K Number
K140270

Validate with FDA (Live)

Date Cleared
2014-08-20

(198 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NEXUS™ MIDLINE CT CATHETER is indicated for short term (less than 30 days) peripheral access to the peripheral venous system for infusion, intravenous therapy and blood sampling. The NEXUS™ MIDLINE CT CATHETER is suitable for use with power injectors.

Device Description

The NEXUS™ MIDLINE CT Catheter is a family of peripherally inserted central catheters made from specially formulated biocompatible medical grade materials. Catheters are packaged in a tray with accessories necessary for a percutaneous microintroducer introduction (Modified Seldinger or Seldinger technique). The device is intended for short term (less than 30days) vascular access. The NEXUS™ MIDLINE CT Catheter is indicated for dwell times shorter than 30 days. The NEXUS™ MIDLINE CT Catheter product line will have catheters in 3Fr, 4 Fr and 5 Fr single lumen, and 4 Fr and 5 Fr dual lumen. The NEXUS™ MIDLINE CT Catheters are supplied in two lengths, approximately 10 and 20 cm long. All catheters are attached to an injectionmolded polyurethane hub that has extension legs with Luer lock fittings for access attachment.

AI/ML Overview

The provided text describes the Nexus Midline CT Catheter and its substantial equivalence to predicate devices, but it does not contain details of a clinical study or a multi-reader multi-case (MRMC) comparative effectiveness study. The information focuses on non-clinical bench testing to demonstrate performance and substantial equivalence.

Here's an analysis based on the provided text, addressing the points where information is available and indicating where it is not:

1. Table of Acceptance Criteria and Reported Device Performance

The document mentions that various bench tests "met predefined acceptance criteria" and "demonstrated equivalence." However, it does not provide the specific numerical acceptance criteria or the numerical reported device performance for each test. It only lists the tests performed and states they were successful.

A qualitative summary of the performance claims based on the text:

Test NameAcceptance CriteriaReported Device Performance
Tensile Strength TestPredefined acceptance criteria (details not provided)Met predefined acceptance criteria
Power Injection PerformancePredefined acceptance criteria (details not provided)Met predefined acceptance criteria
Dynamic Failure TestPredefined acceptance criteria (details not provided)Met predefined acceptance criteria
Static Pressure TestPredefined acceptance criteria (details not provided)Met predefined acceptance criteria
Static Burst TestPredefined acceptance criteria (details not provided)Met predefined acceptance criteria
Life Cycle Power InjectionPredefined acceptance criteria (details not provided)Met predefined acceptance criteria
Catheter Flow Rate TestPredefined acceptance criteria (details not provided)Met predefined acceptance criteria
Air Leakage TestPredefined acceptance criteria (details not provided)Met predefined acceptance criteria
Liquid Leakage TestPredefined acceptance criteria (details not provided)Met predefined acceptance criteria
BiocompatibilityRequirements according to ISO-10993, Part 1, Test profiles for externally communicating, blood-contacting, long-term devices.Test profiles were met
SterilizationRequirements of ISO 11135:2007Requirements were met

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: The document does not specify the sample size for any of the bench tests (e.g., number of catheters tested for tensile strength, burst pressure, etc.).
  • Data Provenance: The data is described as "Nonclinical tests" and "bench test data" performed for "design verification testing." This suggests it's laboratory-based testing of the device itself, not data from human subjects or clinical settings. There is no mention of country of origin as it pertains to clinical data, as this is bench testing. The data is not described as retrospective or prospective clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable as the document describes non-clinical bench testing. There is no mention of human experts being involved in establishing "ground truth" for these engineering and material performance tests. The "ground truth" for such tests would typically be defined by engineering specifications and direct physical measurements.

4. Adjudication method for the test set

This question is not applicable as the document describes non-clinical bench testing. Adjudication methods like 2+1 or 3+1 typically refer to expert consensus in interpreting images or clinical outcomes, which is not relevant here.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done or mentioned. The submission describes non-clinical bench testing for a physical medical device (catheter), not an AI algorithm or a diagnostic tool that would involve human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

No standalone algorithm performance study was done or mentioned. The device is a physical catheter, not a software algorithm.

7. The type of ground truth used

For the non-clinical bench testing, the "ground truth" refers to engineering specifications, direct physical measurements, and adherence to relevant international standards (e.g., ISO 11135:2007 for sterilization, ISO-10993 for biocompatibility). The document states that testing was performed "according to protocols based on the above-referenced guidance document recommendations and additional standards."

8. The sample size for the training set

This question is not applicable as the document describes non-clinical bench testing for a physical medical device. There is no mention of a "training set" in the context of an algorithm.

