Intended Use

The Nipro SafeTouch PSV Scalp Vein Set with Safety Device is intended to insection into a patient's vascular system (for single use) as an indwelling device to administer fluids intravenously or to sample blood. Secondly, it is designed with an active sharp feature that requires physical action by the clinician to prevent needlestick incidents.

Device Description

The SafeTouch PSV Scalp Vein Set with Safety Device is a safety intravascular administration set. It incorporates a safety mechanism that requires physical action by the clinician, to help protect against exposure to blood borne pathogens caused by accidental needlestick injuries after use.

The basic structure of the device consists of a 25G X ¾ inch needle, needle hub with incorporated safety mechanism, wing unit, micro bore tubing, a clamp and female luer connector with cap. It is supplied sterile for single use only, non-pyrogenic, non-toxic, is not made with DEHP and does not contain natural rubber latex.

AI/ML Overview

The provided document is a 510(k) Summary for a medical device (Nipro SafeTouch PSV Scalp Vein Set with Safety Device) and does not contain information about the development or validation of an AI/ML powered device.

Therefore, I cannot extract the requested information regarding acceptance criteria, study details, data provenance, expert ground truth establishment, MRMC studies, standalone algorithm performance, or training set details as none of this information is present in the document.

The document states:

  • "Clinical tests: This submission does not warrant any clinical testing, therefore no clinical testing performed for or provided in this submission." This explicitly confirms that clinical validation studies (which would involve human readers and potentially AI assistance if it were an AI device) were not conducted.
  • The tests conducted were "Non-clinical tests" focused on the physical and mechanical properties of the device (e.g., Gauging test, Liquid leakage test, Mechanical strength of unions, Flow rate, Penetration force of needle, Safety mechanism-Return force after locked, etc.). These are standard engineering and performance tests for medical hardware, not AI model validation.
  • The purpose of this submission is to demonstrate "substantial equivalence" to a predicate device (Nipro SafeTouch Scalp Vein Set - K011297) based on physical characteristics, operational mode, basic scientific technology, and intended use, verified through non-clinical performance testing.

In summary, the request for information on AI/ML validation studies cannot be fulfilled from this document because it describes a traditional medical device, not an AI/ML-powered one.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Nipro Medical Corporation. The logo consists of a stylized, intertwined ribbon-like symbol on the left, followed by the word "NIPRO" in bold, sans-serif font. Below "NIPRO" is the text "MEDICAL CORPORATION" in a smaller, sans-serif font.

FFB 1 4 2014

510(k) Summary: Nipro SafeTouch PSV Scalp Vein Set with Safety Device

807.92(a)(1)

Applicant:

Nipro Medical Corporation 3150 NW 107th Ave. Miami FL 33172 Tel: 305-599-7174 1056186

Contact Person:

Establishment Reg.:

Jessica Oswald-McLeod Director, Quality Assurance & Regulatory Affairs

January 14, 2014 Date of summary preparation:

807.92(a)(2)

Trade Name: Nipro SafeTouch PSV Scalp Vein Set with Safety Device Common Name: Safety Scalp Vein Set Classification Name: catheter, intravascular, therapeutic, short-term less than 30 days Regulation Number: 21 CFR 880.5200 Panel: 80 Product Code: FOZ

807.92(a)(3)

Legally marketed substantial equivalent device: Nipro SafeTouch Scalp Vein Set - K011297

807.92(a)(4)

Description of device:

The SafeTouch PSV Scalp Vein Set with Safety Device is a safety intravascular administration set. It incorporates a safety mechanism that requires physical action by the clinician, to help protect against exposure to blood borne pathogens caused by accidental needlestick injuries after use.

The basic structure of the device consists of a 25G X ¾ inch needle, needle hub with incorporated safety mechanism, wing unit, micro bore tubing, a clamp and female luer connector with cap. It is supplied sterile for single use only, non-pyrogenic, non-toxic, is not made with DEHP and does not contain natural rubber latex.

807.92(a){5)

Indications for Use:

The Nipro SafeTouch PSV Scalp Vein Set with Safety Device is intended to be used for insertion into a patient's vascular system (for single use) as an indwelling device to administer fluids intravenously or to sample blood. Secondly, it is designed with an

{1}------------------------------------------------

active sharp feature that requires physical action by the clinician to prevent needlestick incidents.

807.92(a)(6)

Comparison of technological characteristics:

The syringe is substantially equivalent to the predicate device in the following technological characteristics:

  • . Physical characteristics
  • Operational mode .
  • . Basic Scientific Technology
  • . Intended Use

807.92(b)(1)

Non-clinical tests submitted:

Performance testing was conducted to verify that the device is safe and effective for its intended use. These tests include: Gauging test, Liquid leakage test, Mechanical strength of unions, Flow rate, Closure device (EL luer cap) for leakage test, Penetration force of needle, Pull force at connection, Priming volume, Safety mechanism-Return force after locked, Safety mechanism-Move force after activated, Safety mechanism-Pull force after locked and Individual Package integrity. Other test items include: Transportation, biocompatibility, and shelf-life testing.

These tests along with their associated results and conclusions are included in this submission.

807.92(b)(2)

Clinical tests:

This submission does not warrant any clinical testing, therefore no clinical testing performed for or provided in this submission.

807.92(b)(3)

Conclusions drawn from non-clinical and clinical tests:

The results of the performance testing and the comparison of technological characteristics with the predicate device demonstrate that the Nipro SafeTouch PSV Scalp Vein Set with Safety Device performs equivalent to the predicate device and is safe and effective when used as intended.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The text is written in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 14, 2014

Nipro Medical Corporation C/O Jessica Oswald-McLeod Director. Quality Assurance & Regulatory Affairs 3150 NW 107th ST Miami, FL 33172

Re: K140136

Trade/Device Name: Nipro SafeTouch PSV Scalp Vein Set with Safety Device Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter, therapeutic, short-term less than 30 days Regulatory Class: II Product Code: FOZ Dated: January 14, 2014 Received: January 17, 2014

Dear Ms. Oswald-McLeod:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Oswald-McLeod

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/6 description: The image shows the text "Kwame O. Ulmer for -s FDA". The text is arranged in three lines. The first line contains the name "Kwame O. Ulmer" and the word "for". The second line contains the letters "-s" and the acronym "FDA".

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K140136

Device Name

Nipro SafeTouch PSV Scalp Vein Set with Safety Device

Indications for Use (Describe)

The Nipro SafeTouch PSV Scalp Vein Set with Safety Device is intended to insection into a patient's vascular system (for single use) as an indwelling device to administer fluids intravenously or to sample blood. Secondly, it is designed with an active sharp feature that requires physical action by the clinician to prevent needlestick incidents.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

CALL CORPORATION FOR FOR FOR FOR FOR FOR MANAGER STATULAR CONSTITUTION OF CONSULTION OF CONSULTION OF CONSULTION OF THE CONSTITUTION OF THE

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/4/Picture/13 description: The image shows the logo for the Food and Drug Administration (FDA). The logo is a stylized version of the letters "FDA" in a bold, sans-serif font. The letters are outlined with a thick black line, and there is a thin black line running through the center of each letter. The logo is simple and recognizable, and it is often used on products that have been approved by the FDA.

Digitally signed by Richard C. Chapman Date: 2014.02.12 17:02:33 -05'00'

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).