K Number
K140367

Validate with FDA (Live)

Date Cleared
2014-09-12

(210 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Harbor MedTech Architect™ Px Extracellular Collagen Matrix is indicated for the local management of moderately to heavy exuding wounds, including:

· Partial and full thickness wounds,

  • · Draining wounds,
    · Pressure sores/ulcers,

· Venous ulcers,

· Chronic vascular ulcers,

· Diabetic ulcers,

· Trauma wounds (e.g., abrasions, lacerations, partial thickness burns, skin tears),

· Surgical wounds (e.g., donor sites/grafts, post-laser surgery, post-Mohs surgery, podiatric wounds, dehisced surgical incisions)

Device Description

The Harbor MedTech Architect™ Px Extracellular Collagen Matrix (ECM) is a decellularized equine pericardial device The Architect™ Px ECM is provided sterile for single use only. The device must be rehydrated and rinsed prior to use following the procedure described in the Instructions for Use.

The Architect™ Px ECM is available in multiple sizes. The Architect™ Px ECM is available as a standard dressing and in a fenestrated model with pre-cut slits in the collagen matrix.

AI/ML Overview

This document is a 510(k) summary for the Harbor MedTech Architect™ Px Extracellular Collagen Matrix. It does not describe an AI/ML device or a study involving human readers or ground truth established by experts.

Therefore, many of the requested categories are not applicable to the provided text.

Here's a breakdown of the relevant information from the document:

1. A table of acceptance criteria and the reported device performance

The document does not provide a formal table of acceptance criteria with specific numerical targets. Instead, it describes performance testing results in a narrative way, stating that the device "conforms to the performance specifications established" for its predicate and that "the Architect™ Px ECM tissue has the mechanical strength and integrity to perform as intended."

Criteria CategoryAcceptance Criteria (Implied)Reported Device Performance
Physical AttributesConformance to performance specifications of predicate device."Results of this testing confirmed that the minor changes to the manufacturing process yielded a device that conforms to the performance specifications established for the Architect™ ECM cleared under premarket 510(k) notification K122502."
Mechanical Strength/IntegritySufficient mechanical strength and integrity to perform as intended."Results of the tensile, suture pull-out, staple pull-out and flexibility testing show that the Architect™ Px ECM tissue has the mechanical strength and integrity to perform as intended."
BiocompatibilityAcceptable."Extensive biocompatibility testing... was completed and found to be acceptable." (Note: these studies were not repeated for this submission as materials, processing, packaging, and sterilization had not changed, and the decreased BDDGE concentration did not warrant repeating them.)
Viral InactivationAcceptable."Extensive... viral inactivation testing was completed and found to be acceptable." (Note: these studies were not repeated for this submission as materials, processing, packaging, and sterilization had not changed, and the decreased BDDGE concentration did not warrant repeating them.)
Equivalence to PredicateSubstantially equivalent in indications, materials, manufacturing, and design principles."Harbor MedTech, Inc., demonstrated that... the Architect™ Px Extracellular Collagen Matrix (ECM) is substantially equivalent in indications, materials, manufacturing and design principles to the predicate device... The Architect™ Px Extracellular Collagen Matrix has the following similarities to the predicate devices: has the same intended use, uses the same operating principle, incorporates the same basic design, incorporates the same materials and manufacturing processes, uses the same species of animal tissue."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document describes "verification and validation studies" and "performance bench testing." It does not specify sample sizes for these tests, nor does it provide information on data provenance (country of origin, retrospective/prospective). These appear to be laboratory bench tests on material samples, not clinical data sets from patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is a medical device (collagen matrix), and the testing described is bench testing, not an AI/ML algorithm requiring expert ground truth for image interpretation or diagnosis.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This pertains to expert review for diagnostic algorithms, which is not relevant here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document is for a medical device (collagen matrix) and does not involve AI or human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This document is for a medical device (collagen matrix), not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

Not applicable in the context of diagnostic AI. For this device, the "ground truth" or reference for performance is the established specifications and performance of the predicate device, as confirmed through physical and mechanical testing.

8. The sample size for the training set

Not applicable. This is a medical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established

Not applicable. This is a medical device, not an AI model.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 12, 2014

Harbor MedTech Incorporated Mr. Jerry Mezger Chief Executive Officer 4 Jenner, Suite 190 Irvine, California 92618

Re: K140367

Trade/Device Name: Architect Px Extracellular Collagen Matrix Regulatory Class: Unclassified Product Code: KGN Dated: August 14, 2014 Received: August 18, 2014

Dear Mr. Mezger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140367

Device Name

Architect Px Extracellular Collagen Matrix

Indications for Use (Describe)

The Harbor MedTech Architect™ Px Extracellular Collagen Matrix is indicated for the local management of moderately to heavy exuding wounds, including:

· Partial and full thickness wounds,

  • · Draining wounds,
    · Pressure sores/ulcers,

· Venous ulcers,

· Chronic vascular ulcers,

· Diabetic ulcers,

· Trauma wounds (e.g., abrasions, lacerations, partial thickness burns, skin tears),

· Surgical wounds (e.g., donor sites/grafts, post-laser surgery, post-Mohs surgery, podiatric wounds, dehisced surgical incisions)

