitly listed in the provided 510(k) summary. The summary only broadly states that the device is "adequately designed for the labeled indications for use" and "is substantially equivalent to the predicate device."

Acceptance Criteria (Not explicitly detailed in document)Reported Device Performance
Safety requirements for increased pump rateMet all predetermined acceptance criteria
Performance requirements for increased pump rateMet all predetermined acceptance criteria
Software functionality for increased pump rateMet all predetermined acceptance criteria
Design specifications for increased pump rateMet design specifications
Substantial equivalence to predicate device (K093069)Achieved

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective) for the non-clinical performance testing. It generally refers to "Performance, verification and validation testing."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This information is not applicable as the provided document details non-clinical bench testing for a device modification, not a study involving human expert reading of data to establish ground truth.

4. Adjudication Method for the Test Set:

This information is not applicable as the provided document details non-clinical bench testing for a device modification.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. The submission focuses on non-clinical testing to demonstrate the safety and performance of a device modification and its substantial equivalence to a predicate device. It is not an AI-assisted diagnostic device where human reader improvement would be measured.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

This information is not explicitly stated in the context of "algorithm only" performance. The device is a hemodialysis system, and the testing described is for the system's operational parameters (pump rate) and overall safety/performance. It's not a standalone diagnostic algorithm. The non-clinical testing likely evaluated the device's performance in an automated/standalone manner, but the term 'algorithm' may not be directly applicable in the same way as an AI diagnostic.

7. The Type of Ground Truth Used:

For the non-clinical test/performance testing, the "ground truth" would be established by engineering specifications, regulatory standards, and the performance characteristics of the predicate device. The testing aimed to verify that the modified device's performance aligned with these established benchmarks.

8. The Sample Size for the Training Set:

This information is not applicable. The device is a hemodialysis system, and the modification involves a physical parameter (pump rate) and associated software/hardware, not a machine learning model that requires a "training set."

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable as there is no mention of a training set for a machine learning model.

{0}------------------------------------------------

510(k) Summary

K122051
page 1 of 3

NxStage Medical, Inc. NxStage System One Special 510(k) Device Modification

APR 2 3 2013

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content contained in this 510(k) summary has been provided in conformance with 21 CFR §807.92

Submitter's Information:Name:NxStage Medical, Inc.
Address:350 Merrimack StreetLawrence, MA 01843
FDA EstablishmentOwner/OperatorNumber:9045797
Contact Person:Mary Lou StroumbosRegulatory Affairs Manager
Phone:Fax:(978) 687-4872(978) 687-4750
Manufacturer:NxStage Medical, Inc.350 Merrimack StreetLawrence, MA 01843
FDA EstablishmentRegistration Number:3003464075
Sterilization Site:Steris Isomedix (NxStage C

Cartridge Express) 1000 S. Sarah Place Ontario, CA 91761

Special 510(k) Device Modification NxStage Medical, Inc.

A.

{1}------------------------------------------------

K122051
page 2 of 3

NxStage Medical. Inc. NxStage System One Special 510(k) Device Modification

  • B. Device Name: Trade/Proprietary Name:
    Product Code:

Device Panel:

Common/Usual Name:

Classification Name:

Regulation Number:

Device Classification:

NxStage® System One™

Hemodialysis System High Permeability Hemodialysis System 876.5860 78 KDI Class II Gastroenterology/Urology

C. Substantial Equivalence:

This submission is a Special 510(k) Device Modification as described in the FDA's Guidance document entitled, "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications." In support of this Special 510(k), NxStage has provided certification of compliance to 21 CFR §820.30 Design Control Requirements. Design validation testing was performed to ensure that the NxStage System One with modification to increase the therapy fluid pump rate from 200 ml/min (12 L/hr) to 300 ml/min (18 L/hr) meets design specifications. The NxStage System One has been compared to the legally marketed predicate device as cleared through K093069 (October 23, 2010) and was found to be substantially equivalent.

D. Device Description/Indications for Use:

The NxStage System One is comprised of the NxStage Cycler, an electromechanical control unit and the NxStage Cartridge, a sterile, singleuse extracorporeal blood and fluid management circuit (with or without a preattached high permeability filter) that mounts integrally within the NxStage Cycler. The combined system is designed to deliver hemofiltration. hemodialysis and/or ultrafiltration in an acute or chronic care facility and is also indicated for hemodialysis with or without ultrafiltration in the home. The NxStage System One is also indicated for Therapeutic Plasma Exchange in a clinical environment.

Special 510(k) Device Modification NxStage Medical, Inc.

{2}------------------------------------------------

NxStage Medical, Inc. NxStage System One Special 510(k) Device Modification

Indications for use:

The NxStage System One is indicated for the treatment of acute and chronic renal failure, or fluid overload using hemofiltration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility. The System is also indicated for hemodialysis with or without ultrafiltration in the home.

The NxStage System One is also indicated for Therapeutic Plasma Exchange in a clinical environment.

All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

E. Technological Characteristics:

The proposed device has the same technological characteristics and is similar in design and configuration as compared to the predicate device. The proposed device features a therapy fluid pump rate of 300 ml/min (18 L/hr) and is designed with similar software, components and features also used in the predicate device.

F. Summary of Non-Clinical Test/Performance Testing - Bench

NxStage believes that the information and data provided in this submission clearly describes the proposed device and demonstrates that the device is adequately designed for the labeled indications for use. Performance, verification and validation testing was conducted to characterize performance of the proposed device. This included verification, safety, performance, and software testing. All predetermined acceptance criteria were met. Results of this testing have documented that the proposed NxStage System One with a therapy fluid pump rate of 300 ml/min (18 L/hr) is substantially equivalent to the predicate device and is suitable for the labeled indications for use.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 23, 2013

NxStage Medical, Inc. % Ms. Mary Lou Stroumbos Regulatory Affairs Manager 350 Merrimack Street LAWRENCE MA 01843

Re: K122051

Trade/Device Name: NxStage System One Regulation Number: 21 CFR 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: March 22, 2013 Received: March 25, 2013

Dear Ms. Stroumbos:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Ms. Mary Lou Stroumbos

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): K122051

Device Name:

NxStage System One

Indications for Use:

The NxStage System One is indicated for the treatment of acute and chronic renal failure, or fluid overload using hemofiltration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility. The System is also indicated for hemodialysis with or without ultrafiltration in the home.

The NxStage System One is also indicated for Therapeutic Plasma Exchange in a clinical environment.

All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Herbert | 2 Merner -S

(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K122051 510(k) Number

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”