K Number
K130665

Validate with FDA (Live)

Date Cleared
2013-04-09

(28 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, o Oxazepam, Cocaine, Cocaine, Connelinoids, Contraliae, Methylenedioxymethamphetamine, Morphine, Morphine 2000, Phencyclidine, Oxycodone and Buprenorphine in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine(AMP)1000 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL
Cannabinoids (THC)50 ng/mL
Methamphetamine (MET)1000 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP)300 ng/mL
Morphine 2000 (OPI)2000 ng/mL
Phencyclidine (PCP)25 ng/mL
Oxycodone (OXY)100 ng/mL
Buprenorphine (BUP)10 ng/mL

Configuration of the Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel can consist of any combination of the above listed drug analytes.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method.

The test will yield preliminary positive results when prescription drugs Buprenorphine, Oxazepam, Oxycodone and Secobarbital are ingested, even at or above therapeutic doses. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Device Description

Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Oxazepam, Amphetamine, Secobarbital, Cocaine, Cannabinoids, Methamphetamine, Methylenedioxymethamphetamine, Morphine 2000, Phencyclidine, Oxycodone and Buprenorphine in human urine samples. Wondfo Multi-Drug devices detect each of analytes on different strips.

A positive urine sample will not generate a colored-line for the specific drug tested in the designated region. A negative urine specimen or a urine sample containine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methylenedioxymethamphetamine, Morphine, Morphine, Morphine 2000, Phencyclidine, Oxycodone and Buprenorphine at the concentration below the designated cutoff levels will generate a colored line in the designated test region for the drug. To serve as a test control, a color line will always appear at the control region.

AI/ML Overview

The document describes the Wondfo Multi-Drug Urine Test Cup and Panel, which are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of various drugs in human urine. The acceptance criteria and study details are extracted below.

1. Table of Acceptance Criteria and Reported Device Performance:

The document implicitly defines the acceptance criteria as the detection of the specified drugs at or above defined cut-off levels. The "reported device performance" is not presented in a formal table with specific metrics like sensitivity, specificity, or accuracy compared to a gold standard. Instead, the document states that the device is "substantially equivalent" to legally marketed predicate devices, implying that its performance meets the established standards for such devices. The document does not provide a direct performance study comparing the device against a confirmed ground truth with quantitative metrics for true positives, true negatives, false positives, and false negatives.

Drug (Identifier)Cut-off level (Acceptance Criteria)Reported Device Performance (Implicitly meets predicate device performance)
Amphetamine (AMP)1000 ng/mLQualitatively detects at or above cut-off
Secobarbital (BAR)300 ng/mLQualitatively detects at or above cut-off
Oxazepam (BZO)300 ng/mLQualitatively detects at or above cut-off
Cocaine (COC)300 ng/mLQualitatively detects at or above cut-off
Cannabinoids (THC)50 ng/mLQualitatively detects at or above cut-off
Methamphetamine (MET)1000 ng/mLQualitatively detects at or above cut-off
Methylenedioxymethamphetamine (MDMA)500 ng/mLQualitatively detects at or above cut-off
Morphine (MOP)300 ng/mLQualitatively detects at or above cut-off
Morphine 2000 (OPI)2000 ng/mLQualitatively detects at or above cut-off
Phencyclidine (PCP)25 ng/mLQualitatively detects at or above cut-off
Oxycodone (OXY)100 ng/mLQualitatively detects at or above cut-off
Buprenorphine (BUP)10 ng/mLQualitatively detects at or above cut-off

2. Sample Size and Data Provenance for the Test Set:

The provided document does not contain information regarding the specific sample size used for a test set or the data provenance (e.g., country of origin, retrospective or prospective) for any performance study. The submission is a 510(k) summary demonstrating substantial equivalence, not a detailed clinical study report.

3. Number of Experts and Qualifications for Ground Truth:

The document does not provide information regarding the number of experts used to establish a ground truth or their qualifications. The device is for in-vitro diagnostics, and for such devices, ground truth is typically established by laboratory reference methods.

4. Adjudication Method for the Test Set:

The document does not describe any adjudication method as no specific test set or human interpretation of results is detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study is mentioned or referenced in the provided document. This device is an in-vitro diagnostic test, not a medical imaging or interpretation aid where human reader performance would be a primary metric.

6. Standalone (Algorithm Only) Performance Study:

The document does not detail a "standalone" algorithmic performance study in the context of AI. The device itself is a standalone immunoassay. The information provided is a description of the device and its intended use, along with a comparison to predicate devices, to establish substantial equivalence. Implicitly, the performance referred to is the device's ability to qualitatively detect drugs at the specified cut-off levels.

7. Type of Ground Truth Used:

While not explicitly stated within the provided text, for in-vitro diagnostic devices like drug test cups, the "ground truth" for demonstrating performance is typically established by confirmatory analytical methods such as Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). The document mentions this: "A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method." This indicates that these methods would be used to verify the presence and concentration of drugs in urine samples to establish ground truth for testing the immunoassay's accuracy.

