K Number
K140513

Validate with FDA (Live)

Date Cleared
2014-07-09

(131 days)

Product Code
Regulation Number
870.1340
Age Range
3 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Merit Safety Introducer Needle is used for providing a puncture site in blood vessels for the introduction of vascular access devices. It also incorporates a safety mechanism to minimize needlestick injuries.

Device Description

The Merit Safety Introducer Needle is used for providing a puncture site in blood vessels for the introduction of vascular access devices. It incorporates a safety mechanism/quard to minimize needlestick injuries. The needle consists of a stainless steel one wall cannula and a clear standard female Luer lock hub for immediate bleed-back visualization. The hub of this introducer needle is designed with an ergonomic feel for ease of handling and is offered with a standard hub or a Seldinger shield. The needles are available in 18 gauge and 21 gauge with a usable length of 7 cm. The safety mechanism is color coded for needle gauge identification. The design of the Merit Safety Device Introducer Needle allows clinicians to easily activate the safety mechanism by advancing it to the end of the needle to shield the needle bevel after use. The safety mechanism automatically senses the end of the needle and locks the safety mechanism covering the needle tip which reduces the risk of accidental needlestick injuries by shielding the needle tip. A visual, tactile, and audible confirmation of the locking component over the needle confirms lockout of the safety guard over the needle. The safety mechanism cannot be deactivated and remains protective through disposal into a sharps container. The safety mechanism can be activated over quide wires. The Merit Safety Introducer Needle is a single use device that is supplied sterile and non-pyrogenic for use in the adult population.

AI/ML Overview

The Merit Safety Introducer Needle's acceptance criteria and the study that proves the device meets the acceptance criteria are detailed below. It is important to note that this document is a 510(k) summary for a medical device and thus focuses on demonstrating substantial equivalence to a predicate device rather than comprehensive clinical trial results.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Merit Safety Introducer Needle were based on a risk analysis and compliance with various recognized and unrecognized international standards and FDA guidance documents. These standards cover aspects like device dimensions, material properties, safety mechanism function, biocompatibility, and sterilization. The document reports that the subject device "meets the requirements that are considered essential for its intended use."

The table below summarizes the performance tests conducted, which implicitly define the acceptance criteria for each aspect. The exact quantitative acceptance limits for each test are not provided in this summary, but the general outcome of meeting the requirements is stated.

Acceptance Criteria (Implicitly defined by test)Reported Device Performance
Bench Performance:
DimensionsMeets requirements
Puncture ResistanceMeets requirements
Force to Defeat Safety MechanismMeets requirements
Force to Activate Safety MechanismMeets requirements
Hub to Cannula Bond StrengthMeets requirements
Visual InspectionMeets requirements
Luer GaugingMeets requirements
Luer Liquid Leak TestMeets requirements
Luer Air Aspiration Leak TestMeets requirements
Luer Separation ForceMeets requirements
Luer Unscrewing Torque TestMeets requirements
Luer Ease of AssemblyMeets requirements
Luer Resistance to OverridingMeets requirements
Luer Stress CrackingMeets requirements
Needle PenetrationMeets requirements
Guide Wire CompatibilityMeets requirements
Corrosion ResistanceMeets requirements
Simulated UseMeets requirements
Biocompatibility:
CytotoxicityMeets requirements
SensitizationMeets requirements
IrritationMeets requirements
Acute Systemic ToxicityMeets requirements
PyrogenicityMeets requirements
HematologyMeets requirements
CoagulationMeets requirements
HemolysisMeets requirements
Chemical CharacterizationMeets requirements

2. Sample Size for the Test Set and Data Provenance

The document does not specify the exact sample sizes used for each individual performance test (bench or biocompatibility). The testing was conducted by the manufacturer, Merit Medical Systems, Inc. The data provenance is internal to the manufacturer and would be considered prospective for the purpose of demonstrating substantial equivalence. There is no information regarding country of origin of the data beyond the manufacturer's location in the USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This type of device (introducer needle) does not typically involve human expert interpretation for "ground truth" in the same way an imaging AI device would. The "ground truth" for the performance tests would be defined by the specifications of the device and the standards against which it is tested. For example, for "dimensions," the ground truth is the specified dimension range. For "biocompatibility," the ground truth is the pass/fail criteria for the specific biological tests outlined in ISO 10993. Therefore, no external "experts" were used to establish a subjective ground truth on a test set as would be the case for diagnostic AI.