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Harbor MedTech, Inc. Architect™ Px Extracellular Collagen Matrix

ADMINISTRATIVE INFORMATION:

Date of Preparation:September 3, 2014
Manufacturer Name:Harbor MedTech, Inc.
4 Jenner, Suite 190
Irvine, CA 92618
Telephone:(949) 679-4800
Facsimile:(949) 679-1086
Email:jerrym@harbormedtech.com
Official Correspondent:Jerry Mezger, Chief Executive Officer
Additional Correspondent:Carol White, RA Consultant
21521 Hummingbird Street
Trabuco Canyon, CA 92679
Telephone:(949) 874-2024
Email:carolwhite@cox.net

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary Name:Architect™ Px Extracellular Collagen Matrix
Common Name:Dressing, Wound, Collagen
Classification Regulation:Unclassified
Product Code:KGN
Device Class:Unclassified
Review Panel:General & Plastic Surgery

INTENDED USE

The Harbor MedTech Architect™ Px Extracellular Collagen Matrix is indicated for the local management of moderately to heavy exuding wounds, including:

  • Partial and full thickness wounds, ●
  • Draining wounds,
  • Pressure sores/ulcers,
  • Venous ulcers, ●
  • Chronic vascular ulcers, ●
  • Diabetic ulcers,
  • Trauma wounds (e.g., abrasions, lacerations, partial thickness burns, skin tears), ●
  • Surgical wounds (e.g., donor sites/grafts, post-laser surgery, post-Mohs surgery, podiatric . wounds, dehisced surgical incisions)

{4}------------------------------------------------

DEVICE DESCRIPTION

The Harbor MedTech Architect™ Px Extracellular Collagen Matrix (ECM) is a decellularized equine pericardial device The Architect™ Px ECM is provided sterile for single use only. The device must be rehydrated and rinsed prior to use following the procedure described in the Instructions for Use.

The Architect™ Px ECM is available in multiple sizes. The Architect™ Px ECM is available as a standard dressing and in a fenestrated model with pre-cut slits in the collagen matrix. A table of model numbers and sizes is provided below:

Model NumberDimensionsDescription
HMT-004Px2 cm x 2 cmStandard
HMT-009Px3 cm x 3 cmStandard
HMT-025Px5 cm x 5 cmStandard
HMT-036Px6 cm x 6 cmStandard
HMT-100Px10 cm x 10 cmStandard
HMT-150Px10 cm x 15 cmStandard
HMT-009F-Px3 cm x 3 cmFenestrated
HMT-025F-Px5 cm x 5 cmFenestrated
HMT-036F-Px6 cm x 6 cmFenestrated
HMT-100F-Px10 cm x 10 cmFenestrated
HMT-150F-Px10 cm x 15 cmFenestrated

Due to the natural variability of extracellular collagen matrices. lot to variations may be seen in the handling characteristics of the finished device. The user is instructed to use their professional clinical judgment in selecting the appropriate device for the application of the product.

EQUIVALENCE TO MARKETED PRODUCT

Harbor MedTech, Inc., demonstrated that, for purposes of FDA's regulation of medical devices, the Architect™ Px Extracellular Collagen Matrix (ECM) is substantially equivalent in indications, materials, manufacturing and design principles to the predicate device which has been determined by FDA to be substantially equivalent to preamendment devices.

The Architect™ Px Extracellular Collagen Matrix has the following similarities to the predicate devices:

  • has the same intended use ●
  • uses the same operating principle
  • incorporates the same basic design ●
  • incorporates the same materials and manufacturing processes ●
  • uses the same species of animal tissue ●

{5}------------------------------------------------

The Architect™ Px Extracellular Collagen Matrix differs from the predicate device in the concentration of the BDDGE solution used in the manufacturing process. The concentration of BDDGE used in the proposed device is lower than that of the predicate device.

The Harbor MedTech Architect™ Px ECM is, in our opinion, substantially equivalent to the Harbor MedTech Architect™ ECM K122502.

PERFORMANCE TESTING

A number of verification and validation studies were conducted to assess the physical attributes of the collagen biomaterial and manufacturing processing. Results of this testing confirmed that the minor changes to the manufacturing process yielded a device that conforms to the performance specifications established for the Architect™ ECM cleared under premarket 510(k) notification K122502. Verification and validation studies included Pronase digestion, differential scanning calorimetry (DSC), tissue flexibility and conformity, tensile testing, suture pull-out and staple pull-out testing. Comparative testing was performed with the predicate device: Architect™ Extracellular Collagen Matrix produced by Harbor MedTech, Inc.

Extensive biocompatibility testing and viral inactivation testing was completed and found to be acceptable. These studies were not repeated since the materials, processing, packaging and sterilization have not changed. The decreased concentration of BDDGE fixative used in the new process did not warrant a repeat of these studies.

Results of the tensile, suture pull-out, staple pull-out and flexibility testing show that the Architect™ Px ECM tissue has the mechanical strength and integrity to perform as intended.

Based on the results of the performance bench testing validations and comparison to the predicate device, Harbor MedTech has demonstrated that the Architect™ Px Extracellular Collagen Matrix is equivalent to the predicate device in terms of performance characteristics and will perform as intended for the application of the product.

N/A