8. Sample Size for the Training Set:

The document does not provide information regarding the sample size used for any training set. This is a descriptive document for regulatory submission, not a machine learning development report.

9. How the Ground Truth for the Training Set Was Established:

The document does not provide information regarding how ground truth for a training set was established. Given the nature of a lateral flow immunoassay, a "training set" in the machine learning sense is not typically applicable for its development or validation process as described in this type of regulatory submission. The device's performance relies on established chemical and immunological principles, not on learning from a data set.

{0}------------------------------------------------

SIO (x)SUMMARY

March 6, 2013

012-86-20-32296069

1. Date the summary was prepared:

  1. Submitter's name: Address:

Phone:

Name of contact person:

Joe Shia LSI International Inc. 504 East Diamond Ave., Suite F Gaithersburg, MD 20877 Telephone: 240-505-7880 Fax: 301-916-6213

Guangzhou Wondfo Biotech Co., Ltd. South China University of Technology

Guangzhou, P.R. China 510641

11 130 66 5

3. Name of the device

Common or usual name:

Trade or proprietary name:

Multi-Drug Urine Test Cup Multi-Drug Urine Test Panel Wondfo Multi-Drug Urine Test Cup Wondfo Multi-Drug Urine Test Panel

Classification: All are Class II medical devices with the following various product codes with Code of Federal Regulation references:

Product CodeClassificationRegulation SectionPanel
DKZAmphetamineII21 CFR § 862.3100, AmphetamineTest SystemToxicology(91)
LDJCannabinoidsII21 CFR § 862.3870, CannabinoidsTest SystemToxicology(91)
DIOCocaineII21 CFR § 862.3250, Cocaine andCocaine Metabolites Test SystemToxicology(91)
LAFMethamphetamineII21 CFR § 862.3610,Methamphetamine Test SystemToxicology(91)
DJGMorphine 2000II21 CFR § 862.3650, MorphineTest SystemToxicology(91)
LAFMethylenedioxymethamphetamineII21 CFR § 862.3610,Methamphetamine Test SystemToxicology(91)
DJGMorphineBuprenorphineII21 CFR § 862.3650, MorphineTest SystemToxicology(91)
DJGJXMOxazepamII21 CFR § 862.3170,Benzodiazepine Test SystemToxicology(91)
DJGOxycodoneII21 CFR § 862.3650, Opiate TestSystemToxicology(91)
LCMPhencyclidine.unclassifiedEnzyme ImmunoassayPhencyclidineToxicology(91)
DISSecobarbitalII21 CFR § 862.3150, BarbiturateTest SystemToxicology(91)

APR 0 9 2013

1

{1}------------------------------------------------

4. Description of the device:

Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Oxazepam, Amphetamine, Secobarbital, Cocaine, Cannabinoids, Methamphetamine, Methylenedioxymethamphetamine, Morphine 2000, Phencyclidine, Oxycodone and Buprenorphine in human urine samples. Wondfo Multi-Drug devices detect each of analytes on different strips.

A positive urine sample will not generate a colored-line for the specific drug tested in the designated region. A negative urine specimen or a urine sample containine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methylenedioxymethamphetamine, Morphine, Morphine, Morphine 2000, Phencyclidine, Oxycodone and Buprenorphine at the concentration below the designated cutoff levels will generate a colored line in the designated test region for the drug. To serve as a test control, a color line will always appear at the control region.

5. Intended use of the device:

Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, o Oxazepam, Cocaine, Cocaine, Connelinoids, Contraliae,
Methylenedioxymethamphetamine, Morphine, Morphine 2000, Phencyclidine, Oxycodone and Buprenorphine in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine(AMP)1000 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL
Cannabinoids (THC)50 ng/mL
Methamphetamine (MET)1000 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP)300 ng/mL
Morphine 2000 (OPI)2000 ng/mL
Phencyclidine (PCP)25 ng/mL
Oxycodone (OXY)100 ng/mL
Buprenorphine (BUP)10 ng/mL

Configuration of the Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel can consist of any combination of the above listed drug analytes.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method.