9. How the ground truth for the training set was established

This question is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 9, 2020

Health Line International Corporation John Lincoln Principal Consultant Po Box 2786 St. George, Utah 84771-2786

Re: K140270

Trade/Device Name: Nexus Midline Ct Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: Class II Product Code: FOZ Dated: January 30, 2014 Received: February 3, 2014

Dear John Lincoln:

This letter corrects our substantially equivalent letter of August 20, 2014.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sapana Patel -S

for Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140270

Device Name Nexus™ Midline CT Catheter

Indications for Use (Describe)

The NEXUS™ MIDLINE CT CATHETER is indicated for short term (less than 30 days) peripheral access to the peripheral venous system for infusion, intravenous therapy and blood sampling. The NEXUS™ MIDLINE CT CATHETER is suitable for use with power injectors. For maximum flow rate and maximum pressure that can be used during power injection, please refer to the following:

Catheter SizePriming Volume(ml)Max Labeled FlowRate (ml/sec)*Max Internal CatheterPressure at Max Flow Rate(psi)Rated BurstPressure(psi)**
3Fr Single Lumen0.291167251
4Fr Single Lumen0.375164299
4Fr Dual Lumen0.35/0.353173/173299
5Fr Single Lumen0.385112300
5 Fr Dual Lumen0.37/0.375179/179299
* Pressurized flow rates are determined with pump safety cut-off at 300 psi using viscous fluid at 11.8centipoise (cP) and represents approximate flow capability of power injection of contrast media.
** Max Burst Pressure is the static burst pressure failure point of the catheter when the lumen iscompletely occluded. WARNING: Power injector machine pressure limiting feature may not preventover pressurization of an occluded catheter.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (21 CFR 807.92) for NEXUS™ MIDLINE CT CATHETER

SUBMITTER:

Health Line International Corporation 803 N. 1250 W. - STE 1 Centerville, Utah 84014

ESTABLISHMENT REGISTRATION NUMBER:

3006097687

CONTACT:

Nola L. Benstog QA/RA Director Telephone: 801-773-7798 Fax: 855-228-1336 Email: nbenstog@hlic.net

DATE PREPARED:

January 30, 2014

NAME OF MEDICAL DEVICE:

Proprietary Name:NEXUS™ MIDLINE CT CATHETER
Regulation Name:Intravascular Catheter
Common/Usual Name:Midline Catheter, single and double lumen

DEVICE CLASSIFICATION:

Classification Panel:General Hospital
Regulatory Class:Class II
Product Code:FOZ
Regulation Number:21 CFR 880.5200

PREDICATE DEVICES:

Proprietary Name:SYNERGY™ CT PICC (K101329)
Regulation Name:Percutaneous, Implanted, Long-term Intravascular Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC), single, doubleand triple lumen
Classification Panel:General Hospital
Regulatory Class:Class II
Product Code:L J S
Regulation Number:21 CFR 880.5970

{4}------------------------------------------------

Proprietary Name:UNI-PICCC™ (K130034)
Regulation Name:Percutaneous, Implanted, Long-term Intravascular Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC), single, doubleand triple lumen
Classification Panel:General Hospital
Regulatory Class:Class II
Product Code:L J S
Regulation Number:21 CFR 880.5970
Proprietary Name:Vascu-PICC® and Midline Catheters, Single, Double andTriple Lumen (K091586)
Regulation Name:Percutaneous, Implanted, Long-term Intravascular Catheter
Common/Usual Name:Catheter, Intravascular, Therapeutic, Long-Term
Classification Panel:General Hospital
Regulatory Class:Class II
Product Code:L J S
Regulation Number:21 CFR 880.5970
Proprietary Name:PowerGlide Midline Catheter (K121073)
Regulation Name:Intravascular Catheter
Common/Usual Name:Intravascular Catheter
Classification Panel:General Hospital
Regulatory Class:Class II
Product Code:FOZ
Regulation Number:21 CFR 880.5200

DEVICE DESCRIPTION:

The NEXUS™ MIDLINE CT Catheter is a family of peripherally inserted central catheters made from specially formulated biocompatible medical grade materials. Catheters are packaged in a tray with accessories necessary for a percutaneous microintroducer introduction (Modified Seldinger or Seldinger technique). The device is intended for short term (less than 30days) vascular access.

The NEXUS™ MIDLINE CT Catheter is indicated for dwell times shorter than 30 days. The NEXUS™ MIDLINE CT Catheter product line will have catheters in 3Fr, 4 Fr and 5 Fr single lumen, and 4 Fr and 5 Fr dual lumen. The NEXUS™ MIDLINE CT Catheters are supplied in two lengths, approximately 10 and 20 cm long. All catheters are attached to an injectionmolded polyurethane hub that has extension legs with Luer lock fittings for access attachment.