4. Adjudication Method for the Test Set

Given the nature of the device and the testing performed (bench and biocompatibility), there was no adjudication method (e.g., 2+1, 3+1) involving human readers or interpretation. The performance tests are objective measurements against established standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was conducted or is applicable for this device. This device is an introducer needle with a safety mechanism, not an AI-assisted diagnostic or decision-making tool that would involve human readers interpreting cases.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

This question is not applicable. The Merit Safety Introducer Needle is a physical medical device, not a software algorithm. Therefore, there is no "standalone performance" in the context of an algorithm without human interaction. The safety mechanism is an integral part of the physical device's design.

7. Type of Ground Truth Used

The ground truth used for performance validation was based on:

  • Engineering specifications and design requirements: For bench tests like dimensions, bond strength, Luer connections, and forces.
  • Recognized international standards and FDA guidance documents: For both bench performance (e.g., ISO for needles, Luer fittings, catheters) and biocompatibility (e.g., ISO 10993 series, USP pyrogenicity). These standards define the specific test methods and acceptance criteria.

8. Sample Size for the Training Set

This question is not applicable. This device is a physical medical device and does not involve machine learning algorithms that require training sets of data.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as there is no training set for this device.

{0}------------------------------------------------

5.0 510(k) Summary

:

JUL 09 2014

:

·

GeneralProvisionsSubmitter Name:Address:Telephone Number:Fax Number:Contact Person:Date of Preparation:Registration Number:Merit Medical Systems, Inc.1600 West Merit ParkwaySouth Jordan, UT 84095(801) 208-4789(801) 253-6919Susan ChristensenFebruary 27, 20141721504
Subject DeviceTrade Name:Common/Usual Name:Classification Name:Merit Safety Introducer NeedlesSafety Introducer NeedleCatheter Introducer
PredicateDeviceTrade Name:Classification Name:Premarket Notification:Manufacturer:SecureLoc™ Safety Introducer NeedleCatheter IntroducerK050023Bard Access Systems, Inc.
ClassificationClass II21 CFR § 870.1340FDA Product Code: DYBReview Panel: Cardiovascular
Intended UseThe Merit Safety Introducer Needle is used for providing a puncture sitein blood vessels for the introduction of vascular access devices. It alsoincorporates a safety mechanism to minimize needlestick injuries.

K140513

{1}------------------------------------------------

The Merit Safety Introducer Needle is used for providing a puncture site in blood vessels for the introduction of vascular access devices. It incorporates a safety mechanism/quard to minimize needlestick injuries. The needle consists of a stainless steel one wall cannula and a clear standard female Luer lock hub for immediate bleed-back The hub of this introducer needle is designed with an visualization. ergonomic feel for ease of handling and is offered with a standard hub or a Seldinger shield. The needles are available in 18 gauge and 21 gauge with a usable length of 7 cm. The safety mechanism is color coded for needle gauge identification. The design of the Merit Safety Device Introducer Needle allows clinicians to easily activate the safety Description mechanism by advancing it to the end of the needle to shield the needle bevel after use. The safety mechanism automatically senses the end of the needle and locks the safety mechanism covering the needle tip which reduces the risk of accidental needlestick injuries by shielding the needle tip. A visual, tactile, and audible confirmation of the locking component over the needle confirms lockout of the safety guard over the needle. The safety mechanism cannot be deactivated and remains protective through disposal into a sharps container. The safety mechanism can be activated over quide wires. The Merit Safety Introducer Needle is a single use device that is supplied sterile and non-pyrogenic for use in the adult population. The technological characteristics of the subject Merit Safety Introducer Needles are substantially equivalent to those of the predicate device. The needle is available in the same gauge sizes 18 and 21, same useable length 7 cm, and same hub styles which include with and without a Seldinger shield. The needle incorporates a clear standard female Luer lock hub for immediate bleed-back visualization and the safety mechanism is color coded for needle gauge identification. . The Merit Safety Introducer Needles are similar in clinical use. Comparison to function, materials and use to the predicate SecureLoc™ Predicate Safety Introducer Needles. Device . The Merit Safety Introducer Needles have a safety feature that locks a safety mechanism over the needle tip after the needle is removed from the patient, as does the SecureLoc Safety Introducer Needle predicate device. . The Merit Safety Introducer Needle's safety mechanism lockout can be confirmed by visual means, tactile feel and audible means, as can the SecureLoc Safety Introducer Needle predicate device cited in this submission.

{2}------------------------------------------------

·

/

.

:

:

/

·

:

·

·

.

.