The test will yield preliminary positive results when prescription drugs Buprenorphine, Oxazepam, Oxycodone and Secobarbital are ingested, even at or above therapeutic doses. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

6. Comparison to the predicate device

Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are a "modified" product format derived from the previously FDA-cleared Wondfo single DOA Tests. A summary comparison of features of the Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel and the predicate devices is provided in the following Table

ItemNew DevicesPredicate devices (K112071)
Indication(s) foruseFor the qualitative determination of Amphetamine(AMP), Secobarbital (BAR), Oxazepam (BZO),Cocaine (COC), Cannabinoids (THC),Methamphetamine (MET),Methylenedioxymethamphetamine (MDMA)For the qualitative determination of Cocaine(COC), or Methamphetamine (MET), in humanurine.The configurations of the Predicate devices are

{2}------------------------------------------------

Morphine (MOP), Morphine 2000 (OPI),Phencyclidine (PCP), Oxycodone(OXY) and/orBuprenorphine in human urine.The configurations of the New Devices areavailable in any combination of the above tests.only available in single drug test.
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibody immunochemistry.Same
Type Of TestImmunoassay principles that rely on antigen-antibody interactions to indicate positive ornegative resultSame
ResultsQualitativeSame
Specimen TypeHuman urineSame
Cut Off ValuesAmphetamine (AMP): 1,000 ng/mlSecobarbital(BAR): 300 ng/mlOxazepam (BZO):300 ng/mlCocaine(COC): 300 ng/mlCannabinoids (THC):50 ng/mlMethamphetamine (MET): 1,000 ng/mlMethylenedioxymethamphetamine (MDMA): 500ng/mlMorphine (MOP): 300 ng/mlMorphine 2000 (OPI): 2000 ng/mlPhencyclidine (PCP): 25 ng/mlOxycodone(OXY) : 100 ng/mlBuprenorphine (BUP): 10 ng/mlSame
ConfigurationsCup, dip cardSame
Intended UseOTC Use & Prescription UseSame

Wondfo Multi-Drug Urine Test Cup is a multi-drug test that offers any combination from 2 to 12 drugs of abuse tests while the predicate devices are single-drug test. And the Wondfo Multi-Drug Urine Test Panel is the same as the test dip card format of the predicate devices except that the Wondfo Multi-Drug Urine Test Panel is a multi-drug test that offers any combination from 2 to 12 drugs of abuse tests while the predicate devices are single-drug test.

3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 9, 2013

GUANGZHOU WONDFO BIOTECH CO., LTD. C/O JOE SHIA REGULATORY CONSULTANT 504E DIAMOND AVE:, SUITE F GAITHERSBURG MD 20877

Re: K130665

Trade/Device Name: Wondfo Multi-Drug Urine Test Cup, Wondfo Multi-Drug Urine Test Panel

Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, DKZ, LDJ, DIO, LAF, JXM, LCM, DIS Dated: March 06, 2013 Received: March 12, 2013

Dear Mr. Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of . devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol Benson - S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): K130665

Device Name: Wondfo Multi-Drug Urine Test Cup Wondfo Multi-Drug Urine Test Panel

Indications for Use:

Wondfo Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Morphine 2000, Phencyclidine, Oxycodone and Buprenorphine in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine(AMP)1000 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL
Cannabinoids (THC)50 ng/mL
Methamphetamine (MET)1000 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP)300 ng/mL
Morphine 2000 (OPI)2000 ng/mL
Phencyclidine (PCP)25 ng/mL
Oxycodone (OXY)100 ng/mL
Buprenorphine (BUP)10 ng/mL

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Denise Johnson#byles; -5
2013.04.09 08:334:30 =04:00'

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

K130665 510(k) =======================================================================================================================================================================

Page 1 of 4

{6}------------------------------------------------

510(k) Number (if known): K130665 .

Device Name: Wondfo Multi-Drug Urine Test Cup Wondfo Multi-Drug Urine Test Panel

Indications for Use:

Configuration of the Wondfo Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The test will yield preliminary positive results when prescription drugs Buprenorphine, Oxazepam, Oxycodone and Secobarbital are ingested, even at or above therapeutic doses.

Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Denise Johnson​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​ Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

v v

K130665 510(k) =

{7}------------------------------------------------

510(k) Number (if known): K130665

Device Name: Wondfo Multi-Drug Urine Test Cup Wondfo Multi-Drug Urine Test Panel

Indications for Use:

Wondfo Multi-Drug Urine Test Panel is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Morphine 2000, Phencyclidine, Oxycodone and Buprenorphine in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine(AMP)1000 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL
Cannabinoids (THC)50 ng/mL
Methamphetamine (MET)1000 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP)300 ng/mL
Morphine 2000 (OPI)2000 ng/mL
Phencyclidine (PCP)25 ng/mL
Oxycodone (OXY)100 ng/mL
Buprenorphine (BUP)10 ng/mL

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Denise Johnsonslyless 55
2013.04.09 08:35:118504.00 Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k) K130665

Page 3 of _4

{8}------------------------------------------------

510(k) Number (if known): K130665

Device Name: Wondfo Multi-Drug Urine Test Cup Wondfo Multi-Drug Urine Test Panel

Indications for Use:

Configuration of the Wondfo Multi-Drug Urine Test Panel can consist of any combination of the above listed drug analytes.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The test will yield preliminary positive results when prescription drugs Buprenorphine, Oxazepam, Oxycodone and Secobarbital are ingested, even at or above therapeutic doses. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Denise John ธรุงกรุปyless -S 2013.04.09 08:35:48 204'00'

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k) K130665

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).