{5}------------------------------------------------

INTENDED USE:

The NEXUS™ MIDLINE CT Catheter is intended to be used by medical professionals for shortterm percutaneous access to the venous system for infusion, intravenous therapy and for blood sampling.

The intended use for the NEXUS™ MIDLINE CT Catheter has not changed from that of the SYNERGY™ CT PICC (K101329) and UNI-PICC™ (K130034) predicate devices.

INDICATIONS FOR USE:

The NEXUS™ MIDLINE CT CATHETER is indicated for short term (less than 30 days) peripheral access to the peripheral venous system for infusion, intravenous therapy and blood sampling. The NEXUS™ MIDLINE CT CATHETER is suitable for use with power injectors. For maximum flow rate and maximum pressure that can be used during power injection, please refer to the following:

Catheter SizePriming Volume(ml)Max Labeled FlowRate (ml/sec)*Max Internal CatheterPressure at Max Flow Rate(psi)Rated BurstPressure(psi)**
3Fr Single Lumen0.291167251
4Fr Single Lumen0.375164299
4Fr Dual Lumen0.35/0.353173/173299
5Fr Single Lumen0.385112300
5 Fr Dual Lumen0.37/0.375179/179299

Pressurized flow rates are determined with pump safety cut-off at 300 psi using viscous fluid at 11.8 centipoise (cP) and represents approximate flow capability of power injection of contrast media.

    • Max Burst Pressure is the static burst pressure failure point of the catheter when the lumen is completely occluded. WARNING: Power injector machine pressure limiting feature may not prevent over pressurization of an occluded catheter.

TECHNOLOGICAL COMPARISON TO PREDICATE DEVICES:

New device is compared to Marketed Device? Yes. It is compared to four legally marketed predicate devices.

Does the new device have the same indication statements? Yes the indications for use statement is basically the same as the SYNERGY™ CT PICC (K101329) and the UNI-PICC™ (K130034) predicate devices.

{6}------------------------------------------------

Do the differences alter the intended therapeutic/diagnostic/etc. effect (i.e., deciding may consider impact on safety and effectiveness)? No, the differences do not alter the intended effect of the device.

Does the new device have the same technological characteristics, e.g. design, material, etc.? Yes. The device of this submission, the NEXUS™ MIDLINE CT CATHETER, is identical to that of the predicate devices. In fact, the SYNERGY™ CT PICC and the UNI-PICC™ are manufactured by the same company, Health Line International Corporation, the submitter of this submission. The basic fundamental scientific technology of the device has not changed. There may be minor variations in the contents of the introduction kit components.

Could the new characteristics affect safety or effectiveness? No.

Do the new characteristics raise new types of safety and effectiveness questions? No. There are no new types of safety and effectiveness questions.

Do accepted scientific methods exist for assessing effects of the new characteristics? Yes.

The FDA's Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, dated 3/16/95 was used to determine the appropriate methods for evaluating the device's performance.

Sterilization requirements of ISO 11135:2007, Sterilization of Health Care Products -Requirements for Validation and Routine Control -- Ethylene Oxide Sterilization.

Biocompatibility requirements according to of ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. Test profiles for externally communicating, bloodcontacting, long-term devices will be met.

These and other standards were used to determine the appropriate methods for evaluating the device's performance.

Are performance data available to assess effects of new characteristics? Yes. Verification testing was performed according to protocols based on the above-referenced guidance document recommendations and additional standards.

Nonclinical tests submitted included appropriate biocompatibility tests, as well as performance bench testing. Bench test data including Tensile Strength Test, Power Injection Performance Test, Dynamic Failure Test, Static Pressure Test, Static Burst Test, Life Cycle Power Injection Test, Catheter Flow Rate Test, Air Leakage Test and Liquid Leakage Test all met predefined acceptance criteria, and supported the substantial equivalence determination.

{7}------------------------------------------------

Do performance data demonstrate equivalence? Yes. Performance data gathered in design verification testing demonstrated that the NEXUS™ MIDLINE CT Catheter is substantially equivalent to the noted predicate devices.

CONCLUSION

The NEXUS™ MIDLINE CT Catheter met all established acceptance criteria for performance testing and design verification testing. This testing demonstrated that the NEXUS™ MIDLINE CT Catheter is safe and effective for its intended use, and based on FDA's decision tree is substantially equivalent to the SYNERGY™ CT PICC (K101329), UNI-PICC™ (K130034), Vascu-PICC® and Midline Catheters, Single, Double and Triple Lumen (K091586) and PowerGlide Midline Catheter (K121073).

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).