Safety &PerformanceTestsNo performance standards have been established under Section 514 ofthe Food, Drug and Cosmetic Act for these devices. Performancetesting of the subject Merit Safety Introducer Needles was conductedbased on the risk analysis and based on the requirements of thefollowing recognized and unrecognized international standards andFDA guidance documents:
FDA Guidance for Industry and FDA Staff, Medical Devices withSharps Injury Prevention Features; August 9, 2005. ISO 7864:1993 - Sterile hypodermic needles for single use ISO 7864-2:1993 - Sterile hypodermic needles for single use Part2: Specification for sterile hypodermic needles for single use ISO 594-1:1986 - Conical fittings with 6% (Luer) taper for syringes,needles, and certain other medical equipment Part1: Generalrequirements ISO 594-2:1998 - Conical fittings with 6% (Luer) taper for syringes,needles, and certain other medical equipment Part2: Lock fittings ISO 6009:1992 - Hypodermic needles for single use - Colorcoding for identification ISO 9626:1992 Amendment 1:2001 - Stainless steel needle tubingfor the manufacture of medical devices ISO 11070:1998 - Sterile single use intravascular catheterintroducers ISO 11135-1 First Edition 2007, Sterilization of health careproducts - Ethylene Oxide - part 1: Requirements fordevelopment, validation, and routine control of a sterilizationprocess for medical devices ISO 10993-1:2009, Biological Evaluation of Medical Devices Part1: Evaluation and Testing within a risk management process, andFDA guidance Required Biocompatibility Training and ToxicologyProfiles for Evaluation of Medical Devices, May 1, 1995 ISO 10993-4:2002 (Amd.1:2006), Biological evaluation of medicaldevices - Part 4: Selection of tests for interaction with blood ISO 10993-5:2009, Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicity ISO 10993-7:2008, Biological evaluation of medical devices - Part7: Ethylene oxide sterilization residuals ISO 10993-10:2010, Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization ISO 10993-11:2006, Biological evaluation of medical devices -Part 11: Tests for systemic toxicity ASTM F756-08:2008; Standard Practice for Assessment ofHemolytic Properties of Materials United States Pharmacopeia Section <151>, USP 36 PyrogenTesting, 2013-12-01

.

, -

{3}------------------------------------------------

Performance Testing-Bench

  • Dimensions .
  • Puncture Resistance Testing .
  • Force to Defeat Safety Mechanism Testing .
  • Force to Activate Safety Mechanism Testing t
  • Hub to Cannula Bond Strength Test .
  • Visual Inspection .
  • Luer Gauging Test .
  • Luer Liquid Leak Test .
  • Luer Air Aspiration Leak Test .
  • Luer Separation Force Test .
  • Luer Unscrewing Torque Test .
  • Luer Ease of Assembly Test ●
  • Luer Resistance to Overriding Test .
  • Luer Stress Cracking Test .
  • Needle Penetration Testing .
  • Guide Wire Compatibility .
  • Corrosion Resistance .
  • Simulated Use Testing .

Biocompatibility

  • Cytotoxicity �
  • Sensitization ●
  • Irritation .
  • Acute Systemic Toxicity .
  • Pyrogenicity .
  • Hematology .
  • Coaqulation .
  • Hemolysis t
  • Chemical Characterization .

Summary of Substantial Equivalence

Based on the indications for use, design, safety and performance testing, the subject Merit Safety Introducer Needles meets the requirements that are considered essential for its intended use and is substantially equivalent to the predicate device. SecureLoc Safety Introducer Needle, K050023.

Safety & Performance

Tests cont.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is positioned within a circular border, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" encircling it.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 9, 2014

Merit Medical Systems, Incorporated Ms. Susan Christensen Principal Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, UT 84095

Re: K140513

Trade/Device Name: Merit Safety Introducer Needles Regulation Number: 21 CFR 870.1340 Regulation Name: Introducer, Catheter Regulatory Class: II Product Code: DYB Dated: June 18, 2014 Received: June 19, 2014

Dear Ms. Christensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Christensen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default,htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/8 description: The image shows the name Tejashri Purohit-Sheth, M.D. in two different fonts. Below the name in the second font, it says Clinical Deputy Director. Underneath that, it says DAGRID/ODE/CDRH FOR. The image also contains the FDA logo.

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140513

Device Name

Merit Safety Introducer Needle

Auction of the children theone of expensive parts control on an any a
Indications for Use (Describe)

The Merit Safety Introducer Need for providing a puncture site in blood vessels for the introduction of vascular access devices. It also incorporates a safety mechanism to minimize needlestick injuries.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

極織・無料・FORIFDA USE ONLY (新品) (送料・送料・ Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/6/Picture/12 description: The image shows the logo for the Food and Drug Administration (FDA). The logo is made up of the letters "FDA" in a stylized font. The letters are outlined with a thick black line, and the inside of the letters is filled with a lighter shade of gray. The logo is simple and recognizable, and it is often used on products that have been approved by the FDA.

Digitally signed by Richard C. Chapman -S Date: 2014.07.09 12:09:08 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).